GORE SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL

K053202 · W.L. Gore & Associates · FTL · Dec 7, 2005 · General, Plastic Surgery

Device Facts

Record IDK053202
Device NameGORE SEAMGUARD BIOABSORBABLE STAPLE LINE REINFORCEMENT MATERIAL
ApplicantW.L. Gore & Associates
Product CodeFTL · General, Plastic Surgery
Decision DateDec 7, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.3300
Device ClassClass 2
AttributesTherapeutic

Indications for Use

GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material can be used for reinforcement of staple lines during hysterectomy, lung resection, liver resection, bladder reconstruction, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, small bowel, and spleen procedures. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is also intended to be used for reinforcement of suture-lines and staple-lines (i.e., occlusion of the left atrial appendage during open chest procedures) during cardiac surgery.

Device Story

Bioabsorbable membrane composed of synthetic glycolide and trimethylene carbonate copolymer; designed for use with surgical stapling devices in open procedures. Material supplied in sterile pouches for application to cartridge and anvil jaws of staplers. Provides reinforcement to staple lines and suture lines during soft tissue resection or transection. Used by surgeons in clinical settings to reduce potential for leaks or bleeding at surgical sites. Device is passive; no electronic or mechanical operation; provides physical reinforcement that absorbs over time.

Technological Characteristics

Microporous bioabsorbable membrane; material: synthetic glycolide and trimethylene carbonate copolymer; supplied sterile in polyethylene terephthalate/polyethylene pouches; intended for use with open surgical staplers.

Indications for Use

Indicated for patients undergoing surgical procedures requiring soft tissue transection or resection where staple line reinforcement is necessary, including hysterectomy, lung, liver, bladder, bronchial, bariatric, colon, colorectal, esophageal, gastric, mesenteric, pancreatic, small bowel, splenic, and cardiac surgeries (including left atrial appendage occlusion).

Regulatory Classification

Identification

Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K053202 I/= #### DEC 7 2005 ## Premarket Notification 510(k) Summary #### Applicant: W.L. Gore and Associates Inc. 3250 W. Kiltie Lane Flagstaff, AZ 86001 #### Contact: · Michael E. Ivey #### Date Prepared: November 15, 2005 #### Trade or Proprietary Name: SEAMGUARD® Staple Line Reinforcement Material #### Common or Usual Name: Staple Line Reinforcement Material #### Classification 21 CFR 878.3300, FTL #### Device Predicate: SEAMGUARD® Staple Line Reinforcement Material - K043056 #### Device Description: The GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material for open staplers consists of a bioabsorbable membrane comprosed of a microporous structure of synthetic bioabsorbable glycolide and trimethylenc carbonate copolymer. The GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is to be used with surgical stapling devices. The GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is supplied in sterile polyethylene terphthalate / polyethylene pouches. These pouches contain the necessary material for the cartridge jaw and the anvil jaw of the surgical stapler. {2}------------------------------------------------ ### Statement of Intended Use: GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is indicated for use in surgical procedures in which soft tissue transection with staple line reinforcement is needed. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material can be used for reinforcement of staple lines during hysterectomy, lung resection, liver resection, bladder reconstruction, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, small bowel, and spleen procedures. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is also intended to be used for reinforcement of suture-lines and staple-lines (i.e., occlusion of the left atrial appendage during open chest procedures) during cardiac surgery. ### Substantial Equivalence: A variety of tests, assessments, and comparisons demonstrate that the SEAMGUARD® Staple Line Reinforcement Material for Open Staplers is substantially equivalent to its predicate in terms of composition, design, intended use, mode of operation and performance attributes. {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the eagle. Public Health Service DEC 7 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Michael Ivey Regulatory Affairs Medical Products Division W. L. Gore & Associates, Inc. 3450 West Kiltie Lane Flagstaff, Arizona 86003-2400 Re: K053202 Trade/Device Name: GORE SEAMGUARD® Staple Line Reinforcement Material for Open Staplers Regulation Number: 21 CFR 878.3300 Regulation Name: Surgical mesh Regulatory Class: II Product Code: FTL Dated: November 15, 2005 Received: November 16, 2005 Dear Mr. Ivey: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ #### Page 2- Michael Ivey This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Barbara Buelum Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K053202 Device Name: GORE SEAMGUARD® Staple Line Reinforcement Material for Open Staplers Indications for Use: The GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is indicated for use in surgical procedures in which soft tissue transection or resection with staple line reinforcement is needed. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material can be used for reinforcement of staple lines during hysterectomy, lung resection, liver resection, bladder reconstruction, bronchial, bariatric, colon, colorectal, esophagus, gastric, mesentery, pancreas, small bowel, and spleen procedures. GORE SEAMGUARD® Bioabsorbable Staple Line Reinforcement Material is also intended to be used for reinforcement of suture-lines and staple-lines (i.e., occlusion of the left atrial appendage during open chest procedures) during cardiac surgery. | Prescription Use | X | |-----------------------------|--------| | (Part 21 CFR 801 Subpart D) | | | | AND/OR | | Over-The-Counter Use | | | (21 CFR 801 Subpart C) | | (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Sign-Off) for mxm Division of General, Restoration and Neurological Devices n - 3202
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...