The Masimo SET Rad-57 cm and m Pulse CO-Oximeters and accessories are indivated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO), pulse rate (measured by an SpOn sensor), carboxyhemoglobin, and/or methemoglobin saturation (measured by an SpCO/SpMet sensor) for adult, pediative, and neonaal patients in hospitals, hospital-type facilities, mobile, and home environments.
Device Story
Handheld pulse CO-oximeter; uses multi-wavelength LED sensors (400-1000nm) to measure light absorption in capillary beds (fingertip, hand, foot). Masimo SET signal extraction technology processes photoplethysmography signals to calculate SpO2, pulse rate, SpCO, and SpMet. Device displays numeric values, perfusion index, and signal quality (SIQ) on LED screen. Used by clinicians or patients in hospitals, mobile, or home environments. Provides continuous monitoring to assist in clinical assessment and intervention decisions. Benefits include noninvasive, real-time detection of hemoglobin species and pulse rate, even during motion or low perfusion.
Clinical Evidence
Clinical studies on healthy adult volunteers under induced hypoxia (70-100% SpO2) compared device readings against laboratory CO-oximeter blood samples. Validated for SpO2 accuracy during motion/no-motion and low perfusion. SpCO (1-40%) and SpMet (1-15%) accuracy validated against CO-oximeter. Results show performance within specified accuracy ranges (±2-3 digits for SpO2, ±3 digits for SpCO, ±1 digit for SpMet).
Technological Characteristics
Handheld device; 4 AA battery powered. Multi-wavelength LED sensor (400-1000nm). Microprocessor-controlled. LED 7-segment display. Biocompatible materials per ISO-10993-1. Connectivity via patient cables. No external power/ground. Averaging modes 2-16 seconds. Sensitivity modes: Max, Normal, APOD.
Indications for Use
Indicated for continuous noninvasive monitoring of functional arterial oxygen saturation (SpO2), pulse rate, carboxyhemoglobin (SpCO), and methemoglobin (SpMet) in adult, pediatric, and neonatal patients. Suitable for use during motion and no-motion conditions, and for patients with good or poor perfusion in hospital, mobile, and home settings.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
# 1053477
#### 17 11 1 6
011
| Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>(714) 297-7000<br>FAX (714) 297-7001 |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Contact: | James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance |
| Date Summary Prepared: | December 13, 2005 |
| Trade Name | Masimo Rainbow SET* Rad-57 cm and m Pulse CO-Oximeters and accessories |
| Common Name | Pulse Oximeter and Sensor |
| Classification Name | Oximeter (74DQA)<br>Transducer and Electrode Cable (including connector) (74DSA)<br>Carbon monoxide test system (JKS)(862.3220) |
| Substantially Equivalent Devices | Masimo SET Rad-57 Pulse CO-Oximeter and accessories<br>510(k) Number K042536<br>Radiometer America, Inc OSM3 Hemoximeter<br>510(k) Number – K853990 |
## Description of Masimo Rainbow SET® Rad-57 cm and m Pulse CO-Oximeters
The Rad-57 cm and m Handheld Pulse CO-Oximetesr with Rainbow technology are noninvasive, arterial oxygen saturation and pulse rate monitor that also measure Carboxyhemoglobin. The Rad-ST featurena multicologies and Indicelosmallion display that continuously displays numeric values for SpO2 and pulse rate, a Low Signal (Q Indicator (Low SIQ) included LED) indicator bars for Perfusion Index (PI) and Carboxyhemoglobin saturation (%SpCO) and/or Methemoglobin saturation(%21). alarm status, alarm silence, battery life and SpCO/SpMet sensor connected. The Masino SET Rad-57 cm and and Photo (10-Oximeters are intended to be used with Masimo's LNOP and LNCS series of oximelery sensors and patient ables and Masimo is Rainbow SpCO/SpMet sensors and Rainbow cables.
Features and Benefits Clinically proven Masimo SETTN technology performance Applicable for use on neonate, infant, pediatric and adult patients Proven for accurate monitoring in motion and low perfusion environments Sp()2, pulse rate, alarm, Perfusion Index, % SpCO and/or SpMet displays Low Signal IQ (SIQ) indicator Lightweight, convenient handheld design Up to 8 hours continuous use on 4 "AA" alkaline batteries Visual battery life indicator
Masimo Corporation -
{1}------------------------------------------------
Audible Alarm for sensor-off and low battery Alarms for Hi/Low saturation and Hi/Low pulse rate, SpCO, SpMet FastSat™ (for SpO2 measurement) Three sensitivity levels - Max, Normal and APOD™ (for SpO2 measurement) 72 hours of trending memory Adjustable alarm volume Adjustable averaging 2 to 16 seconds
#### Intended use
The Masimo SET" Rad-57 cm and m Pulse CO-Oximeters and accessories are indivated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO), pulse rate (measured by an SpOn sensor), carboxyhemoglobin, and/or methemoglobin saturation (measured by an SpCO/SpMet sensor) for adult, pediative, and neonaal patients in hospitals, hospital-type facilities, mobile, and home environments.
#### Indications for use
The Masimo SET® Rad-57 cm and m Pulse CO-Oximeters and accessories are indicated for the continuous noninvasive montoring of functional oxygen saturation of arterial hemoglobin (SpO-) and pulse rate (measured by an SpO-sensor) and carboxyhenoglobin and methemoglobin saturation or methemoglobin (measured by an SpCO/SoMet sepsor). The Masimo SFFT Rad-57 cm and m Pulse CO-Oximeters and accessories are indicated for use with adult, pediatric, and neouatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
#### Principles of Operation
#### SpO2 General Description
Palse oxinctry is a continuous and non-invasive method of measuring the level of arterial oxygen saturation in blood. The measurement is taken by placing a sensor on a patient, usually on the fingertip for adults, and the hand or foot for neovates. The sensor connects to the pulse oximent with a patient cable. The sensor collects signal data from the patient and sculs it to the instrument. The instrument displays the calculated data in two ways: 1) as a percent value for arterial oxygen saturation (Sp()2). and 2) as a pulse rate (PR).
### SpCO And SpMet General Description
Pulse CO-oximetry is a continuous and non invasive method of measuring the levels of carbon monoxide concentration (SpCO) and oxidized hemoglobin concentration (SpMet) in arterial blood. It velies on the same principles of pulse oximetry to make its SpCO/SpMet measurements. The measurements are taken by placing a sensor on a patient, usually on the fingertip for adults. The sensor connects directly to the pulse CO-oximetry instrument or with a patient cable. The sensor collects signal data from the patient and sends it to the instrument. The instrument displays the calculated data as percentage value for the SpCO and the SpMet. The Rad 57 cm is a combined SpO2, SpCO, and SpMet monitor with the same setup as that of a pulse oximeter and can display percentage values for SpCO and SpMet as well as SpO2 and pulse rate.
Pulse oximetry is governed by the following principles
- 1. Oxyhemoglobin (oxygenated blood), deoxyhemoglobin (non-oxygenated blood), carboxyhemoglobin (blood with carbon monoxide content), and methemoglobin (blood with oxidized hemoglobin content) species differ in their absorption of visible and infrared light.
- The amount of arterial blood in tissue changes with your pulse (photoplethysography). Therefore, is the amount of light absorbed by the varying quantities of arterial blood changes as well.
The Rad-57 cm handheld Pulse CO-Oximeter uses a multi-wavelength sensor to distinguish between oxygenated blood, deoxygenated blood, blood with carbon monoxide, and blood with oxidized hemoglobin content. Signal data is obtained by passing various visible and infrared lights (LED's, 400 to 1000nm) through a capillary bed (for example, a fingertip, a hand, a foot) and measuring changes in light absorption during the blood pulsatile cycle.
{2}------------------------------------------------
The photodetector receives the light, converts it into an electronic signal and sends it to the Rad-57 cm for calculation.
Once the Rad-57 cm receives the signal from the sensor, it utilizes Masimo SET signal extraction technology to calculate the patient's functional oxygen saturation, fractional concentrations of carboxyhemoglobin and methemoglobin, and pulse rate. The SpCO and the SpMet measurements rely on multiwavelength calibration equations to estimate the percentages of carboxyhemoglobin and methemoglobin in arterial blood.
#### Method of Operation
The Masimo Rainbow SET% Rad-57 cm or m Pulse CO-Oximeter is turned on. An oximetery sensor is attached to a patient's finger and one end of a patient cable is connected to the sensor and the other end connected to the Rad-57 cm or m Pulse CO-Oximeter.
The monitor will begin continuously displaying the patient's pulse rate, and SpO2 value. The practitioner can then use the information that is continuously displayed on the montor to help assess the condition of the patient and as an aide in decermining if any intervention is required by the practitioner.
Once the practitioner determines the patient no longer requires monitoring, the cable is disconnected from the oximetry sensor is removed (and disposed of if it is a single use device), and the power to the monitor is turned off.
#### Power Source
The Masimo Rainbow SET® Rad-57 cm and m Pulse CO-Oxincters are powcred by 4 AA batteries with an operating time of 8 hours".
#### Specifications and Operating Ranges
| Range | | |
|----------|-----------------------------------------------------------|-----------------------|
| | Oxygen Saturation (% SpO2) | 1% - 100% |
| | Carboxyhemoglobin Saturation (% SpCO) | 1-99% |
| | Methemoglobin Saturation (% SpMet) | 1-99.9% |
| | Pulse Rate (bpm) | 25 - 240 |
| | Perfusion | 0.02% - 20% |
| Accuracy | | |
| | Oxygen Saturation (% SpO2) - During No Motion Conditions1 | |
| | Adults, Pediatrics | 70% - 100% ± 2 digits |
| | | 0% - 69% unspecified |
| | Neonates | 70% - 100% ± 3 digits |
| | | 0% - 69% unspecified |
| | Oxygen Saturation (% SpO2) - During Motion Conditions2,3 | |
| | Adults, Pediatrics | 70% - 100% ± 3 digits |
| | | 0% - 69% unspecified |
| | Neonates3 | 70% - 100% ± 3 digits |
| | | 0% - 69% unspecified |
| | Carboxyhemoglobin Saturation (% SpCO)4 | 0% - 40% ± 3 digits |
| | Methemoglobin Saturation (% SpMet)4 | 0% - 15% ± 1 digits |
| | Pulse Rate (bpm) - During No Motion Conditions1 | |
| | Adults, Pediatric, Neonates | 25 to 240 ± 3 digits |
{3}------------------------------------------------
| Pulse Rate (bpm) - During Motion Conditions2,3<br>Adults, Pediatric, Neonates | 25 to 240 ± 5 digits |
|-------------------------------------------------------------------------------|----------------------------------------------------------------|
| Low Perfusion Performance4 | |
| > 0.02% Pulse Amplitude<br>and % Transmission > 5% | Oxygen Saturation (% SpO2) ± 2 digits<br>Pulse Rate ± 3 digits |
| Resolution | |
| Oxygen Saturation (% SpO2) | 1% |
| Carboxyhemoglobin Saturation (% SpCO), digital display | 1% |
| Methemoglobin Saturation (% SpMet), digital display | 0.1% |
| Carboxyhemoglobin Saturation (% SpCO), continuous bar display | 5% |
| Carboxyhemoglobin Saturation (% SpMet), continuous bar display | .5, 1-5, 7.5, 10, 15, >20% |
| Pulse Rate (bpm) | 1 |
#### Interfering Substances
Carboxyhemoglobin may erroneously increase oxygen saturation readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or any substance containing dyes, that change usual arterial pigmentation may cause erroncous readings.
#### Power
Internally powered by 4 "AA" Alkaline batteries
#### Isolation
No external power or ground connection, internally powered only
#### Environmental
| Operating Temperature | 41°F to 104°F (5°C to +40°C) |
|-----------------------|------------------------------------------------------------------------------|
| Storage Temperature | -40°F to + 158°F (-40°C to -70°C) |
| Relative Humidity | 5% to 95% noncondensing |
| Operating Altitude | 500 mbar to 1060 mbar pressure<br>-1.000 ft to 18.000 ft (-304 m to 5.486 m) |
#### Circuitry
Microprocessor controlled Automatic self-test of oximeter when powered on Automatic setting of parameters Automatic alarm messages
#### Display
Type Data Displayed
## LED, 7-segment
Pulse Rate, SpO2 %, % SpCO, % SpCO bar, %SpMet, %SpMet bar, alarm status, alarm silenced status, low Signal IQ, battery statusand SpCO/SpMet connected
]
#### Audio indicators
Adjustable volume audible pulse: OFF and 33% to 100% in 3 steps Adjustable volume audible alarm tone: levels and 33% to 100% in 3 steps Alarm silence (120 seconds); all mute (continuous silence) Sensor condition alarms System failure and battery low alarms
#### Physical characteristics
Dimensions: 6.2" x 3.0" x 1.4" (15.8 cm x 7.6cm x 3.6 cm) Weight: 13oz. (0.32 kg)
{4}------------------------------------------------
Modes
Averaging mode: Sensitivity
2, 4, 8, 10, 12, and 16 seconds Normal, APOD, and MAX
- The Masimo SET Technology with LNOP Adt sensors has been validated for no motion accuracy in human blood l studies on healthy adult volunteers in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-oximeter and ECG montor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- The Masimo SET Technology with LNOP Adt sensors has been validated for motion accuracy in human blood 2 studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an anylitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-oximeter and ECG monitor. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- r ; The Masimo SET Technology with LNOP Neo sensors has been validated for neonatal motion accuracy in human blood studies on healthy adult volunteers in induced hypoxia studies while performing rubbing and tapping motions, at 2 to 4 Hz at an anplitude of 1 to 2 cm and a non-repetitive motion between 1 to 5 Hz at an amplitude of 2 to 3 cm in induced hypoxia studies in the range of 70-100% SpO2 against a laboratory CO-oximeter and ECG monitor. 1% has been added to the saturation accuracy to account for the effects of fetal hemoglobin. This variation equals plus or minus one standard deviation which encompasses 68% of the population ..
- 4 The Masino SET Rainbow Technology with Rainbow DC-dc sensors have been validated in human blood on healthy adult volunteers against a laboratory CO-oximeter from 1-40% for carboxyhemoglobin and 1-15% for methemoglobin. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- 5 The Masimo SET Technology has been validated for low perfusion accuracy in bench top testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70 to 100%. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
- This represents approximate run time at lowest indicator brightness, using a new, fully charged battery. ు.
### Environmental Testing
Applicable environmental testing per the Reviewers Guidance for Premarket Submissions - November 1993, i.e. electrical. mechanical and environmental were performed and all tests passed.
#### Biocompatibility Testing
All patient contact materials were tested as Surface Devices with skin contact duration (>24 hr to 30 days) as defined ISO-10993-1: 1992 Biological Evaluation of Medical Devices - Part 1: Guidance on Selection of Tests. All patient contacting material passed.
#### Nonclinical tests performed that support a determination of substantial equivalence.
The Masimo Rainbow SET% Rad-57 cm and in Pulse CO-Oximeter and accessories was subjected to bench testing using a simulator that determined the performance accuracy of the instruments against the range of saturation and pulse rates that both devices specify.
The results of the bench testing showed that the Masimo Rainbow SET® Rad-57 cm and m Pulse CO-Oximeter and accessories returned the same saturation accuracy values within ± 2 digits and pulse within ± 3 digits when compared to the simulators used.
### Clinical tests performed that support a determination of substantial equivalence.
{5}------------------------------------------------
Clinical studies were performed using the Masimo Rainbow SET® Rad-57 cm and m Pulse CO-Oximeter on healthy adult volunteer subjects during no motion conditions who were subjected to a progressive induced hypoxia and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter.
The Masino SET Rainbow Technology with Rainbow DC-dc sensors have been validated in human blood on healthy adult volunteers against a laboratory CO-oximeter from 1-40% for carboxyhemoglobin and 1-15% for methenoglobin. This variation equals plus or minus one standard deviation which encompasses 68% of the population.
#### Conclusions
The results of the environmental testing demonstrated that the Masimo Rainbow SET® Rad-57 cm and m Pulse CO-Oximeters and accessories met the requirements of Reviewers Guidance for Premarket Submissions - November 1993.
The results of the biocompatibility testing demonstrates the all patient contacting material met the requirements of ISO-1093-1. 1992 Biological Evaluation of Medical Devices - Part 1: Guidance on Selection of Tests for Surface Devices with skin contact for prolonged contact duration (>24 hr to 30 days).
The results of the bench testing demonstrates that the Masimo Rainbow SET® Rad-57 cm and ni Pulse CO-Oximeters meet them performance requirements.
The results of the clinical testing demonstrates that the Masimo Rainbow SET® Rad-57 cm and m Pulse CO-Oximeters and accessories meet their performance requirements during no motion conditions and low perfusion conditions.
The non-clinical and clinical testing performed demonstrates that the Masimo Rainbow SET & Rad-57 cm and m Pulse CO-Oximeters and accessories are safe, effective, and performs as well as the predicate device, the Masino SET® Rad-57 Pulse CO-Oximeter, and therefore, it is substantially equivalent to the Masimo SET® Rad-57 Pulse CO-Oximeter.
{6}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
MAR 16 2008
Mr. James J. Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 40 Parker Irvine, California 92618
Re: K053477
> Trade/Device Name: Masimo SET Rad 57 cm and m Pulse CO-Oximeters Regulation Number: 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: December 13, 2005 Received: December 14, 2005
Dear Mr. Cronin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 - Mr. James J. Cronin
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set Orth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 594-4618. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html
Sincerely yours,
Oare
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
## Section 4 - Indications for Use
#### 510(k) Number (if known):
Masimo SET Rad 57 cm and m Pulse CO-Oximeters Device Name:
#### Indications For Use:
The Masimo SET® Rad-57 cm Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) momently of famoronal on Jyon and methemoglobin (measured by an SpCO/SpMet sensor). The Masimo SET® Rad-57 cm Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.
The Masimo SET® Rad-57 m Pulse CO-Oximeter and accessories are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2-sensor) momering of antinental ation or methemoglobin (measured by an SpCO/SpMet sensor). The Masimo SET® Rad-57 m Pulse CO-Oximeter and accessories are indicated for use with adult, pediatric, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospital-type facilities, mobile, and home environments.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use __ (Per 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
thl Mhal
General Hosp
ental Devices
1053477
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.