AURA 70 WATT BIPOLAR ELECTROSURGICAL COAGULATORS, MODELS 28-2000 AND 28-2600
K052203 · Kirwan Surgical Products, Inc. · GEI · Oct 20, 2005 · General, Plastic Surgery
Device Facts
Record ID
K052203
Device Name
AURA 70 WATT BIPOLAR ELECTROSURGICAL COAGULATORS, MODELS 28-2000 AND 28-2600
Applicant
Kirwan Surgical Products, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 20, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
General purpose solid state bipolar generator to supply RF signal to electrosurgical hand-pieces used on soft body tissue where a wide range of tissue types, patient conditions and load impedances are encountered. Where applicable (i.e., Auraplus, 28-2600) a peristaltic irrigation pump controls flow rate individually or simultaneously while coagulating.
Device Story
Aura 70 Watt Bipolar Electrosurgical Coagulator (Models 28-2000 and 28-2600) is a solid-state generator providing RF energy for soft tissue coagulation. The 28-2600 (Auraplus) includes an integrated peristaltic irrigation pump. The device operates via footswitch activation; it interfaces with electrosurgical hand-pieces through standard 4mm female banana jacks. It features microprocessor-controlled power delivery and membrane keypads for setting adjustments. Used in clinical settings by healthcare professionals, the device provides controlled RF energy to achieve hemostasis. The microprocessor-based design simplifies construction and power management compared to previous models. The device benefits patients by providing reliable, adjustable coagulation during surgical procedures.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with ANSI/AAMI/IEC 60601-1-2 (General safety), IEC 60601-2-2 (High frequency surgical equipment), and ANSI/AAMI HF 18 (Electrosurgical devices). Software validation was performed in accordance with FDA guidance. No clinical data was required or provided.
Indicated for use as a general purpose solid state bipolar generator to supply RF signals to electrosurgical hand-pieces for coagulation of soft body tissues across a wide range of tissue types, patient conditions, and load impedances.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
80-1170, Malis™ Bipolar Electrosurgical System CMC-III
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo for KIRVAN SURGICAL PRODUCTS. The logo includes a medical symbol to the left of the company name. The image also includes the date OCT 20 2005 and the text "Page 1 of 2".
# 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS
{as required by 21 CFR, section 807.92( c )}
#### FOR
### MODELS 28-2000 & 28-2600, AURA, 70 WATT SERIES BIPOLAR ELECTROSURGICAL COAGULATORS
Common name: Electrical Surgical unit (ESU) / Electrosurgical Generator.
Classification name: Electrosurgical Cutting and Coagulation Device and Accessories (§878.4400)
Product code: GEI
The model 28-2000, Aura, 70 Watt Bipolar Electrosurgical Coagulator and the model 28-2600, Auraplus, 70 Watt Irrigating Bipolar Electrosurgical Coagulator model 20 2.000, haraphid state bipolar generators used to supply the RF signal to electrosurgical hand-pieces used on soft body tissues where a wide range of tissue types, patient conditions, and load impedances are encountered. Where applicable (i.e., Auraplus) a peristaltic irrigation pump controls flow rate individually or simultaneously while coagulating.
Technological safety and effectiveness is established by the fact that the Aura 70, like its predicates, offers well-proven, basic electrosurgical unit technology. Its fike to productivation and solid state circuitry delivers low voltage RF/bipolar energy, interfacing with electrosurgical accessories via its standard 4mm female banana jacks. The Aura 70 differs technologically from its predicates with only minor operational features designed to enhance user interface. The following are some of those features.
- While operating on both 115 and 230 VAC mains supply as its predicates, . the Aura 70 does not require the placement of an external jumper in order to select between the two different voltage levels.
Kirwan Surgical Products, Inc. 180 Enterprise Drive Marshfield, MA 02050 Phone: (781) 834-9500 Fax: (781) 834-0022 Contact: Kevin P. Prario,Regulatory Affairs Manager Date prepared: 8/5/2005
{1}------------------------------------------------
052203
Page 2 of 2
- Activation of the Aura 70 is via footswitch only, eliminating the optional s hand switch activation of its 28-1000 predicate, which is viewed as a littleused option and expendable in order to simplify construction, and limit size.
- While similarly relying on solid-state electronics, the Aura 70 has been 트 greatly simplified over some of its predicates by the use of microprocessor technology and modular medical-grade power supply.
- The Aura 70 uses membrane keypads instead of potentiometers (knobs) . as adjustable controls for tonal volume and power level settings.
Performance safety has been tested in accordance with, and found to comply with, the requirements of the applicable sections of the following standards and quidelines:
- ANSI/AAMI/IEC 60601-1-2 (2001), Medical Electrical Equipment Part 1; 를 General Requirements for the Safety.
- IEC 60601-2-2 (1998), Medical Electrical Equipment Part 2; particular . Requirements for the safety of High Frequency Surgical Equipment.
- ANSI/AAMI HF 18 (2001), American National Standard for Electrosurgical Devices
- ANSI/AAMI/ISO 14971:2000, Medical devices Application of risk . management to medical devices.
- General Principles of Software Validation; Final Guidance for Industry and 트 FDA Staff (issued January 11, 2002).
- Guidance for FDA Reviewers and Industry. Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices (issued May 29, 1998).
#### Safety and hazard analysis has determined that the hazard conditions for the Aura 70 range in the low-to-moderate level and for this reason are acceptable.
Therefore, the Aura 70 is substantially equivalent in intended use, technological safety and effectiveness, and performance to the following predicates;
- 28-1000, Kirwan 50 Watt Bipolar Generator, l
- 26-2500, Aura, 20 Watt Bipolar Electrosurgical Coagulator, 내
- RF111, Pegasys Electrosurgical Generator, and 에
- 80-1170, Malis™ Bipolar Electrosurgical System CMC-III. 트
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized eagle with three heads, representing health, human services, and well-being. The eagle is surrounded by the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" in a circular arrangement.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 2 0 2005
Kevin P. Prario Regulatory Affairs Manager Kirwan Surgical Products, Inc. 180 Enterprise Drive Marshfield, Massachusetts 02050
Re: K052203
Trade/Device Name: Models 28-2000 and 28-2600, Aura 70 Watt Bipolar Electrosurgical Coagulators Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: August 12, 2005 Received: August 16, 2005
Dear Mr. Prario:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
#### Page 2- Kevin P. Prario
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter with and in yourse FDA finding of substantial equivalence of your device to a legally premative noticated on "Tresults in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please if you deale office of Compliance at (240) 276-0115. Also, please note the regulation entitled, Conider the Ories of Comparket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small other general informational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely vours.
Barbara Buemo
to
Mark N. Melkerson Acting Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
## Indications for Use
510(k) Number (if known): _KOS2203
Device Name:
Models 28-2000 and 28-2600, Aura 70 Watt Bipolar Electrosurgical Coagulators
Indications for Use:
General purpose solid state bipolar generator to supply RF signal to General purpose oona otals used on soft body tissue where a wide electrosurgiour hardent conditions and load impedances are runge of thouse type applicable (i.e., Aurap/us, 28-2600) a peristaltic irrigation pump controls flow rate individually or simultaneously while coagulating.
X Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use __ (21 CFR 801 Subpart C)
(Please Do not WRITE Below This LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Cailane theund foruken
(Divis . Restorative. Division of Gener and Neurological Devices
Page 1 of 1
510(k) Number J052203
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.