← Product Code [NAD](/productcode/NAD) · K060046

# DERMASTREAM (K060046)

_Enzysurge , Ltd. · NAD · Feb 27, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K060046

## Device Facts

- **Applicant:** Enzysurge , Ltd.
- **Product Code:** [NAD](/productcode/NAD.md)
- **Decision Date:** Feb 27, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4020
- **Device Class:** Class 1
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The DermaStream™ is an occlusive wound dressing which permits the introduction of other topical wound treatments such as irrigating solutions, antimicrobial and enzymatic debriding solutions, suspensions, and other solutions. It is intended to provide a moist healing environment and to allow debridement to facilitate the normal wound healing process.

## Device Story

DermaStream™ is a sterile, single-use, disposable occlusive wound dressing. It functions by creating a moist environment over a wound while allowing the introduction of topical treatments (irrigating, antimicrobial, or enzymatic debriding solutions) and drainage. Unlike vacuum-assisted closure devices that use electrical power, DermaStream™ relies on gravity for solution delivery and drainage. It is used in clinical settings to facilitate the normal wound healing process through controlled topical application.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Occlusive wound dressing; sterile, single-use, disposable. Operates via gravity-fed fluid delivery and drainage. No electrical components or software.

## Regulatory Identification

An occlusive wound dressing is a nonresorbable, sterile or non-sterile device intended to cover a wound, to provide or support a moist wound environment, and to allow the exchange of gases such as oxygen and water vapor through the device. It consists of a piece of synthetic polymeric material, such as polyurethane, with or without an adhesive backing. This classification does not include an occlusive wound dressing that contains added drugs such as antimicrobial agents, added biologics such as growth factors, or is composed of materials derived from animal sources.

## Predicate Devices

- KCI Wound Cell Transparent Wound Dressing ([K020781](/device/K020781.md))
- 3M Tegaderm Transparent Dressing ([K973036](/device/K973036.md))
- 3M Tegaderm Transparent Dressing ([K901845](/device/K901845.md))
- KCI V.A.C. Instillamat ([K021501](/device/K021501.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) Summary

K060046

Pursuant to CFR 807.92, the following 510(k) Summary is provided:

| 1. (a) | Submitter's<br>Corporate<br>Address:               | MedicSense, Ltd.<br>14 Imber<br>Kiriat Arie, Petach Tikva, Israel 49511<br>www.medicsense.com                                                                                                                                                                                                                                                                                |
|--------|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. (b) | Manufacturer<br>Address:                           | EnzySurge, Ltd.<br>63 Ha ' Odem St. POB 1<br>Shoham, Israel 73142                                                                                                                                                                                                                                                                                                            |
|        | Mfg. Phone:                                        | 972-3-979-6344                                                                                                                                                                                                                                                                                                                                                               |
|        | Contact Person:                                    | Dr. Allon Leibovitz, CEO                                                                                                                                                                                                                                                                                                                                                     |
|        | Date:                                              | January 4, 2006                                                                                                                                                                                                                                                                                                                                                              |
| 2.     | Device &<br>Classification<br>Name:                | Dressing, wound and burn occlusive (Class 1), Product Code NAD,<br>21 CFR 878.4020- Trade-name of device: DermaStream™                                                                                                                                                                                                                                                       |
| 3.     | Predicate Device:                                  | KCI Wound Cell Transparent Wound Dressing (K020781)<br>3M Tegaderm Transparent Dressing (K973036 & K901845)<br>KCI V.A.C. Instillamat (K021501)                                                                                                                                                                                                                              |
| 4.     | Description:                                       | The DermaStream™ is an occlusive wound dressing which permits the<br>introduction of topical wound treatments. It is provided as a sterile, single-<br>use, disposable device.                                                                                                                                                                                               |
| 5.     | Intended Use:                                      | The DermaStream™ is an occlusive wound dressing which permits the<br>introduction of other topical wound treatments such as irrigating solutions,<br>antimicrobial and enzymatic debriding solutions, suspensions, and other<br>solutions. It is intended to provide a moist healing environment and to allow<br>debridement to facilitate the normal wound healing process. |
| 6.     | Comparison of<br>Technological<br>Characteristics: | With respect to technology, the DermaStream™ is substantially equivalent<br>to its predicate devices. This is because it is a combination of the<br>fundamental features and characteristics of its predicate devices. The<br>primary difference is the ability to provide a solution and drainage to the                                                                    |

protected wound via gravity versus a vacuum provided by electrical power

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circle around the eagle. The eagle is black, and the text is also black. The logo is simple and recognizable.

Public Health Service

FEB 2 7 2006

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

EnzySurge, Ltd. c/o Mr. George J. Hattub, RAC, CQE Senior Staff Consultant MedicSense, Ltd. 291 Hillside Avenue Somerset, Massachusetts 02726

Re: K060046

Trade/Device Name: DermaSteam™ Regulation Number: 21 CFR 878.4020 Regulation Name: Occlusive wound dressing Regulatory Class: I Product Code: NAD Dated: January 4, 2006 Received: January 6, 2006

Dear Mr. Hattub:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roviewed your and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assossion in the Merced 1976, the enactment date of the Medical Device Amendments, or to conniner of the rear reclassified in accordance with the provisions of the Federal Food, Drug, de neces mat na vo been require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, dierefore, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it your de subject to such additional controls. Existing major regulations affecting your device n may of sucject to Cool of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA ean or the mirther announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of acrisou that i bring that your device complies with other requirements of the Act that I Dr Has/Intations and regulations administered by other Federal agencies. You must or any I vith all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI I rart 007), idoling (21 (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2 - Mr. Hattub

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter will and w yours of substantial equivalence of your device to a legally premation fouriedion." - e results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: DermaStream™

Indications For Use: The DermaStream™ is an occlusive wound dressing which permits the introduction of topical wound treatments such as irrigating solutions, antimicrobial and enzymatic debriding solutions, suspensions, and other solutions. It is intended to provide a moist healing environment and allow debridement to facilitate the normal wound healing process.

Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

> Concurrence of C Qffice of Device Evaluation (ODE) - son of General, Restorative. Page 1 of 1 nd Neurological Devices

510(k) Number

K060046

---

**Source:** [https://fda.innolitics.com/device/K060046](https://fda.innolitics.com/device/K060046)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
