FUSION CURING LIGHT

K052593 · Dentlight, Inc. · EBZ · Sep 28, 2005 · Dental

Device Facts

Record IDK052593
Device NameFUSION CURING LIGHT
ApplicantDentlight, Inc.
Product CodeEBZ · Dental
Decision DateSep 28, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6070
Device ClassClass 2
AttributesTherapeutic, 3rd-Party Reviewed

Intended Use

The FUSION Curing Light is a dental curing light that is intended for photopolymerization of dental resins, restorative composite materials, and orthodontic brackets, bonding, and sealing that are photo-polymerized in the 440 – 480 nm wavelength range.

Device Story

FUSION Curing Light is a battery-powered, cordless dental device for photopolymerization of VLC materials. Input: electrical energy from Li-Ion battery. Operation: blue LED technology emits light in 440-480nm range; user selects curing programs via push buttons to control intensity and exposure duration. Output: focused light beam delivered via probe tip. Used in dental clinics by dentists/staff. Benefits: faster curing of resins and activation of bleaching agents; increased curing distance provides operational flexibility; ergonomic size/weight improves handling. Output used by clinician to harden restorative materials or activate whitening agents.

Clinical Evidence

Bench testing only. Testing included optical power, resin curing time, curing range, optical wavelength, and software/hardware verification and validation. Compliance with IEC 60601 and 60601-1-2 for electrical safety and electromagnetic compatibility.

Technological Characteristics

Cordless, battery-powered handpiece with single LED probe tip. Wavelength: 440-480nm. Power: Li-Ion rechargeable battery. Base unit for charging/stand. Includes eye shield, disposable caps, and disinfectant barrier sleeves. Software-controlled curing programs. Complies with IEC 60601 and 60601-1-2.

Indications for Use

Indicated for photopolymerization of dental resins, restorative composite materials, and orthodontic brackets, bonding, and sealing sensitive to 440-480nm light in dental patients.

Regulatory Classification

Identification

An ultraviolet activator for polymerization is a device that produces ultraviolet radiation intended to polymerize (set) resinous dental pit and fissure sealants or restorative materials by transmission of light through a rod.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K052593 ### 510(k) Summary for FUSION Curing Light #### 1. SPONSOR DentLight Inc. 4404 Breckinridge Blvd. Suite 500 Richardson, TX 75082 Contact Person: Richard Liu Tel: 972-889-8857 Date Prepared: September 9, 2005 ### 2. DEVICE NAME Proprietary Name: FUSION curing light Common/Usual Name: dental curing light Classification Name: ultraviolet activator for polymerization (872.6070) #### 3. PREDICATE DEVICES Dentsply SmartLite PS Pen-Style LED curing light (K041372) Southern Dental Industries Radii (K030568) #### 4. DEVICE DESCRIPTION The DentLight's FUSION Curing Light is a battery-powered cordless unit designed for curing VLC materials whose initiator systems are sensitive to light in the 440-480nm wavelength range of the visible spectrum. The unit is based on blue LED (light emitting diode) technology for light generation in the desired wavelength and Li-Ion rechargeable batteries. The FUSION Curing Light includes: - A handpiece with a focusing lens focusing light emitted from a single LED probe tip and controlled by built-in control electronics - · A detachable rechargeable battery unit - · A base unit that plugs into main power as the handpiece stand - · An eye shield and disposable cap at probe tip - · Disposable disinfectant barrier sleeves on the probe tube {1}------------------------------------------------ #### 5. INTENDED USE The FUSION LED curing light is indicated for curing camphorquinone-based visible light cured (VLC) dental materials. ## 6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANDTIAL EQUIVALENCE The FUSION LED Curing Light is substantially equivalent to K041372 and K030568 in intended use and operation. FUSION Curing Light offers multiple curing programs for photopolymerization of dental resins and activation of bleaching materials. These program differ in the intensity of the light delivered and the length of the light exposure. The operational principles of the proposed and predicate devices are identical. The operator chooses the appropriate program by a selection of the push buttons. The major difference between the proposed FUSION Curing Light and the predicate curing lights is the intensity of the light power delivered, the longer range of curing distance, size and weight. The increased light intensity of the proposed FUSION Curing Light allows the resin curing and tooth whitening agent activation to be conducted at faster speed. The longer range of curing distance enables. Horre flexibility for operators. The size and weight is a benefit to constant patient operations and counter space. # 7. PERFORMANCE TESTING AND COMPLIANCE The following testing was conducted to evaluate the functionality performance of the proposed FUSION Curing Light: - Optical Power Testing - Resin curing time . - Curing range . - Optical wavelength . - Software and hardware verification and validation . The FUSION is designed to comply with electrical safety and electromagnetic compatibility and will comply with electrical safety requirement established.by IEC 60601 and 60601-1-2. We believe the similarity of the FUSION LED Curing Light to the legally marketed predicate devices and the performance data provided support the safety and effectiveness of the FUSION LED Curing Light for the indicated use. {2}------------------------------------------------ # Third Party Review Quality Assessment | Section 1 - Submission Information | |------------------------------------| |------------------------------------| | 510(k) No.: | K052593 | |------------------------------------|------------------------| | Third Party Organization: | Reg Tech Services, LLC | | Third Party's Primary Reviewer(s): | Mark Job | | ODE/OIVD Division: | DAGID | | Branch/Team: | DEDB | | Section 2 - 510(k) Decision | |-----------------------------| |-----------------------------| | Third party recommendation: | SE <span style="text-decoration: underline;">✓</span> NSE ______ Other (specify): ______ | |-----------------------------|------------------------------------------------------------------------------------------| | ODE/OIVD final decision: | SE <span style="text-decoration: underline;">✓</span> NSE ______ Other (specify): ______ | Section 3 – Assessment of Third Party Review | Review Element | Rating (check one) | | | |---------------------------------------------------------------------------------------------------------------------------------------------|--------------------|----------------|----------------| | | Adequate | Minor Issue(s) | Major Issue(s) | | a. Determination of device eligibility for third party review | ✓ | | | | b. Extent of pre-submission consultation with ODE/OIVD division | ✓ | | | | c. Organization and format of review documentation | ✓ | | | | d. Determination of 510(k) administrative completeness (screening review) | ✓ | | | | e. Summary of device characteristics, intended use, and performance (including accessories, if applicable) and reason for 510(k) submission | ✓ | | | | f. Comparison to legally marketed devices—identification and analysis of key similarities and differences | ✓ | | | | g. Rationale for conclusions and recommendation | ✓ | | | | h. Use of guidance documents and standards | ✓ | | | | i. Resolution of 510(k) deficiencies and FDA requests for additional information | ✓ | | | | j. Scope of reviewer expertise and use of consulting reviewers | ✓ | | | | k. Other (specify): | | | | Comments (explanation of ratings/issues): ____________________________________________________________________________________________________________________________________ . Section 4 - ODE/OLD Assessor Information Assessed by: *MSPR* Date: *09/26/05*. Tel. No.: *301-827-5253* ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . ______________________________________________________________________________________________________________________________________________________________________________ Routing: Division -- Clip completed assessment (this page only) to inside front cover of 510(k). DMC-Forward this page only to Eric Rechen, POS/ODF, Rm. 120J, Corp. Blvd. (HFZ-402). {3}------------------------------------------------ Image /page/3/Picture/2 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged in a circular fashion around the eagle. SEP 2 8 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DentLight, Incorporated C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services, LLC 1394 25th Strect, NW Buffalo, Minnesota 55313 Re: K052593 Trade/Device Name: FUSION Curing Light Regulation Number: 21 CFR 872.6070 Regulation Name: Ultraviolet Activator for Polymerization Regulatory Class: II Product Code: EBZ Dated: September 19, 2005 Received: September 21, 2005 Dcar Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general approvisions of the Act. The general controls provisions of the Act include eequirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting (1 Nr.), it may be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Mr. Job Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. or mo ret of any with all the Act's requirements, including, but not limited to: registration r od listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice and ifing (2) - 2011 forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a premiurket notification - 11 device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), n free contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Anthony D. Watson for CKL Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use 510(k) Number (if known): K052593 Device Name: FUSION Curing Light Indications For Use. The FUSION Curing Light is a dental curing light that is intended for photopolymerization of dental resins, restorative composite materials, and orthodontic brackets, bonding, and sealing that are photo-polymerized in the 440 – 480 nm wavelength range. X Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Colleen for MSR (Division Sign-Off) (Division Sign-Off) Division of Anesthesiology, General Hospital, Division of Anesthesiology, General Devices Division of Anesthoolors of Device Infection Control 510(k) Number. K0512593
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...