K052213 · Brainlab AG · HAW · Nov 2, 2005 · Neurology
Device Facts
Record ID
K052213
Device Name
NAVIGATION SW HIP 3.1 ON CI
Applicant
Brainlab AG
Product Code
HAW · Neurology
Decision Date
Nov 2, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
BrainLAB VectorVision hip is intended to be an intraoperative image guided localization system. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space either on a patient's preoperative image data being processed by a VectorVision or a Ci™ workstation or on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, X-ray, MR based model of the system aids the surgeon to accurately navigate a hip endoprothesis to the preoperatively or intraoperatively planned position. Example orthopedic surgical procedures include but are not limited to: Total Joint Replacement (TJR) Revision surgery of TJR Tumor resection and bone/joint reconstruction
Device Story
VectorVision hip is an intraoperative image-guided navigation system for orthopedic surgery. It tracks surgical instruments (e.g., cup reamer, cup inserter, stem rasp, bone saw) using a passive marker sensor system. Inputs include preoperative patient image data (CT, X-ray, MR) or intraoperatively acquired bone surface landmarks. The system reconstructs 3D mechanical axes and planes of the femur and pelvis, allowing surgeons to align implants relative to patient anatomy. Used in the OR by surgeons to navigate tools and implants to planned positions. Output is visual navigation data displayed on a workstation, assisting in precise implant placement. Benefits include improved accuracy in joint replacement and reconstruction procedures.
Clinical Evidence
No clinical data provided. Safety and effectiveness were established through bench testing and validation of the system's design and development procedures.
Technological Characteristics
Intraoperative image-guided localization system; utilizes passive marker sensor tracking; workstation-based; supports CT, X-ray, and MR image data; 3D anatomical modeling via landmark acquisition; software-based navigation of surgical tools and implants.
Indications for Use
Indicated for patients requiring stereotactic orthopedic surgery where rigid anatomical structures (skull, long bone, vertebra) can be referenced to CT, X-ray, or MR models. Applicable for Total Joint Replacement (TJR), revision TJR, tumor resection, and bone/joint reconstruction.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
NOV - 2 2005
# KOSTALS
## 510 (k) Summary of Safety and Effectiveness for VectorVision® hip
## Manufacturer:
Address:
BrainLAB AG Ammerthalstrasse 8 85551 Heimstetten Germany Phone: +49 89 99 15 68 0 Fax: +49 89 99 15 68 33
| Contact Person: | Mr. Per Persson |
|-----------------|-----------------|
|-----------------|-----------------|
Summary Date: October 7th, 2005
#### Device Name:
Trade name:
DePuy hip Navigation Software
Common/Classification Name:
VectorVision, BrainLAB Image Guided Surgery System / Instrument, Stereotaxic
Predicate Device: Vector Vision® hip 3.0(K 040368) Kolibri™ Image Guided Surgery System (K 014256)
Device Classification Name: Instrument, Stereotaxic Requlatory Class: Class II
#### Intended Use:
BrainLAB VectorVision hip is intended to be an intraoperative image guided localization system. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space either on a patient's preoperative image data being processed by a VectorVision or a Ci™ workstation or on an individual 3D-model of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, X-ray, MR based model of the system aids the surgeon to accurately navigate a hip endoprothesis to the preoperatively or intraoperatively planned position.
Example orthopedic surgical procedures include but are not limited to:
Total Joint Replacement (TJR) Revision surgery of TJR Tumor resection and bone/joint reconstruction
{1}------------------------------------------------
#### Device Description:
BrainLAB VectorVision®hip is intended to enable operational planning and navigation in orthopedic surgery. It links a surgical instrument, tracked by flexible passive markers to virtual computer image space on an individual 3D-model of the patient's bone, which is either based on a patients preoperative image data or generated through acquiring multiple landmarks on the bone surface. VectorVision® hip uses the registered landmarks to navigate the needed surgical tools like cup reamer, cup inserter, stem rasp, bone saw and the implant to the planned position. If no implant data is available it is possible to provide information in order to achieve a generally targeted alignment relative to the bone orientation as defined by the operating surgeon.
VectorVision® hip allows 3-dimensional reconstruction of the relevant mechanical axes and planes of femur and pelvis and alignment of the implants. The VectorVision® hip software has been designed to read in data of implants and tools if provided by the implant manufacturer and offers to individually choose the prosthesis during each surgery. The VectorVision® hip software registers the patient data needed for planning and navigating the surgery intra-operatively within the CT free module. The System can be used to generally align tool orientations according to the anatomy described and defined by the landmarks acquired by the surgeon. Using the CT based module the patient data can be used additionally for surgery, the patient data is then is provided by the CT data stored on a standard data storage media.
#### Substantial equivalence:
VectorVision® hip has been verified and validated according to BrainLAB's procedures for product design and development. The validation proves the safety and effectiveness of the system. The information provided by BrainLAB in this 510 (k) application was found to be substantially equivalent with the predicate device Vector Vision® hip 3.0(K 040368) and Kolibri™ Image Guided Surgery System (K 014256).
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services, USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" arranged around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV - 2 2005
Per Persson Quality Manager BrainLab AG Ammerthalstrasse 8 85551 Heimstetten
Re: K052213
Trade/Device Name: VectorVision hip Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: October 7, 2005 Received: October 11, 2005
Dear Mr. Persson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Barbara BuehrmD
tr
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
### Indications for Use
510(k) Number (if known): K052213
Device Name: VectorVision hip
Indications For Use:
BrainLAB VectorVision hip is intended to be an intraoperative image quided localization system. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space either on a patient's preoperative image data being processed by a VectorVision or a Citt workstation or on an individual 3Dmodel of the patient's bone, which is generated through acquiring multiple landmarks on the bone surface. The system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, a long bone, or vertebra, can be identified relative to a CT, X-ray, MR based model of the anatomy. The system aids the surgeon to accurately navigate a hip endoprothesis to the preoperatively or intraoperatively planned position.
Example orthopedic surgical procedures include but are not limited to:
Total Joint Replacement (TJR) Revision surgery of TJR Tumor resection and bone/joint reconstruction
Prescription Use × (Per 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Barbara michu
Colatu
| CDRH, Office of Device Evaluation (ODE) |
|------------------------------------------------------------|
| (Division Sign-Off) |
| Division of General, Restorative, and Neurological Devices |
| Page | 1 of 1 |
|------|--------|
|------|--------|
| 510(k) Number | K052213 |
|---------------|---------|
|---------------|---------|
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.