← Product Code [MVN](/productcode/MVN) · K052832

# SAUFLON NEW DAY (METHAFILCON A) SOFT (HYDROPHILIC) DAILY DISPOSABLE CONTACT LENSES (K052832)

_Sauflon Pharmaceuticals, Ltd. · MVN · May 15, 2006 · Ophthalmic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K052832

## Device Facts

- **Applicant:** Sauflon Pharmaceuticals, Ltd.
- **Product Code:** [MVN](/productcode/MVN.md)
- **Decision Date:** May 15, 2006
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 886.5925
- **Device Class:** Class 2
- **Review Panel:** Ophthalmic

## Indications for Use

The Sauflon New Day® (methafilcon A) Soft (hydrophilic) Visibility tinted contact lens is a daily disposable lens indicated for the correction of the refractive ametropia (myopia and hyperopia) and astigmatism in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Dioptres that does not interfere with visual acuity.

## Device Story

Hemispherical flexible soft contact lens; composed of methafilcon A (hydrophilic polymer of HEMA and methacrylic acid cross-linked with ethyleneglycol dimethacrylate); 55% water content; aqua visibility tint; includes UV-absorbing monomer. Designed for daily disposable wear; eliminates need for cleaning/care regimes. Used by patients for vision correction; prescribed by eye care professionals. Lens covers cornea and portion of adjacent sclera. Benefits include convenience and reduced risk of contact lens-related problems due to limited daily exposure time.

## Clinical Evidence

Bench testing only. Toxicology testing (cytotoxicity, systemic toxicity, ocular irritation) confirmed non-toxicity. Leachability studies (residual monomers and color additives) showed no significant levels of leachable substances. Physical, optical, and chemical properties compared to predicate and found equivalent.

## Technological Characteristics

Material: Methafilcon A (HEMA/methacrylic acid copolymer); 55% water content. Dk: 22.0 x 10^-11 (cm2/sec)(ml O2/ml x mmHg) at 35°C. Refractive index: 1.40. Aspheric design. UV-blocking monomer. Sterilization: Steam (autoclave). Packaging: Polypropylene blister with aluminum foil. Connectivity: None.

## Regulatory Identification

A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.

## Predicate Devices

- SAUFLON 55UV ([K013649](/device/K013649.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

MIY 15 2005

K052832

# 510(K) SUMMARY AS REQUIRED BY SECTION 807.92(c)

#### SUBMITTER INFORMATION:

Sauflon Pharmaceuticals Ltd. Company Name:

49 - 53 York Street Address: Twickenham Middlesex TW1 3LP 020 8322 4231 Phone:

020 8891 2833 Fax:

Ms Tanya Fair Contact Person:

DATE SUMMARY PREPARED: 22nd September 2005

DATE SUMMARY REVISED: 11th May 2006

DEVICE NAME:

(methafilcon Soft Day A) SAUFLON New Trade Name: (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens Contact Lens Common Name: CLASS II (21 CFR 886.5925) CODE - MVN;LPL Classification: SOFT (HYDROPHILIC) CONTACT LENS

### SUBSTANTIAL EQUIVALENCE:

SAUFLON New Day® (methafilcon A) Soft (hydrophilic) Visibility Tinted Daily Disposable Contact Lens are substantially equivalent to SAUFLON 55UV (methafilcon A) Soft (hydrophilic) Visibility Tinted Contact Lenses for Daily Wear that received market clearance pursuant to K013649 and are currently marketed in the USA.

#### DESCRIPTION of the DEVICE:

.

The SAUFLON New Day® (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lenses are available in an aqua visibility tint. The lens material (methafilcon A) is a hydrophilic polymer of 2-hydroxyethyl methacrylate

{1}------------------------------------------------

(HEMA) and methacrylic acid, which is cross-linked with ethyleneglycol dimethacrylate. When hydrated the lens consists of 45% HEMA and 55% water by weight of saline immersed in normal saline. The lens is visibility tinted aqua with weight of summe inimelod in his meller on the well and one UV absorbing monomer is used to block UV radiation. Sauflon New Day lenses are daily disposable lenses.

The average transmittance characteristics are less than 10% in the UVB range of 280 to 315nm and less than 40% in the UVA range of 315-380nm.

The SAUFLON New Day Daily Disposable Lens is a hemispherical flexible shell, which covers the cornea and a portion of the adjacent sclera, with the following dimensions:

| Chord Diameter:                                        | 14.0mm to 15.0mm                                                                                                                             |
|--------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|
| Centre Thickness @ -3.00DS:                            | 0.07mm                                                                                                                                       |
| Base Curve:                                            | 8.70mm                                                                                                                                       |
| Powers:                                                | -20.00 to +20.00<br>Presently available:<br>-0.50 to -6.00 (0.25DS steps)<br>-6.00 to -10.00 (0.50DS steps)<br>+0.50 to +6.00 (0.25DS steps) |
| The physical/optical properties of the lens are:       |                                                                                                                                              |
| Refractive Index:                                      | 1.40 +/- 0.32                                                                                                                                |
| %Transmittance @ 590nm                                 | 94.61                                                                                                                                        |
| % Transmittance @ 280-315nm                            | 9.41                                                                                                                                         |
| % Transmittance @ 316-380nm                            | 36.00                                                                                                                                        |
| Surface Character:                                     | Hydrophilic                                                                                                                                  |
| Water Content:                                         | 55%                                                                                                                                          |
| Specific Gravity                                       | 1.09                                                                                                                                         |
| Oxygen Permeability (Dk):                              | $22.0 x 10^{-11} (cm^2 / sec)$                                                                                                               |
| (Fatt Method for determination of oxygen permeability) | (ml O2 / ml x mm Hg) at 35°C                                                                                                                 |

{2}------------------------------------------------

The table overleaf Table 7.1 Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft (Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft Disposable + Inted Contact Lens for Daily Wear shows the rationale for (11) and the Saufloon the predicate device Sauflon 55UV and the Sauflon New Day Lens.

This substantial equivalence is based on the lens having the same indications for use, being equivalent in terms of performance and sharing the same chemical composition while utilising the same manufacturing method, sterilisation and packaging processes. Wheelers material Methafilcon A utilised by both predicate and test device was shown to be non-toxic in all cytotoxicity, systemic toxicity and ocular irritation tests as well as no residual monomers or dyes being present during leachability studies. as no reciente device Sauflon 55UV has an established shelf-life of 5 years based on both accelerated and real time stability studies presented in K013649 The rationale for these studies being applicable to the Sauflon New Day is that the same materials and processes are used for manufacture and packaging and there is no reason to believe that this lens cannot exceed the already established shelf-life. The lens do vary in design in that the Sauflon 55UV lens is of spherical design while the Sauflon New Day lens is aspheric. This design alteration does not affect the safety or the efficacy of the lens. The design engineering drawing for the aspheric design is included in Section 8 of this submission.

Due to the varying modality of the predicate i.e. a daily wear lens and subject lens i.e. a daily disposable lens the labelling of the package insert does vary in that the package insert for a daily disposable is not required to contain as much detail as a daily wear lens requires. Items such as a cleaning and care regime for the lenses are not required as the New Day lens is disposed of following use.

The package insert does however contain all necessary information for the safe and effective use of a daily disposable lens including lens handling, insertion, removal and precautions.

The final draft labelling for the Sauflon New Day lens blister pack, carton, patient instructions and professional fitting and information guide are included in Section 5 of this submission. The patient instructions for the predicate device are included in Section 6 of this submission.

{3}------------------------------------------------

| Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft<br>(Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) Visibility<br>Tinted Contact Lens for Daily Wear. |                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                 |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                                                                                                                                                                                                                          | PREDICATE DEVICE - SAUFLON 55UV (K013649)                                                                                                                                                                                                                         | SUBJECT DEVICE - SAUFLON NEW DAY                                                                                                                                                                                                                                |
| LENS DESCRIPTION                                                                                                                                                                                                                                                                         | (Methafilcon A) Soft (Hydrophilic) Visibility Tinted<br>Contact Lens                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                                       |
| INDICATIONS FOR USE                                                                                                                                                                                                                                                                      | Myopia, Hyperopia and Astigmatism                                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                       |
| MODALITY                                                                                                                                                                                                                                                                                 | Daily Wear                                                                                                                                                                                                                                                        | Daily Disposable wear                                                                                                                                                                                                                                           |
| LABELLING                                                                                                                                                                                                                                                                                | Follows the recommendations given in the guidance<br>document for Daily Wear Soft Contact lenses issued by<br>FDA in May 1994.                                                                                                                                    | Due to the fact that these are daily disposable lenses and not<br>daily wear lenses the labelling has been tailored for this and<br>areas such as care regime or replacement programmes of the<br>lenses are not relevant and therefore have not been included. |
|                                                                                                                                                                                                                                                                                          | Labelling has been provided for lens blister, carton, Patient<br>Instructions and Professional Fitting and Information guide<br>suitable for a daily wear lens.                                                                                                   | Labelling has been provided for lens blister, carton, Patient<br>Instructions and Professional Fitting and Information guide<br>suitable for a daily disposable lens.                                                                                           |
| LENS MATERIAL                                                                                                                                                                                                                                                                            | Methafilcon A                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                                       |
| WATER CONTENT                                                                                                                                                                                                                                                                            | 55%                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                                       |
| %TRANSMITTANCE @590NM                                                                                                                                                                                                                                                                    | 94.61%                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                       |
| %TRANSMITTANCE @280-<br>315NM                                                                                                                                                                                                                                                            | 9.41%                                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                       |
| % TRANSMITTANCE @316<br>80NM                                                                                                                                                                                                                                                             | 36.00%                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                       |
| DK @35°C (EDGE<br>CORRECTED)                                                                                                                                                                                                                                                             | 22.0 x 10-11                                                                                                                                                                                                                                                      | Identical                                                                                                                                                                                                                                                       |
| POWERS                                                                                                                                                                                                                                                                                   | -20.00 to +20.00 D                                                                                                                                                                                                                                                | Identical                                                                                                                                                                                                                                                       |
| Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft<br>(Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) Visibility<br>Tinted Contact Lens for Daily Wear. |                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                          | PREDICATE DEVICE - SAUFLON 55UV (K013649)                                                                                                                                                                                                                         | SUBJECT DEVICE -SAUFLON NEW DAY                                                                                                                                                                                                                                 |
| COLOUR                                                                                                                                                                                                                                                                                   | Aquamarine Visibility                                                                                                                                                                                                                                             | Identical                                                                                                                                                                                                                                                       |
| REFRACTIVE INDEX                                                                                                                                                                                                                                                                         | 1.4020                                                                                                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                       |
| TENSILE STRENGTH                                                                                                                                                                                                                                                                         | 1.47                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                                       |
| MODULUS                                                                                                                                                                                                                                                                                  | 0.52                                                                                                                                                                                                                                                              | identical                                                                                                                                                                                                                                                       |
| ELONGATION AT BREAK                                                                                                                                                                                                                                                                      | 280                                                                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                                       |
| TOUGHNESS                                                                                                                                                                                                                                                                                | 1.39                                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                                       |
| LENS DESIGN                                                                                                                                                                                                                                                                              | Spherical                                                                                                                                                                                                                                                         | Aspherical                                                                                                                                                                                                                                                      |
| CHEMICAL COMPOSITION<br>OF FINISHED LENS                                                                                                                                                                                                                                                 | Methafilcon A which is a copolymer of 2-<br>hydroxyethylomethacrylate and methacrylic acid cross<br>linked with ethyleneglycol dimethacrylate.<br>with benzophenone type UV absorbing monomer, 2-(4-benzoyl-3-hydroxyphenoxy) ethyl acrylate.                     | Identical                                                                                                                                                                                                                                                       |
| RAWMATERIAL<br>SPECIFICATIONS                                                                                                                                                                                                                                                            | Copies of all the raw material specifications were included<br>and approved in K013649                                                                                                                                                                            | Identical                                                                                                                                                                                                                                                       |
| MANUFACTURING PROCESS                                                                                                                                                                                                                                                                    | Cast Moulding                                                                                                                                                                                                                                                     | Identical                                                                                                                                                                                                                                                       |
| Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft<br>(Hydrophilic) Daily Disposable Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) Visibility<br>Tinted Contact Lens for Daily Wear. |                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                          | PREDICATE DEVICE - SAUFLON 55UV (K013649)                                                                                                                                                                                                                         | SUBJECT DEVICE - SAUFLON NEW DAY                                                                                                                                                                                                                                |
| TINTING PROCESS                                                                                                                                                                                                                                                                          | 'In monomer' tinting                                                                                                                                                                                                                                              | Identical                                                                                                                                                                                                                                                       |
| TINT USED                                                                                                                                                                                                                                                                                | CI Reactive Blue Dye 4<br>CI Reactive Yellow Dye 86                                                                                                                                                                                                               | Identical                                                                                                                                                                                                                                                       |
| PACKAGING MATERIALS                                                                                                                                                                                                                                                                      | Injected molded polypropylene blisters covered by<br>aluminium foil laminate and blister strips are packed into<br>printed cartons                                                                                                                                | Identical                                                                                                                                                                                                                                                       |
| LENS PACKING SOLUTION                                                                                                                                                                                                                                                                    | 0.9%w/v sodium chloride solution buffered at pH 7.5 to<br>7.8 with sodium hydrogen carbonate and containing<br>0.005% w/v poloxamer 407                                                                                                                           | Identical                                                                                                                                                                                                                                                       |
| PACKAGING METHOD                                                                                                                                                                                                                                                                         | Hermetically sealed blister pack                                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                       |
| STABILITY TESTS                                                                                                                                                                                                                                                                          | Real time studies (5 years at 20°C) and accelerated stability<br>studies (45°C) carried out.<br>All parameters including sterility remained within<br>specification.                                                                                              | These studies are applicable to the New Day lens.                                                                                                                                                                                                               |
| LEACHABILITY<br>STUDIES-RESIDUAL<br>MONOMERS                                                                                                                                                                                                                                             | Residual Monomers were analysed by HPLC. The levels of<br>monomers were below detection limits                                                                                                                                                                    | As the material used in the two lenses i.e. Methafilcon A and<br>the composition of the lenses are identical this leachability<br>data is therefore applicable.                                                                                                 |
|                                                                                                                                                                                                                                                                                          | Table 7.1. Comparison of Physical / Optical Properties and Processes for the SAUFLON NEW DAY (methafilcon A) Soft (Hydrophilic) Visibility Tinted contact lens VS. Sauflon 55UV (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens for Daily Wear. |                                                                                                                                                                                                                                                                 |
|                                                                                                                                                                                                                                                                                          | PREDICATE DEVICE - SAUFLON 55UV (K013649)                                                                                                                                                                                                                         | SUBJECT DEVICE - SAUFLON NEW DAY                                                                                                                                                                                                                                |
| LEACHABILITY STUDIES -<br>COLOUR ADDITIVES                                                                                                                                                                                                                                               | No detectable amounts of the dyes were found in the extract tested.                                                                                                                                                                                               | As the dyes used in the New Day lens is identical to those used for Sauflon 55UV lens the leachability data is therefore applicable.                                                                                                                            |
| STERILISATION METHOD                                                                                                                                                                                                                                                                     | Sterilised in an autoclave by Steam Sterilisation                                                                                                                                                                                                                 | Identical                                                                                                                                                                                                                                                       |
| SHELF-LIFE                                                                                                                                                                                                                                                                               | Shelf-life is 5 years based on stability studies                                                                                                                                                                                                                  | Identical                                                                                                                                                                                                                                                       |
| TOXIOCOLOGY OF THE LENS                                                                                                                                                                                                                                                                  | The lens material Methafilcon A was shown to be non-toxic in all cytotoxicity, systemic toxicity and ocular irritation tests.                                                                                                                                     | The identical lens material is utilised therefore the toxicity data is applicable.                                                                                                                                                                              |
| TOXICOLOGY OF THE<br>BLISTER PACKS                                                                                                                                                                                                                                                       | The moulded polypropylene blister material was shown to be non-toxic in all cytotoxicity, systemic toxicity and ocular irritation tests.                                                                                                                          | The identical materials are utilised in the packaging of the Sauflon New Day lens therefore the toxicity data is applicable.                                                                                                                                    |

{4}------------------------------------------------

.

、

:

.

·

{5}------------------------------------------------

and the comments of the comments of

,

{6}------------------------------------------------

:

,

{7}------------------------------------------------

### PRECLINICAL TESTING

The results of toxicology testing (cytotoxicity, acute systemic toxicity and acute ocular irritation) show the lens material methafilcon A to be non-toxic and nonirritating. Furthermore, the results of residual monomer and colour leachability testing demonstrate that the respective extracts did not contain significant levels of leachable colour or residual monomers.

The physical optical, and chemical properties of the SAUFLON New Day® Daily Disposable Lenses (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens are equivalent to those of Sauflon 55 UV (methafilcon A) Soft Hydrophilic Visibility Tinted Contact Lens for Daily Wear. This lens is in group 4, Ionic, high water content polymers as established by the FDA and located in the Guidance Document for Daily Wear Contact Lenses, Revised Edition May 1994.

The lens will be sterilised and packaged in the same manner as previously cleared in K013649. This lens will also be sterility released by parametric release, as cleared in K013649.

### INDICATIONS FOR USE

The SAUFLON New Day® Daily Disposable Lenses (methafilcon A) Soft (Hydrophilic) Visibility Tinted Contact Lens is a daily disposable lens indicated for the correction of the refractive ametropia (myopia and hyperopia) and astigmatism in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Dioptres that does not interfere with visual acuity.

The predicate device Sauflon 55UV lens is prescribed for daily wear however the Sauflon New Day is a Daily Disposable lens.

The New Day lens is a convenient alternative to a daily wear lens as the patient can dispose of the lens at the end of the day and then insert a brand new pair the next morning.

This eliminates the need for the patient to carry out a cleaning regime for the lens and the risk of experiencing contact lens related problems is reduced due to the limited exposure time of using one particular set of lenses.

{8}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/8/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes representing the agency's mission to promote health, well-being, and human services. The eagle is encircled by the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAY 15 2006

Sauflon Pharmaceuticals, Ltd. c/o Ms. Tanya Fair 49-53 York Street Twickenham, Middlesex TW1 3LP, United Kingdom

Re: K052832

. K022032
Trade/Device Name: Sauflon New Day® (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens

Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) Contact Lenses Regulatory Class: II Product Code: MVN; LPL Dated: May 4, 2006 Received: May 4, 2006

Dear Ms. Fair:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for association to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{9}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (301) 827-8910. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

MB Egerthuis - MS

Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{10}------------------------------------------------

# INDICATIONS FOR USE STATEMENT

510(k) Number (if known): K052832

Device Name:

SAUFLON New Day (methafilcon A) Soft (Hydrophilic) Visibility Tinted Daily Disposable Contact Lens

Indications For Use:

The Sauflon New Day® (methafilcon A) Soft (hydrophilic) Visibility tinted contact lens is a daily disposable lens indicated for the correction of the refractive ametropia (myopia and hyperopia) and astigmatism in aphakic and not-aphakic persons with non-diseased eyes that may exhibit astigmatism up to 2.00 Dioptres that does not interfere with visual acuity.

## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of the CDRH, Office of Device Evaluation (ODE)

| Prescription Use<br>(Per 21 CFR 801.109) | <div> <span style="text-decoration: overline;">✓</span> </div>                | OR | Over-The Counter |
|------------------------------------------|-------------------------------------------------------------------------------|----|------------------|
|                                          | Ming-chner Shi                                                                |    |                  |
|                                          | (Division Sign-Off)<br>Division of Ophthalmic Ear,<br>Nose and Throat Devises |    |                  |
| 510(k) Number                            | K052832                                                                       |    |                  |

---

**Source:** [https://fda.innolitics.com/device/K052832](https://fda.innolitics.com/device/K052832)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
