OPTO GLOBAL DIGITAL FUNDUS CAMERA SYSTEM MODEL ADS 1.5
K053044 · Opto Global, Inc. · HKI · Jan 24, 2006 · Ophthalmic
Device Facts
Record ID
K053044
Device Name
OPTO GLOBAL DIGITAL FUNDUS CAMERA SYSTEM MODEL ADS 1.5
Applicant
Opto Global, Inc.
Product Code
HKI · Ophthalmic
Decision Date
Jan 24, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.1120
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Opto Global Digital Fundus Camera System Model ADS 1.5 is intended to be used to capture, archive, and display digital images of the eye, particularly the retina obtained through the use of an ophthalmic camera (fundus camera).
Device Story
System captures, archives, and displays digital retinal images; comprises digital sensor head, image capture card, computer, monitor, and proprietary 'OPTO Global Capture' software. Used by ophthalmologists or photographers in clinical settings to document retinal pathologies. User views patient eye through fundus camera; light source illuminates retina; digital camera captures images. Software allows image acquisition, monitoring, manipulation, and storage on hard drive or removable media (CD/DVD). Output includes individual images or proof sheets (collages) for viewing or printing. Facilitates clinical decision-making by providing visual documentation for monitoring and treating eye diseases.
Clinical Evidence
Bench testing only. Software validation tests performed to confirm system functionality and performance standards comparable to predicate devices.
Technological Characteristics
System includes digital sensor head, computer interface circuit board, CPU, monitor, keyboard, and mouse. Operates as a digital ophthalmic camera system with image storage. Software platform provides image acquisition, storage, analysis, and retrieval. Connectivity via standard PC hardware. No specific material standards or AI/ML architecture specified.
Indications for Use
Indicated for capturing, archiving, recalling, and displaying black and white and color images of the retina and surrounding areas to aid in the diagnosis and monitoring of eye diseases viewable and photographable non-invasively. Prescription use only.
Regulatory Classification
Identification
An ophthalmic camera is an AC-powered device intended to take photographs of the eye and the surrounding area.
Special Controls
*Classification.* Class II (special controls). The device, when it is a photorefractor or a general-use ophthalmic camera, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
Imagenet Digital Ophthalmic Imaging (Cannot Locate)
50X, 50 EX Fundus camera (Cannot Locate)
Submission Summary (Full Text)
{1}------------------------------------------------
K053044
JAN 2 4 2006
# 510(k) Summary
### Name of device:
OPTO GLOBAL Digital Fundus Camera System Model ADS 1.5
Common or usual name: Camera, Ophthalmic (AC Powered)
#### Classification Name:
Ophthalmic Camera (per 21 CFR.886.1120) Device, Storage, Images, Ophthalmic (per 21 CFR.892.2010 Device, Communication, Images, Ophthalmic (per CFR 21 892.2020)
#### Product Code:
HKI, NFF, NFG
Submitter: OPTO Global, Inc.
4021 Daggett Drive Granite Bay, CA 95746 Phone: (916) 783-3223 Facsimile: (916) 782-5847 Contact Person: Mark T. Fukuhara Date Prepared: 20 August 2005
#### Predicate Devices
| Trade Name | Manufacturer | 510 (k) |
|-----------------------------------------------------------------------|-----------------------------------|---------------|
| Visucam | Carl Zeiss Ophthalmic Systems Inc | K021787 |
| FF450 plus Visupac System | Carl Zeiss Jena GmbH | K011877 |
| Imagenet Digital Ophthalmic Imaging*<br>*(Sold in the U.S. for years) | Topcon Corp | Cannot Locate |
| 50X, 50 EX Fundus camera*<br>*( Sold in the U.S. for years) | Topcon Corp | Cannot Locate |
Sec 12: 1 of 4
{2}------------------------------------------------
### Intended Use
The Opto Global Digital Fundus Camera System Model ADS 1.5 is intended to be used to capture, archive, and display digital images of the eye, particularly the retina obtained through the use of an ophthalmic camera (fundus camera).
### Device Description
The OPTO Global ADS 1.5 Digital Fundus Camera System is an automated imaging device used in conjunction with a digital camera that requires ninimal intervention during the capture of an image. The system is simple to use and requires minimal training for a user to become proficient with the system. Like the listed predicate devices, the Opto Global Digital Fundus Camera System Model ADS 1.5 is comprised of a digital imaging camera, computer hardware, and a software platform intended to be used to store images captured by the fundus camera. Thus the Opto Global Digital Fundus Camera System Model ADS 1.5 has the same intended use and indications as the listed predicate devices.
### Substantial Equivalence
The Opto Global Digital Fundus Camera System Model ADS 1.5 and the predicate devices listed all have the same intended use: to capture and archive images of the retina taken with a fundus camera. The Opto Global Digital Fundus Camera System Model ADS 1.5 and the predicate devices listed have equivalent principles of operation and technological characteristics. Each of the devices is a digital ophthalmic camera system and image storage device. The user views the patient's retina through a fundus camera. A light source is used to illuminate the repacency the fundus triggers the digital cameras to capture images of the retina. These digitized images are then archived for future use or record.
The Opto Global Digital Fundus Camera System Model ADS 1.5 and all of the listed predicate devices use, and are operated by, PC's with keyboards and a or and operated mouse. While there may be some very minor differences in types of processor (Intel Pentium IV or AMD Athelon) processors speed, and software platform, these minor differences do not raise any new issues of safety or effectiveness and do not affect the imaging capabilities of the Opto Global Digital Fundus Camera System Model ADS 1.5 or any of the predicate devices listed.
{3}------------------------------------------------
The Opto Global Digital Fundus Camera System Model ADS 1.5 and all of the predicate devices have the same basic software functions: image acquisition, storage, analysis, and retrieval. The principle differences can be found incon, graphical user interface.
Most software platforms found in the predicate devices are windows based and effectively function using different screens for different functions, and drop down menus and graphic buttons to control, or perform, various tasks in the crop ure, archive, and recall of images.
These slight differences do not raise new or additional questions regarding safety or efficacy in the OPTO Global Digital Fundus Camera System Model ADS 1.5 r other predicate systems. OPTO Global Inc. has performed several software validation tests the results of which clearly indicate that the Opto Global Digital Fundus Camera System Model ADS 1.5 meets comparable system and software standards exhibited by the predicate devices listed.
The OPTO Global Digital Fundus Camera System Model ADS 1.5 and all of the predicate devices listed are operated in the same manner. The Ophthalmolngist, or Photographer, views the patients eye through an ophthalmic camera (ffundyis camera) with a digital camera used to capture, manipulate and archive images. using the graphical user software interface. The Opto Global Digital Fundus Camera System Model ADS 1.5 and all of the predicate devices listed use the system software to manipulate images into produce proof sheets (photo collages of the patients eye captured during the procedure) the software also carries the capabilities of printing individual images or proof sheets, view individual photos or proof sheets on the computer monitor, and to archive selected captured images onto the computers hard drive or several types of removable media such in standard CD's or DVD's.
# Performance characteristics
The Opto Global Digital Fundus Camera System Model ADS 1.5 is comprised of the following components: A digital sensor head (digital camera) a computer interface circuit board (digital image capture card), and connecting cables. These components are then combined and sold together with our "OPTO Global Capture" proprietary imaging software and a computer (CPU) monitor, keyboard and mouse.
{4}------------------------------------------------
This total system with imaging software provides acquisition and hardware control capabilities used to take digital pictures of the retina which are then transferred via the digital camera and connecting cable to the computer system transment can be viewed, modified, stored or printed. The Opto Global Digital Funchs Camera System Model ADS 1.5 is intended to capture and store images of the Retina, and it is also indicated for use as an ophthalmic camera for individuals where documentation of the retina is important to follow and treat certain pathologies.
The Opto Global Digital Fundus Camera System Model ADS 1.5 software interface allows images from the fundus camera to be acquired, monitored, stored and retrieved. Imaging, focusing, and camera orientation in relation to the relina are controlled by the user. With verification and monitoring, the software allows the user to monitor, capture, and process, images thus verifying the device is operating correctly. The image viewed through the fundus camera and captured by the digital camera is then stored as individual images in a non compressed state on the hard drive of the computer to be displayed electronically as the user requires it.
## Conclusion
The Opto Global Digital Fundus Camera System Model ADS 1.5 has the same intended use, indications, and very similar principals of operation to the predicate devices listed. The Opto Global Digital Fundus Camera System Model ADS 1.5 also has similar technological characteristics (hardware and software) to the predicate devices listed.
The minor differences between the Opto Global Digital Fundus Camera System Model ADS 1.5 and those of the listed predicate devices do not raise any new questions of safety or of effectiveness in comparison to the predicate devices. Thus the Opto Global Digital Fundus Camera System Model ADS 1.5 is substantially equivalent to legally marketed ophthalmic camera systems.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of a human figure, represented by three curved lines, positioned to the right. Encircling the figure is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 4 2006
OPTO Global, Inc. C/O Lorinda Badalian Senior Project Engineer Underwriters Laboratories, Inc. 455 E. Trimble Rd San Jose, CA 95131-1230
Re: K053044
Trade/Device Name: OPTO Global Digital Fundus Camera System Model ADS 1.5 Regulation Number: 21 CFR 886.1120 Regulation Name: Ophthalmic Camera Regulatory Class: Class II Product Code: HKI; NFF; NFF; NFG Dated: January 10, 2006 Received: January 11, 2006
Dear Ms. Badalian:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Lorinda Badalian
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M.B.Egelman, MD
Malvina B. Eydelman, M.D Acting Division Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## Indications for Use
510(k) Number (if Known): K053044
Device Name: OPTO Global Digital Fundus Camera System Model ADS 1.5
#### Indications for Use :
The OPTO Global Digital Fundus Camera System Model ADS 1.5 is intended to be used to capture, archive, recall and display black and white, and color images of the retina, as well as surrounding areas, to aid in the diagnosis and monitoring of diseases of the eye that may be viewed and photographed non-invasively.
Prescription Use (Per CFR 21 801.109)
or
Over-the-counter Use__________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH Office of Device Evaluation (ODE) |
|-------------------------------------------------------|
| (Division Sign-Off) |
Division of Ophthalmic Ear,
Nose and Throat Devises
10(k) Number KOS 306
Sec 10: 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.