← Product Code [LNH](/productcode/LNH) · K053009

# GE 0.35T SIGNA OVATION WITH EXCITE MR SYSTEM (K053009)

_GE Medical Systems · LNH · Nov 17, 2005 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K053009

## Device Facts

- **Applicant:** GE Medical Systems
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Nov 17, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis. Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles

## Device Story

Open, whole-body 0.35T permanent magnet MRI system; acquires 2D single-slice, multi-slice, and 3D volume images; utilizes Fourier transformation techniques for image reconstruction. Operated by trained clinicians in clinical settings; supports interventional procedures via open design and MR-compatible accessories (e.g., biopsy needles, in-room displays). Input: proton density, T1, T2, and flow signals. Output: transverse, sagittal, coronal, and oblique anatomical images. Clinicians interpret images to assist in diagnosis. Modification from predicate includes motorized lateral table movement.

## Clinical Evidence

Bench testing only. Evaluated against NEMA performance standards and IEC 601-1 and IEC 601-2-33 safety standards for magnetic resonance equipment.

## Technological Characteristics

0.35T permanent magnet MRI system; supports 2D/3D Fourier transformation image reconstruction; motorized lateral table movement; complies with IEC 601-1 and IEC 601-2-33 standards.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- Signa® 0.35T® Ovation with Excite System ([K033504](/device/K033504.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K053009

#### GE Healthcare

General Electric Company P.O. Box 414. Milwaukee. WI 53201

NOV 1 7 2005

# 510(k) Summary

This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.92(c).

| Submitter:      | GE Healthcare<br>PO Box 414<br>Milwaukee, WI 53201             |
|-----------------|----------------------------------------------------------------|
| Contact Person: | Larry A. Kroger Ph.D.<br>Senior Regulatory Programs<br>Manager |
| Telephone:      | 262- 544-3894                                                  |
| Fax:            | 262- 548-4768                                                  |
| Date Prepared:  | October 25, 2005                                               |

### Device Name:

GE 0.35T Signa® Ovation with Excite Magnetic Resonance System Magnetic Resonance Diagnostic System, 21 CFR 892.1000, 90-LNH

#### Marketed Device:

The GE 0.35T Signa® Ovation with Excite MR System is substantially equivalent to the currently marketed Signa® 0.35T® Ovation with Excite System (K033504) with the main differences being the addition of a motorized control circuit for the lateral movement of the table.

## Device Description:

The 0.35T Signa® Ovation with Excite Magnetic Resonance System is a modification to the 0.35T Signa® MFO/i Magnetic Resonance System (K033504) which utilizes a permanent magnet to acquire 2D single-slice and multi-slice, and 3D volume images. The 0.35T Signa® Ovation with Excite Magnetic Resonance System features a permanent magnet operating at 0.35T. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences. Images are acquired and reconstructed using 2D and 3D Fourier transformation techniques. The system is intended for high-resolution anatomical applications, short scan times, and multinuclear spectroscopy.

#### Indications for Use:

The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined and surrounded by a circular pattern. The logo is black and white.

General Electric Comnania P.O. Box 414. Milwaukee. WI 53201

the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Due to the 'open' design of the 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles

#### Comparison with Predicate Device:

The GE 0.35T Signa® Ovation with Excite MR System is a modification of the Signa® Ovation with Excite system (K033504) with the main differences being the addition of motorized drive control for lateral table movement.

## Summary of Studies:

The 0.35T Signa® Ovation with Excite Magnetic Resonance System was evaluated to the appropriate NEMA performance standards as well as the IEC 601-1 International Medical Equipment Safety standard and IEC 601-2-33 Particular Requirements for Safety of Magnetic Resonance Equipment for Medical Diagnosis. The 0.35T Signa® Ovation with Excite Magnetic Resonance System is comparable to the currently marketed Signa® 0.35T Ovation Magnetic Resonance System (K033504).

#### Conclusion:

It is the opinion of GE that the 0.35T Signa® Ovation with Excite Magnetic Resonance System is substantially equivalent to the 0.35T Signa® Ovation with Excite Magnetic Resonance System. Usage of the 0.35T Signa® Ovation with Excite Magnetic Resonance System does not result in any new potential hazards.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows a black and white seal. The seal is circular and contains the symbol of the United States Department of Health and Human Services (HHS). The HHS symbol is an abstract representation of an eagle with three lines forming its body and wings. The text "HEALTH & HUMAN SERVICES - USA" is written around the perimeter of the circle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 7 2005

Larry A. Kroger, Ph.D. Senior Regulatory Programs Manager GE Healthcare W-400 P.O. Box 414 MILWAUKEE WI 53201

Re: K053009 Trade/Device Name: GE 0.35T Signa® Ovation with Excite MR System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: October 25, 2005 Received: October 26, 2005

Dear Dr. Kroger:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If vour device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act s requirements. including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820); and if applicable, the clectronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{3}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) I mis letter notification. The FDA finding of substantial equivalence of your device to a legally prematicated predicated on one of a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the 60 : 77 / - 1 St mall Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Brigdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular shape. The letters and the circle are in white, set against a black background, creating a high-contrast design. The logo is a well-known symbol associated with the General Electric company.

GE Healthcare

General Fleriric Compan P.O. Box 414, Mihraukee, W1 53201

# STATEMENT OF INTENDED USE

510(k) Number (if known):

510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________

# Device Name: GE 0.35T Signa® Ovation with Excite MR System

# Indications for Use

The 0.35T Signa® Ovation with Excite Magnetic Resonance System is an open, whole body scanner designed to support improved higher resolution imaging and shorter scan times. The Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System is indicated for use as a diagnostic imaging device to produce transverse, sagittal, coronal and oblique images of the internal structures and organs of the entire body, including, but not limited to, the musculocskeletal, vascular, cardiac, and neuro systems. The images produced by the Signa 0.35T Signa® Ovation with Excite Magnetic Resonance System reflect the spatial distribution of protons (hydrogen nuclei) exhibiting magnetic resonance. The NMR properties that determine the image appearance are proton density, spin-lattice relaxation time (T1), spin-spin relaxation time (T2) and flow. When interpreted by a trained physician, these images provide information that can be useful in determining a diagnosis.

Due to the 'open' design of the Signa 0.35T Signa® Ovation with Excite may also be used for imaging during interventional procedures when performed with MR compatible devices such as, in-room display and MR safe biopsy needles

Daniel A. Lupyan

(Division Sign-Off) Division of Reproductive, Abdomin and Radiological Devices 510(k) Number

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use × (Per 21 CFR 801-109)

OR

Over-The-Counter Use

E-2

---

**Source:** [https://fda.innolitics.com/device/K053009](https://fda.innolitics.com/device/K053009)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
