ORTHOGLIDE MEDICAL KNEE IMPLANT AND INSTRUMENTS, MODELS 9063, 9064, 9065, 9060, 9062, 9057, 9056
K053094 · Advanced Bio-Surfaces, Inc. · HSH · Feb 6, 2006 · Orthopedic
Device Facts
Record ID
K053094
Device Name
ORTHOGLIDE MEDICAL KNEE IMPLANT AND INSTRUMENTS, MODELS 9063, 9064, 9065, 9060, 9062, 9057, 9056
Applicant
Advanced Bio-Surfaces, Inc.
Product Code
HSH · Orthopedic
Decision Date
Feb 6, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3590
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Orthoglide Medial Knee Implant is intended for use in the osteoarthritic knee, where a substantial amount of cartilage has been lost as a result of the disease. The device is indicated for uncemented use in the treatment of moderate degeneration of the medial compartment of the knee (grade II-IV chondromalacia) with no more than minimal degeneration (grade I-II chondromalacia, no loss of joint space) in the lateral and patellofemoral compartments in patients with osteoarthritis.
Device Story
Orthoglide Medial Knee Implant is a metallic hemi-knee resurfacing prosthesis for the medial compartment. Implanted via minimally invasive surgery between the tibial plateau and femoral condyle. Device geometry and ligament tension maintain position. Functions as an articulating surface with low friction and high durability to relieve pain, improve knee alignment (valgus position), and restore joint spacing. Used by surgeons in clinical settings. Benefits include pain relief and preservation of joint function through mechanical realignment and load distribution.
Clinical Evidence
No clinical data. Evidence consists of bench testing (tensile elongation, load deflection, cyclic fatigue resistance, material consistency) and cadaver evaluations verifying surgical technique, instruments, stability, angular correction, and range of motion preservation. Biocompatibility (ISO 10993) and sterility (EN550) standards met.
Technological Characteristics
Cobalt-Chrome-Molybdenum alloy construction. Uncemented hemi-knee resurfacing prosthesis. Designed for medial compartment placement. Conforms to ISO 10993 (biocompatibility) and EN550 (sterility) standards.
Indications for Use
Indicated for patients with osteoarthritis requiring treatment for moderate medial compartment knee degeneration (grade II-IV chondromalacia), provided lateral and patellofemoral compartments show no more than minimal degeneration (grade I-II chondromalacia, no joint space loss).
Regulatory Classification
Identification
A knee joint tibial (hemi-knee) metallic resurfacing uncemented prosthesis is a device intended to be implanted to replace part of a knee joint. The device limits minimally (less than normal anatomic constraints) translation in one or more planes. It has no linkage across-the-joint. This prosthesis is made of alloys, such as cobalt-chromium-molybdenum, and is intended to resurface one tibial condyle. The generic type of device is limited to those prostheses intended for use without bone cement (§ 888.3027).
Predicate Devices
Sulzer Orthopedics, Inc. Unicondylar Interpositional Spacer (K003269)
ITI, Knee Interpositional Mini-Repair system (KIMRS) (K033242)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the text "LABS" in large, bold, sans-serif font. A trademark symbol is located to the right and slightly above the "S". Below "LABS", smaller text reads "APPLIED BIO SURFACES, INC." The text is black against a white background.
## 510(k) Summary
# Advanced Bio-Surfaces, Inc.
# Orthoglide® Medial Knee Implant
510(k) Notification K053094
## MANUFACTURER INFORMATION
- Advanced Bio-Surfaces, Inc. Name & 5909 Baker Road Address: Minnetonka, MN 55345 USA
Summary Prepared: November 2, 2005
- Contact: Philip B. Jarvi Director, Q.A. / Regulatory Affairs 952-912-5400 phone 952-912-5410 fax
## DEVICE INFORMATION
Trade Name: Orthoglide® Medial Knee Implant
Classification Name: 21CFR 888.3590 - Knec joint tibial (hemi-knee) metallic resurfacing uncemented prosthesis
Product Code: HSH
Common / Usual Name: Hemi- knee prosthesis
Substantial equivalence: The ABS Orthoglide® Medial Knee Implant is substantially equivalent to FDA approved predicate devices with regard to indications for use, materials, technological characteristics and surgical techniques. These predicate devices
- are: Sulzer Orthopedics, Inc. Unicondylar Interpositional Spacer (K003269)
ITI, Knee Interpositional Mini-Repair system (KIMRS) (K033242)
{1}------------------------------------------------
Device Description: The Orthoglide Medial Knee Implant is placed in the medial compartment of the knee between the tibial plateau and femoral condyle by means of minimally invasive surgery. The instruments are intended to aid in the surgical preparation of the implant site and implant placement. The Orthoglide is made of a Cobalt-Chrome-Molybdenum Alloy. Device geometry and ligament tension combine to keep the implant in place. The implant covers the tibial plateau. The device is designed to improve the alignment of the knee, returning the joint to a more valgus position. Realignment of the knee distributes the weight-bearing forces across the joint and helps restore the normal relationships of the articular surfaces and the surrounding capsular, ligamentous and muscular structures.
The device is designed to help relieve pain by providing an articulating surface with a low coefficient of friction and high durability. Device geometry improves knee alignment and joint spacing. The device surface is smooth and when wet, is intended to mimic the lubricious surface previously provided by the articular cartilage.
- Intended Use: The Orthoglide Medial Knee Implant is intended for use in the osteoarthritic knee, where a substantial amount of cartilage has been lost as a result of the disease. The device is indicated for uncemented use in the treatment of moderate degeneration of the medial compartment of the knee (grade II-IV chondromalacia) with no more than minimal degeneration (grade I-II chondromalacia, no loss of joint space) in the lateral and patellofemoral compartments in patients with osteoarthritis.
{2}------------------------------------------------
Substantial Equivalence: The ABS Orthoglide Medial Knee Implant is substantially equivalcnt to FDA approved predicate devices with regard to indications for use, materials, technological characteristics and surgical techniques. Orthoglide offers no additional risks to the patient and the materials and manufacturing methods add no new or additional safety concerns. The Orthoglide as well as the predicates are submitted without clinical information. The substantially equivalent predicate devices are:
Sulzer Orthopedics, Inc. Unicondylar Interpositional Spacer (K003269)
ITI, Knee Interpositional Mini-Repair system (KIMRS) (K033242)
Testing: The Orthoglide Implant has been developed, verified and validated in compliance with a comprehensive design process. The corresponding Bench Testing has been accomplished according to a Master Test Plan for the device. The test plan shows the required test, the source of the requirement, the protocol and the report. If the method is from a standard, the voluntary standard is referenced. The bench testing has demonstrated the physical attributes and durability of the Orthoglide by tensile elongation, load deflection, cyclic fatigue resistance, material consistency and stability and processing control. Cadaver testing has verified the surgical technique and instruments. Cadaver evaluations have also verified the desired physiological effects of stability, angular correction and range of motion preservation. The Orthoglide also demonstrated that it meets internationally recognized standards for
biocompatibility (ISO 10993), sterility (EN550), and conformance to material specifications. Clinical testing was not used to determine substantial equivalence.
Summary: Based on the evidence of substantial equivalence the Orthoglide is considered to safe and effective and will perform as well or better than the referenced predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
Public Health Service
2006 FEB 6
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Philip B. Jarvi Director, Q.A./Regulatory Affairs Advanced Bio-Surfaces, Inc. 5909 Baker Road Minnetonka, Minnesota 55345
Re: K053094
Trade/Device Name: Orthoglide Medial Knee Implant Regulation Number: 21 CFR 888.3590 Regulation Name: Knee joint tibial (hemi-knee) metallic resurfacing uncemented prosthesis Regulatory Class: II Product Code: HSH Dated: January 19, 2006 Received: January 20, 2006
Dear Mr. Jarvi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 – Mr. Philip B. Jarvi
CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CI IT rail 0077, incomig (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation on begin marketing your device as described in your Section 510(k) I mo loter witification. The FDA finding of substantial equivalence of your device to a legally premated notification - results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H Millikan
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K053094
Device Name: Orthoglide® Medial Knee Implant
Indications for Use:
The Orthoglide Medial Knee Implant is intended for use in the osteoarthritic knee, where substantial amounts of cartilage have been lost as a result of the disease. The device is indicated for uncemented use in the treatment of moderate degeneration of the medial compartment of the knee (grade II-IV chondromalacia) with no more than minimal degencration (grade I-II chondromalacia, no loss of joint space) in the lateral and patellofemoral compartments in patients with ostcoarthritis.
Prescription Use Yes (Part 21 CFR 801 Subpart D)
=... = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = = =
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark N Mulkerso
Sign-Off) Divisio: Cone Restorative Division and Neu nical bevi Page 1 of 1 S10(k) Number
(Posted November 13, 2003)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.