← Product Code [ITI](/productcode/ITI) · K052706

# A-200 POWERED WHEELCHAIR (K052706)

_Otto Bock Healthcare LP · ITI · Dec 27, 2005 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K052706

## Device Facts

- **Applicant:** Otto Bock Healthcare LP
- **Product Code:** [ITI](/productcode/ITI.md)
- **Decision Date:** Dec 27, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.3860
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The A-200 is a rear wheel drive powered wheelchair with caster front wheels for indoor use (primarily) but can be used outdoors as well. These wheclchairs provide mobility to physically challenged persons. The wheelchair can be moved by the user opcrating the Penny and Giles (P&G) VSI Electric Wheelchair Control System that is connected to the Micro Motor. The wheelchair is steered by different rotation of the rear wheels.

## Device Story

A-200 Powered Wheelchair is a rear-wheel-drive mobility device for indoor/outdoor use. Input: user interaction via Penny and Giles (P&G) VSI Electric Wheelchair Control System. Transformation: controller signals drive Micro Motor; steering achieved via differential rotation of rear wheels; electronic regenerative disc brakes provide deceleration. Output: physical mobility for user. Operated by patient; used in home/community settings. Benefits: provides independent mobility for physically challenged individuals.

## Clinical Evidence

Bench testing only. Device tested to EN 12184, ISO 7176 series, and ANSI/RESNA WA Vol. 2 Section 21 (EMC) standards. Results confirmed the test sample fulfilled all requirements.

## Technological Characteristics

Rear-wheel-drive powered wheelchair with closed 'U' frame. Features P&G VSI Electric Wheelchair Control System, Micro Motor, and electronic regenerative disc brakes. Tested to EN 12184, ISO 7176, and ANSI/RESNA standards. No software algorithm class specified.

## Regulatory Identification

A powered wheelchair is a battery-operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.

## Predicate Devices

- Evantgarde/Endeavor/A-100 ([K000739](/device/K000739.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Otto Bock®

### QUALITY FOR LIFE

DEC 2 7 2005

# 510(k) SUMMARY of SAFETY and EFFECTIVENESS

## A. General Information

3. Common Name:

5. Product Code:

B.

4.

| 1. Submitter's Name:    | OTTO BOCK HealthCare LP                                         |
|-------------------------|-----------------------------------------------------------------|
| 2. Address:             | Two Carlson Parkway N., Suite 100<br>Minneapolis, MN 55447-4467 |
| 3. Telephone:           | 763-489-5105                                                    |
| 4. Contact Person:      | Bob Clarke                                                      |
| 5. Date Prepared:       | August 15, 2005                                                 |
| 6. Registration Number: | 2182293                                                         |
| Device                  |                                                                 |
| 1. Name:                | A-200 Powered Wheelchair                                        |
| 2. Trade Name:          | A-200 Powered Wheelchair                                        |

Powered wheelchair

 Classification Name: Powered wheelchair

ן דיין

6. Class: II

7. Regulation Number: 890.3860

{1}------------------------------------------------

Otto Bock®

### QUALITY FOR LIFE

## C. Identification of Legally Marketed Devices

| 1. Name:         | Evantgarde/Endeavor/A-100 |
|------------------|---------------------------|
| 2. K Number:     | K000739                   |
| 3. Date Cleared: | October 24, 2000          |

### D. Description of the Device

The A-200 Powered Wheelchair is a rear wheel drive powered wheelchair, manufactured in Germany at production facilities of OTTO BOCK Health Carc. The A-200 has a closed "U" frame, controlled by a P&G Controller, electronic regenerative disc brakes and Micro Motor.

#### E. Intended Use Statement

The A-200 is a rear wheel drive powered wheelchair with caster front wheels for indoor use (primarily) but can be used outdoors as well. These wheclchairs provide mobility to physically challenged persons. The wheelchair can be moved by the user opcrating the Penny and Giles (P&G) VSI Electric Wheelchair Control System that is connected to the Micro Motor. The wheelchair is steered by different rotation of the rear wheels.

## F. Technological Characteristics Summary

The A-200 Wheelchair is substantially equivalent to the OTTO BOCK Evantgarde/Endeavor/A-100 Wheelchair, cleared on October 24, 2000 as K000739.

Each wheelchair is a powered wheelchair for the active user, with a rigid frame and similar characteristics.

The A-200 was tested by TÜV Product Service to the following standards:

- . EN 12184
- . ISO 7176 - Series
- ANSI/RESNA WA Vol. 2 Section 21 Amendments 1998 for EMC .

with the conclusion that "the test sample fulfills the requirements."

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known): To be determined

Device Name: A-200 Powered Wheelchair

Indications for Use:

Children State State Case Case Cases Canada Caracter Comparentation Comments of Children Comments of Children Comments of Children

- Provide mobility to persons physically challenged and limited to sitting positions. .
Prescription Use (Part 21 CFR 801 Subpart D)

-

AND/OR

Over-The-Counter Use _ X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Page I of I

V

V

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 7 2005

Mr. Bob Clarke Otto Bock HealthCare LP Two Carlson Parkway N., Suite 100 Minneapolis, Minnesota 55447-4467

Re: K052706/S1

Trade/Device Name: A-200 Powered Wheelchair Regulation Number: 21 CFR 890.3860 Regulation Name: Powered Wheelchair Regulatory Class: II Product Code: ITI Dated: November 11, 2005 Received: November 15, 2005

Dear Mr. Clarke:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your booken v r (c) f the device is substantially equivalent (for the indications for referenced above and have to legally marketed predicate devices marketed in interstate commerce use suited in the enactment date of the Medical Device Amendments, or to devices that prior to way 20, 1970, and cordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general therefore, manier the act include requirements for annual registration, listing of devices, good controls provisions of thabeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be may be subject to sam additions, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that I lease of advisou mar 271 i onethat your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all r carates statutes and registerial of the limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set fabering (21 CFR (21 600) 3 300) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2 - Mr. Clarke

This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to obgin hanteeing , " .
premarket notification. The FDA finding of substantial equivalence of your device to a legally prematication: "The FDATManig sification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrise of Compliance at (240) 276-0120. Also, please note the regulation entitled, Coliact the Office of Octification" (21CFR Part 807.97). You may obtain other IMISolanding by responsibilities under the Act from the Division of Small Manufacturers, general information on your tespension its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html

Sincerely yours,

Barbaco Buehm

Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### Indications for Use

510(k) Number (if known): To be determined

Device Name: A-200 Powered Wheelchair

Indications for Use:

- Provide mobility to persons physically challenged and limited to sitting positions. .
Prescription Use (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH Office of Device Evaluation (ODE)

Barbara BuemMD & MKM
(Division Sign Off)

Division of General, Restorative, and Neurological Devices i

Page I of I

**510(k) Number** k052706/Si

.

---

**Source:** [https://fda.innolitics.com/device/K052706](https://fda.innolitics.com/device/K052706)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
