← Product Code [GAP](/productcode/GAP) · K053638

# MANI NEEDLE & SUTURE PACK (SILK) (K053638)

_Mani, Inc. · GAP · May 19, 2006 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K053638

## Device Facts

- **Applicant:** Mani, Inc.
- **Product Code:** [GAP](/productcode/GAP.md)
- **Decision Date:** May 19, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5030
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The MANI Needle & Suture Pack (Silk) is intended for use in short term soft tissue approximation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological tissue.

## Device Story

Device consists of stainless steel needle and silk suture; used for short-term soft tissue approximation. Operated by surgeons in clinical settings, including ophthalmic surgery. Provides mechanical means for tissue closure. Benefits patient by facilitating wound approximation. Not for cardiovascular or neurological applications.

## Clinical Evidence

No clinical data; bench testing only.

## Technological Characteristics

Stainless steel needle; silk suture material. Manual surgical instrument. Non-absorbable. Sterile.

## Regulatory Identification

Natural nonabsorbable silk surgical suture is a nonabsorbable, sterile, flexible multifilament thread composed of an organic protein called fibroin. This protein is derived from the domesticated species Bombyx mori (B. mori ) of the family Bombycidae. Natural nonabsorbable silk surgical suture is indicated for use in soft tissue approximation. Natural nonabsorbable silk surgical suture meets the United States Pharmacopeia (U.S.P.) monograph requirements for Nonabsorbable Surgical Suture (class I). Natural nonabsorbable silk surgical suture may be braided or twisted; it may be provided uncoated or coated; and it may be undyed or dyed with an FDA listed color additive.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Alcon Laboratories Needles and Sutures ([K760158](/device/K760158.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

## 510(k) SUMMARY MANI Needle and Suture Pack MANI, Inc.

This 510(k) summary of safety and effectiveness for the MANI Needle and Suture Pack (Silk) is submitted in accordance with the requirements of SMDA and follows Office of Device Evaluation guidance concerning the organization and content of a 510(k) summary.

| Applicant:           | MANI, Inc.                                                                                                                                                                                                                                              |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:             | 743 Nakaakutsu Takanezawa-Machi<br>Tochigi 329-1234 Japan (Headquarters)                                                                                                                                                                                |
| Contact Person:      | David J. Bloch<br>Regulatory Counsel                                                                                                                                                                                                                    |
| Telephone:           | (202) 414-9209 (telephone)<br>(202) 414-9209 (fax)                                                                                                                                                                                                      |
| Preparation Date:    | January 2005                                                                                                                                                                                                                                            |
| Device Trade Name:   | MANI Needle and Suture Pack (Silk)                                                                                                                                                                                                                      |
| Common Name:         | Guide Needle, Surgical; Non-<br>Absorbable Suture, Silk.                                                                                                                                                                                                |
| Classification Name: | Guide Needle, Surgical; (see 21 C.F.R. § 878.4493)<br>Non-Absorbable Suture, Silk<br>(21 C.F.R. § 878.5030)<br>Product Code: GAP                                                                                                                        |
| Predicate Devices:   | Alcon Laboratories Needles and Sutures, 510(k) # 760158.                                                                                                                                                                                                |
| Device Description:  | The MANI Needle & Suture Pack (Silk) consists of a<br>stainless steel needle and silk suture, for use in short term<br>soft tissue approximation, including use in ophthalmic<br>surgery, but not for use in cardiovascular and neurological<br>tissue. |
| Intended Use:        | The MANI Needle & Suture Pack (Silk) is intended for use in<br>short term soft tissue approximation, including use in                                                                                                                                   |

: 上一篇: 上一篇:

K053638

1/2

{1}------------------------------------------------

ophthalmic surgery, but not for use in cardiovascular and neurological tissue.

CONCLUSIONS: Based on the foregoing and other information in this application, MANI, Inc. believes that the MANI Needle & Suture Pack (Silk) is substantially equivalent to its claimed predicates under conditions of intended use.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with outstretched arms, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" arranged in a circular pattern around the figure. The text is in all caps and is written in a sans-serif font. The logo is black and white.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 1 9 2006

MANI, Inc. % ReedSmith Mr. David J. Bloch, Esq. 1301 K. Street, NW Suite 1100 - East Tower Washington, District of Columbia 20005-3373

Re: K053638

Trade/Device Name: MANI Needle & Suture Pack (Silk) Regulation Number: 21 CFR 878.5030 Regulation Name: Natural nonabsorbable silk surgical suture Regulatory Class: II Product Code: GAP Dated: April 24, 2006 Received: April 26, 2006

Dear Mr. Bloch:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. David J. Bloch, Esq.

forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html

Sincerely yours,
Mark N. Milliken

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use

510(k) Number (if known):

Device Name: MANI Needle & Suture Pack (Silk)

Indications For Use:

大きなくている人気をなると、同時のためますが、

The MANI Needle & Suture Pack (Silk) is intended for use in short term soft tissue approximation, including use in ophthalmic surgery, but not for use in cardiovascular and neurological tissue.

the state the the production of the season of the state of the collection of the comments of

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert Lemer no

(Division Sign-Division of General, Restorative, and Neurological Devices

**510(k) Number** K053638

---

**Source:** [https://fda.innolitics.com/device/K053638](https://fda.innolitics.com/device/K053638)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
