LUMINARY CANNULATOR BIDEFLECTABLE CATHETER WITH LUMEN

K052575 · St Jude Medical · DRA · Jan 19, 2006 · Cardiovascular

Device Facts

Record IDK052575
Device NameLUMINARY CANNULATOR BIDEFLECTABLE CATHETER WITH LUMEN
ApplicantSt Jude Medical
Product CodeDRA · Cardiovascular
Decision DateJan 19, 2006
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1280
Device ClassClass 2

Indications for Use

The Luminary Cannulator™ Bideflectable Catheter with Lumen is indicated to provide a pathway for delivery and support of transvenous devices and fluids to the Coronary Sinus and coronary vasculature of the heart. The Luminary Cannulator™ Bideflectable Catheter with Lumen can also be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intra-vascular sites when minimizing blood loss is essential.

Device Story

Luminary Cannulator™ is a steerable, bideflectable catheter designed for Cardiac Resynchronization Therapy (CRT) and arrhythmia evaluation. Device features two deflectable curves at the distal tip (large for CS cannulation, small for venous subselection) and two platinum ring electrodes for electrical confirmation of CS access. Operated by a proximal handle with an actuator mechanism and automatic steering lock; includes a hemostasis valve system with sideport for aspiration/fluid injection. Used by physicians in clinical settings to provide a pathway for guidewires (up to 0.035") and fluids. Enables precise navigation within the coronary sinus and venous branches; facilitates temporary pacing and sensing of electrical activity. Benefits include improved access to target cardiac vessels and reduced blood loss during procedures.

Clinical Evidence

No human clinical trials. Evidence includes bench testing comparing subselection performance against Medtronic Attain Prevail catheter in a human heart model, and an in-vivo study in two adult canines evaluating CS cannulation and branch subselection. Biocompatibility testing performed per ISO 10993. Device integrity (mechanical, functional, packaging) testing confirmed performance requirements met.

Technological Characteristics

Bideflectable steerable catheter; 108 cm length; 0.035" guidewire compatibility. Features two distal platinum ring electrodes (1 mm width, 10 mm spacing). Includes proximal handle with deflection actuator, hemostasis valve, and electrical connector. Sterilized via Ethylene Oxide (EtO). Biocompatible materials per ISO 10993.

Indications for Use

Indicated for patients requiring transvenous device/fluid delivery to the coronary sinus or coronary vasculature, and for patients undergoing evaluation of cardiac arrhythmias where blood loss minimization is required.

Regulatory Classification

Identification

A steerable catheter is a catheter used for diagnostic and monitoring purposes whose movements are directed by a steering control unit.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JAN 1 9 2006 510(k) Summary As required by 21 CFR 807.92(c) KD52575 510(k) Number: #### Date Prepared: September 16, 2005 #### Submitter Information: | Submitter's Name/ | St. Jude Medical | |-------------------|---------------------------| | Address: | 14901 DeVeau Place | | | Minnetonka, MN 55345-2126 | Contact Person: Patrice Stromberg Sr. Regulatory Affairs Specialist Tel: 952-351-1496 Fax: 952-930-9481 pstromberg@sim.com #### Device Information: Trade Name: | | Bideflectable Catheter with Lumen | |----------------------|-----------------------------------| | Common Name: | Steerable Catheter | | Classification Name: | Steerable Catheter | | Class: | Class II, 21 CFR 870.1280 | ## Predicate Devices: - St. Jude Medical Reflexion Cannulator™ Steerable Electrophysiology . Catheter with Lumen (K040165) - Medtronic Attain™ Prevail® 6228CTH Steerable Catheter Set (K031211) . Luminary Cannulator™ ## Device Description: The Luminary Cannulator™ Bideflectable Catheter with Lumen is designed for Cardiac Resynchronization Therapy procedures and can be used with St. Jude Medical (SJM) Catheter Delivery Systems. The Luminary Cannulator™ Bideflectable Catheter with Lumen features two deflectable curves in the catheter tip: a large curve for Coronary Sinus (CS) cannulation and a small curve for venous subselection. The large curve for cannulation comes in Large Curl and Extra Large Curl sizes for a wide range of heart failure patients, including those with enlarged atria. The small curve facilitates subselection of venous branches of the CS. {1}------------------------------------------------ The Luminary Cannulator™ Bideflectable Catheter with Lumen has an overall length of 108 cm (80 cm working length). The internal lumen of the catheter accommodates up to 0.035" guidewires. The Luminary Cannulator™M Bideflectable Catheter with Lumen has two distal platinum ring electrodes 1 mm in widths, which are spaced 10 mm apart. The bipolar electrode tip configuration provides electrical confirmation of CS access. The distal active tip is manipulated remotely by a control handle with actuator mechanism located at the proximal end of the catheter. An automatic steering lock holds the tip deflection position without the need for an extra locking mechanism. The proximal handle with deflection actuator contains a hemostasis valve system with sideport and stopcock for aspiration and fluid injection and an electrical connector for sensing of electrical activity and temporary pacing. Valve Bypass Tools, packaged with the Luminary Cannulator™ Bideflectable Catheter with Lumen, are used to direct compatible guidewires through the hemostasis valve from either direction (through the valve, or through distal tip), as well as for forceful manual injection of fluids. #### Intended Use/Indications for Use: The Luminary Cannulator™ Bideflectable Catheter with Lumen is indicated to provide a pathway for delivery and support of transvenous devices and fluids to the Coronary Sinus and coronary vasculature of the heart. The Luminary Cannulator™ Bideflectable Catheter with Lumen can also be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intra-vascular sites when minimizing blood loss is essential. ## Summary of Testing ## Performance Testing: To support the expansion of the indications for use, comparison bench testing was conducted with the St. Jude Medical Luminary Cannulator'[M Bideflectable Catheter with Lumen and the Medtronic 7F Attain Prevail 6228 CTH Steerable Catheter System. The purpose of the test was to compare the ability of the Luminary Cannulator™ Bideflectable Catheter with Lumen to subselect a branch off the Coronary Sinus or the Great Cardiac Vein with that of the Medtronic Attain Prevail, using a bench model of the human heart. Sub-selection of the branches in the heart model was successful using the Luminary Cannulator™ Bideflectable Catheter with Lumen. An In-Vivo study of the mechanical performance characteristics of the Luminary Cannulator™ Bideflectable Catheter with Lumen were evaluated in two adult canines. The study evaluated the Luminary Cannulator™ Bideflectable Catheter with Lumen for CS cannulation, subselection of a target branch off the CS / Great Cardiac Vein, and provided analysis of the vessels of the heart through which it passed immediately after the test procedure. Performance of the Luminary {2}------------------------------------------------ Cannulator™ Bideflectable Catheter with Lumen was satisfactory and met all the study objectives, including the ability to sub-select a branch off the Coronary Sinus or great cardiac vein. Device integrity testing was performed to support the equivalency of the Luminary Cannulator™ Bideflectable Catheter with Lumen to the predicate devices. Testing included mechanical, functional, and packaging testing. The Luminary Cannulator™ Bideflectable Catheter with Lumen met all spccified design and performance requirements. Biocompatibility. Biocompatibility testing in accordance with ISO 10993 was provided. The material used in the Luminary Cannulator™ Bideflectable Catheter with Lumen has been demonstrated to be biocompatible. Sterilization Validation: The Luminary Cannulator™ Bideflectable Catheter with Lumen will be sterilized using the existing validated Ethylene Oxide (EtO) sterilization process. The results of the tests provide reasonable assurance that the device has been designed and tested to assure conformance to the requirements for its indications for use. The Luminary Cannulator™ Bideflectable Catheter with Lumen uses similar technology and has similar intended uses, materials and dimensional characteristics to the predicate devices. #### Statement of Equivalence: Through the data and information presented, St. Jude Medical considers the Luminary Cannulator™ Bideflectable Catheter with Lumen to be substantially equivalent to the St. Jude Medical Reflexion Cannulator™ with Lumen (K040165) and Medtronic Attain™ Prevail® 6228 CTH Steerable Catheter Set (K031211). {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features a stylized depiction of an eagle with three curved lines representing its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle symbol. Public Health Service JAN 1 9 2006 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 St. Jude Medical c/o Ms. Patrice Stromberg Sr. Regulatory Affairs Specialist 14901 De Veau Place Minnetonka, MN 55345-2126 Re: K052575 Luminary Cannulator™ Bideflectiable Catheter with Lumen Regulation Number: 21 CFR 870.1280 Regulation Name: Steerable Catheter Regulatory Class: II Product Code: DRA Dated: December 22, 2005 Received: December 27, 2005 #### Dear Ms. Stromberg: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Ms. Patrice Stromberg Please be advised that FDA's issuance of a substantial equivalence determination does not mean r that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) Premarket notification. The FDA finding of substantial equivalence of your device to a legally premated predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Douna D. Valines - Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ## INDICATIONS FOR USE 510(K) Number (if known): K052575 Device Name: Luminary Cannulator™ Bideflectiable Catheter with Lumen Indications for Use: The Luminary Cannulator™ Bideflectable Catheter with Lumen is indicated to provide a pathway for delivery and support of transvenous devices and fluids to the coronary sinus and coronary vasculature of the heart. The Luminary Cannulator™ Bideflectable Catheter with Lumen can also be used in the evaluation of a variety of cardiac arrhythmias from endocardial and intravascular sites when minimizing blood loss is essential. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Donna R. Vochner (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number ko 5 7 Page 1 of 1
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...