← Product Code [MQP](/productcode/MQP) · K052096

# NUBIC SPINAL IMPLANT (K052096)

_Signus Medical, LLC · MQP · Aug 16, 2005 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K052096

## Device Facts

- **Applicant:** Signus Medical, LLC
- **Product Code:** [MQP](/productcode/MQP.md)
- **Decision Date:** Aug 16, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3060
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The NUBIC™ Spinal Implant is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to LS) and is intended for use with supplemental internal fixation. The NUBIC™ may be implanted singularly or in pairs.

## Device Story

NUBIC™ Spinal Implant is a vertebral body replacement device for the thoracolumbar spine. It is implanted by a surgeon to replace bone resected due to tumor or trauma. The device is used in conjunction with supplemental internal fixation systems (e.g., plates or rods) to provide stability. It may be implanted singularly or in pairs. The device functions as a structural spacer to restore spinal column height and support. It is intended for prescription use only.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on design, material, and performance characteristic similarities to the predicate device.

## Technological Characteristics

Vertebral body replacement implant for thoracolumbar spine. Materials and performance characteristics are equivalent to the predicate device (K043316).

## Regulatory Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

## Predicate Devices

- RABEA™ ([K043316](/device/K043316.md))

## Reference Devices

- SIGNUS CONKLUSION System
- DePuy AcroMed Kaneda SR
- DePuy AcroMed University Plate
- DePuy AcroMed M2
- DePuy AcroMed ISOLA
- DePuy AcroMed VSP
- DePuy AcroMed Moss
- DePuy AcroMed TiMX
- DePuy AcroMed Profile

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image contains the words "Public Health Service" in bold, black font. The words are stacked on top of each other, with "Public Health" on the top line and "Service" on the bottom line. The background of the image is white.

Image /page/0/Picture/2 description: The image is a seal for the Department of Health and Human Services (HHS) in the USA. The seal is circular and contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" around the top half of the circle. In the center of the seal is a stylized symbol of three human profiles facing to the right, stacked on top of each other.

AUG 2 5 2005

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Signus Medical, LLC. c/o Ms. Tracy Gray Principal Consultant Alquest, Inc. 4050 Olson Memorial Highway, Suite 350 Minneapolis, Minnesota 55422

Re: K052096

Trade/Device Name: NUBIC™ Spinal Implant Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: II Product Code: MQP Dated: July 29, 2005 Received: August 4, 2005

Dear Ms. Gray:

This letter corrects our substantially equivalent letter of August 16, 2005. The Company title has been changed to reflect the correct owner of the 510(k), Signus Medical. LLC., as indicated in your 510(k) submission.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Ms. Tracy Gray

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to continue marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson

Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use Page

510(k) Number (if known):

Device Name: NUBIC™ Spinal Implant

Indications for Use:

The NUBIC™ Spinal Implant is indicated for use to replace a vertebral body that has been resected or excised due to tumor or trauma/fracture. The device is intended for use as a vertebral body replacement in the thoracolumbar spine (from T1 to LS) and is intended for use with supplemental internal fixation.

GS2046

The NUBIC™ may be implanted singularly or in pairs.

The supplemental internal fixation systems that may be used with the NUBIC™ Spinal Implant include, but are not limited to, SIGNUS CONKLUSION System, DePuy AcroMed titanium plate or rod systems (Kaneda SR, University Plate, M2, ISOLA, VSP, Moss, TiMX, and Profile).

Prescription Use X (Per 21 CFR 801 Subpart D) AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices

| SIGNUS Medical LLC. 510(k) Number  | K052096           |
|------------------------------------|-------------------|
| NUBIC <sup>TM</sup> Special 510(k) | Appendix B Page 2 |

{3}------------------------------------------------

## 510(k) Summary

{4}------------------------------------------------

|             | Implanted using the same surgical techniques and equipment; Used in conjunction with the same types of supplemental internal fixation systems; The same manufacturing environment; The same sterilization process; and The same packaging configurations.              In summary, the NUBICTM, as described in this submission is, in the opinion of Signus Medical, LLC., substantially equivalent to the predicate device. |
|-------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Conclusion: | The NUBICTM as modified in this submission, is substantially equivalent to the predicate device, RABEATM cleared under K043316. This conclusion is based upon the similarities of the devices in terms of functional design, indication for use, principles of operation, materials, and performance characteristics.                                                                                                         |

---

**Source:** [https://fda.innolitics.com/device/K052096](https://fda.innolitics.com/device/K052096)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
