SURGIFROST 6CM, 7CM AND 10 CM CRYOSURGICAL DEVICES PLUS FROSTBYTE CLAMP AND CRYOSURGICAL CONSOLE

K053436 · Cryocath Technologies, Inc. · OCL · Jan 13, 2006 · General, Plastic Surgery

Device Facts

Record IDK053436
Device NameSURGIFROST 6CM, 7CM AND 10 CM CRYOSURGICAL DEVICES PLUS FROSTBYTE CLAMP AND CRYOSURGICAL CONSOLE
ApplicantCryocath Technologies, Inc.
Product CodeOCL · General, Plastic Surgery
Decision DateJan 13, 2006
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The SurgiFrost® 7cm and 10cm Cryosurgical Devices plus FrostByte" Clamp and Cryosurgical Console is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The SurgiFrost® 6cm, 7cm and 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console freezes the target tissue and blocks the electrical conduction pathways by creating an inflammatory response and cryonecrosis.

Device Story

System comprises cryosurgical console, gas tank regulator, and single-use disposable probes (SurgiFrost 6cm, 7cm, 10cm) and FrostByte clamp. Operates via Joule-Thomson effect using inert argon gas to deliver extreme cold to target cardiac tissue. Used in surgical settings by physicians to create cryonecrosis and inflammatory response, effectively blocking aberrant electrical conduction pathways. Console powered by 120/230 VAC. System provides controlled freezing to facilitate cardiac ablation, aiding in arrhythmia management.

Clinical Evidence

Bench testing only. No clinical data provided.

Technological Characteristics

Cryosurgical unit utilizing inert argon gas for cooling. System includes console, gas regulator, single-use disposable probes, and clamp. Operates on 120/230 VAC. Mechanical cooling via gas expansion. No software-based algorithms or electronic signal processing described.

Indications for Use

Indicated for minimally invasive cardiac surgical procedures, including treatment of cardiac arrhythmias, in patients requiring tissue ablation to block electrical conduction pathways.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JAN 1 3 2006 ## 510(k) Summary ## CryoCath Technologies Inc. SurgiFrost® 7cm, 10 cm Cryosurgical Devices plus FrostByte" Clamp and Cryosurgical Console #### 1. SPONSOR CryoCath Technologies Inc. 16771 Chemin Ste-Marie Kirkland, Quebec H9H 5H3, CANADA Contact Person: Flor del Pilar Arana, Director Regulatory Affairs and Official Correspondent Telephone: 514-694-2380 ext 358 Date Prepared: December 7th, 2005 #### 2. Device Name | Device Trade Name: | SurgiFrost® 6cm, 7cm and 10cm Surgical<br>CryoAblation System plus FrostByte™ Clamp | |------------------------|-------------------------------------------------------------------------------------| | Common/Usual Name: | Cryosurgical system | | Classification Name: | Cryosurgical Unit and Accessories | | Device Classification: | Class II | #### 3. PREDICATE DEVICES CryoCath Technologies - SurgiFrost® Cryosurgical System (K021010 & K040690) Endocare Inc. - CryoCare Cardiac Surgical System (K011040) CryoMedical Sciences - AccuProbe 450, Accuprobe 550/530, Accuprobe 600 series (K973190) Boston Scientific - Cobra Bipolar System (K023288) Atricure Inc. - Atricure Bipolar Coagulation System (K020919) Medtronic Inc. - Cardioblate™ Radiofrequency Ablation System (K013392) {1}------------------------------------------------ ### 4. DEVICE DESCRIPTION The SurgiFrost® 7cm and 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console are intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The SurgiFrost® 7cm and 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console freezes the target tissue and blocks the electrical conduction pathways by creating an inflammatory response or cryonecrosis. # The Intended Use does not change as a result of this special 510(k) submission. The SurgiFrost® 7cm and 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console and accessories act together as a System. The system is composed of the following components: - Cryosurgical Control Panel and Power Cord - Cryosurgical Console - Tank Carrier - Gas Tank Regulator and Wrench - Gas Hose - SurgiFrost® Cryosurgical Device (Single Use) - FrostByte™ Clamp (Single Use) - Footswitch (optional) The SurgiFrost® Cryosurgical System consists of a compact, easy-to-operate Cryosurgical Console and associated accessories that include SurgiFrost® 7cm and 10cm Cryosurgical Probes to deliver cold temperatures to the targeted tissue. The control console operates off standard 120/230 VAC (60/50 Hz) wall power and utilizes inert argon gas. The Cryosurgical Console controls a single-use, disposable SurgiFrost® Cryosurgical Probe. ### 5. INTENDED USE The SurgiFrost® 7cm, 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console are intended for minimally invasive treatment of cardiac arrhythmias. The CryoCath SurgiFrost® System freezes the target tissue and blocks the electrical conduction pathways by creating an inflammatory response and cryonecrosis. {2}------------------------------------------------ ### Technological Characteristics and Substantial Equivalence ୍. The SurgiFrost® 7cm, 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console is substantially equivalent to the predicate devices based on intended use, design, materials, technological, and operational characteristics. The differences between the CryoCath devices are minor and raise no new issues of safety or effectiveness. ### 7. Performance Testing Information submitted in this premarket notification for the SurgiFrost® 7cm, 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console is substantially equivalent to the predicate devices listed. {3}------------------------------------------------ Image /page/3/Picture/12 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 FEB 2 1 2008 CryoCath Technologies. Inc. c/o Ms. Flor del Pilar Arana, MBA Director, Regulatory Affairs 16771 Chemin Ste-Marie Kirkland, Quebec Canada H9H 5H3 Re: K053436 Trade Name: SurgiFrost® 6 cm, 7 cm and 10 cm CryoSurgical Devices plus FrostByte™ Clamp Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: Class II (two) Product Code: OCL Dated: December 8, 2005 Received: December 9, 2005 Dear Ms. Arana: This letter corrects our substantially equivalent letter of January 13, 2006. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Flor del Pilar Arana, MBA Please be advised that FDA's issuance of a substantial equivalence determination does not mean r hat FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/dsma/dsmamain.html Sincerely yours. el ome Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use K053436 510(k) Number (if known): Device Name: SurgiFrost® 6cm, 7cm and 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console Indications For Use: The SurgiFrost® 7cm and 10cm Cryosurgical Devices plus FrostByte" Clamp and Cryosurgical Console is intended for minimally invasive cardiac surgical procedures, including the treatment of cardiac arrhythmias. The SurgiFrost® 6cm, 7cm and 10cm Cryosurgical Devices plus FrostByte™ Clamp and Cryosurgical Console freezes the target tissue and blocks the electrical conduction pathways by creating an inflammatory response and cryonecrosis. Prescription Use X AND/OR Over-The-Counter Use (21CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Bhumman Ilvision Sign-Off) Alvision of Cardiovascular Devices 510(k) Number k. 253736 Page 1 of_____________________________________________________________________________________________________________________________________________________________________ Special 510(k) Confidential CryoCath Technologies Inc., December 7th, 2005 Page vi SurgiFrost® Surgical Cryoablation System
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...