← Product Code [KNW](/productcode/KNW) · K053151

# RUBICOR MAGIC BREAST BIOPSY DEVICE (K053151)

_Rubicor Medical, Inc. · KNW · Nov 18, 2005 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K053151

## Device Facts

- **Applicant:** Rubicor Medical, Inc.
- **Product Code:** [KNW](/productcode/KNW.md)
- **Decision Date:** Nov 18, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1075
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** 3rd-Party Reviewed

## Indications for Use

The Rubicor Magic ™ Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses. The Rubicor Magic ™ Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality. The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

## Device Story

Sterile, single-use percutaneous biopsy device; 10-gauge stainless steel coring cannula; sharp stylet. Handle contains DC motor, 9V battery, drive mechanism for cannula rotation/advancement, and sample collection chamber. Operated by clinician during breast biopsy procedures. Device captures tissue core samples for histologic examination. Provides partial or complete removal of imaged abnormalities. Benefits patient by facilitating diagnostic tissue sampling for suspected breast abnormalities.

## Clinical Evidence

Bench testing only. In-vitro performance testing demonstrated safety and effectiveness for intended function. Biocompatibility testing met ISO 10993-1 requirements.

## Technological Characteristics

10-gauge, 8 cm stainless steel coring cannula; sharp stylet. Electromechanical drive (DC motor, 9V battery). Single-use, sterile. Biocompatibility per ISO 10993-1. No software or connectivity.

## Regulatory Identification

A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Predicate Devices

- Ethicon Mammotome® Hand-Held System ([K991980](/device/K991980.md))
- Sanarus Cassie™ Rotational Core Biopsy System ([K042136](/device/K042136.md))
- BIP VacuFlash® Biopsy System ([K024089](/device/K024089.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510 (k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Date Prepared: October 18, 2005

510(k) number: 1 1 65315 15) | _______________________________________________________________________________________________________________________________________________

Applicant Information: Rubicor Medical, Inc.

849 Veterans Blvd. Redwood City, CA 94063

| Contact Person: | Ary Chernomorsky |
|-----------------|------------------|
| Phone Number:   | (650) 556-1070   |
| Fax Number:     | (650) 556-1821   |

# Device Information:

Classification: Class II Rubicor Magic TM Breast Biopsy Device Trade Name: Biopsy Instrument (21 CFR 876.1075) Classification Name:

### Equivalent Device:

The subject device is substantially equivalent in intended use and/or method of operation to the Ethicon Mammotome® Hand-Held System (K991980), the Sanarus Cassie™ Rotational Core Biopsy System (K042136), and the BIP VacuFlash® Biopsy System (K024089).

# Device Description:

The Rubicor Magic Breast Biopsy Device is a sterile, single-use percutaneous biopsy device. The working end of the device includes an 8 cm long, 10 gauge, stainless steel coring cannula and a sharp stylet that runs through the center of the cannula and extends from its distal end. The handle of the device contains an actuation button, two indicator lights, a removable sample collection chamber, a drive mechanism for advancing and rotating the coring cannula and transporting the core specimen to the collection chamber, a DC motor, and a 9 V battery.

### Intended Use:

The Rubicor Magic Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses.

The Rubicor Magic TM Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

: :

{1}------------------------------------------------

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

# Test Results:

#### Performance

Results of in-vitro testing demonstrate that the Rubicor Magic Breast Biopsy Device is safe and effective for its intended function.

#### Biocompatibility

The materials used in the Rubicor Magic TM Breast Biopsy Device meet the requirements of ISO 10993-1.

#### Summary:

Based on the intended use, product, and performance information provided in this notification, the subject device has been shown to be substantially equivalent to the currently marketed and unmodified predicate device.

{2}------------------------------------------------

#### Indication for Use Statement

| 510(k) Number (if known): | K053151 |
|---------------------------|---------|
|---------------------------|---------|

Device Name:

Rubicor Magic TM Breast Biopsy Device

Indications for Use:

The Rubicor Magic ™ Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses.

The Rubicor Magic ™ Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

### Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use V (Per 21 CFR 801.109)

OR

Over-the Counter Use

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines above them.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 8 2005

Rubicor Medical, Inc. c/o Morten Simon Christensen Underwriters Laboratories, Inc. 1655 Scott Boulevard Santa Clara, California 95050

Re: K053151

Trade/Device Name: Rubicor Magic "" Breast Biopsy Device Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: II Product Code: KNW Dated: November 8, 2005 Received: November 10, 2005

Dear Mr. Christensen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

#### Page 2- Morten Simon Christensen

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark McMillan

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indication for Use Statement

510(k) Number (if known):

KOSSISI

Device Name:

Rubicor Magic ™ Breast Biopsy Device

Indications for Use:

The Rubicor Magic ™ Breast Biopsy Device is intended for diagnostic sampling of breast tissue during breast biopsy procedures. It is to be used for diagnostic purposes only and is not intended for therapeutic uses.

The Rubicor Magic ™ Breast Biopsy Device is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality.

The extent of histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore, the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation(ODE) |
|-------------------------------------------------------|
|-------------------------------------------------------|

(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices

| 510(k) Number | K053151 |
|---------------|---------|
|---------------|---------|

| ription Use | OR | Over-the Counter Use |
|-------------|----|----------------------|
|-------------|----|----------------------|

Prescr (Per 21 CFR 801.109)

Rubicor Medical, Inc.

---

**Source:** [https://fda.innolitics.com/device/K053151](https://fda.innolitics.com/device/K053151)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
