The MAKO Surgical Unicondylar Knee is a single use device intended for cemented reconstruction of the medial or lateral femur and corresponding tibial surface of moderately disabled and/or painful knee resulting from osteoarthritis, traumatic arthritis provided there is evidence of sufficient bone to seat the implant. Indications for use include moderate joint impairment from painful arthritis (osteo and/or post-traumatic), and as an alternative to tibial osteotomy in patients with unicompartmental arthritis.
Device Story
MAKO Surgical Unicondylar Knee System is a cemented, single-use knee prosthesis; replaces medial or lateral femorotibial surfaces. Components include CoCr femoral component and UHMWPE tibial component with radiographic marker pin. System includes manual surgical instrumentation (sizers, templates, trials, drills, impactors) for implantation. Used by surgeons in clinical settings for patients with moderate joint impairment from arthritis. Provides mechanical reconstruction of the knee joint; restores joint function; alleviates pain associated with unicompartmental arthritis.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Materials: Femoral component (CoCr per ASTM F-75), Tibial component (UHMWPE per ASTM F-648 Type 2), Radiographic marker (Titanium wire per ASTM F-1341). Design: Non-constrained, cemented, metal/polymer femorotibial prosthesis. Congruency: 20-27%. Sterilization: Gamma radiation for implants; prevacuum steam for instrumentation. Packaging: Double sterile barriers; tibial component in Argon gas.
Indications for Use
Indicated for patients with moderate joint impairment from painful osteoarthritis or post-traumatic arthritis requiring cemented reconstruction of the medial or lateral femur and corresponding tibial surface; alternative to tibial osteotomy for unicompartmental arthritis; requires sufficient bone stock to seat the implant.
Regulatory Classification
Identification
A knee joint femorotibial metal/polymer non-constrained cemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral condylar resurfacing component or components made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components made of ultra-high molecular weight polyethylene and are intended for use with bone cement (§ 888.3027).
Predicate Devices
Stelkast Corporation's Unicondylar Knee System (K032824)
Submission Summary (Full Text)
{0}------------------------------------------------
FEB 2 3 2006
Image /page/0/Picture/2 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, sans-serif font, with the words "SURGICAL CORP" in a smaller font size underneath. Above the word "MAKO" is a graphic element that resembles a stylized mountain range or a shark fin, created with a pixelated effect. The overall design is simple and modern, conveying a sense of precision and innovation.
2901 simms street
lei 954 927 2044
fax 954.927.0446
## EXHIBIT E - 510(K) SUMMARY
hollywood, florida 33020
| Submitter: | MAKO Surgical Corp. |
|----------------------------|------------------------------------------------------------------------------------|
| Address: | 2901 Simms Street, Hollywood, FL, 33020 |
| Phone number / Fax Number: | (Ph) 954-927-2044 x105; (F) 954-927-0446 |
| Contact Person: | William F. Tapia |
| Date Prepared: | December 30, 2005 |
| Proprietary Name: | MAKO Surgical Unicondylar Knee System |
| Common Name: | Knee joint femorotibial metal/polymer non-constrained cemented prosthesis |
| Classification Name / #: | Class II; 21 CFR 888.3520 |
| Product Code: | 87 HSX – Knee joint femorotibial metal/polymer non-constrained cemented prosthesis |
Substantial Equivalence: The MAKO Surgical Unicondylar Knee System is substantially equivalent to Stelkast Corporation's Unicondylar Knee System (K032824).
| Feature | Stelkast Unicondylar Knee System (K032824) / MAKO Surgical Corp.'s<br>Unicondylar Knee System |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use/Indications for<br>Use | The MAKO Surgical Unicondylar Knee is a single use device intended for<br>cemented reconstruction of the medial or lateral femur and corresponding tibial<br>surface of moderately disabled and/or painful knee resulting from osteoarthritis,<br>traumatic arthritis provided there is evidence of sufficient bone to seat the<br>implant. Indications for use include moderate joint impairment from painful<br>arthritis (osteo and/or post-traumatic), and as an alternative to tibial osteotomy in<br>patients with unicompartmental arthritis. |
| Implant Components | CoCr femoral component. UHMWPE tibial component. Radiographic marker pin in tibial component |
| Sizes | Femoral components available in 5 sizes (left and right component for each size). Tibial components available in 5 sizes with 3 thicknesses (6.5mm, 7.5mm, 8.5mm). Thinnest section of 6.5mm tibial component is 6.0mm. |
| Congruency | 20 – 27% congruency |
| Materials | Femoral component – CoCr per ASTM F-75 Tibial component – UHMWPE per ASTM F-648 Type 2. Radiographic marker pin – titanium wire per ASTM F-1341. |
| Instrumentation | Provided separately in a re-usable/sterilizable tray. Tray includes various tools<br>(e.g., sizers, templates, trials, drill, gage, impactors, inserters, extractors) used<br>during surgery. |
| Sterilization and Packaging | Sterilization: Implant and tibial components – gamma radiation Instrumentation – prevacuum steam Packaging: Both femoral and tibial components are supplied in double sealed containers maintaining double sterile barriers. Tibial component is packaged in Argon gas environment. |
| Biocompatibility | Both devices are made of materials for surgical implant applications per ASTM<br>standards listed above. |
Description: The MAKO Surgical Unicondylar Knee System is for use in reconstruction of either the medial or Description. The MANO Ourface. The MAKO Surgical Unicondylar Knee System consists of the following basic components. These components are intended for cemented, one-time use.
- CoCr distal femoral component .
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters. Below the word "MAKO" is the text "SURGICAL CORP" in smaller letters. Above the word "MAKO" is a graphic that resembles a mountain range.
fax 954.327.0446 hollywood, florida 35020 lei 954,927 2044
- UHMWPE tibial resurfacing component -- includes radiographic pin .
2001 simms street
Standard unicondylar knee instrumentation is also provided with the system. This includes for example femoral Standard unicondylar knee instrumentation is also provides, this leveling guide, impactor,
sizers, femoral templates with handles & peg drill assembly, tibial inotrument trav sizers, femoral templates with handles & peg unlinessentialy, tible in the same of the many of the matrument tray.
Intended Use/Indications for Use: The MAKO Surgical Unicondylar Knee is a single use device intended for Intended Uselindications for USE. The MANO Oulgiour Onloondyna mobile in the materaly disabled cemented reconstruction of the medial of lateral leman and sorrosponding tamily in the sudence of sufficient bone to and/or paintil Knee resulting from osteoarinhis, traditions and a thritis (osted and/or postseal the inipiant. Thurcations for asc inolde moderate your the with unicompartmental arthritis.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three stylized lines representing its wings and body. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 3 2006
Mr. William F. Tapia Director of Regulatory/Quality/Clinical Affairs Mako Surgical Corp. 2901 Simms Street Hollywood, California 33020
K060017 Re:
Trade/Device Name: Unicondylar Knee System Regulation Number: 21 CFR 888.3520 Regulation Name: Knee joint femorotibial metal/polymer non-constrained cemented prosthesis Regulatory Class: II Product Codes: HSX Dated: February 14, 2006 Received: February 15, 2006
Dear Mr. Tapia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to eoniner to pror of they 2011-11-11 accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{3}------------------------------------------------
Page 2 – Mr. William F. Tapia
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set a CFR Part 807), labeling (21 CFR Part 807), good managed and if applicable, the electronic
forth in the quality systems (QS) regulation (21 CFR Part 820); and 11 GPD 1960 form in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Declients of Cevice as described in your Section 510(k)
This letter will allow you to begin marketing your device as described in your a I his letter will anow you to begin mankeling your antial equivalence of your device to a legally premarket notification. The PDA muding of backand of your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acries for your 201) 276-0120. Also, please note the regulation entitled, Comact the Office of Comphance at (210) = i Ecation" (21CFR Part 807.97). You may obtain Misolanding Uy Telefone to premailsonibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-204 or Manufacturers, international and Oolisanhoon varian of the more of the new online with he
Sincerely yours,
Heckin Leuer (w)
Image /page/3/Picture/5 description: The image shows the name and title of Mark N. Melkerson, M.S., who is the Acting Director of the Division of General, Restorative and Neurological Devices. The image also mentions the Office of Device Evaluation and the Center for Devices and Radiological Health. The text is aligned to the left and is written in a clear, readable font.
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, sans-serif font, with the words "SURGICAL CORP" in a smaller font size underneath. Above the word "MAKO" is a stylized graphic that resembles a shark fin or a wave. The logo is black and white.
... kr 184 827 0449.... .............................................................................................................................................................................. . . . .......? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . #1 12: #RE #MACC . . . . . .
## EXHIBIT G
## INDICATIONS FOR USE
510(k) Number (if known): K060017
Device Name: MAKO Surgical Unicondylar Knee System
Indications for Use:
The MAKO Surgical Unicondylar Knee is a single use device intented for cemented reconstruction of the resultiv The MAKO Surgical Unicondyial Khee is a single dse act noc rice neable disabled and/or paintil knee resulting
medial or lateral femur and corresponding tibial surface of mode medial or lateral femur and corresportung tible Sunder disable in the implant.
from osteoarthritis, traumatic arthritis provided there is evidence of seat the implic Indications for use include moderate joint impaiment from painfilis (osteo and/or post-traumatic), and molcations for use include moderate joint impartmental arthrilis.
| Prescription Use | AND/OR | Over-the-Counter Use |
|-----------------------------|--------|------------------------|
| X | | |
| (Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence(Division Sign-Off) Evaluation (ODE)
Division of General, Restorative.
and Neurological Devices
510(k) Number_
the Countor I lea
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.