FIDIS VASCULITIS, MODEL MX007

K053012 · Biomedical Diagnostics (Bmd) SA · MOB · Mar 2, 2006 · Immunology

Device Facts

Record IDK053012
Device NameFIDIS VASCULITIS, MODEL MX007
ApplicantBiomedical Diagnostics (Bmd) SA
Product CodeMOB · Immunology
Decision DateMar 2, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Indications for Use

The FIDIS™ VASCULITIS* test is a semi-quantitative homogenous fluorescent-based microparticles immunoassay using flow cytometry readings. The test system is used to detect in patient serum samples the presence of anti-neutrophil cytoplasm antibodies (ANCA) directed against Myeloperoxidase (MPO) and Serine Proteinase 3 (PR3); and anti-glomerular basement membrane (GBM) antibodies. The results of the FIDIS™ VASCULITIS* test are to be used in conjunction with the clinical findings and other laboratory tests to aid in the diagnosis of various primary systemic small vessel vasculitis. The presence of anti-MPO and anti-PR3 antibodies is associated with primary systemic small vessel vasculitis: Wegener's granulomatosis, Churg Strauss syndromes, microscopic periarteritis and idiopathic crescentic glomerulonephritis; and the presence of anti-GBM antibodies is associated with Goodpasture's syndrome. FIDIS™ VASCULITIS* kit is to be used on the FIDIS Analyser, MLX-BOOSTER Software and Washer. FIDIS™ VASCULITIS* kit could be used with the CARIST™ system (diluting and dispensing device). This test is for in vitro diagnostic use.

Device Story

Multiplexed immunoassay detects anti-MPO, anti-PR3, and anti-GBM antibodies in patient serum. Antigens coupled to color-coded microspheres; patient serum added to microplate wells; autoantibodies bind to antigens. Unbound antibodies removed via filtration wash; phycoerythrin-labeled anti-human IgG added to bind antibody-antigen complexes. FIDIS Analyzer (flow cytometer) discriminates bead sets and measures fluorescence intensity. MLX-Booster software interpolates antibody titers against calibration system. Used in clinical laboratories; results interpreted by clinicians alongside clinical findings to aid vasculitis diagnosis. CARIST system provides automated sample dilution/dispensing.

Clinical Evidence

No clinical data provided; the document is a 510(k) clearance notification focusing on the device's intended use and regulatory status.

Technological Characteristics

Multiplexed fluorescent immunoassay. Components: color-coded microspheres covalently coupled to MPO, PR3, and GBM antigens; phycoerythrin-labeled anti-human IgG conjugate. Platform: FIDIS Analyzer (Luminex 100 flow cytometer). Software: MLX-Booster. Sample: Serum. Format: 96-well microplate with filtration membrane. Detection: Fluorescence intensity measurement. Automated processing via CARIST system optional.

Indications for Use

Indicated for professional, prescription-only in vitro diagnostic use to detect anti-MPO, anti-PR3, and anti-GBM antibodies in patient serum. Aids in the diagnosis of primary systemic small vessel vasculitis (Wegener's granulomatosis, Churg-Strauss syndrome, microscopic periarteritis, idiopathic crescentic glomerulonephritis) and Goodpasture's syndrome.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Submission Summary (Full Text)

{0} 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k053012 B. Purpose for Submission: New Device C. Measurand: Anti-Myeloperoxidase antibodies (MPO) Anti-Serine Proteinase 3 antibodies (PR3) Anti-Glomerular Basement Membrane antibodies (GBM) D. Type of Test: Semi-quantitative E. Applicant: Biomedical Diagnostics S.A. (bmd) F. Proprietary and Established Names: FIDIS™ VASCULITIS G. Regulatory Information: 1. Regulation section: 21 CFR 866.5660, Multiple autoantibodies immunological test system 2. Classification: II 3. Product codes: MOB, Test system, antineutrophil cytoplasmic antibodies (ANCA) MVJ, Devices, Measure, Antibodies to Glomerular Basement Membrane (GBM) 4. Panel: Immunology 82 H. Intended Use: 1. Intended use(s): The FIDIS™ VASCULITIS* test is a semi-quantitative homogenous fluorescent-based microparticles immunoassay using flow cytometry readings. The test system is used to detect in patient serum samples the presence of anti-neutrophil cytoplasm antibodies (ANCA) directed against Myeloperoxidase (MPO) and Serine Proteinase 3 (PR3); and anti-glomerular basement membrane (GBM) antibodies. The results of the FIDIS™ VASCULITIS* test are to be used in conjunction with the clinical findings and other laboratory tests to aid in the diagnosis of various primary systemic small vessel vasculitis. The presence of anti-MPO and anti-PR3 antibodies is associated with primary systemic small vessel vasculitis: Wegener's granulomatosis, Churg Strauss syndromes, microscopic periarteritis and idiopathic crescentic glomerulonephritis; and the presence of anti-GBM antibodies is associated with Goodpasture's syndrome. FIDIS™ VASCULITIS* kit is to be used on the FIDIS Analyser, MLX-BOOSTER Software and Washer. FIDIS™ VASCULITIS* kit could be used with the CARIST™ system (diluting and dispensing device). This test is for in vitro diagnostic use. *Detection of the serologic markers for primary systemic small vessel vasculitis (ANCA) and for Goodpasture's syndrome (GBM). 2. Indication(s) for use: {1} Same as intended use 3. Special conditions for use statement(s): The device is for prescription use only. 4. Special instrument requirements: FIDISTM 100 Analyzer (Luminex $100^{\mathrm{TM}}$ ) FIDISTM MLX-Booster Software (k050286) FIDISTM Washer (k041002) Ultrasonic bath # I. Device Description: Each device contains the following: distinct uniform size color-coded microspheres (each microsphere set is covalently coupled to one of the following antigens: MPO, PR3, and GBM) (ready to use); calibrator (ready to use); positive control (to be diluted); negative control (to be diluted); goat anti-human IgG coupled to phycoerythrin (ready to use); concentrated PBS-Tween (to be diluted with distilled water); one 96 wells microplate including a filtering membrane and a lid. # J. Substantial Equivalence Information: 1. Predicate device name(s): QuantaLiteTM MPO QuantaLiteTM PR3 QuantaLiteTM GBM 2. Predicate 510(k) number(s): k981330 (MPO) k981328 (PR3) k984336 (GBM) 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | New Device | Predicate Device | | | FIDISTM VASCULITIS | Individual INOVA ELISA Assays for autoantibodies to MPO, PR3, and GBM | | Intended use | Individual determination of IgG antibodies to MPO, PR3, GBM | Same | | Antigen | Purified MPO, PR3, GBM | Same | | Sample type | Serum | Same | | Type of test | Semi-quantitative | Same | | Platform | 96 well plates | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Technology | Flow Cytometry | ELISA | | Assay Format | Multiplexed | Individual analyzes | | Sample dilution | 1:200 | 1:101 | | Substrate | None | TMB | | Enzyme-Conjugate | Phycoerythrin | HRP | | Detection method | Fluorescence | Colorimetric | {2} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Solid surface (antigen coated) | Color-coded microspheres | Microwells | | Instrument | Luminometer (Luminex 100™) | Spectrophotometer | | Result | Negative: < 20 AU/mL | Negative: ≤ 20 Units | | Interpretation for each specificity: | Equivocal: 20-25 AU/mL | Weak Pos: 21-30 Units | | Anti-MPO, Anti-PR-3, Anti-GBM Antibody | Positive: >25 AU/mL | Positive: > 30 units | **K. Standard/Guidance Document Referenced (if applicable):** None provided. **L. Test Principle:** FIDISTM VASCULITIS is a multiplexed semi-quantitative, fluorescent immunoassay performed on the FIDISTM Analyzer (Luminex 100™) with the MLX Booster software. Each antigen (MPO, PR3, and GBM) is covalently coupled to an individual set of microspheres through its surface functional groups. The different sets of antigen-coupled microspheres are mixed together to constitute the final microspheres reagent and put into wells of a microtiter plate. Prediluted controls and diluted patient sera are added to separate wells allowing autoantibodies to bind to the immobilized antigens on the beads. After incubation, a wash step through a filtration process will remove the unbound antibodies. Then a phycoerythrin labeled anti-human IgG is added to each well and binds to patient autoantibodies/antigen complexes on the microspheres. The samples are subsequently measured in the FIDISTM Analyzer (flow cytometer) which discriminates the different bead sets as well as measures the fluorescent intensity of the conjugate on each bead. For each sample, the antibody titer for each antigenic specificity is interpolated against a calibration system. **M. Performance Characteristics (if/when applicable):** 1. Analytical performance: a. Precision/Reproducibility: Three samples (weak, moderate and high titers) were analyzed 10 times (for each specificity) in one run for the intra-assay study. The same three samples were analyzed 4 times in 6 different runs for the inter-assay study. The intra-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were: 1.7% to 5.3%; 2.8% to 10.9%; and 3.8% to 5.7% respectively. The inter-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were: 6.3% to 8.1%; 10.1% to 13.7%; and 7.1% to 9.1% respectively. In addition to the above studies, three samples with antibody concentrations close to the cut-off were assayed 10 times in the same run and had %CV of 4.5% to 9.5%; 4.2% to 6.6%; and 3.0 to 12.7% for Anti-MPO, Anti-PR3 and Anti-GBM respectively. For inter-assay reproducibility, three samples close to the cut-off were assayed 4 times in six different runs, had CV of 10.1% to 16.0%, 5.6% to 13.1%, and 4.3% to 11.4% for Anti-MPO, Anti-PR3 and Anti-GBM respectively (see table below). {3} | | Within-run (Tested10X in the same run) | | Between-run (Tested 4X in 6 different runs) | | | --- | --- | --- | --- | --- | | Antigen | Mean value | CV (%) | Mean value | CV (%) | | MPO | 28.1 | 9.5 | 29.7 | 10.1 | | | 30.7 | 7.4 | 32.0 | 16.0 | | | 34.7 | 4.5 | 33.2 | 11.3 | | | 42 | 1.7 | 39 | 6.3 | | | 171 | 5.3 | 162 | 8.1 | | | 546 | 4.0 | 507 | 7.4 | | PR3 | 29 | 5.4 | 27 | 13.7 | | | 29.1 | 6.6 | 28.4 | 10.3 | | | 41.3 | 4.2 | 45.5 | 5.6 | | | 101 | 2.8 | 89 | 11.0 | | | 178.5 | 5.8 | 175.1 | 13.1 | | | 688 | 10.9 | 652 | 10.1 | | GBM | 30.6 | 3.0 | 31.5 | 4.3 | | | 32.8 | 12.7 | 36.5 | 11.4 | | | 46.3 | 7.5 | 53.9 | 9.3 | | | 49 | 3.8 | 45 | 9.1 | | | 113 | 2.7 | 106 | 6.3 | | | 204 | 5.7 | 190 | 7.1 | The reproducibility performance of FIDISTM VASCULITIS with the CARIS system (diluting and dispensing device) was determined as follows: For the intra-assay study, three samples were analyzed 10 times in one run for each specificity. The same 3 samples were analyzed 4 times in 6 different runs for the inter-assay study. The intra-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were: $6.3\%$ to $8.8\%$ ; $4.7\%$ to $7.0\%$ ; and $4.1\%$ to $4.1\%$ respectively. The inter-assay CV ranges on Anti-MPO, Anti-PR3, and Anti-GBM were: $9.1\%$ to $11.8$ ; $3.9\%$ to $7.0\%$ ; and $4.3\%$ to $4.8\%$ respectively (see table below). | | Within-run (tested 10X in same run) | | Between-run (tested 4X in 6 different runs) | | | --- | --- | --- | --- | --- | | Antigen | Mean value | CV (%) | Mean value | CV (%) | | MPO | 49.5 | 6.3 | 43.8 | 9.1 | | | 54.6 | 8.8 | 64.9 | 11.8 | | PR3 | 26.6 | 7.0 | 33.1 | 7.9 | {4} 5 | | Within-run (tested 10X in same run) | | Between-run (tested 4X in 6 different runs) | | | --- | --- | --- | --- | --- | | | 203.1 | 4.7 | 204.7 | 3.9 | | GBM | 46.7 | 4.1 | 48.6 | 4.3 | | | 61.9 | 4.1 | 62.8 | 4.8 | Lot to lot reproducibility: Two lots were analyzed for reproducibility using 3 positive samples (2 near cut off) and 9 negative samples. The acceptable criterion is 25%. The lowest variability was 2% and highest variability was 15%. Both lots were within the 25% variability. b. Linearity/assay reportable range: Linearity is not claimed in this assay. c. Traceability, Stability, Expected values (controls, calibrators, or methods): There are no reference standard for anti-MPO, anti-PR3 and anti-GBM. The calibrators and controls (positive and negative) were prepared in-house and assigned arbitrary units/mL (AU/mL) during the development process. d. Detection limit: Not applicable. e. Analytical specificity: Interference study: Twenty eight samples were selected for evaluation of potential interference and crossreactivity: 3 Cryoglobulinemia, 8 Complement, 2 IgG monoclonal immunoglobulins, 3 IgM monoclonal immunoglobulins, 7 Rheumatoid Factor, 3 citrated plasma, and 2 hemolyzed samples. Twenty seven samples were negative and one RF sample was positive for PR3 (see table below). The package insert states to avoid hemolytic, lipemic, icteric samples or samples with abnormal concentration of IgG and/or complement levels or samples with rheumatoid factor. f. Assay cut-off: The cut-off value of &gt;25 AU/mL was based on testing 65 samples: 37 normal blood donor sera and 28 selected samples with potential biologic interferences. With this cut-off value, 37 normal donor and 28 selected samples (100%) were negative for MPO and GBM; and 37 normal donor and 27 selected samples (98.5%) were negative for PR3. 2. Comparison studies: a. Method comparison with predicate device: Testing was performed on 182 samples: 117 positive for one or more ANCA parameters or positive for GBM; and 65 negative samples. No information about age, gender, and clinical diagnosis was available. Equivocal results were considered negative. The positive, negative and total percent agreements for Anti-MPO, Anti-PR3, and Anti-GBM were shown in tables below. | Anti-MPO | QuantaLite™ MPO | | | | | --- | --- | --- | --- | --- | | | | Positive | Negative | Total | | FIDISTM VASCULITIS MPO | Positive | 31 | 3 | 34 | | | Negative | 2 | 146 | 148 | | | Total | 33 | 149 | 182 | Positive percent agreement: 93.9% (31/33) (95% CI: 85.8% to 100%) {5} 6 Negative percent agreement: 98% (146/149) (95% CI: 95.7% to 100%) Overall percent Agreement: 97.3% (177/182) (95% CI: 94.9% to 99.6%) | Anti-PR3 | QuantaLite™ PR3 | | | | | --- | --- | --- | --- | --- | | | | Positive | Negative | Total | | FIDISTM VASCULITIS PR3 | Positive | 40 | 3 | 43 | | | Negative | 7 | 132 | 139 | | | Total | 47 | 135 | 182 | Positive percent agreement: 85.1% (40/47) (95% CI: 74.9% to 95.3%) Negative percent agreement: 97.8% (132/135) (95% CI: 95.3% to 100%) Overall percent Agreement: 94.5% (172/182) (95% CI: 91.2% to 97.8%) | Anti-GBM | QuantaLite™ GBM | | | | | --- | --- | --- | --- | --- | | | | Positive | Negative | Total | | FIDISTM VASCULITIS GBM | Positive | 21 | 0 | 21 | | | Negative | 0 | 161 | 161 | | | Total | 21 | 161 | 182 | Positive percent agreement: 100% (21/21) Negative percent agreement: 100% (161/161) Overall percent Agreement: 100% (182/182) Comparison of manual preparation and the automated CARISTM System: Results of the comparison studies are summarized below. | Anti-MPO | Manual FIDIS | | | | --- | --- | --- | --- | | | | Positive | Negative | | CARISTM FIDIS | Positive | 5 | 0 | | | Negative | 0 | 26 | Positive percent agreement: 100% (5/5) Negative percent agreement: 100% (26/26) Overall agreement: 100% (31/31) | Anti-PR3 | Manual FIDIS | | | | --- | --- | --- | --- | | | | Positive | Negative | | CARISTM FIDIS | Positive | 7 | 0 | | | Negative | 0 | 24 | Positive percent agreement: 100% (7/7) Negative percent agreement: 100% (24/24) Overall agreement: 100% (31/31) | Anti-GBM | Manual FIDIS | | | | --- | --- | --- | --- | | | | Positive | Negative | | CARISTM FIDIS | Positive | 7 | 0 | | | Negative | 0 | 24 | Positive percent agreement: 100% (7/7) Negative percent agreement: 100% (24/24) Overall agreement: 100% (31/31) {6} Linear Regression analysis of antigenic specificities with CARISTM. ![img-0.jpeg](img-0.jpeg) ![img-1.jpeg](img-1.jpeg) ![img-2.jpeg](img-2.jpeg) b. Matrix comparison: Serum is the only recommended matrix. 3. Clinical studies: {7} a. Clinical sensitivity and specificity: Not applicable. b. Other clinical supportive data (when a. is not applicable): Not applicable. 4. Clinical cut-off: Same as assay cut-off. 5. Expected values/Reference range: Expected values in the normal population should be negative. Table of incidence on Diseases with PR3, MPO and GBM antibodies | Diseases | PR3 | MPO | GBM | Reference | | --- | --- | --- | --- | --- | | Wegener's granulomatosis | 30-90% | | | 14 | | | 50-90% | | | 15 | | Churg Strauss Syndrome | 35% | 35% | | 14 | | | 31% | 7% | | 15 | | Microscopic periarteritis | 20-40% | 50% | | 14 | | | 25-30% | 36% | | 15 | | Idiopathic glomerulonephritis | 20-40% | 50% | | 14 | | | | 80% | | 15 | | Goodpasture's syndrome or anti-GBM nephritis | | | 98% | 16 | (14) MOLLOY P. ANCA and Associated disease: Update. PSA Consult, 2000, vol III, 5. (15) SANCHEZ-LALLOYER N. Anti-neutrophil cytoplasmic antibodies. Spectra Biologie, 1993; 93/3: 38-42. (16) ROSSERT J. Goodpasture's disease. Orphanet encyclopedia, 2002: 1-4. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...