K053628 · Lumenis, Inc. · GEX · Apr 7, 2006 · General, Plastic Surgery
Device Facts
Record ID
K053628
Device Name
LIGHTSHEER DUET LASER SYSTEM
Applicant
Lumenis, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Apr 7, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
LightSheer® Pulsed Diode Array Laser Systems are indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer® Pulsed Diode Array Laser Systems are intended for use on all skin types (Fitzpatrick skin types I – VI), including tanned skin. LightSheer® Duet™ Laser Systems with LightSheer® ET™ Laser Handpiece is intended for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. The LightSheer® Duet™ Laser System with LightSheer® ET™ Laser Handpiece is also intended for hair removal, permanent hair reduction, and the treatment of benign pigmented lesions and leg veins. LightSheer® Duet™ Laser Systems with LightSheer® HS™ Laser Handpiece is intended for the treatment of benign vascular and pigmented lesions, hair removal, and permanent hair reduction.
Device Story
Non-invasive aesthetic laser system; delivers pulsed infrared light (790-950 nm). Two handpieces: ET (9x9 mm tip, water-cooled, 10-100 J/cm2) and HS (22x35 mm diode array, vacuum-assisted, 3-5.2 J/cm2). Operated by physicians in clinical settings. Console-based; umbilical cables connect handpieces. Activation requires redundant handpiece enable and trigger buttons. ET handpiece uses active water cooling; HS handpiece uses vacuum to reduce skin heating. Physician controls energy settings via console LCD. Output targets vascular/pigmented lesions and hair follicles; thermal energy induces therapeutic effect. Benefits include permanent hair reduction and lesion treatment.
Clinical Evidence
No clinical data provided; substantial equivalence based on technological characteristics and performance specifications compared to predicate devices.
Technological Characteristics
Pulsed diode array laser; 790-950 nm wavelength (800 nm nominal). ET handpiece: 9x9 mm spot, water-cooled, 5-400 ms pulse duration, 10-100 J/cm2 fluence. HS handpiece: 22x35 mm spot, vacuum-assisted, 10-30 ms pulse duration, 3-5.2 J/cm2 fluence. Console-based with LCD interface. Redundant activation controls. Electrical energy source.
Indications for Use
Indicated for patients of all skin types (Fitzpatrick I-VI), including tanned skin, requiring treatment for vascular lesions (angiomas, hemangiomas, telangiectasia), benign pigmented lesions, pseudofolliculitis barbae, hair removal, and permanent hair reduction.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lumenis, Inc. LightSheer® Pulsed Diode Array Laser System
AesThera Corporation AIP™
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Lumenis. The logo consists of a stylized letter "L" inside of a circle, followed by the word "LUMENIS" in all capital letters. Below the company name is the tagline "Enhancing life. Advancing technology."
APR
K053628
# 510(k) Summary
| Applicant: | Lumenis, Inc |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| Address: | 2400 Condensa Street<br>Santa Clara, California 95051<br>USA |
| Contact Person: | Connie Hoy<br>Global Director of Regulatory Affairs<br>and Quality Assurance |
| Contact Numbers: | (408) 764-3303 Phone<br>(408) 764-3500 Fax |
| Preparation Date: | December 28, 2005 |
| Device Trade Name: | LightSheer® Duet™ Laser System |
| Common Name: | Pulsed Diode Array Laser |
| Classification Name: | Laser surgical instrument for use in General and<br>Plastic Surgery and in Dermatology (see 21 CFR<br>878.4810).<br>Product Code: GEX<br>Panel: 79 |
| Legally-Marketed Predicate Device: | Lumenis, Inc. LightSheer® Pulsed Diode Array<br>Laser System and AesThera Corporation AIP™ |
Intense Pulsed Light System.
Lomenis Inc. 2400 Condensa Street Santa Clara, CA 95051 USA t 408.764.3000 1 800.635.1313 1 408.764.3999 ww.lumenis.com
{1}------------------------------------------------
K053628
### System Description:
The LightSheer Duet Laser System is a non-invasive aesthetic laser. The system delivers pulsed infrared laser light with a wavelength ranging from 790 - 950 nm (800 nm nominal) and has two unique treatment handpieces. One handpiece is the LightSheer® ET™ handpiece, which delivers laser energy through a 9 x 9 mm tip up to 90 J maximum. The settings for this handpiece are selectable pulse duration from 5 – 400 ms, selectable fluence from 10 - 100 J/cm2 and a pulse repetition rate up to 3 Hz maximum. The second handpiece is the LightSheer® HS™ handpiece, which delivers laser energy from a 22 x 35 mm diode array up to 45 J maximum. The settings for this handpiece are pulse duration from 10 - 30 ms, selectable fluence from 3 - 5.2 J/cm2 and multiple pulsing up to 3 pulses.
The complete system consists of a console and two handpieces connected to the system by umbilical cables. In standard use, the handpiece is pressed against the patient's skin and a pulse of light is delivered. To initiate energy output, the system requires redundant activation of the handpiece enable button and the handpiece trigger button while the system is in the Ready mode. The LightSheer ET handpiece tip is watercooled to provide active skin cooling. The LightSheer HS handpiece tip uses vacuum and lower laser energy densities which reduces skin heating. The physician is able to control the settings of laser energy from the LCD display on the main console.
## Intended Use of the Device:
LightSheer® Pulsed Diode Array Laser Systems are indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer® Pulsed Diode Array Laser Systems are intended for use on all skin types (Fitzpatrick skin types I – VI), including tanned skin.
LightSheer® Duet™ Laser Systems with LightSheer® ET™ Laser Handpiece is intended for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. The LightSheer® Duet™ Laser System with LightSheer® ET™ Laser Handpiece is also intended for hair removal, permanent hair reduction, and the treatment of benign pigmented lesions and leg veins.
LightSheer® Duet™ Laser Systems with LightSheer® HS™ Laser Handpiece is intended for the treatment of benign vascular and pigmented lesions, hair removal, and permanent hair reduction.
{2}------------------------------------------------
s 36/i
Image /page/2/Picture/2 description: The image shows the Lumenis logo. The logo consists of a stylized letter "L" inside a circle on the left, followed by the word "LUMENIS" in a sans-serif font. Below the word "LUMENIS" is the tagline "Enhancing Life. Advancing technology."
### Substantial Equivalence:
The specifications and indications for use of the LightSheer Duet Laser System are the same or very similar to those of the claimed pr dicate devices. The LightSheer Duet Laser System has the same or very similar indications for use for which the claimed predicates have been cleared. Based on the icregoing, the LightSheer Duet Laser System is substantially equivalent to the legally-1 1arketed claimed predicate devices for the purposes of this 510(k) submission.
Lumenis Inc. 2400 Condensa Street Santa Clara, CA 95051 DSA t 408.764.3000 1 800.635.131 3 1-408.764.3000 www.lumenis.com
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines representing its body and wings. The eagle is positioned within a circle, and the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is written around the circumference of the circle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
7 2006 APR
Lumenis, Inc. c/o Ms. Connie Hoy Global Director of Regulatory Affairs and Quality Assurance 2400 Condensa Street Santa Clara, California 95051
Re: K053628
Trade/Device Name: LightSheer® Duet™Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: March 1. 2006 Received: March 3, 2006
Dear Ms. Hoy:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 – Ms. Connie Hoy
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely vours.
Sincerely yours,
Mark N. Melkerson
Mark N. Melke Director
Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): _ İC 5 14 20
Device Name: LightSheer® Duet™ Laser System
Indications for Use:
LightSheer® Pulsed Diode Array Laser Systems are indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer® Pulsed Diode Array Laser Systems are intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
LightSheer® Duet™ Laser Systems with LightSheer® ET™ Laser Handpiece is indicated for the treatment of vascular lesions, including angiomas, hemangiomas, telangiectasia and other benign vascular lesions, and the treatment for pseudofolliculitis barbae. The LightSheer® Duet™ Laser System with LightSheer® ET™ Laser Handpiece is also indicated for hair removal, permanent hair reduction, and the treatment of benign pigmented lesions and leg veins.
LightSheer® Duet™ Laser Systems with LightSheer® HS™ Laser Handpiece is indicated for the treatment of benign vascular and pigmented lesions, hair removal, and permanent hair reduction.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use | (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
eamoll
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 1 of 1 -
510(k) Number_ Kos 3628
Lumenis 510(k) Submission: LightSheer Duet Laser System
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.