Acro Rapid Benzoylecgonine (Cocaine) Urine test is a lateral flow, one-step immunoassay for the qualitative detection of benzoylecgonine in human urine at a cutoff of 300 ng/mL. The test is used to obtain a visual qualitative result and is intended for laboratory use only. This assay provides only preliminary result. Clinical consideration and professional judgment must be applied to a drug test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectroscopy (GC/MS) analysis is preferred.
Device Story
Lateral flow, one-step immunoassay for qualitative detection of Benzoylecgonine in human urine; 300 ng/mL cutoff. Device used in laboratory settings by professional personnel. Input is human urine sample; output is visual qualitative result (presence/absence of analyte). Preliminary result requires confirmation via alternate chemical method, preferably GC/MS. Clinical decision-making relies on professional judgment and confirmatory testing.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Lateral flow, one-step immunoassay. Qualitative visual readout. Cutoff concentration: 300 ng/mL. Designed for laboratory use.
Indications for Use
Indicated for the qualitative detection of Benzoylecgonine (cocaine metabolite) in human urine at a 300 ng/mL cutoff. Intended for laboratory use only to obtain preliminary results requiring confirmation by GC/MS.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
Related Devices
K994403 — INSTANT-VIEW COCAINE (BENZYLECGONINE) URINE DIP STRIP TEST · Alfa Scientific Designs, Inc. · Aug 14, 2000
K994398 — INSTANT-VIEW COCAINE (BENZYLECGONINE) URINE CASSETTE TEST · Alfa Scientific Designs, Inc. · Jul 27, 2000
K972384 — AQ ONE STEP COCAINE TEST · Bionike Laboratories, Inc. · Aug 6, 1997
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a caduceus or a staff with intertwined elements.
Public Health Service
MAY 5 2006 Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Acro Biotech LLC. c/o Gregory Lee, Ph.D. 9500 7th Street, Unit M Rancho Cucamonga, CA 91730
Re: k053035
> Trade/Device Name: Acro Rapid Benzoylecgonine (Cocaine) Urine Test Regulation Number: 21 CFR§862.3250 Regulation Name: Cocaine and cocaine metabolite test system Regulatory Class: Class II Product Code: DIO Dated: January 15, 2006 Received: March 6, 2006
Dear Dr. Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements af the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{1}------------------------------------------------
Page 2 -
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of sybstantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please not the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its tollar de (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Alberto G.
Alberto Gutierrez, Ph.D. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Premarket Notification
Acro Biotech LLC. Acro Rapid Benzoylecgonine (Cocaine) Urine Test
## II. FDA INDICATIONS FOR USE FORM
## 510(K) NUMBER (IF KNOWN): k053035
Device Name: Acro Rapid Benzoylecgonine (Cocaine)Urine Test
Indications for Use
Acro Rapid Benzoylecgonine (Cocaine) Urine Test is a lateral flow, one-step immunoassay for the qualitative detection of Benzoylecgonine in hyman wipe at a cutoff of 300 ng/mL The test is used to obtain a visual qualitative result and is intended for laboratory use only.
This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to a drug test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/ Mass Spectroscopy (GC/MS) anchiate is preferred.
> Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off
Office of In Vitao Diagnostic Device Evaluation and Safety
510(k) k053035
Acro Biotech, LLC., 9500 7th Street, Unit M, Rancho Cucamonga CA 91730 Tel: 909-466-6892, Fax: 909-466-6857
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.