BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFUROXIME (STREP) 0.125 - 4UG/ML
K052270 · Becton, Dickinson & CO · LON · Oct 5, 2005 · Microbiology
Device Facts
| Record ID | K052270 |
| Device Name | BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - CEFUROXIME (STREP) 0.125 - 4UG/ML |
| Applicant | Becton, Dickinson & CO |
| Product Code | LON · Microbiology |
| Decision Date | Oct 5, 2005 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1645 |
| Device Class | Class 2 |
Intended Use
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus, Enterococcus, and Streptococcus. This premarket notification is for the addition of the antimicrobial agent cefuroxime at concentrations of 0.125-4 µg/mL to Streptococcus ID/AST or AST only Phoenix panels. Cefuroxime has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent. Active In Vitro and in Clinical Infections Against: Streptococcus pneumoniae Streptococcus pyogenes (and other streptococci)
Device Story
System uses sealed, self-inoculating polystyrene trays with 136 micro-wells containing dried reagents; organism suspension (0.5 McFarland) prepared in AST-S broth with redox indicator. Instrument continuously incubates panels at 35°C; performs automated colorimetric oxidation-reduction readings every 20 minutes to detect growth. System interprets biochemical/antimicrobial reactions to provide ID and MIC values. Software-driven 'EXPERT' system applies CLSI-derived rules for final interpretation. Used in clinical microbiology laboratories; operated by trained laboratory personnel. Output provides MIC values and categorical interpretations (SIR) to assist clinicians in selecting appropriate antibiotic therapy for bacterial infections.
Clinical Evidence
Clinical performance evaluated by comparing BD Phoenix results to CLSI reference broth microdilution method across multiple US sites. Metrics included Essential Agreement (EA) and Category Agreement (CA). Intra-site reproducibility >90%, inter-site reproducibility >95% for streptococcal isolates.
Technological Characteristics
System consists of automated instrument, software, and molded polystyrene panels (136 wells). Uses broth microdilution method with redox indicator for growth detection. Incubation at 35°C with automated optical reading every 20 minutes. Connectivity: Automated system.
Indications for Use
Indicated for in vitro rapid identification and antimicrobial susceptibility testing of Streptococcus species from pure culture. Prescription use only.
Regulatory Classification
Identification
A fully automated short-term incubation cycle antimicrobial susceptibility system is a device that incorporates concentrations of antimicrobial agents into a system for the purpose of determining in vitro susceptibility of bacterial pathogens isolated from clinical specimens. Test results obtained from short-term (less than 16 hours) incubation are used to determine the antimicrobial agent of choice to treat bacterial diseases.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA.”
Predicate Devices
- VITEK® System (PMA No. N50510)
- BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321)
- BD Phoenix™ Automated Microbiology System with Ofloxacin (K020323)
- BD Phoenix™ Automated Microbiology System with Levofloxacin (K020322)
Related Devices
- K023527 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - NORFLOXACIN - 0.25-16 UG/ML · Becton, Dickinson & CO · Dec 23, 2002
- K060493 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM - VANCOMYCIN-GP DETECTION OF VRSA. · Becton, Dickinson & CO · Mar 29, 2006
- K033889 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM FOR USE WITH THE ANTIMICROBIAL AGENT CHLORAMPHENICOL 1-32 UG/ML GRAM POSITIVE · Becton, Dickinson & CO · Feb 3, 2004
- K061327 — BD PHOENIX AUTOMATED MICROBIOLOGY SYSTEM-AMOXICILLIN-CLAVULANATE (GN) 0.5/0.25-32/16 UG/ML, AMPICILLIN-SULBACTAM (GN) 0. · Becton, Dickinson & CO · Jul 3, 2006
- K151320 — BD Phoenix Automated Microbiology System-Ertapenem 0.0625-8 mcg/ml · Becton, Dickinson and Company · Jan 15, 2016
Submission Summary (Full Text)
{0}------------------------------------------------
OCT 5 - 2005
#### 510(k) SUMMARY
| SUBMITTED BY: | Becton, Dickinson and Company<br>7 Loveton Circle<br>Sparks, MD 21152<br>Phone: 410-316-4287<br>Fax: 410-316-4499 |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| CONTACT NAME: | Monica Giguere,<br>RA Specialist |
| DATE PREPARED: | September 29, 2005 |
| DEVICE TRADE NAME: | BD Phoenix™ Automated Microbiology System -<br>Cefuroxime 0.125-4 µg/mL |
| DEVICE COMMON NAME: | Antimicrobial susceptibility test system-short incubation |
| DEVICE CLASSIFICATION: | Fully Automated Short-Term Incubation Cycle Antimicrobial<br>Susceptibility Device, 21 CFR 866.1645 |
| PREDICATE DEVICES: | VITEK® System (PMA No. N50510) and BD Phoenix™<br>Automated Microbiology System with Gatifloxacin (K020321,<br>May 23, 2002), Ofloxacin (K020323, April 14, 2002), and<br>Levofloxacin (K020322, March 27, 2002). |
| INTENDED USE: | The BD Phoenix™ Automated Microbiology System is<br>intended for the rapid identification and <i>in vitro</i> antimicrobial<br>susceptibility testing of isolates from pure culture of most<br>aerobic and facultative anaerobic Gram-negative and Gram-<br>positive bacteria of human origin. |
### DEVICE DESCRIPTION:
The BD Phoenix Automated Microbiology System (Phoenix System) is an automated system for the rapid identification (ID) and antimicrobial susceptibility testing (AST) of clinically relevant bacterial isolates. For testing Streptococcus species the system includes the following components:
- . BD Phoenix instrument and software.
- BD Phoenix panels containing biochemicals for organism ID testing and antimicrobial agents . or AST determinations.
- BD Phoenix ID Broth used for performing ID tests and preparing AST Broth inoculum. .
- BD Phoenix AST-S Broth used for performing AST tests only. .
- BD Phoenix AST-S Indicator solution added to the AST Broth to aid in bacterial growth . determination.
{1}------------------------------------------------
The Phoenix panel is a sealed and self-inoculating molded polystyrene tray with 136 micro-wells The I nooms parel to a couler anisms for susceptibility testing must be a pure culture and preliminarily identified as a Gram-negative or Gram-positive isolate. For each isolate, an inoculation equivalent to a 0.5 McFarland standard is prepared in Phoenix ID broth.
The Phoenix AST method is a broth based microdilution test. The Phoenix System utilizes a redox indicator for the detection of organism growth in the presence of an antimicrobial agent. indicator for the detection of erganism go as well as bacterial turbidity are used in the determination Mrasurements of enanges to T panel configuration contains several antimicrobial agents with a wide range of two-fold doubling dilution concentrations.
The instrument houses the panels where they are continuously incubated at a nominal temperature of 35℃. The instrument takes readings of the panels every 20 minutes. The readings are interpreted to give an identification of the isolate, minimum inhibitory concentration (MIC) values and category interpretations, S, I, R or N (susceptible, intermediate, resistant or not susceptible).
## DEVICE COMPARISON:
The BD Phoenix™ Automated Microbiology System demonstrated substantially equivalent performance when compared with the CLSI reference broth microdilution method. This premarket notification provides data supporting the use of the BD Phoenix™ Automated Microbiology System Streptococcus ID/AST or AST only Phoenix panels with this antimicrobial agent.
# SUMMARY OF SUBSTANTIAL EQUIVALENCE TESTING:
The BD Phoenix™ Automated Microbiology System has demonstrated substantially equivalent performance when compared to the CLSI reference broth microdilution method (AST panels prepared according to NCCLS M7). The system has been evaluated as defined in the FDA Draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003.
## Site Reproducibility
Intra- and inter-site reproducibility of this antimicrobial agent in the BD Phoenix System was mated at three sites using a panel of streptococcal isolates. Each site tested the isolates in triplicate on three different days using one lot of Streptococcus Phoenix panels containing this antimicrobial agent and associated reagents.
The results of the study demonstrate for the this antimicrobial agent there was an overall intra-site reproducibility of greater than 90% and an overall inter-site reproducibility greater than 95% for the streptococcal isolates tested.
{2}------------------------------------------------
## Clinical Studies
Clinical, stock and challenge isolates were tested across multiple geographically diverse sites across the United States to demonstrate the performance of the Phoenix antimicrobial susceptibility test with the Streptococcus Phoenix Panel format containing this antimicrobial agent. Phoenix System results for Challenge set isolates were compared to the expected results. Phoenix System results for clinical isolates were compared to the results obtained from the CLSI reference broth microdilution method.
The performance of the Phoenix System was assessed by calculating Essential Agreement (EA) and Category Agreement (CA) to expected/reference results for all isolates tested. Essential Agreement (EA) occurs when the BD Phoenix™ Automated Microbiology System agrees exactly or within + one two-fold dilution to the reference result. Category Agreement (CA) occurs when the BD Phoenix™ Automated Microbiology System agrees with the reference method with respect to the FDA categorical interpretive criteria (susceptible, intermediate, resistant or nonsusceptible).
Table 1 summarizes the performance for the isolates tested in this study.
#### Performance of BD Phoenix System for Streptococcal Organisms by Drug Table 1:
| Antimicrobia<br>- RELLER AST BEY LETT LES PER<br>-------------------------------------------------------------------------------------------------------------- | oncentration<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>FREE IS COLLECTION CALL ALL THE CLEAN THE | Concession and Cities of Cities Concession Compressional Concession<br>119<br>11 12 | Company of the Art Concession of Children Company of Children Company of<br>10/0<br>ह | and and the first and the state of the same of the states of the states of the states of the states of the states of the status of the status of the status of the status of t | A minimum on considered and consisted on and consisted for the first of<br>(0)p) |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1<br>efüroxime | 100<br>רד/ע) וו<br>-<br>ﻟﺴﻠ 1 1 1 | Acres of the Street Frances of Concession Company of Children<br>020<br>100 | 07 0<br>1 4 4 4 | 01<br>1 | 07 2<br> |
## Conclusions Drawn from Substantial Equivalence Studies
The data collected from the substantial equivalence studies demonstrate that testing on the BD Phoenix™ Automated Microbiology System with this antimicrobial agent is substantially equivalent as outlined in the FDA draft guidance document, "Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems; Guidance for Industry and FDA," February 5, 2003. Technological characteristics of this system are substantially equivalent to those used in the VITEK® system, which received approval by the FDA under PMA number N50510 and BD Phoenix™ Automated Microbiology System with Gatifloxacin (K020321, May 23, 2002), Ofloxacin (K020323, April 14, 2002), and Levofloxacin (K020322, March 27, 2002).
{3}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Ms. Monica Evelyn Giguere Regulatory Affairs Specialist BD Diagnostics Systems Becton, Dickinson and Company 7 Loveton Circle Sparks, MD 21152
Re: k052270
> Trade/Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent cefuroxime (0.125-4 ug/mL) - Steptococcus ID/AST or AST only Phoenix Panels Regulation Number: 21 CFR 866.1645 Regulation Name: Fully Automated Microbiology system Regulatory Class: Class II Product Code: LON Dated: August 17, 2005 Received: August 19, 2005
OCT 5 - 2005
Dear Ms. Giguere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240)276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours.
Sale, a Hom
Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Page 1 of 1
510(k) Number: KO52270
Device Name: BD Phoenix™ Automated Microbiology System for use with the antimicrobial agent cefuroxime (0.125-4 ug/mL) - Streptococcus ID/AST or AST only Phoenix Panels,
Indications for Use:
The BD Phoenix™ Automated Microbiology System is intended for in vitro quantitative determination of antimicrobial susceptibility by minimal inhibitory concentration (MIC) of most Gram-negative aerobic and facultative anaerobic bacteria isolates from pure culture for Enterobacteriaceae and Non-Enterobacteriaceae and most Gram-positive bacteria isolates from pure culture belonging to the genera Staphylococcus . Enterococcus, and Streptococcus.
This premarket notification is for the addition of the antimicrobial agent cefuroxime at concentrations of 0.125-4 µg/mL to Streptococcus ID/AST or AST only Phoenix panels. Cefuroxime has been shown to be active in vitro against most strains of microorganisms listed below, as described in the FDA-approved package insert for this antimicrobial agent.
#### Active In Vitro and in Clinical Infections Against:
Streptococcus pneumoniae Streptococcus pyogenes (and other streptococci)
Prescription Use 1 (Per 21 CFR 801.109)
Over-the-Counter Use
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Lueddi tu. Pool
Division Sign-Off
Office of In Vitro Diagnostic Device
Evaluation and Safety
K05 2270