K052435 · Pneumrx, Inc. · KNW · Nov 3, 2005 · Gastroenterology, Urology
Device Facts
Record ID
K052435
Device Name
PNEUMRX FAMILY OF BIOPSY NEEDLES AND KITS
Applicant
Pneumrx, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Nov 3, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The PneumRx Family of Biopsy Needles and Kits is intended for use in obtaining bone, vertebral body, and bone marrow biopsy specimens, and single and multiple tissue biopsy specimens of tissues or lesions, partial or complete removal of imaged abnormalities, in soft tissues, for applying radionuclide sources into the body, and for aspirating, draining, or injecting various fluids or contents (e.g., sclerosing agents, vasoconstrictors, anesthetics, ethanol, saline, heparin, and other historically used, approved, and/or generally accepted medical fluids in accordance with the judgment and training of the medical practitioner). The PneumRx Family of Biopsy Needles and Kits is indicated for use in obtaining bone, vertebral body, and/or bone marrow biopsy using coring, cutting or aspiration, as well as providing and maintaining surgical site access with coaxial biopsy needle kits. They are also indicated for use in percutaneous, open surgical, fine needle, tissue core, coaxial, and aspiration biopsy/ tissue sampling of soft tissues for: Microscopic, histologic, diagnostic, evaluation; Fluoroscopic, ultrasound (US), computed tomography (CT)/x-ray, mammographic imaging, MRI (magnetic resonance imaging) and/or direct visualization; Guiding biopsy needles to the target lesion; Avoiding punctures in non-penetrable anatomic obstacles (e.g., bone, spine) or vital structures (e.g., blood vessels, nerves); Obtaining multiple biopsy, core biopsy, and aspiration biopsy tissue samples with partial or complete removal of the imaged abnormality; Aspirating, draining, or injecting various fluids or contents in soft tissues, including introduction of sclerosing agents or vasoconstrictors to control actual or potential bleeding; Pericardiocentesis, transhepatic cholorangiography, and other applications requiring injection or aspiration of fluids; Lumbar puncture, spinal fluid sampling, local anesthetic injection to provide regional nerve block/anesthesia, and amniocentesis; Manual application of a radionuclide source into or on the body for radiation therapy; Including, but not limited to, the following anatomical locations and lesions: Pleural cavity, Lung, Lymph nodes, Thyroid, Adrenals, Soft tissue organs of the abdomen and thorax, Abdominal cavity/ abdominal soft tissue masses, Kidney, Liver, Spleen, Pancreas, Prostate, Tumors, Cysts, Fluid spaces, Soft tissue lesions or tissue spaces, Breast: The PneumRx Family of Biopsy Needles and Kits is indicated to provide breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality; The extent of a histologic abnormality cannot be reliably determined from its mammographic appearance. Therefore the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Device Story
PneumRx Family of Biopsy Needles and Kits consists of standard and steerable biopsy needles, straight and steering stylets, a remote control steering handle, and procedure accessories (drapes, syringes, etc.). Device used by medical practitioners in clinical settings for percutaneous or open surgical biopsy, fluid aspiration/injection, and radionuclide placement. Steering handle allows navigation to target lesions while avoiding vital structures or non-penetrable obstacles. Output is tissue specimen or fluid sample for diagnostic evaluation. Benefits include precise tissue sampling and site access for therapeutic interventions.
Clinical Evidence
Bench testing only.
Technological Characteristics
System includes standard and steerable biopsy needles, stylets, and a remote control steering handle. Accessories include drapes, gauze, syringes, scalpels, and depth markers. Designed for percutaneous or open surgical use under various imaging modalities (CT, US, MRI, fluoroscopy).
Indications for Use
Indicated for patients requiring bone, vertebral body, bone marrow, or soft tissue biopsy/sampling, fluid aspiration/injection, or radionuclide source placement. Applicable to various anatomical sites including lung, breast, liver, kidney, prostate, and others. Used for diagnostic, histologic, or therapeutic purposes under imaging guidance (US, CT, MRI, fluoroscopy).
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
Ethicon Endo-Surgery, Inc. Mammotome Biopsy System (K992813)
Medical Device Technologies, Inc. Tru-Core I Reusable Biopsy Instrument (K990839)
Medical Tech. Introducer Needle for Biopsy Needles (K981889)
United States Endoscopy Group, Inc. Coaxial Needle (K971842)
Parallax Medical Clearview Plus Bone and Vertebral Body Biopsy Needles (K022169)
Stryker Bone and Vertebral Body Biopsy Kit and Stryker Capture Bone and Vertebral Body Biopsy Kit (K032943)
MD Technologies InterV Brand SnareLok Bone Marrow Biopsy Needle (K043523)
Orthovita, Inc. Imbibe Bone Marrow Aspiration Needle (K050795)
Medical Device Technologies, Inc. Chiba, Spinal, Automatic Cutting, General Purpose Introducer, Techna-Cut Biopsy, Super-Core Biopsy, Breast Localization and Simon Breast Localization Needles (K980211)
{1}------------------------------------------------
# NOV - 3 2005
#### 510(k) Summary for the PneumRx Family of Biopsy Needles and Kits Attachment 32:
#### I. General Information
| Submitter: | PneumRx, Inc.<br>530 Logue Avenue<br>Mountain View, CA 94043 |
|---------------------------|--------------------------------------------------------------|
| Contact Person: | Anne C. Worden<br>VP, Regulatory & Clinical Affairs |
| Summary Preparation Date: | October 24, 2005 |
#### II. Names
| Device Names: | PneumRx Family of Biopsy Needles and Kits |
|---------------|-------------------------------------------|
|---------------|-------------------------------------------|
Primary Classification Names: Biopsy Needle and Needle Set
#### III. Predicate Devices
- Cook® Sidearm Core Tissue Biopsy Device/Set (K041544) .
- Biopsy Sciences, LLC. Maxi-Cell Biopsy Needles (K021847) .
- Boston Scientific Corp. Interject Injection Therapy Needle (K012864) .
- . ISPG, Inc. Chiba Needle (K012358)
- . INRAD, Inc. AccuCore Core Biopsy Needle (K981166)
- INRAD, Inc. AccuCore Single Action Core Biopsy Device (K000612) .
- INRAD, Inc. Co-axial Introducer Needle (K981721) .
- Ethicon Endo-Surgery, Inc. Mammotome Biopsy System (K992813) .
- Medical Device Technologies, Inc. Tru-Core™ I Reusable Biopsy Instrument . (K990839)
- Medical Tech. Introducer Needle for Biopsy Needles (K981889) .
- United States Endoscopy Group, Inc. Coaxial Needle (K971842) .
- Parallax Medical Clearview Plus Bone and Vertebral Body Biopsy Needles . (K022169)
- Stryker Bone and Vertebral Body Biopsy Kit and Stryker Capture Bone and . Vertebral Body Biopsy Kit (K032943)
- MD Technologies InterV Brand SnareLok Bone Marrow Biopsy Needle (K043523) .
- . Orthovita, Inc. Imbibe Bone Marrow Aspiration Needle (K050795)
- Medical Device Technologies, Inc. Chiba, Spinal, Automatic Cutting, General . Purpose Introducer, Techna-Cut Biopsy, Super-Core Biopsy, Breast Localization and Simon Breast Localization Needles (K980211)
- . MD Manan™ Soft Tissue Biopsy Needles (K980122)
- . Daum Corp. Aspiration-Biopsy Needle (K974575)
- Promex/ US Biopsy Prostate Seeding Needle (K973184) ●
- . Klein-Baker Med. Neo-Care Lumbar Puncture Kit (K970997)
{2}------------------------------------------------
- . Allegiance Healthcare Clear Hub Spinal Needle (K982269)
- SIMS Spinal Anesthesia Needle (K983858) ●
- Ferguson Medical SmartGuide (K963894) .
- � US Biopsy Automated Core Biopsy Device (K994272 and K011270)
- . INRAD, Inc. Aspiration Biopsy Needle (K number unknown)
- . Cook Chiba Aspiration Biopsy Needles (K851957)
- . MD Manan™ Co-Axial Introducer Needles (K980004)
- . Manan™M MD Super-Core Biopsy Needle (K974814)
- MD Manan™ Techna-Cut Biopsy Needle (K974766) t
#### IV. Product Description
The PneumRx Family of Biopsy Needles and Kits are comprised of the following main components:
- Standard and steerable biopsy needles in a range of sizes; ◆
- Standard straight and steering stylets compatible with a range of steerable and . standard biopsy needles;
- . Remote Control Steering Handle accessory:
- Procedure accessories including drapes, gauze sponges, syringes, scalpels, depth . markers, and prep tray.
#### V. Indications for Use
The PneumRx Family of Biopsy Needles and Kits is intended for use in obtaining bone, vertebral body, and bone marrow biopsy specimens, and single and multiple tissue biopsy specimens of tissues or lesions, partial or complete removal of imaged abnormalities, in soft tissues, for applying radionuclide sources into the body, and for aspirating, draining, or injecting various fluids or contents (e.g., sclerosing agents, vasoconstrictors, anesthetics, ethanol, saline, heparin, and other historically used, approved, and/or generally accepted medical fluids in accordance with the judgment and training of the medical practitioner).
The PneumRx Family of Biopsy Needles and Kits is indicated for use in obtaining bone. vertebral body, and/or bone marrow biopsy using coring, cutting or aspiration, as well as providing and maintaining surgical site access with coaxial biopsy needle kits.
They are also indicated for use in percutaneous, open surgical. fine needle, tissue core, coaxial, and aspiration biopsy/ tissue sampling of soft tissues for:
- . Microscopic, histologic, diagnostic, evaluation
- Fluoroscopic, ultrasound (US), computed tomography (CT)/x-ray, mammographic . imaging, MRI (magnetic resonance imaging) and/or direct visualization
- . Guiding biopsy needles to the target lesion
- Avoiding punctures in non-penetrable anatomic obstacles (e.g., bone, spine) or vital . structures (e.g., blood vessels, nerves)
- . Obtaining multiple biopsy, core biopsy, and aspiration biopsy tissue samples with partial or complete removal of the imaged abnormality
{3}------------------------------------------------
- . Aspirating, draining, or injecting various fluids or contents in soft tissues, including introduction of sclerosing agents or vasoconstrictors to control actual or potential bleeding
- Pericardiocentesis, transhepatic cholorangiography, and other applications requiring . injection or aspiration of fluids
- Lumbar puncture, spinal fluid sampling, local anesthetic injection to provide regional . nerve block/anesthesia, and amniocentesis
- . Manual application of a radionuclide source into or on the body for radiation therapy
- Including, but not limited to, the following anatomical locations and lesions: .
- Pleural cavity ム
- ♪ Lung
- > Lymph nodes
- Thyroid レ
- 必 Adrenals
- Soft tissue organs of the abdomen and thorax >
- Abdominal cavity/ abdominal soft tissue masses v
- Kidney や
- v Liver
- Spleen レ
- Pancreas >
- Prostate >
- Tumors >
- Cysts ア
- Fluid spaces P
- > Soft tissue lesions or tissue spaces
- > Breast:
- The PneumRx Family of Biopsy Needles and Kits is indicated to provide . breast tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality;
- The extent of a histologic abnormality cannot be reliably determined from its . mammographic appearance. Therefore the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
#### VI. Rationale for Substantial Equivalence
The PneumRx Family of Biopsy Needles and Kits shares the same or similar indications for use, device operation, overall technical and functional capabilities, and therefore is substantially equivalent for use in obtaining single and multiple tissue biopsy specimens of tissues or lesions for microscopic, histologic, diagnostic, evaluation, for partial or complete removal of imaged abnormalities, in soft tissues, for applying radionuclide sources into the body, and for aspirating, draining, or injecting various fluids or contents to the predicate devices.
In addition, comparative performance test data demonstrated adequate device performance.
{4}------------------------------------------------
#### VII. Safety and Effectiveness Information
The review of the indications for use and technical characteristics provided demonstrates that the PneumRx Family of Biopsy Needles and Kits is substantially equivalent to the predicate devices.
## VIII. Conclusion
The PneumRx Family of Biopsy Needles and Kits was found to be substantially equivalent to the predicate devices.
The PneumRx Family of Biopsy Needles and Kits shares identical indications for use, similar design features, and functional features with, and thus is substantially equivalent to, the predicate devices.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS) in the United States. The logo features a stylized eagle with three curved lines representing its wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" is arranged in a circular pattern around the eagle symbol. The logo is black and white.
NOV - 3 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Anne C. Worden Vice President, Regulatory & Clinical Affairs PneumRx, Inc. 530 Logue Avenue Mountain View, California 94043
Re: K052435
Trade/Device Name: PneumRx Family of Biopsy Needles and Kits Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology-biopsy instrument Regulatory Class: II Product Code: KNW, FCG, GAA, GDM, DWO Dated: September 2, 2005 Received: September 6, 2005
Dear Ms. Worden:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Barbara Buehrig
for
Mel N. Miller
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K052435
#### PneumRx Family of Biopsy Needles and Kits Device Name:
Indications for Use:
The PneumRx Family of Biopsy Needles and Kits is intended for use in obtaining bone, vertebral body, and bone marrow biopsy specimens, and single and multiple tissue biopsy specimens of tissues or lesions, partial or complete removal of imaged abnormalities, in soft tissues, for applying radionuciide sources into the body, and for aspirating, or injecting various fluids or contents (e.g., sclerosing agents, vasoconstrictors, anesthetics, ethanol, saline, heparin, and other historically used, and/or generally accepted medical fluids in accordance with the judgment and training of the medical practitioner).
The PneumRx Family of Biopsy Needles and Kits is indicated for use in obtaining bone, vertebral body, and/or bone marrow biopsy using coring, cutting or aspiration, as well as providing and maintaining surgical site access with coaxial biopsy needle kits.
They are also indicated for use in percutaneous, open surgical, fine needle, tissue core, coaxial, and aspiration biopsy/ tissue sampling of soft tissues for:
- . Microscopic, histologic, diagnostic, evaluation
- Fluoroscopic, ultrasound (US), computed tomography (CT)/x-ray, mammographic imaging, . MRI (magnetic resonance imaging), and/or direct visualization
- . Guiding biopsy needles to the target lesion
- Avoiding punctures in non-penetrable anatomic obstacles (e.g., bone, spine) or vital � structures (e.g., blood vessels, nerves)
- Obtaining multiple biopsy, core biopsy, and aspiration biopsy tissue samples with partial or . complete removal of the imaged abnormality
- Aspirating, draining, or injecting various fluids or contents in soft tissues, including . introduction of sclerosing agents or vasoconstrictors to control actual or potential bleeding
- Pericardiocentesis, transhepatic cholorangiography, and other applications requiring injection . or aspiration of fluids
- Lumbar puncture, spinal fluid sampling, local anesthetic injection to provide regional nerve . block/anesthesia, and amniocentesis
- Manual application of a radionuclide source into or on the body for radiation therapy .
| Prescription Use<br>Image: Checkmark | AND/OR | Over-The-Counter Use |
|--------------------------------------|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Vartare Buellin
al, Restorative, and Neurological Devices
Page 1 of 2
**510(k) Number** K052435
{8}------------------------------------------------
### Indications for Use Statement
510(k) Number (if known): K052435
Device Name: PneumRx Family of Biopsy Needles and Kits
Indications for Use: - Continued from previous page -
- Including, but not limited to, the following anatomical locations and lesions; .
- Pleural cavity v
- > Lung
- > Lymph nodes
- > Thyroid
- > Adrenals
- Soft tissue organs of the abdomen and thorax >
- > Abdominal cavity/ abdominal soft tissue masses
- ♪ Kidney
- > Liver
- > Spleen
- ン Pancreas
- い Prostate
- > Tumors
- > Cysts
- > Fluid spaces
- 5 Soft tissue lesions or tissue spaces
- ン Breast:
- The PneumRx Family of Biopsy Needles and Kits is indicated to provide breast . tissue samples for diagnostic sampling of breast abnormalities. It is designed to provide breast tissue for histologic examination with partial or complete removal of the imaged abnormality;
- The extent of a histologic abnormality cannot be reliably determined from its . mammographic appearance. Therefore the extent of removal of the imaged evidence of an abnormality does not predict the extent of removal of a histologic abnormality (e.g., malignancy). When the sampled abnormality is not histologically benign, it is essential that the tissue margins be examined for completeness of removal using standard surgical procedures.
Prescription Use > (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Buelind
Division of General. Restorative, and Neurological Devices
Page 2 of 2
510(k) Number K050435
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.