PORT-A-CATH, PORT-A-CATH II, PROPORT AND P.A.S. PORT IMPLANTABLE VENOUS ACCESS SYSTEMS WITH DUAL LAYER CATHETER
K060036 · Smiths Medical MD, Inc. · LJT · May 17, 2006 · General Hospital
Device Facts
Record ID
K060036
Device Name
PORT-A-CATH, PORT-A-CATH II, PROPORT AND P.A.S. PORT IMPLANTABLE VENOUS ACCESS SYSTEMS WITH DUAL LAYER CATHETER
Applicant
Smiths Medical MD, Inc.
Product Code
LJT · General Hospital
Decision Date
May 17, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5965
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PORT-A-CATH® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling. PORT-A-CATH® II Systems with Dual Layer Catheter are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling. ProPort® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling. P.A.S. PORT® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling.
Device Story
Totally implantable venous access systems consisting of a portal, catheter, and connector; designed for repeated venous access for medication/fluid delivery and blood sampling. Portals are titanium, polysulfone/titanium, or plastic; catheters feature a dual-layer design with a Polyether block amide outer lining and an aliphatic polyurethane inner lining. Components are assembled during implantation in the chest or arm. Systems utilize ULTRA-LOCK® or SlideLock connectors. Replaces previous single-layer polyurethane catheters. Used by clinicians to provide long-term vascular access, facilitating therapy delivery and reducing the need for repeated venipuncture, thereby benefiting patients undergoing chronic treatments.
Clinical Evidence
No clinical data. Safety and effectiveness established via bench testing (in-vitro) and in-vivo performance evaluation of the new catheter material. Biocompatibility testing was performed on all system components with passing results.
Technological Characteristics
Totally implantable venous access system. Portal materials: titanium, polysulfone, or plastic. Catheter: dual-layer construction (Polyether block amide outer lining, aliphatic polyurethane inner lining). Connectors: ULTRA-LOCK® or SlideLock. Sterilization: unchanged from predicate. No software or electronic components.
Indications for Use
Indicated for patients requiring repeated vascular access for injection, infusion therapy, or venous blood sampling.
Regulatory Classification
Identification
A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.
Predicate Devices
PORT-A-CATH® Implantable Venous Access Systems
PORT-A-CATH® II Implantable Venous Access Systems
ProPort® Implantable Venous Access Systems
P.A.S. PORT® Implantable Venous Access Systems
Submission Summary (Full Text)
{0}------------------------------------------------
PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Summary of Safety and Effectiveness
MAY 17 2006
#### GENERAL INFORMATION
| Applicant's Name and Address: | Smiths Medical MD, Inc.<br>1265 Grey Fox Road<br>St. Paul, MN 55112 |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Melanie Hess RN, BSN<br>Regulatory and Clinical Affairs Associate |
| Common/Usual Name: | Implantable Access System |
| Proprietary Name: | PORT-A-CATH®, PORT-A-CATH® II,<br>ProPort®, and P.A.S. PORT® Implantable<br>Venous Access Systems with Dual Layer<br>Catheter |
| Equivalence Device Comparison: | PORT-A-CATH®, PORT-A-CATH® II,<br>ProPort®, and P.A.S. PORT® Implantable<br>Venous Access Systems |
#### II. DEVICE DESCRIPTION
### PORT-A-CATH® Implantable Venous Access Systems with Dual Layer Catheter
The PORT-A-CATH® Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems and are designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids nutritional solutions and for the sampling of venous blood. The PORT-A-CATH® venous systems can be placed in the chest.
The PORT-A-CATH® Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems consisting of the following components; a portal. catheter and connector. The components are assembled during implantation. The portals are available in titanium standard or Low Profile", and the ULTRA-LOCK® catheter connector is utilized in the PORT-A-CATH® systems. The catheter is a Polyether block amide outer lining and an aliphatic polyurethane inner lining. This catheter will replace the current single layer polyurethane catheters on the equivalent Smiths Medical MD, Inc. PORT-A-CATH® Implantable Venous Access Systems, The systems are offered in either a kit or a tray configuration.
Confidential
! . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . :
.
1999 18
January 03, 2006
{1}------------------------------------------------
### PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Summary of Safety and Effectiveness
#### PORT-A-CATH® II Implantable Venous Access Systems with Dual Layer Catheter
The PORT-A-CATH® II Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems and are designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids nutritional solutions and for the sampling of venous blood. The PORT-A-CATH® II venous systems can be placed in the chest.
The PORT-A-CATH® II Implantable Venous Access Systems with Dual Laver Catheter are totally implantable access systems consisting of the following components: a portal, catheter and connector. The components are assembled during implantation. The portals are available in polysulfone/titanium standard or Low Profile", and the ULTRA-LOCK® catheter connector is utilized in the PORT-A-CATH® II systems. The catheter is a Polyether block amide outer lining and an aliphatic polyurethane inner lining. This catheter will replace the current single layer polyurethane catheters on the equivalent Smiths Medical MD, Inc. PORT-A-CATH® II Implantable Venous Access Systems. The systems are offered in either a kit or a tray configuration.
#### ProPort® Implantable Venous Access Systems with Dual Layer Catheter
The ProPort® Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems and are designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids nutritional solutions and for the sampling of venous blood. The ProPort venous systems can be placed in the chest.
The ProPort® Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems consisting of the following components: a portal, catheter and connector. The components are during implantation. The portals are available in plastic standard or Low Profile", and the SlideLock connector is utilized in the ProPort® systems. The catheter is a Polyether block amide outer lining and an aliphatic polyurethane inner lining. This catheter will replace the current single layer polyurethane catheters on the equivalent Smiths Medical MD, Inc. ProPort® Implantable Venous Access Systems. The systems are offered in either a kit or a tray configuration.
## P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter
The P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems and are designed to permit repeated access to the venous system for the parenteral delivery of medications, fluids nutritional solutions and for the sampling of venous blood. The P.A.S. PORT® venous systems can be placed in the arm or chest.
The P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter are totally implantable access systems consisting of the following components: a portal, catheter and connector. The components are assembled during implantation. The portals
Confidential
Page 2 of 5
January 03, 2006
5
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the text "page 3 of 5" on the top line and "K0600036" on the second line. The text is handwritten in black ink on a white background. The handwriting is somewhat messy, but the text is still legible.
## PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Summary of Safety and Effectiveness
are available in titanium or plastic/titanium standard or Low Profile™, and the ULTRA-LOCK® catheter connector is utilized in the P.A.S. PORT® systems. The catheter is a Polyether block amide outer lining and an aliphatic polyurethane inner lining. This catheter will replace the current single layer polyurethane catheters on the equivalent Smiths Medical MD, Inc. P.A.S. PORT® Implantable Venous Access Systems. The systems are offered in either a kit or a tray configuration.
#### III. INTENDED USE OF THE DEVICE
## PORT-A-CATH® Implantable Venous Access Systems with Dual Layer Catheter
PORT-A-CATH® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
### PORT-A-CATH® II Implantable Venous Access Systems with Dual Layer Catheter
PORT-A-CATH® II Systems with Dual Layer Catheter are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling.
#### ProPort® Implantable Venous Access Systems with Dual Layer Catheter
ProPort® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling.
# P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter
P.A.S. PORT® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling.
#### IV. DEVICE COMPARISON
# PORT-A-CATH® Implantable Venous Access Systems with Dual Layer Catheter
The PORT-A-CATH® Implantable Venous Access Systems with Dual Layer Catheter was compared to and found to be substantially equivalent to the following commercially available predicate systems with respect to indications for use and performance features PORT-A-CATH® Implantable Venous Access Systems.
Confidential
Page 3 of 5
January 03, 2006
1. 1. 1. 1. 1.
.
{3}------------------------------------------------
page 4 of 5
K060036
## PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Summary of Safety and Effectiveness
## PORT-A-CATH® II Implantable Venous Access Systems with Dual Layer Catheter
The PORT-A-CATH® II Implantable Venous Access Systems with Dual Layer Catheter was compared to and found to be substantially equivalent to the following commercially available predicate systems with respect to indications for use and performance features: PORT-A-CATH® II Implantable Venous Access Systems.
## ProPort® Implantable Venous Access Systems with Dual Layer Catheter
The ProPort® Implantable Venous Access Systems with Dual Layer Catheter was compared to and found to be substantially equivalent to the following commercially available predicate systems with respect to indications for use and performance features: ProPort® Implantable Venous Access Systems.
## P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter
The P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter was compared to and found to be substantially equivalent to the following commercially available predicate systems with respect to indications for use and performance features: P.A.S. PORT® Implantable Venous Access Systems.
#### SUMMARY OF STUDIES V.
### A. Functional Testing
In-vitro testing was conducted in accordance with the FDA guidance, "Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters," dated March 16, 1995. All testing met product specification.
In-vivo testing of the new catheter material of the PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter was conducted to evaluate product performance. All testing met product specification.
Biocompatibility testing was conducted on system components. All testing passed.
Packaging and Sterilization systems are unchanged from predicate devices.
## B. Clinical Studies
Human clinical studies were deemed not necessary to evaluate the safety or effectiveness of the PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter.
Confidential
Page 4 of 5
・・・
{4}------------------------------------------------
**SMITHS MEDICAL MD, INC.**
510(k) Premarket Notification
## PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Summary of Safety and Effectiveness
#### C. Conclusions Drawn from the Studies
Based upon the information provided, the PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter is safe, effective and performs to established specifications.
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Confidential
January 03, 2006
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/2 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo consists of a stylized graphic of three abstract human figures connected by flowing lines, arranged in a triangular shape. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the graphic.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## MAY 17 2006
Ms. Melanie Hess Regulatory and Clinical Affairs Associate Smiths Medical MD, Incorporated 1265 Grey Fox Road St. Paul. Minnesota 55112
Re: K060036
Trade/Device Name: PORT-A-CATH® Systems with Dual Layer Catheter, PORT-A-CATH® II Systems with Dual Layer Catheter, ProPort ® Systems with Dual Layer Catheter, P.A.S. Port® Systems with Dual Layer Catheter Regulation Number: 21 CFR 880.5965 Regulation Name: Subcutaneous, Implanted Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT Dated: January 3, 2006 Received: January 5, 2006
Dear Ms. Hess:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
#### Page 2 - Ms. Hess
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Chiu Lin, Ph.D.
Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Indications for Use
510(k) Number (if known): TBD
Device Name: PORT-A-CATH® Systems with Dual Layer Catheter
Indications for Use:
"PORT-A-CATH® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling."
Prescription Use X 21 CFR 801.109)
OR
Over-The Counter Use Per
Device Name: PORT-A-CATH® II Systems with Dual Layer Catheter
Indications for Use:
"PORT-A-CATH" II Systems with Dual Layer Catheter are indicated when patient therapy requires repeated vascular access for injection or infusion therapy and/or venous blood sampling."
Prescription Use X 21 CFR 801.109)
OR
Over-The Counter Use Per
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthony D. Vincent
an Sign-City) of sion of Anesthesiology, General Hospital,
action Control, Dental Devices
: * * (x) Number: _ Kybopp 36
{8}------------------------------------------------
## PORT-A-CATH®, PORT-A-CATH® II, ProPort®, and P.A.S. PORT® Implantable Venous Access Systems with Dual Layer Catheter Indications for Use
510(k) Number (if known): TBD
Device Name: ProPort® Systems with Dual Layer Catheter
Indications for Use:
"ProPort® Systems with Dual Layer Catheter are indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling."
Prescription Use X 21 CFR 801.109)
OR
Over-The Counter Use Per
Device Name: P.A.S. Port® Systems with Dual Layer Catheter
Indications for Use:
"P. A.S. PORT Systems with Dual Layer Catheter are indicated when patient therapy requires repeated venous access for injection or infusion therapy and/or venous blood sampling."
Prescription Use X 21 CFR 801.109)
OR
Over-The Counter Use Per
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Clutos Vime
on of Anesthesiology, General Hospital, Control, Dental Devices
Number: K 469936
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.