VALIDATE CARDIAC MARKER (CM1) CALIBRATION VERIFICATION TEST SET
K053247 · Maine Standards Co. · JJY · Jan 13, 2006 · Clinical Chemistry
Device Facts
Record ID
K053247
Device Name
VALIDATE CARDIAC MARKER (CM1) CALIBRATION VERIFICATION TEST SET
Applicant
Maine Standards Co.
Product Code
JJY · Clinical Chemistry
Decision Date
Jan 13, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1660
Device Class
Class 1
Intended Use
The VALIDATE® Cardiac Marker Calibration Verification Test Sets are used by trained laboratory professionals for quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for CK-MB and Myoglobin. The VALIDATE® Cardiac Marker Calibration Verification Test Sets are used by trained laboratory professionals. They are not intended for use as routine quality control materials or as calibration materials. They are not intended for use with systems employing reflectance spectroscopy
Device Story
VALIDATE® Cardiac Marker Calibration Verification Test Sets consist of six levels of human serum matrix containing purified CK-MB and myoglobin. Used by laboratory professionals in clinical settings to verify calibration, linearity, and reportable ranges of chemistry analyzers. Not for routine quality control or calibration. Input: test material processed by chemistry analyzer; Output: quantitative recovery values compared against expected ranges to assess system performance. Benefits: ensures accuracy and linearity of cardiac marker testing.
Clinical Evidence
No clinical data. Bench testing only. Stability assessed via real-time and accelerated studies. Value assignment performed via triplicate assays across multiple runs with control material monitoring.
Technological Characteristics
Human serum-based calibration verification material. Liquid, ready-to-use form. 6 levels including zero. Storage at -10 to -20°C. No specific ASTM standards or software algorithms described.
Indications for Use
Indicated for trained laboratory professionals to verify calibration, validate reportable ranges, and determine linearity of automated, semi-automated, and manual chemistry systems for CK-MB and Myoglobin analytes.
Regulatory Classification
Identification
A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.
Predicate Devices
Bio-Rad Liquichek Cardiac Markers Control LT (k040277)
Related Devices
K102617 — AUDIT MICROCV CARDIAC MARKERS LINEARITY SET · Aalto Scientific, Ltd. · Jan 10, 2011
K090475 — VALIDATE CM2 CALIBRATION VERIFICATION / LINEARITY TEST SET, MODEL 402 · Maine Standards Co. · Apr 16, 2009
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the words "JAN 13, 2006" at the top. Below that is the word "maine" above a horizontal line. Below the line is the word "standards" and below that is the word "COMPANY" with each letter separated by a space.
## 510(k) SUMMARY
"This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92."
"The assigned 510(k) number is: K 053247 ."
| Submitter: | Maine Standards Company |
|------------|-----------------------------|
| Address: | 765 Roosevelt Trail |
| | Windham, ME 04062 |
| Telephone: | 207-892-1300 |
| Fax: | 207-892-2266 |
| Contact: | Christine Beach, Dir. RA/QA |
Summary prepared on: November 7, 2005
| Device classification name: | Multi-Analyte Controls, All Kinds (Assayed and Unassayed) |
|-----------------------------|----------------------------------------------------------------|
| Device description: | Quality control material (assayed and unassayed)* |
| Proprietary Name: | VALIDATE® Cardiac Markers Calibration Verification Test<br>Set |
| Regulation Number: | 21 CFR 862.1660 |
| Product Code: | JJY* |
*Note: There is no FDA product code for calibration / linearity materials. Therefore, as with previous submissions by Maine Standards and other calibration verification / linearity manufacturers, JJY has been selected as the "best fit" FDA code for this product.
Regulatory Class: Class I
#### Predicate Device:
- 1. Bio-Rad Liquichek Cardiac Markers Control LT (K040277), Bio-Rad Laboratories, Irvine, CA
- 2. DOCUMENT CK-MB CAL-VER, Microgenics Corporation, Fremont, CA Note: While the package insert for this commercially available product states that it is intended for in vitro diagnostic use in the determination of linearity and calibration verification for CK-MB and the accompanying worksheet provides lot specific analyte values, a k number could not be identified through the CDRH database.
Device description: VALIDATE Cardiac Markers Calibration Verification Test Sets are human based calibration verification materials containing multiple levels used to establish the relationship between theoretical operation and actual performance of the included analytes. There exists a linear relationship among each set of solutions.
{1}------------------------------------------------
Intended use: The VALIDATE Cardiac Markers Calibration Verification Test Sets are used by trained laboratory professionals for quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for the analytes listed on the package insert. These materials are not intended for use as routine quality control materials or as calibration materials.
### Comparison of VALIDATE® CM 1 Calibration Verification Test Set to the predicate devices:
Table 1 compares characteristics of the VALIDATE® CM1 Calibration Verification Test Set With those of the DOCUMENT CK-MB CAL ●VER and Bio-Rad Liquichek Cardiac Markers Control LT.
| | VALIDATE®<br>CM1 | DOCUMENT®<br>CK-MB CAL·VER | Bio-Rad<br>Liquichek™<br>Cardiac Markers<br>Control LT |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------|
| Catalog # | CM1 | M-116 | |
| Intended<br>Use | For in vitro diagnostic use<br>in quantitatively verifying<br>calibration, validating<br>reportable ranges, and<br>determining linearity in<br>automated, semi-<br>automated and manual<br>chemistry analyzers. | For use in the quantitative<br>determination of linearity,<br>calibration verification,<br>verification of analytical<br>measurement range<br>(AMR), and clinical<br>reportable range (CRR) of<br>manual and automated<br>chemistry analyzers. | For use as an<br>assayed quality<br>control serum to<br>monitor the<br>precision of<br>laboratory testing<br>procedures |
| Analytes | CK-MB, MYO | CK-MB | CK-MB, MYO,<br>Digitoxin,<br>Homocysteine,<br>NT-ProBNP,<br>Troponin-I,<br>Troponin-T |
| Matrix | Human Serum | Human Serum | Human Serum |
| Number of<br>Levels | 6 including zero | 6 | 3 |
| Preparation | Liquid, ready to use | Liquid, ready to use | Liquid, ready to use |
| Packaging | 6 x 3.0mL | 6 x 5.0ml | 6 x 3.0mL |
| Stability | Until expiration | Until expiration | 10 days after<br>opening |
| Storage | -10 to -20°C | -10 to -20°C | -10 to -20°C |
# TABLE 1 - Comparison of Devices
The performance of VALIDATE® CM1 Calibration Verification Test Set solutions on the Abbott AxSym instrument system as compared to DOCUMENT® CK-MB CAL·VER and Bio-Rad Liquichek Cardiac Markers Control LT has been shown to be substantially equivalent using preproduction lots of VALIDATE® CM1 Calibration Verification Test Sets.
{2}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the edge. In the center of the seal is a stylized image of an eagle with its wings spread. The eagle is made up of three thick, curved lines that form the body and wings, and two ribbon-like lines that form the tail.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
JAN 1 3 2006
Ms. Christine Beach Director, QA/RA Maine Standards Co. 765 Roosevelt Trail Suite 9A Windham, ME 04062
Re: k053247
Trade/Device Name: VALIDATE® Cardiac Marker Calibration Verification Test Sets Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I Product Code: JJY Dated: November 10, 2005 Received: November 21, 2005
Dear Ms. Beach:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{3}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto Guts
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use
K 053247 510(k) Number (if known):
VALIDATE® Cardiac Marker Calibration Verification Test Sets Device Name:
Indications For Use:
The VALIDATE® Cardiac Marker Calibration Verification Test Sets are used by trained laboratory professionals for quantitatively verifying calibration, validating reportable ranges, and determining linearity in automated, semi-automated and manual chemistry systems for CK-MB and Myoglobin.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
R. Chelen for Carol Benson
ision Sign-Off
Office of In Vitro Diagnostic Device Evaluation and Safety
Page 1 of _ 1 ___
K 053277
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.