The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system Physiological parameter data includes ECG, invasive blood pressure, noninvasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctor's office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.
Device Story
TRAM Module functions as a component of GE host monitoring systems; collects physiological data (ECG, blood pressure, pulse oximetry, cardiac output, temperature, respiration) from patients; processes and stores data; transmits data to bedside monitors for display. Used in professional medical facilities (hospitals, clinics, surgical centers) across various units (OR, ICU, PARR). Operated by licensed healthcare practitioners or trained personnel. Provides continuous monitoring during intra-hospital transport; facilitates seamless transition between transport and bedside monitoring. Output allows clinicians to observe real-time patient status, supporting clinical decision-making and patient safety during movement.
Clinical Evidence
Bench testing only. Quality assurance measures included risk analysis, requirements/design reviews, unit-level module verification, system-level integration testing, final validation, performance testing, safety testing, and environmental testing.
Technological Characteristics
Multi-parameter modular system; processes ECG, invasive/noninvasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. Designed for integration with GE host monitoring systems. Class II device.
Indications for Use
Indicated for adult, pediatric, and neonatal patients requiring uninterrupted physiological monitoring (ECG, invasive/noninvasive blood pressure, pulse oximetry, cardiac output, temperature, respiration) during intra-hospital transport or bedside monitoring in professional medical facilities.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
Predicate Devices
GEMS-IT Transport Remote Acquisition Module (TRAM) (K011000)
Submission Summary (Full Text)
{0}------------------------------------------------
05-3/21
# Premarket Notification 510(k) Summary
As required by section 807.92
| Date: | February 1, 2006 | | APR | 5 | 2006 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------|--|-----|---|------|
| Submitter: | GE Medical Systems Information Technologies<br>8200 West Tower Avenue<br>Milwaukee, WI 53223 USA | | | | |
| Contact Person: | Joel Kent<br>GE Healthcare<br>86 Pilgrim Road<br>Needham, MA 02492 USA<br>Phone: (781) 449-8685<br>Fax: (781) 433-1344 | | | | |
| Device: Trade Name: | TRAM Module | | | | |
| Common/Usual Name: | Physiological Patient Monitor (Multi-parameter Module) | | | | |
| Classification Names: | The following Class II classifications appear applicable: | | | | |
| | 21 CFR 870.1025 Detector and Alarm, Arrhythmia<br>21 CFR 870.1025 Monitor, Physiological, Patient (With<br>Arrhythmia Detection Or Alarms) | | | | |
| | 21 CFR 868.2375 Breathing Frequency Monitor | | | | |
| | 21 CFR 870.1100 Monitor, Blood Pressure, Indwelling | | | | |
| | 21 CFR 870.1130 Noninvasive Blood Pressure Measurement<br>System | | | | |
| | 21 CFR 870.1100 Blood Pressure Alarm | | | | |
| | 21 CFR 870.1425 Programmable Diagnostic Computer | | | | |
| | 21 CFR 870.1915 Probe, Thermodilution | | | | |
| | 21 CFR 870.2340 Electrocardiograph | | | | |
| | 21 CFR 880.2910 Monitor, Temperature (with probe) | | | | |
| | 21 CFR 870.2300 Monitor, Cardiac (Incl. Cardiotachometer &<br>rate alarm) | | | | |
| | 21 CFR 870.2700 Oximeter, Pulse | | | | |
| Predicate Devices: | K011000 GEMS-IT Transport Remote Acquisition Module<br>(TRAM) | | | | |
{1}------------------------------------------------
Device Description: The TRAM Module is part of a multi-parameter modular system that measures and processes a patient's physiologic parameters. The TRAM Module works as a component of GE Medical Systems Information Technologies host monitoring systems and does not function on its own. The TRAM Module collects a patient's physiological data and sends it to a GEMS IT bedside monitor for display. TRAM modules incorporate different monitoring capabilities based on their configuration. The TRAM Module can also be used as a transport monitor when used with the Transport Remote Acquisition Module Patient Monitoring System.
#### Intended Use:
The TRAM Module is intended for use under the direct supervision of a licensed healthcare practitioner, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system Physiological parameter data includes ECG, invasive blood pressure, noninvasive blood pressure, pulse oximetry, cardiac output, temperature, and respiration. The TRAM Module acquires, processes and stores information regarding these parameters. The device is intended for use in a professional medical facility, such as hospital, clinic, surgical center or doctor's office. The TRAM System can be used in multiple areas such as operating room (OR), post anesthesia recovery (PARR), critical care, surgical intensive care, respiratory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.
Technology:
The TRAM Module employs the same functional scientific technology as its predicate devices.
{2}------------------------------------------------
## Test Summary:
The TRAM Module and its host patient monitoring system comply with the voluntary standards as detailed in Section 9 of this submission. The following quality assurance measures were applied to the development of the TRAM Module:
- Risk Analysis .
- . Requirements Reviews
- Design Reviews .
- Testing on unit level (Module verification) ●
- . Integration testing (System verification)
- Final acceptance testing (Validation) +
- Performance testing .
- . Safety testing
- Environmental testing .
### Conclusion:
The results of these measurements demonstrated that the TRAM Module are as safe, as effective, and perform as well as the predicate device.
{3}------------------------------------------------
Image /page/3/Picture/10 description: The image shows a circular logo with the words "U.S. DEPARTMENT OF HEALTH AND HUMAN SERVICES" written around the outside. Inside the circle is a symbol that looks like three stylized birds in flight. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
5 2006 APR
GE Medical Systems Information Technologies c/o Mr. Joel Kent GE Healthcare 86 Pilgrim Road Needham, MA 02492
Re: K053121 Trade Name: TRAM Module Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-segment Measurement and Alarm) Regulatory Class: Class II (two) Product Code: MHX Dated: February 10, 2006 Received: February 13, 2006
Dear Mr. Kent:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Mr. Joel Kent
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bhemmmon for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): K053121
Device Name: TRAM Module
Indications for Use:
The TRAM Module is intended for use under the direct supervision of a licensed The Treationer, or by personnel trained in its proper use. The TRAM Module is intended to provide uninterrupted monitoring of physiologic parameter data on adult, pediatric and neonatal patients during transport from one area of the hospital or facility, and monitoring system, to another. During non-transport monitoring, the TRAM Module functions in the bedside monitoring system.
Physiological parameter data includes ECG, invasive blood pressure, non-invasive ma I fryslorogical paramoter data merades output, temperature, and respiration. The TRAM Module acquires, processes and stores information regarding these parameters.
The device is intended for use in a professional medical facility, such as hospital, clinic, I he device is mentator's office. The TRAM System can be used in multiple areas such surgiour of Colors (OR), post anesthesia recovery (PARR), critical care, surgical as operating Toon. (ory intensive care, coronary care, medical intensive care, pediatric intensive care, or neonatal intensive care. The device is intended for use as part of a transport monitoring system for intra-hospital transport.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Blymmumor
gn-Off)
Cardiovascular Devices
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.