IMMULITE/IMMULITE 1000 TURBO INTACT PTH

K053533 · Diagnostic Products Corporation · CEW · Feb 3, 2006 · Clinical Chemistry

Device Facts

Record IDK053533
Device NameIMMULITE/IMMULITE 1000 TURBO INTACT PTH
ApplicantDiagnostic Products Corporation
Product CodeCEW · Clinical Chemistry
Decision DateFeb 3, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1545
Device ClassClass 2

Indications for Use

For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers—for the quantitative measurement of intact parathyroid hormone parathyrin, PTH) test system in EDTA plasma or serum. It is intended as an aid in the differential diagnosis of hypercalcemia and hypocalcemia and can be used intraoperatively.

Device Story

Turbo Intact PTH Assay is a quantitative, chemiluminescent, immunometric assay for measuring intact parathyroid hormone (PTH) in EDTA plasma or serum. Used on IMMULITE and IMMULITE 1000 analyzers in clinical laboratory settings. Operates via solid-phase sandwich immunoassay using goat polyclonal anti-PTH (44-84) capture antibody and alkaline phosphatase-conjugated anti-PTH (1-34) detection antibody. Provides rapid results (15-minute turnaround) to assist surgeons intraoperatively in verifying successful resection of hypersecreting parathyroid tissue. Clinicians use the assay to monitor PTH decline (typically >50% decrease from baseline) post-excision, which correlates with successful tumor removal and achievement of postoperative normocalcemia. The device aids in surgical decision-making, potentially reducing the need for frozen section analysis.

Clinical Evidence

Clinical studies conducted at Mayo Clinic (n=47) and Washington University/Barnes Jewish Hospital (n=49 study, n=55 control). Primary endpoint: ≥50% decrease in iPTH from baseline post-resection. Results: 96% of Mayo patients showed >50% decrease; 87% returned to normal iPTH levels. Washington University study showed 90% of study group achieved normocalcemia vs 89% control; study group had statistically significant reduction in frozen section use (p<0.0001).

Technological Characteristics

Solid-phase chemiluminescent immunometric assay. Components: beads coated with goat polyclonal anti-PTH (44-84) antibody; alkaline phosphatase-conjugated goat polyclonal anti-PTH (1-34) antibody. Energy source: IMMULITE/IMMULITE 1000 analyzer platform. Connectivity: Integrated with analyzer software. Sterilization: Not applicable (reagent kit).

Indications for Use

Indicated for quantitative measurement of intact parathyroid hormone (PTH) in EDTA plasma or serum to aid in differential diagnosis of hypercalcemia and hypocalcemia, and for intraoperative monitoring during parathyroidectomy.

Regulatory Classification

Identification

A parathyroid hormone test system is a device intended to measure the levels of parathyroid hormone in serum and plasma. Measurements of parathyroid hormone levels are used in the differential diagnosis of hypercalcemia (abnormally high levels of calcium in the blood) and hypocalcemia (abnormally low levels of calcium in the blood) resulting from disorders of calcium metabolism.

Submission Summary (Full Text)

{1} # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k053533 B. Purpose for Submission: Changes to labeling and modified Indications for Use C. Measurand: Intact parathyroid hormone D. Type of Test: Quantitative, Chemiluminescent, Immunometric assay E. Applicant: Diagnostic Products Corporation F. Proprietary and Established Names: IMMULITE/IMMULITE 1000 Turbo Intact PTH Assay G. Regulatory Information: 1. Regulation section: 21 CFR 862.1545 – Parathyroid hormone test system 2. Classification: Class II 3. Product code: CEW 4. Panel: Chemistry (75) {2} H. Intended Use: 1. Intended use(s): See Indications for Use below 2. Indication(s) for use: For *in vitro* diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers—for the quantitative measurement of intact parathyroid hormone parathyrin, PTH) test system in EDTA plasma or serum. It is intended as an aid in the differential diagnosis of hypercalcemia and hypocalcemia and can be used intraoperatively. 3. Special conditions for use statement(s): This device is for prescription use. 4. Special instrument requirements: IMMULITE and IMMULITE 1000 analyzers I. Device Description: The Turbo Intact PTH Kit is supplied with the following: - 50 Turbo Intact PTH Units. Each unit contains one bead coated with affinity purified goat polyclonal anti-PTH (44-84) antibody - 1 Turbo Intact Reagent Wedge. 7.5 mL alkaline phosphatase (bovine calf intestine) conjugated to affinity purified goat polyclonal anti-PTH (1-34) antibody in buffer with preservative. - 2 vials Intact PTH Adjustors (Low and High) of lyophilized synthetic human Intact PTH in a buffered matrix J. Substantial Equivalence Information: 1. Predicate device name(s): IMMULITE Turbo Intact PTH Assay 2. Predicate 510(k) number(s): k992105 {3} # 3. Comparison with predicate: | | Similarities | | | --- | --- | --- | | Item | Predicate | Device | | Assay Principle | Solid phase, chemiluminescent immunometric assay | Same | | Antibodies | Goat polyclonal anti-PTH (44-84) as the capture antibody and goat polyclonal anti-PTH (1-34) conjugated to alkaline phosphatase as the detection antibody | Same | | Analytical Sensitivity | 4.0 pg/mL | Same | | Reference: Range | 8-74 pg/mL | Same | | | Differences | | | --- | --- | --- | | Item | Predicate | Device | | Indications for Use | For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers—for the quantitative measurement of intact parathyroid hormone (parathyrin, PTH) test system in EDTA plasma or serum. It is intended as an aid in the differential diagnosis of hypercalcemia and hypocalcemia. | For in vitro diagnostic use with the IMMULITE and IMMULITE 1000 Analyzers—for the quantitative measurement of intact parathyroid hormone (parathyrin, PTH) test system in EDTA plasma or serum. It is intended as an aid in the differential diagnosis of hypercalcemia and hypocalcemia and can be used intraoperatively. | | Limitations | It is not intended for the diagnosis of hypercalcemia and hypocalcemia. | It is not intended for the diagnosis or monitoring of hypercalcemia and hypocalcemia. EDTA sample is the preferred sample type for rapid turnaround time particularly in the intraoperative setting | {4} | Instructions | PTH assays are used intraoperatively to verify a decline in PTH production after resection. When used intraoperatively, it has been recommended that two or more samples be collected at least 5-10 minutes apart. | PTH assays are used intraoperatively to verify a decline in PTH production after resection. When used intraoperatively, it has been recommended that intact PTH concentrations be measured at baseline before exploration and then at 5-10 minutes post tumor excision with a 50% decrease in values observed if all hypersecreting tissue has been removed. | | --- | --- | --- | # K. Standard/Guidance Document Referenced (if applicable): None referenced # L. Test Principle: IMMULITE/IMMULITE 1000 Turbo Intact PTH is a solid phase chemiluminescent immunometric assay utilizing a goat polyclonal anti-PTH (44-84) antibody as the capture antibody and a goat polyclonal anti-PTH (1-34) antibody conjugated to alkaline phosphatase as the detection antibody. IMMULITE/IMMULITE 1000 Turbo mode processes samples at an accelerated rate, with time to first result being 15 minutes and subsequent samples are generated every 45 seconds. # M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: Performance Characteristics were established in k992105 b. Linearity/assay reportable range: Performance Characteristics were established in k992105. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Performance Characteristics were established in k99105. {5} d. Detection limit: Performance Characteristics were established in k992105. e. Analytical specificity: Performance Characteristics were established in k992105. f. Assay cut-off: Performance Characteristics were established in k992105. 2. Comparison studies: a. Method comparison with predicate device: Performance Characteristics were established in k992105. b. Matrix comparison: Performance Characteristics were established in k992105. 3. Clinical studies: a. Clinical Sensitivity: Not applicable b. Clinical specificity: Not applicable c. Other clinical supportive data (when a. and b. are not applicable): Clinical studies were performed at the Mayo Clinic¹, Rochester, MN (Kao, PC et al.) and Washington University School of Medicine/Barnes Jewish Hospital², St. Louis, MO (Johnson, LR, et. al) to evaluate the Immulite/Immulite 1000 Turbo Intact PTH assay results on patients undergoing parathyroidectomy. These assay results were compared to clinical data and reported in the following peer reviewed journals: 1. Kao PC, van Heerden JA, Farley DR, et al. Intraoperative monitoring of parathyroid hormone with a rapid automated assay that is commercially available. Annal of Clin &amp; Lab Science 2002; 32(3):244-251. {6} 2. Johnson LR, Doherty G, Lairmore T, et al. Evaluation of the performance and clinical impact of a rapid intraoperative parathyroid hormone assay in conjunction with preoperative imaging and concise parathyroidectomy. Clin Chem 2001; 47(5):919-925. **Study #1:** The Mayo Clinic Study was conducted between July 1999 and October 2001 using the currently marketed Immulite/Immulite 1000 Turbo Intact PTH assay. EDTA blood specimens were collected from forty seven patients undergoing parathyroidectomy at the following time intervals: baseline (prior to removing the parathyroid gland) and 5, 10, and 20 minutes after parathyroid resection. Turbo Intact PTH results on ninety-four specimens were immediately analyzed. These results were compared to the in-house rapid assay and generated the following correlation: $y = 1.047x + 0.186$ ; $r = 0.968$ . 45 of 47 patients $(96\%)$ had their plasma PTH levels decrease to $&lt; 25\%$ of the baseline levels. In 41 of 47 patients $(87\%)$ , the PTH value decreased to $&lt; 5\mathrm{pmol} / \mathrm{L}$ (provisional reference range) within 20 minutes after tumor excision. The assay met the generally accepted criterion of $a &gt; 50\%$ decrease from baseline (before incision) with complete tumor removal. **Study #2:** The Washington University Barnes Jewish Hospital Study evaluated the IMMULITE and IMMULITE 1000 Turbo iPTH assay intraoperatively in 49 patients and compared clinical outcomes to a "control" group of 55 patients that underwent parathyroidectomies without intraoperative PTH determinations. This study also employed the generally accepted guideline that $\geq 50\%$ decrease in the iPTH value from baseline suggests complete tumor removal. Samples were drawn in the operating room before incision and during the period 10-12 minutes after excision of suspected diseased parathyroid gland tissue. Of the 49 patients in the study group, 46 had $\geq 50\%$ decrease in iPTH values within the first three post-resection samples. When the control and study groups with similar sex, age, and diagnoses were compared, 44/49 (90%) of the patients in the study group and 49/55 (89%) of the control patients achieved normocalcemia postoperatively. The surgery guided by intraoperative IMMULITE/IMMULITE Turbo Intact PTH measurements had identical physiologic outcomes. Frozen section use in the experimental group was statistically significantly less $(p &lt; 0.0001)$ than in the control group. The assay met the generally accepted criterion of $a &gt; 50\%$ decrease from baseline (before incision) with complete tumor removal. 4. Clinical cut-off: {7} Not applicable 5. Expected values/Reference range: Performance Characteristics were established in k992105. **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...