LANX VBR SYSTEM

K052384 · Lanx, LLC · MQP · Oct 7, 2005 · Orthopedic

Device Facts

Record IDK052384
Device NameLANX VBR SYSTEM
ApplicantLanx, LLC
Product CodeMQP · Orthopedic
Decision DateOct 7, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Lanx VBR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system.

Device Story

Lanx VBR is a spinal intervertebral body fixation orthosis; consists of flat, trapezoidal, rectangular, and curved PEEK implants with titanium position markers; hollowed central area accommodates bone graft; upper/lower surfaces feature transverse slots/grooves for stability. Used in thoracolumbar spine (T1-L5) to replace resected/excised vertebral bodies due to tumor or trauma/fracture; restores height of collapsed vertebrae; achieves anterior decompression of spinal cord/neural tissues; restores biomechanical integrity of spinal column. Requires supplemental internal fixation. Implanted by surgeons in clinical/OR settings. Benefits include structural support and stabilization of the spinal column following resection or fracture.

Clinical Evidence

Bench testing only. Fatigue and static testing performed according to accepted engineering and scientific principles. Results demonstrate performance equivalent to comparison devices. Biocompatibility information provided via device material Master File (MAF).

Technological Characteristics

Materials: Implant-grade PEEK with titanium position markers; instruments made of stainless steel. Design: Hollowed central area for bone graft; transverse slots/grooves on surfaces for fixation. Sterilization: Non-sterile; requires high-temperature steam sterilization (270°F/134°C for 18 minutes).

Indications for Use

Indicated for patients requiring vertebral body replacement or partial vertebrectomy in the thoracolumbar spine (T1-L5) due to tumor, trauma, or fracture. Requires use with supplemental internal fixation.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ | 510(k) Summary Information<br>Premarket Notification, Section 510(k) | Lanx LLC<br>SEPT 17, 2005 | |----------------------------------------------------------------------|---------------------------| |----------------------------------------------------------------------|---------------------------| Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92 #### 1. Device Name: | Trade Name: | Lanx VBR | |-----------------|----------------------------------------------| | Common: | | | Name(s): | Vertebral body replacement | | Classification: | | | Name(s): | Spinal intervertebral body fixation orthosis | # 2. Establishment Name & Registration Number: | Name: | Lanx LLC | |---------|------------| | Number: | 3004485144 | #### 3. Classification(s): Sec. 888.3060 Spinal intervertebral body fixation orthosis. (a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vetebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fareaded or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct "smay back," scoliosis (lateral curvature of the spine), or other conditions, | Device Class: | Class II for the requested indications | |-----------------------|----------------------------------------------| | Classification Panel: | Orthopaedic and Rehabilitation Devices Panel | | Product Code(s): | MQP | # Equivalent Predicate Device: Lanx LLC believes that the Lanx VBR is substantially equivalent to the following: 1. K003709, Mesh Cage, Surgical Dynamics 2. K001340, Stackable Cage, DePuy 3. K040928, PEEK Cage, Interpore/Cross 4. K042268, EBI CAS Spine Spacer System, EBI. L.P. Equivalence can be seen in the design, material composition, surgical technique and intended use. ## 5. Device Description: 4. Implants: The Lanx VBR implants consists of a series of flat, trapezoidal, rectangular and curved shaped The device is offered in different size/configurations to better approximate the anatomical implants. variation observed in different vertebral levels and/or patient anatomy. The devices are manufactured with a hollowed out central area to accommodate bone graft material The upper and lower surfaces are manufactured in such a way that a series of transverse slots or grooves are formed to improve stability and fixation once the device is inserted. Materials: all implants are made from implant grade PEEK with titanium position markers, Instruments: The instruments are made from instrument grade stainlent post Indications for Use. The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spire (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of diseased {1}------------------------------------------------ vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Lanx VBR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For fixstion is secured to may be most be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system. Testing Summary. Fatigue and static testing is complete. Samples were tested according to accepted engineering and scientific principals. Test results demonstrate that the system can be expected to perform in a manner equivalent to the comparison device material Master File (MAF) recorded with the FDA provided additional biocompatibility information. #### 6. Applicant Name & Address: Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501 ## 7. Company Contact: Michael Fulton, M.D. Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501 ## 8. Submission Correspondent: Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax david(@fda-help.com #### 9. Performance Standards: United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized of tills device do not exist Standard Operating Process certification and qualification and qualification procedures, Quality, Systems Regulations, ISO materials standards and ISO 9000 series quality regulations, LANX, LLC, also meets appropriate general controls authorized under Sections 501, 502, 516, 516, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act, ## I O. Sterilization Information: The Lanx Spinal Fixation System is supplied "NON-STERILE" and must be sterilized prior to use. The recommended sterlization process is high temperature sterilization. The referenced sterilization cycle produces a Sterility Assurance Level (SAL) of att least 106 The validated cycle is: Method: Steam Cycle: Gravity Temperature: 270°F (134°C) Exposure Time: 18 minutes {2}------------------------------------------------ | 510(k) Summary Information | Lanx LLC | |----------------------------------------|---------------| | Premarket Notification, Section 510(k) | SEPT 17, 2005 | Regulatory Authority: Safe Medical Devices Act of 1990, 21 CFR 807.92 #### 1. Device Name: Lanx VBR Trade Name: Common Name(s): Vertebral body replacement Classification Spinal intervertebral body fixation orthosis Name(s): ## 2. Establishment Name & Registration Number: | Name: | Lanx LLC | |---------|------------| | Number: | 3004485144 | ## 3. Classification(s): Sec. 888.3060 Spinal intervertebral body fixation orthosis. (a) Identification. A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral odiaes. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or rimed at each eye-type screw. The device is used to apply force to a series of vertebrae to correct "sway back," scoliosis (lateral curvature of the spine), or other conditions. | Device Class: | Class II for the requested indications | |-----------------------|----------------------------------------------| | Classification Panel: | Orthopaedic and Rehabilitation Devices Panel | | Product Code(s): | MQP | ## 4. Equivalent Predicate Device: Lanx LLC believes that the Lanx VBR is substantially equivalent to the following: - 1. K003709, Mesh Cage, Surgical Dynamics - 2. K001340, Stackable Cage, DePuy - 3. K040928, PEEK Cage, Interpore/Cross - 4. K042268, EBI CAS Spine Spacer System, EBI. L.P. Equivalence can be seen in the design, material composition, surgical technique and intended use ## 5. Device Description: Implants: The Lance VBR implants consists of a series of flat, trapezoidal, rectangular and curved shaped The device is offered in different size/configurations to better approximate the anatomical implants. variation observed in different vertebral levels and/or patient anatomy. The devices are manufactured with a hollowed out central area to accommodate bone graft material. The upper and lower surfaces are manufactured in such a way that a series of transverse slots or grooves are formed to improve stability and fixation once the device is inserted. Materials: all implants are made from implant grade PEEK with titanium position markers, Instruments: The instruments are made from instrument grade stainlens steel. Indications for Use. The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body themely when with the thoracolumbar spine (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of diseased {3}------------------------------------------------ vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Land VRR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental fixation. Supplemental internal fixation is required to properly utilize this system. Testing Summary. Fatigue and static testing is complete. Samples were tested according to accepted engineering and scientific principals. Test results demonstrate that the system can be expected to perform in a manner equivalent to the comparison device. Additionally, a device material Master Five (MAF) recorded with the FDA provided additional biocompatibility information. #### 6. Applicant Name & Address: Lanx、LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501 ## 7. Company Contact: Michael Fulton, M.D. Lanx, LLC 1155 Alpine Avenue, Suite 320 Boulder, CO 80304 303-443-7500 Fax: 303-443-7501 ## 8. Submission Correspondent: Mr. David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill, CA 94523-3389 925.356.2640 - 925.356.2654 - fax david@fda-help.com #### 9. Performance Standards: United States Food and Drug Administration mandated performance standards for this device do not exist. Various voluntary performance standards are utilized overaliands of tilliged include ASTM, Surtame Rocedures, vendor & process certification and qualification procedures, Qualification procedures, Qualification procedures, Qualification procedures, Qualification pr Systems Regulations, ISO materials standards and ISO 9000 series quality regulations. LANX, LLC, also meets appropriate general controls authorized under Sections 501, 516, 516, 518, 519, and 520 of the Food, Drug, and Cosmetic Act. ## 1 0. Sterilization Information: The Lanx Spinal Fixation System is supplied "NON-STERILE" and must be sterilized prior to use. The recommended sterilization process is high temperature sterilization of sterilization. The referenced sterilization cycle produces a Sterility Assurance Level (SAL) of at least 106 The validated cycle is: Method: Steam Cycle: Gravitv Temperature: 270°F (134°C) Exposure Time: 18 minutes {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff entwined with three snakes, representing healing and medicine. The logo is surrounded by the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" in a circular arrangement. Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 7 - 2005 Lanx, LLC c/o David W. Schlerf Buckman Company, Inc. 200 Gregory Lane, Suite C-100 Pleasant Hill. California 94523 Re: K052384 Trade/Device Name: Lanx VBR Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: August 26, 2005 Received: August 30, 2005 Dear Mr. Schlerf: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {5}------------------------------------------------ Page 2- David W. Schlerf This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Signature: Mark McMullen Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Page 1 of 1 510(k) Number: K052384 LanxVBR Device Name(s): Indications for Use: The Lanx VBR System is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from TI to LS) The Lanx VBR System may also be used in the thoracolumbar spine (i.e., Ti- L5) for partial replacement (i.e., partial vertebrectomy) of a diseased vertebral body resected or excised for the treatment of tumors in order to achieve anterior decompression of the spinal cord and neural tissues, and to restore the height of a collapsed vertebral body. The Lanx VBR System is also indicated for treating fractures of the thoracic and lumbar spine. The Lanx VBR System is designed to restore the biomechanical integrity of the anterior, middle, and posterior spinal column. For either indication the system must be used with supplemental internal fixation. Supplemental internal fixation is required to properly utilize this system. Over-The-Counter Use Prescription Use X OR PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NECESSARY Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of General, Restorative, and Neurological Devices 510(k) Number K052384 (Per 21 CFR 801.109) (Optional format 1-2-96) ાં ર
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...