MEDIQ VERSION 1.0

K052963 · Clario Medical Imaging, Inc. · LLZ · Dec 5, 2005 · Radiology

Device Facts

Record IDK052963
Device NameMEDIQ VERSION 1.0
ApplicantClario Medical Imaging, Inc.
Product CodeLLZ · Radiology
Decision DateDec 5, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.2050
Device ClassClass 2
AttributesSoftware as a Medical Device

Indications for Use

MedIQ is a post-processing, productivity software package designed to assist radiologists in the analysis of dynamic CT and MR images. The software provides supplemental information and visual enhancement of time/intensity changes extracted from CT and MR temporal datasets. Single or multi-slice datasets, using standard acquisition protocols, are used for input. The software displays the temporal variation in dynamic data as a surface map that enhances visualization of pixel intensity and time/intensity changes extracted from CT and MR temporal data sets. The results may be displayed as either a grayscale enhancement or a color overlay on the selected image. MedIQ also may be used to perform post-processing analysis of multimodality, digital images, including MR, CT, X-Ray, PET and Nuclear Medicine images. The software analysis tools may be applied to image subtractions, reformatted images, multiplanar reformats, and maximum intensity projections. The software package includes tools to allow the radiologist to manipulate and fly-through images for enhanced visualization, and a z-axis kinematic motion feature. When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of MedIQ analysis. The MedIQ software is not intended for use for primary interpretation of digital mammography images.

Device Story

MedIQ Version 1.0 is a software package for radiologists to analyze multi-modality digital radiology images (CT, MR, X-Ray, PET, Nuclear Medicine). Input consists of single or multi-slice datasets. The software applies 2D/3D image processing techniques on a pixel-by-pixel basis to perform visual enhancement, including image subtractions, reformatted images, multiplanar reformats, and maximum intensity projections. It generates surface maps of temporal variations in dynamic data, displayed as grayscale enhancements or color overlays. The software includes tools for image manipulation, rotation, and fly-through visualization, plus a z-axis kinematic motion feature. Used in clinical settings (PACS, workstations, or stand-alone PCs), it assists physicians in screening and diagnosis by providing supplemental visual information. It is not intended for primary interpretation of digital mammography; patient management decisions must not rely solely on its output.

Clinical Evidence

No clinical data provided. Substantial equivalence is based on bench testing, risk analysis, and design verification/validation processes.

Technological Characteristics

Software-based image processing system. Compatible with DICOM standards. Operates on Windows 2000/XP platforms. Functions include 2D/3D image processing, pixel-by-pixel analysis, surface mapping, and kinematic motion features. Distributed as SDK or application for PACS, workstations, or stand-alone PCs.

Indications for Use

Indicated for radiologists to assist in the analysis of dynamic CT and MR images, and post-processing of multimodality digital images (MR, CT, X-Ray, PET, Nuclear Medicine). Not intended for primary interpretation of digital mammography.

Regulatory Classification

Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ K052963 #### DEC & 2005 # 510(k) Summary of Safety and Effectiveness This 510(k) summary of safety and effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87 (h). ## 1. Identification of Submitter: 7 | Submitter: | Clario Medical Imaging, Inc | |----------------|---------------------------------------------| | Address: | 725 N. State Street<br>Bellingham, WA 98225 | | Phone: | 360-312-5200 | | Fax: | 360-312-5201 | | Contact: | Patricia A. Milbank | | Title: | Regulatory Consultant | | Phone: | 425-894-9733 | | Fax: | 425-865-9023 | | Date Prepared: | October 15, 2005 | #### 2. Identification of Product: | Trade Name: | MedIQ Version 1.0 | |----------------------|---------------------------------------------| | Common Name: | Picture Archiving and Communications System | | Regulation Number: | 21 CFR 892.2050 | | Classification Name: | Image Processing System | | Product Code: | LLZ | | Manufacturer: | Clario Medical Imaging, Inc.<br>725 N. State Street<br>Bellingham, WA 98225 | |---------------|-----------------------------------------------------------------------------| |---------------|-----------------------------------------------------------------------------| ### 3. Marketed Devices MedIQ Version 1.0 is substantially equivalent to the legally marketed predicate devices listed below: | Model: | SharpView | |-----------------|-----------------------| | Manufacturer: | ContextVision | | 510 (k) Number: | K993082 | | Model: | Vitrea 2, Version 3.7 | | Manufacturer: | Vital Images | | 510 (k) Number: | K043333 | {2}------------------------------------------------ | Model: | FuncTool 2000, designed for Advantage Workstation | |-----------------|---------------------------------------------------| | Manufacturer: | GE Medical Systems | | 510 (k) Number: | K960265 | | Model: | MIStar | | Manufacturer: | Apollo Medical Imaging Technology Pty. Ltd. | | 510 (k) Number: | K043350 | | Model: | Syngo Colonography Software Package | | Manufacturer: | Siemens Medical Systems | | 510 (k) Number: | K030982 | | Model: | Realtime 3D Software Package | | Manufacturer: | Siemens Medical Systems | | 510 (k) Number: | K973010 | | Model: | Fly Through PACS Software | | Manufacturer: | Siemens Medical Systems | | 510 (k) Number: | K971717 | #### 4. Device Description: MedIO Version 1.0 is a software package designed to assist the radiologist in interpretation of multi-modality, digital radiology images, including dynamic CT and MR image data sets. The software applies standard 2D and 3D image processing techniques on a pixel-by-pixel basis to analyze grayscale data and perform visual enhancement of user-selected images. The software may be applied to image subtractions, reformatted images, multiplanar reformats, and maximum intensity projections. Single or multi-slice data sets may be used as input. The results of the analysis are displayed as a surface map that enhances visualization of pixel intensity data provided by the image. The results may be displayed in grayscale enhancement or as a color overlay. The software includes standard workflow tools that allow the radiologist to manipulate, rotate and fly-though images for enhanced visualization of the digital output. The software application consists of proprietary software developed by Clario and is a Windows 2000/XP, DICOM-compatible platform. The software is designed to be distributed as an SDK or as an application for installation on DICOM-compatible OEM imaging systems, PACS, workstations, stand-alone PCs, or embedded in software applications cleared for use in medical imaging. The device also may be distributed by Clario as a stand-alone PC product. {3}------------------------------------------------ The MedIQ user interface is designed to follow typical clinical workflow patterns to process, review, and analyze digital images. #### 5. Indications for Use MedIQ is a post-processing, productivity software package designed to assist radiologists in the analysis of dynamic CT and MR images. The software provides supplemental information and visual enhancement of time/intensity changes extracted from CT and MR temporal datasets. Single or multi-slice datasets, using standard acquisition protocols, are used for input. The software displays the temporal variation in dynamic data as a surface map that enhances visualization of pixel intensity and time/intensity changes extracted from CT and MR temporal data sets. The results may be displayed as either a grayscale enhancement or a color overlay on the selected image. MedIQ also may be used to perform post-processing analysis of multimodality, digital images, including MR, CT, X-Ray, PET and Nuclear Medicine images. The software analysis tools may be applied to image subtractions, reformatted images, multiplanar reformats, and maximum intensity projections. The software package includes tools to allow the radiologist to manipulate and fly-through images for enhanced visualization, and a z-axis kinematic motion feature. When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of MedIQ analysis. The MedIQ software is not intended for use for primary interpretation of digital mammography images. #### 6. Comparison with Predicate Devices MedIQ and the cited predicate devices are designed to assist radiologists in analyzing pixel intensity data from multi-modality, digital images and performing post-processing analysis of selected images using standard software techniques. The output of these software packages is intended to provide supplemental information and enhanced visual analysis of the digital images selected by the user for processing. {4}------------------------------------------------ The GE FuncTool 2000 and MIStar software products perform postprocessing analysis of dynamic CT and/or MR images to enhance the radiologist's visual interpretation of time/intensity changes observed in temporal data sets obtained during the administration of contrast agents. The output is displayed as curves or color parametric images from user-specified processing of algorithms. The Siemens software packages, including Syngo, Realtime 3D, and Fly Through, provide functional features that are equivalent to MedIQ, allowing the user to view and manipulate images in 3D in real time, to make measurements on images and to provide 3D surface shading of selected images. The cited predicate devices allow easy selection, review, processing, archiving, printing and media interchange of multi-modality images from a variety of diagnostic imaging systems. #### 7. Conclusions MedIQ Version 1.0 provides functionality that is substantially equivalent to the cited predicate software devices. The potential hazards have been studied and controlled as part of the product development process, including risk analysis, test and design considerations, and planned verification and validation testing processes. There are no new safety issues raised by this device. {5}------------------------------------------------ Image /page/4/Picture/1 description: The image shows a circular logo with a stylized bird-like symbol in the center. Three curved lines emanate from the bird, resembling wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES" is arranged around the circumference of the circle, following its curve. The logo appears to be a simplified representation of a government agency or organization related to health and human services. DEC 5 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Patricia A. Milbank Regulatory Consultant Clario Medical Imaging, Inc. 725 N. State Street BELLINGHAM WA 98225 #### Re: K052963 Trade/Device Name: MedIQ Version 1.0 Regulatory Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 18, 2005 Received: October 21, 2005 Dear Ms. Milbank: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. {6}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html. Sincercly yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ K05 2963 #### 6 # Indication(s) for Use Statement To be assigned by FDA 510(k) Number: MedIQ Version 1.0 Device Name: Indications for Use: MedIQ is a post-processing, productivity software package designed to assist radiologists in the analysis of dynamic CT and MR images. The software provides supplemental information and visual enhancement of time/intensity changes extracted from CT and MR temporal datasets. Single or multi-slice datasets, using standard acquisition protocols, are used for input. The software displays the temporal variation in dynamic data as a surface map that enhances visualization of pixel intensity and time/intensity changes extracted from CT and MR temporal data sets. The results may be displayed as either a grayscale enhancement or a color overlay on the selected image. MedIQ also may be used to perform post-processing analysis of multimodality, digital images, including MR, CT, X-Ray, PET and Nuclear Medicine images. The software analysis tools may be applied to image subtractions, reformatted images, multiplanar reformats, and maximum intensity projections. The software package includes tools to allow the radiologist to manipulate and fly-through images for enhanced visualization, and a z-axis kinematic motion feature. When interpreted by a skilled physician, this device provides information that may be useful in screening and diagnosis. Patient management decisions should not be made based solely on the results of MedIQ analysis. The MedIQ software is not intended for use for primary interpretation of digital mammography images. Prescription Use (Part 21 CFR 801 Subpart D) Image /page/6/Picture/12 description: The image shows the words "AND/OR" with the "AND" portion crossed out. The slash between the two words is also crossed out. This indicates a choice between the two options, with the selection leaning towards "OR" due to the cancellation of "AND". The image is simple and focuses on the logical operator. Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Daniel A. Segrestor vision Sinn-( )F vision of Renroduct d Radiological Devic Original 510(k) PreMarket No Clario Medical Imaging.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...