DIMENSION CARDIOPHASE HIGH SENSITIVITY C-REACTIVE PROTEIN CALIBRATOR

K053104 · Dade Behring, Inc. · JIS · Dec 23, 2005 · Clinical Chemistry

Device Facts

Record IDK053104
Device NameDIMENSION CARDIOPHASE HIGH SENSITIVITY C-REACTIVE PROTEIN CALIBRATOR
ApplicantDade Behring, Inc.
Product CodeJIS · Clinical Chemistry
Decision DateDec 23, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1150
Device ClassClass 2

Intended Use

The Dimension® CCRP Calibrator is an in vitro diagnostic product intended to be used to calibrate the Dimension® CardioPhase® high sensitivity C-reactive protein (Cat. No. RF434) method for the Dimension® clinical chemistry system with the heterogeneous immunoassay module. This product was designed to meet the needs of users to assure accurate results over the assay range of this method.

Device Story

Liquid bovine serum albumin-based calibrator; contains human C-reactive protein in levels 2-5. Used in clinical chemistry laboratories to calibrate Dimension® clinical chemistry systems for hsCRP assays. Instrument automatically performs sampling, reagent delivery, mixing, and processing. Calibrator ensures accurate results across assay range (0.1–15.0 mg/L). Human source material tested negative for HIV 1/2, HBV, and HCV.

Clinical Evidence

No clinical data. Bench testing only. Stability studies conducted over 13 months comparing product stored at 2-8°C against -70°C reference. Acceptance criteria: linear regression slopes non-significant (p > 0.05) or drift within ±5% for non-zero samples and ±0.05 mg/L for zero samples.

Technological Characteristics

Liquid bovine serum albumin-based matrix; human-derived CRP. 5-level calibration set. Standardized to IFCC/BCR/CAP reference preparation (CRM470). Designed for use with Dimension® clinical chemistry systems. Stability validated via real-time, freeze/thaw, and stressed temperature protocols.

Indications for Use

Indicated for use as an in vitro diagnostic calibrator for the Dimension® CardioPhase® high sensitivity C-reactive protein (hsCRP) assay on Dimension® clinical chemistry systems. No specific patient population or contraindications defined.

Regulatory Classification

Identification

A calibrator is a device intended for medical purposes for use in a test system to establish points of reference that are used in the determination of values in the measurement of substances in human specimens. (See also § 862.2 in this part.)

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K053104 # DEC 2 3 2005 Page 1 of 4 # 510(k) Summary for the Dimension® CardioPhase® high sensitivity CRP Calibrator (CCRP Calibrator - RC434) ### A. 510(k) Number: B. Analyte: C-Reactive Protien Calibrator C. Type of Test: Calibrator Material ### D. Applicant: Dade Behring Inc., P.O. Box 6101, Newark, DE 19714-6101 Contact: Andrea M. Tasker, Regulatory Affairs and Compliance Manager (302) 631-9454 E. Proprietary and Established Names: Dimension® CardioPhase® high sensitivity CRP Calibrator (CCRP Calibrator- RC434) ### F. Regulatory Information: - 1. Regulation section: 21 CFR § 862-1150 CALIBRATOR - 2. Classification: Class II - 3. Product Code: JIS CALIBRATORS, PRIMARY - 4. Panel: CLINICAL CHEMISTRY ### G. Intended Use: 1. Intended use(s): The Dimension® CCRP Calibrator is an in vitro diagnostic product intended to be used to calibrate the Dimension® CardioPhase® high sensitivity C-reactive protein (Cat. No. RC434) method for the Dimension® clinical chemistry system with the heterogeneous immunoassay module. This product was designed to meet the needs of users to assure accurate results over the assay range of this method. ### 2. Indication(s) for use: The Dimension® CCRP Calibrator is an in vitro diagnostic product intended to be used to calibrate the Dimension® CardioPhase® high sensitivity C-reactive protein (Cat. No. RC434) method for the Dimension® clinical chemistry system with the heterogeneous immunoassay module. This product was designed to meet the needs of users to assure accurate results over the assay range of this method. {1}------------------------------------------------ - 3. Special condition for use statement(s): none - 4. Special instrument Requirements: none ### H. Device Description: The high sensitivity C-reactive protein Calibrator is a liquid bovine serum albumin-based product. Levels 2 -5 contain a human C-reactive protein. # I. Substantial Equivalence Information: - 1. Predicate device name(s): N Rheumatology Standard SL on the BN Systems. - 2. Predicate K number(s): K964527 - 3. Comparison with predicate: | Similarities | | | |--------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Item | Device<br>Dimension® CardioPhase®<br>(CCRP) Calibrator | Predicate<br>N Rheumatology Standard<br>SL | | Intended Use | The Dimension® CCRP<br>Calibrator is an in vitro<br>diagnostic product intended to<br>be used to calibrate the<br>Dimension® CardioPhase®<br>high sensitivity C-reactive<br>protein method for the<br>Dimension® clinical chemistry<br>system with the heterogeneous<br>immunoassay module. | Establishment of reference<br>curves for the<br>immunonephelometric<br>determination of rheumatoid<br>factors (RF), anti-streptolysin<br>O (ASL) and C-reactive<br>protein (CRP) using the BN<br>Systems. | | Traceability | IFCC/BCR/CAP<br>CRM 470 | IFCC/BCR/CAP<br>CRM 470 | {2}------------------------------------------------ | Differences | | | |------------------------------------|----------------------------------------------------|----------------------------------------------------------------------------------| | Item | Device | Predicate | | | Dimension® CardioPhase®<br>hsCRP (CCRP) Calibrator | N Rheumatology Standard<br>SL | | Matrix | liquid bovine serum albumin-<br>based | human sera based | | Number of levels | Multi-Point<br>5 levels provided | Multi-Point<br>1 level provided<br>(instrument prepares<br>6 standard dilutions) | | Reference Curve<br>Fit Calculation | Logit | 4 Parameter Logit Log | # J. Standard/Guidance Document Referenced: #### 1. Guidance; Review Criteria for Assessment of C-Reactive Protein (CRP), High Sensitivity C Reactive Protein (hsCRP) and Cardiac C-Reactive Protein (cCRP) Assays Document issued on: September 22, 2005 | 2. Standards; | | |----------------|-----------------------------------------------------------------------------------------------------| | GP22-A | Continuous Quality Improvement Essential Management Approaches | | CEN 13640 | Stability testing of In-Vitro Diagnostic Devices | | ISO 14971:2000 | Medical devices -Application of risk management to medical<br>devices | | ISO 15223 | Medical devices - Symbols to be used with medical device labeling<br>and information to be supplied | #### K. Test Principle: The Dimension® CCRP Calibrator is an in vitro diagnostic product intended to be used to calibrate the Dimension® CardioPhase® high sensitivity C-reactive protein (Cat. No. RF434) method for the Dimension® clinical chemistry system with the heterogeneous immunoassay module. The hsCRP method is a one step enzyme immunoassay based on the "sandwich" principle. {3}------------------------------------------------ ### L. Performance Characteristics: ### 1. Stability Calibrator shelf life is determined by comparing results of the product stored at 4°C with product stored at -70°C to ensure that analytical system drift is dissociated with calibrator drift. Linear regression across the shelf life interval for each test sample versus test day will have non-significant slopes (p> 0.05) or observed drift within +/-5% for non-zero samples and +/-0.05 mg/L at 0 mg/L CRP over a 12 month interval The shelf life of the product will be 6 months after the successful completion of 7 months of real time stability on three lots of product. ### 2. Traceability: The assigned values of the CCRP calibrator are standardized to the International Federation of Clinical Chemistry (IFCC) International Reference Preparation for Plasma Proteins, the Community Bureau of Reference (BCR) and the College of American Pathologists (CAP). The basis of this international standardization is the IFCC/BCR/CAP reference preparation for 14 human serum proteins (Lot No. 91/0619=CRM470=RPPHS 91/0619) (lot V). ### 3. Value Assignment Three Dimension® clinical chemistry analyzers are calibrated with the approved CCRP Masterpool. The acceptable recovery of an approved CCRP Calibrator lot must be obtained. Test calibrator levels are then tested on three separate analyzers with different CCRP flex lots. The grand mean of all 5 replicate test means of all 9 curves is the value assignment for each Calibrator Level. {4}------------------------------------------------ # DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Image /page/4/Picture/2 description: The image shows the logo for the Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract image of an eagle with three lines representing its wings and tail feathers. DEC 2 3 2005 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Ms. Andrea M. Tasker Regulatory Affairs and Compliance Dade Behring Inc. Glasgow Business Community PO Box 6101, Building 500 Newark, DE 19714-6101 Re: k053104 K033104 Trade/Device Name: Dimension® CardioPhase® high sensitivity C-reactive protein Calibrator (CCRP Calibrator -RC434) Regulation Number: 21 CFR 862.1150 Regulation Name: Calibrator Regulatory Class: Class II Product Code: JIS Dated: November 2, 2005 Received: November 4, 2005 Dear Ms. Tasker: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your becared on the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for assession to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, thereforey isions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {5}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin maxions of sybstantial equivalence of your device to a legally premation notification "Te results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or you desire specific mostion and advertising of your device, please contact the Office of In or quotions on the program and Safety at (240) 276-0484. Also, please note the v Itto Diagnoutied, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the r ou may other generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Alberto G Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### Indications For Use Statement 510(k) Number (if known): k053104 ### Device Name: Dimension® CardloPhase® high sensitivity C-reactive protein Calibrator (CCRP Calibrator - RC434) ### Indications for Use: The Dimension® CCRP Calibrator is an in vitro diagnostic product intended to be used to calibrate the Dimension® CardioPhase® high sensitivity C-reactive protein (Cat. No. RF434) method for the Dimension® clinical chemistry system with the heterogeneous immunoassay module. This product was designed to meet the needs of users to assure accurate results over the assay range of this method. Ann Chappie Division Office of in Vier Tingnoma ರಿಕಾರಿ ಕಾರದಿಂದರೆಯ ಮನೆ ಕಿಡ್ಡೆತಿ / 510(k) K053104 × Prescription Use (Per 21 CFR 801 Subpart D) AND/OR Over-the-counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of -In Vitro Diagnostic Devices (OIVD)
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...