← Product Code [DRG](/productcode/DRG) · K052961

# THE PHILIPS MP20, MP30, MP40, MP50, MP60, MP70, AND MP90 INTELLIVUE PATIENT MONITORS, WITH 802.11 WIRELESS LAN (K052961)

_Philips Medical Systems · DRG · Nov 18, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K052961

## Device Facts

- **Applicant:** Philips Medical Systems
- **Product Code:** [DRG](/productcode/DRG.md)
- **Decision Date:** Nov 18, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.2910
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. The MP40 and MP50 are additionally intended for use in MP20, MP30, transport situations within healthcare environments. ST Segment monitoring is restricted to adult patients only. The transcutaneous gas measurement (tcp02 / tcpCQ2) is restricted to neonatal patients only.

## Device Story

IntelliVue MP20-MP90 patient monitors provide real-time physiological data acquisition, processing, and alarming. Device accepts inputs from various sensors (ECG, blood pressure, SpO2, gas analyzers, etc.) to monitor patient status. Modification adds 802.11 wireless LAN capability for data transmission. Used by healthcare professionals in clinical settings and during transport. Output displayed on integrated screen for clinician review; facilitates clinical decision-making through continuous monitoring and alarm generation. Benefits include improved patient oversight and mobility within healthcare facilities.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance, functionality, and reliability were established through system-level testing, performance testing, and safety testing derived from hazard analysis. Results demonstrated that the modified device meets all specifications and performance claims of the predicate devices.

## Technological Characteristics

Multi-parameter patient monitors (MP20-MP90) with integrated 802.11 wireless LAN adapter. Supports various physiological sensors (ECG, SpO2, NIBP, gas analysis). Connectivity via wireless network. Class II device. Software-based signal processing and alarming.

## Regulatory Identification

A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.

## Predicate Devices

- [K014159](/device/K014159.md)
- [K021778](/device/K021778.md)
- [K030038](/device/K030038.md)
- [K032858](/device/K032858.md)
- [K040304](/device/K040304.md)
- [K040183](/device/K040183.md)
- [K040357](/device/K040357.md)
- [K041235](/device/K041235.md)
- K41741
- [K042845](/device/K042845.md)
- [K050141](/device/K050141.md)
- [K050762](/device/K050762.md)
- [K051106](/device/K051106.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K052961

## NOV 1 8 2005

510 (k) Summary

This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21 C.F.R. §807.92.

1. The submitter of this premarket notification is: Tapan D. Shah Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1099 Tel: 978 659 7681 Fax: 978 685 5624 e-mail: tapan.shah@philips.com

This summary was prepared on October 14, 2005.

- 2. The names of the devices are the Philips MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors with 802.11 Wireless LAN Classification names are as follows:

| Device Panel                             | Classification | ProCode | Description                                                                 |
|------------------------------------------|----------------|---------|-----------------------------------------------------------------------------|
| Circulatory<br>System Devices<br>(12625) | §870.1025, II  | DSI     | Detector and alarm, arrhythmia                                              |
|                                          | §870.1025, II  | MLD     | Monitor, ST Segment with Alarm                                              |
|                                          | §870.1025, II  | MHX     | Monitor, Physiological, Patient<br>(with arrhythmia detection or<br>alarms) |
|                                          | §870.1100, II  | DSJ     | Alarm, Blood Pressure                                                       |
|                                          | §870.1110, II  | DSK     | Computer, Blood Pressure                                                    |
|                                          | §870.1130, II  | DXN     | System, Measurement, Blood-<br>Pressure, Non-Invasive                       |
|                                          | §870.1435, II  | DXG     | Computer, Diagnostic, Pre-<br>Programmed, Single-Function                   |
|                                          | §870.1915, II  | KRB     | Probe, Thermodilution                                                       |
|                                          | §870.2060, II  | DRQ     | Amplifier and Signal<br>Conditioner, Transducer Signal                      |
|                                          | §870.2300, II  | DRT     | Monitor, Cardiac (incl.<br>Cardiotachometer & Rate Alarm)                   |
|                                          | §870.2300, II  | MSX     | System, Network and<br>Communication, Physiological<br>Monitors             |
|                                          | §870.2340, II  | DPS     | Electrocardiograph                                                          |
|                                          | §870.2340, II  | MLC     | Monitor, ST Segment                                                         |
|                                          | §870.2350, II  | DRW     | Electrocardiograph, Lead<br>Switching Adapter                               |
|                                          | §870.2370, II  | KRC     | Tester, Electrode, Surface,<br>Electrocardiograph                           |
|                                          | §870.2450, II  | DXJ     | Display, Cathode-Ray Tube,<br>Medical                                       |
|                                          | §870.2600, I   | DRJ     | System, Signal Isolation                                                    |
|                                          | §870.2700, II  | DQA     | Oximeter                                                                    |
|                                          | §870.2770, II  | DSB     | Plethysmograph, Impedance                                                   |
|                                          | §870.2800, II  | DSH     | Recorder, Magnetic tape,<br>Medical                                         |
|                                          | §870.2810, I   | DSF     | Recorder, Paper Chart                                                       |
|                                          | §870.2850, II  | DRS     | Extravascular Blood Pressure<br>Transducer                                  |
|                                          | §870.2900, I   | DSA     | Cable, Transducer and<br>Electrode, incl. Patient<br>Connector              |

{1}------------------------------------------------

| Anesthesiology<br>and Respiratory<br>Therapy (12624)                          | \$870.2910, II | DRG | Transmitters and Receivers,<br>Physiological Signal,<br>Radiofrequency       |
|-------------------------------------------------------------------------------|----------------|-----|------------------------------------------------------------------------------|
|                                                                               | \$868.1400, II | CCK | Analyzer, Gas, Carbon Dioxide,<br>Gaseous-Phase                              |
|                                                                               | \$868.1500, II | CBQ | Analyzer, Gas, Enflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                                               | \$868.1500, II | NHO | Analyzer, Gas, Desflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                                               | \$868.1500, II | NHP | Analyzer, Gas, Sevoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)   |
|                                                                               | \$868.1500, II | NHQ | Analyzer, Gas, Isoflurane,<br>Gaseous-Phase (Anesthetic<br>Concentration)    |
|                                                                               | \$868.1620, II | CBS | Analyzer, Gas, Halothane,<br>Gaseous-Phase (Anesthetic<br>Concentration)     |
|                                                                               | \$868.1700, II | CBR | Analyzer, Gas, Nitrous Oxide,<br>Gaseous-Phase (Anesthetic<br>Concentration) |
|                                                                               | \$868.1720, II | CCL | Analyzer, Gas, Oxygen, Gaseous-<br>Phase                                     |
|                                                                               | \$868.2375, II | BZQ | Monitor, Breathing Frequency                                                 |
|                                                                               | \$868.2480, II | LKD | Monitor, Carbon Dioxide,<br>Cutaneous                                        |
|                                                                               | \$868.2500, II | KLK | Monitor, Oxygen, Cutaneous, for<br>Infant not under Gas Anesthesia           |
| General<br>Hospital and<br>Personal Use<br>(12520)<br>Neurological<br>(12513) | \$880.2910, II | FLL | Thermometer, Electronic,<br>Clinical                                         |
|                                                                               | \$882.1400, II | GWR | Electroencephalograph                                                        |

- 3. The modified devices are substantially equivalent to previously cleared Philips devices marketed pursuant to K014159, K021778, K030038, K032858, K040304, K040183 and K040357, K041235, K41741, K042845, K050141, K050762, and K051106.
- 4 . The modification is a change that creates, an optional, 802.11 wireless network adapter for the Philips MPxx IntelliVue patient monitors.
- 5. The modified devices have the same intended use as the legally marketed predicate devices. They are intended for the monitoring, recording, and alarming of multiple physiological parameters of adults, pediatrics, and neonates in healthcare environment and during transport within healthcare environments.
- The modified devices have the same technological characteristics 6. as the legally marketed predicate devices.
- 7. Verification, validation, and testing activities establish the performance, functionality, and reliability characteristics of the modified device with respect to the predicate. Testing involved system level tests, performance tests, and safety

{2}------------------------------------------------

testing from hazard analysis. Pass/Fail criteria were based on the specifications cleared for the predicate device and test results showed substantial equivalence. The results demonstrate that the Philips IntelliVue Patient Monitor meets all reliability requirements and performance claims.

{3}------------------------------------------------

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

NOV 1 8 2005

Philips Medical Systems c/o Mr. Tapan D. Shah Regulatory Affairs Engineer 3000 Minuteman Road Andover, MA 01810-1099

Re: K052961

Trade Name: MP20, MP30, MP40, MP50, MP60, MP70, and MP90 IntelliVue Patient Monitors with 802.11 Wireless LAN Regulation Number: 21 CFR 870.2910 Regulation Name: Radiofrequency Physiological Signal Transmitters and Receivers Regulatory Class: Class II (two) Product Code: DRG Dated: October 19, 2005 Received: October 21, 2005

Dear Mr. Shah:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Page 2 - Mr. Tapan D. Shah

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Bhimmataylor

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known):

The Philips MP20, MP30, MP40, MP50, MP60, MP70,and Device Name: MP90 IntelliVue Patient Monitors, with 802.11 Wireless LAN

Indications for Use: Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients. Intended for monitoring, recording and alarming of multiple physiological parameters of adults, pediatrics and neonates in healthcare environments. The MP40 and MP50 are additionally intended for use in MP20, MP30, transport situations within healthcare environments.

ST Segment monitoring is restricted to adult patients only.

The transcutaneous gas measurement (tcp02 / tcpCQ2) is restricted to neonatal patients only.

Prescription Use yes-AND/OR (Part 21 CFR 801 Subpart D)

Over-The-Counter Use No-(21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Bhimma

ાર

---

**Source:** [https://fda.innolitics.com/device/K052961](https://fda.innolitics.com/device/K052961)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
