ANALGESIA GAS MACHINE, MODEL DIGITAL ULTRA

K052335 · Accutron, Inc. · BSZ · Nov 23, 2005 · Anesthesiology

Device Facts

Record IDK052335
Device NameANALGESIA GAS MACHINE, MODEL DIGITAL ULTRA
ApplicantAccutron, Inc.
Product CodeBSZ · Anesthesiology
Decision DateNov 23, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5160
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Digital Ultra Analgesia Gas Machine is intended for use in nitrous oxide-oxygen sedation systems for delivering to a patient a mixture of nitrous oxide and oxygen gases with a maximum nitrous oxide concentration of 70%.

Device Story

The Accutron Digital Ultra Analgesia Gas Machine is a medical gas delivery system used in dental offices and hospitals for conscious sedation. It receives input from oxygen and nitrous oxide gas sources. The device features digital controls allowing the attending physician or dentist to precisely set the percentage of nitrous oxide and total gas flow. It incorporates mechanical and fail-safe features to ensure oxygen concentration remains at or above 30% of total flow and prevents nitrous oxide flow if oxygen supply is absent. The device provides a mixture of gases to the patient via a delivery system. It is operated by trained clinicians to manage patient sedation levels, potentially benefiting the patient by providing controlled analgesia during procedures.

Clinical Evidence

Bench testing only. The device was evaluated against voluntary standards including EN/IEC 60601-1, EN/IEC 60601-1-2, ISO 5356-1, ASTM F1054-01, and CGA standards for gas safety and connectivity. No clinical data was required or provided.

Technological Characteristics

Gas-powered anesthesia/analgesia machine. Features digital flow and concentration metering. Safety features: 30% minimum oxygen concentration, mechanical gas mix-up prevention, and nitrous oxide fail-safe. Conforms to ISO 5356-1 (conical connectors), ASTM F1054-01 (conical fittings), and CGA C-9/V-5 (gas container/connection standards). Complies with IEC 60601-1 and 60601-1-2 for electrical safety and EMC.

Indications for Use

Indicated for patients requiring nitrous oxide-oxygen conscious sedation in dental offices or hospitals. Contraindicated for use by personnel not qualified in conscious sedation procedures.

Regulatory Classification

Identification

A gas machine for anesthesia is a device used to administer to a patient, continuously or intermittently, a general inhalation anesthetic and to maintain a patient's ventilation. The device may include a gas flowmeter, vaporizer, ventilator, breathing circuit with bag, and emergency air supply. A gas machine for analgesia is a device used to administer to a patient an analgesic agent, such as a nitrous oxide-oxygen mixture (maximum concentration of 70 percent nitrous oxide).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Accutron, Inc. Digital Ultra Analgesia Gas Machine NOV 2 3 2005 510(k) Number: K 052335 > Date: Page 1 of 2 # 510(k) Summary #### Introduction This summary is intended to comply with requirements of the SMDA and 21CFR§807.92. FDA may make this summary available to the public within 30 days following a finding of substantial equivalence. # 510(k) Applicant Accutron, Inc. 1733 West Parkside Lane Phoenix, AZ, 85027 USA #### 510(k) Correspondent Robert N. Clark, President and Senior Consultant Medical Device Regulatory Advisors 13605 West 700 Ave., Golden, CO 80401 USA #### Date Prepared August 24, 2005 #### Trade Name of Device Accutron Digital Ultra Analgesia Gas Machine #### Classification Name Gas-Machine, Anesthesia #### 510(k) Classification 21CFR§868.5160 Product Code: BSZ #### Device Description The Accutron Digital Ultra Analgesia Gas Machine precisely meters oxygen and nitrous oxide medical gases for conscious sedation of patients in dental offices and hospitals. With the device, the attending physician or dentist is able to set both the percentage of nitrous oxide and the total flow to a desired level of sedation and flow. The Digital Ultra Analgesia Gas Machine has been designed with built in safety features that prevent the level of oxygen gas from falling below 30% of total flow, mechanical features that prevent the mix-up of gases, and fail-safe features to prevent the flow of nitrous oxide in the absence of oxygen gas. {1}------------------------------------------------ Accutron, Inc. Digital Ultra Analgesia Gas Machine | 510(k) Number: | | |----------------|--| | Date: | | | Page 2 of 2 | | ### Intended Use The Digital Ultra Analgesia Gas Machine is intended for use in nitrous oxide-oxygen sedation systems for delivering to a patient a mixture of nitrous oxide and oxygen gases with a maximum nitrous oxide concentration of 70%. # Predicate Devices K970163 Technical Medical Products Model 20 Alpha MX / Model 30 Ultra PC manufactured by Accutron Inc., Phoenix AZ. K945722 Matrx MDM, RA manufactured by Matrx Medical, Inc., Orchard Park, New York. ### Voluntary Standards EN 60601-1 / IEC 60601-1, "Medical Electrical Equipment - Part 1 : General Requirements for Safety" EN 60601-1-2 / IEC 60601-1-2, "Medical electrical equipment -- Part 1-2: General requirements for safety – Collateral standard: Electromagnetic compatibility – Requirements and tests" ISO 5356-1, "Anesthetic and respiratory equipment -- conical connectors" ASTM F1054-01, "Standard specification for conical fittings" CGA C-9:1988, "Standard color marking of compressed gas containers for medical use" CGA V-5:2000, "Diameter-Index Safety System (non-interchangeable low pressure connections for medical gas applications)" # ADA Recommendations The Accutron Digital Ultra Analgesia Gas Machine has been designed to meet the recommendations of the American Dental Association. #### Risk Management This device has been designed to either completely eliminate or mitigate known health hazards associated with the use of the device. Health hazard risk reduction has been accomplished by rigorous application of a risk management program. The user must be qualified in conscious sedation procedures, and must be familiar with all labeling and instructions for use associated with the device. Accutron Inc. believes that the Digital Ultra Analgesia Gas Machine is safe and effective when used as instructed by knowledgeable and trained personnel, and is substantially equivalent to the legally marketed predicate devices. {2}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three abstract shapes resembling birds in flight or flowing lines, stacked on top of each other. NOV 2 3 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Accutron, Incorporated C/O Mr. Robert N. Clark President & Senior Consultant Medical Device Regulatory Advisors 13605 West 7th Avenue Golden, Colorado 80401-4604 Re: K052335 Trade/Device Name: Accutron Digital Ultra Analgesia Gas Machine Regulation Number: 868.5160 Regulation Name: Gas Machine for Anesthesia or Analgesia Regulatory Class: II Product Code: BSZ Dated: November 14, 2005 Received: November 15, 2005 Dear Mr. Clark: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ #### Page 2 - Mr. Clark Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits vour device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {4}------------------------------------------------ # Indications for Use 510(k) Number (if known): K052335 Accutron Digital Ultra Analgesia Gas Machine Device Name: Indications For Use: To be used in nitrous oxide-oxygen sedation systems for delivering to a patient a mixture of nitrous oxide and oxygen gases with a maximum nitrous oxide concentration of 70%. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Cucy Syllom on Sim-Old on of Anesthesiology, General Hospital, . Collion Control, Dental Devices .. ) Number .. Page 1 of 1 K052335/S1 Page 5 of 205
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...