K053485 · Mds Nordion, Inc. · IWB · Mar 9, 2006 · Radiology
Device Facts
Record ID
K053485
Device Name
THERATRON EQUINOX, MODELS 80 CM AND 100 CM
Applicant
Mds Nordion, Inc.
Product Code
IWB · Radiology
Decision Date
Mar 9, 2006
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
Device Story
Cobalt-60 teletherapy system for cancer treatment; delivers gamma radiation via source head, collimator, and gantry. System includes asymmetric jaw collimator, motorized wedge, and collision detection. Operated by healthcare professionals in radiation therapy clinics. Provides fixed or moving beam therapy. New control system, graphic user interface, and parameter display facilitate treatment setup and execution. Automated diagnostics and in-room monitors assist operator. Output is controlled radiation dose; intended to minimize risk to patient and operator. Mechanical structure and radiological shielding remain consistent with predicate; design updates focus on ergonomics, modern appearance, and improved control functionality.
Clinical Evidence
Bench testing only. Performance validated against functional specifications in a simulated operating environment. Compliance with safety standards EN 60601-1, EN 60601-1-2, EN 60601-1-4, EN 60601-2-11, and EN 61217 demonstrated.
Technological Characteristics
Cobalt-60 radionuclide radiation therapy system. Features: asymmetric jaw collimator (independent X/Y movement), motorized wedge, collision detection, and in-room display monitors. Control system includes GUI, hardware-linked control panel, and automated diagnostic capabilities. Complies with EN 60601-1 (general safety), EN 60601-1-2 (EMC), EN 60601-1-4 (programmable electrical medical systems), EN 60601-2-11 (gamma beam therapy), and EN 61217 (coordinates/movements).
Indications for Use
Indicated for the treatment of cancer via gamma radiation delivery in a clinical setting under the direction of healthcare professionals.
Regulatory Classification
Identification
A radionuclide radiation therapy system is a device intended to permit an operator to administer gamma radiation therapy, with the radiation source located at a distance from the patient's body. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts (including beam-limiting devices), and accessories.
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for MDS Nordion. The logo consists of the text "MDS Nordion" in a bold, sans-serif font. Below the text is the tagline "Science Advancing Health" in a smaller, italicized font. There is a curved line above the text.
MAR 9 2006
K053485
# 510(k) SUMMARY
| Date Summary Prepared | December 7, 2005 |
|-----------------------------------|------------------------------------------------------------------------------------------------------------------|
| Submitted by | MDS Nordion<br>447 March Road<br>Ottawa, Ontario K2K 1X8<br>Canada<br>Tel. (613) 592-3400<br>Fax. (613) 592-2006 |
| Contact Person | Mr. Ross Kachaniwsky<br>Director, Quality & Regulatory Affairs |
| Device Name | Theratron Equinox 80<br>Theratron Equinox 100 |
| Common Name | Cobalt Teletherapy Device |
| Classification Name | Radionuclide Radiation Therapy System |
| Legally Marketed Predicate Device | Theratron Elite 80 and 100 |
### Description of Device
The Theratron Equinox consists of a source head, collimator, gantry, main frame, base, controls and a pendulum or beam-stopper style counterweight. The design of this device is similar to predicate device Theratron Elite 80 and 100. For a more detailed description, please see Section 4.
### Intended Use of Device
The unit is intended to be used in:
- Delivering the intended dose at a specified position; .
- Delivering the radiation in accordance with the selected relationship of the radiation to the patient . (fixed or moving beam therapy, beam modifying device, etc.);
- . Delivering the radiation without causing unnecessary risk to the patient, the operator, other persons, or the immediate environment.
#### Summary of Technological Characteristics
The Theratron Equinox 80 and 100 is substantially equivalent to the predicate device (k983917).
The design change is to provide a new 'modern' appearance to the previous design using existing major components of the predicate device, and adds some new functionality that allows customers to treat patients with modern radiotherapy techniques while not changing the intended use of the device.
Changes include a new:
- · Control System: the system is easy to maintain and diagnosis of problems is automated as much as possible
- · Parameter Display: for displaying all gantry and collimator parameters relevant to treatment setup
- · Control Panel: providing all hardware links to the unit that cannot be software controlled by the User Interface e.g. emergency stop
- . Graphic User Interface: providing for efficient treatment setup and execution
- · Hand Control modified for modern appearance and ergonomic improvement over previous design
{2}------------------------------------------------
- · Covers: The unit has a more modern appearance
- · Collision Detection Device -- stops motion of the gantry when a collision is detected
- · Asymmetric jaws collimator X and Y jaws can move independently
- · Motorized wedge -- a wedge affixed to the inside of the collimator can be moved to the in or out position
- · In-room display monitors
The major components of the Theratron Elite, including the head, gantry, main frame, base, and pendulum/beam stopper counterweight have had minor modifications to accommodate the above changes.
There are no changes to the mechanical structure or radiological shielding of the head.
The irradiation source and radioactivity of the cobalt-60 source remains unchanged as does the source drawer mechanism.
The control system has been designed to meet the same intended use as the current model.
### Safety & Effectiveness
The safety of the Theratron Equinox 80 and 100 is equivalent or better than the predicate device.
In terms of safety, the Theratron Equinox 80 and 100 are designed to comply with
- EN 60601-1 (1995), Medical Electrical Equipment. Part 1; General requirements for safety, and .
- EN 60601-1-2 (2004), Medical Electrical Equipment, Part 1;General requirements for safety; . Electromagnetic Compatibility - Requirements for Tests
- EN 60601-1-4 (2001) Medical electrical equipment Part 1-4: General requirements for safety -. Collateral Standard: Programmable electrical medical systems
- EN 60601-2-11 (2004) Medical electrical equipment Part 2: Particular requirements for the safety of . qamma beam therapy equipment
- EN 61217 (2000) Radiotherapy equipment -- Coordinates, movements and scales .
The performance of the device was tested against a set of functional specifications in an environment that simulated, as much as possible, the actual operating environment. Validation testing demonstrated that the device is as safe and effective as the predicate device.
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
#### Public Heatth Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
2006 MAR 9
2000
Mr. Ross Kachaniwsky Director, Quality Assurance MDS Nordion 447 March Road Ottawa, ON K2K 1X8 CANADA
Re: K053485 Trade/Device Name: Theratron Equinox 80 and Theratron Equinox 100 Regulation Number: 21 CFR 892.5750 Regulation Name: Radionuclide radiation therapy system Regulatory Class: II Product Code: IWB Dated: February 3, 2006 Received: February 7, 2006
Dear Mr. Kachaniwsky:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act 's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Nancy C. Brigeton
Naney C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for MDS Nordion. The logo consists of the text "MDS Nordion" in a bold, sans-serif font. Below the text is the tagline "Science Advancing Health" in a smaller, italicized font. The logo is simple and professional, and it conveys the company's focus on science and health.
# INDICATIONS FOR USE
**510(k) Number:** K053485
Device Name:
Theratron Equinox 80 Theratron Equinox 100
## Indications For Use:
A Cobalt Teletherapy unit is a device by which gamma radiation is delivered for the treatment of cancer under the direction of health care professionals in a radiation therapy clinic.
Prescription Use
(Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
Nancy C. Broxton
Division Sign-Off
Jivision Sign-Off) / Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.