K052067 · Multigon Industries, Inc. · JAF · Aug 29, 2005 · Radiology
Device Facts
Record ID
K052067
Device Name
DOPPLER GUIDED PROCTOSCOPE, MODEL 500H
Applicant
Multigon Industries, Inc.
Product Code
JAF · Radiology
Decision Date
Aug 29, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1540
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The 500H is a Doppler guided proctoscope which is used to detect blood vessels supplying hemorrhoids and for performing HAL (Hemorrhoid Arterial Ligation) for Class II and Class III hemorrhoids. It is to be used by physicians in hospitals, clinics, and physicians offices by prescriptions or doctors orders.
Device Story
Device consists of disposable proctoscope with removable fiber optic light pipe and removable 8 MHz continuous wave (CW) Doppler probe; paired with external detector unit containing loudspeaker and light source. Physician inserts proctoscope into rectum; rotates device to locate hemorrhoidal artery via audible arterial blood flow signals; maximizes signal to ensure artery is positioned under probe and within proctoscope slit. Physician performs ligation; rotates device 360 degrees to confirm all branches ligated. Used in hospitals, clinics, and physician offices. Provides audible feedback only; no display or automated calculation. Benefits patient by enabling targeted arterial ligation for hemorrhoid treatment.
Clinical Evidence
Bench testing only. Device underwent safety and performance testing to ensure compliance with functional specifications and industry standards (UL 2601-1, IEC 60601-1-2, IEC 60601-2-37). Acoustic output measurements were performed and confirmed to be below pre-amendment levels for vascular applications (Track 1 device).
Technological Characteristics
8 MHz CW Doppler transducer; fiber optic illumination; disposable proctoscope. Operates as a standalone unit with loudspeaker output. Designed to meet UL 2601-1, IEC 60601-1-2, IEC 60601-2-37, and AIUM/NEMA UD 2/UD 3 standards. No software or firmware.
Indications for Use
Indicated for physicians to detect hemorrhoidal arteries and perform Hemorrhoid Arterial Ligation (HAL) in patients with Class II and Class III hemorrhoids.
Regulatory Classification
Identification
A nonfetal ultrasonic monitor is a device that projects a continuous high-frequency sound wave into body tissue other than a fetus to determine frequency changes (doppler shift) in the reflected wave and is intended for use in the investigation of nonfetal blood flow and other nonfetal body tissues in motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Model KM-25 (k951449)
Hemo-Dop (k041915)
Reference Devices
Cuda Products Corp/ Sun Optics Light Source Model I-150 (k893491)
Cuda Products Corp/ Sunoptics fiber optic cable (k901035)
Multigon Model 500A Vascular Spectrum Analyzer with CW Doppler (k860435)
Submission Summary (Full Text)
{0}------------------------------------------------
Ko52067 12
# AUG 2 9 2005
500H 510k Submission
500H Doppler Guided Proctoscope Multigon Industries, Inc.
#### SUMMARY
This summary of 510k safety and effectiveness information is being submitted in accordance with 21CFR part 807.92
1. Submitters name, address, phone number, contact person and preparation date:
Name: Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 ext. 27 914 376 6111 Fax: Email: wstern@multigon.com Responsible person: William Stern
Official correspondent:
William Stern Multigon Industries, Inc. 1 Odell Plaza Yonkers, N.Y. 10701 Phone: 914 376 5200 ext. 27 914 376 6111 Fax: Email: wstern@multigon.com
Date of Preparation: 7/1/05
2. Proprietary Name: Model 500H Doppler Guided Proctoscope
Common /Usual Name: Non Fetal Ultrasonic Monitor Diagnostic Ultrasound Transducer Classification Name: 21 CFR892.1540 Non Fetal Ultrasonic Monitor 21 CFR892.1570 Diagnostic Ultrasound Transducer
{1}------------------------------------------------
500H 510k Submission
500H Doppler Guided Proctoscope Multigon Industries,Inc.
Classification Number: 90JAF 90ITX
3. Substantially Equivalent Devices
Multigon Industries, Inc. believes that the Model 500H Doppler Guided Proctoscope and Light source and transducers is substantially equivalent to the following cleared devices:
| Trade or Proprietary Name | Manufacturer | 510(k) Number |
|---------------------------|--------------------------------|---------------|
| Model KM-25 | Koven Technology Inc. | k951449 |
| Hemo-Dop | DWL Elektronische Systeme,GmbH | k041915 |
The 500H includes Cuda Products Corp/ Sun Optics Light Source Model I-150 510k # k893491 and Cuda Products Corp/ Sunoptics fiber optic cable 510k# k901035 repackaged for convenience in the 500H.
The 500H includes a CW 8 mHz Doppler which is an updated version of the 8 mHz Doppler electronics module of the model Multigon Model 500A Vascular Spectrum Analyzer with CW Doppler 510k# k860435 which has been repackaged as the 8 mHz CW Doppler without the spectrum analyzer in the 500H for convenience. The 500H has the functionality only of producing the 8 mHz CW Doppler and insonating blood vessels. It enables the user to listen to the 8 mHz CW Doppler sound without any display, or calculation function.
Comparison of Multigon Model 500H Doppler Guided Proctoscope and the two predicate devices cited above: Koven Technology, Inc. Model KM-25 and DWL Elektronische Systeme GmbH Hemo-Dop
| 500H | KM-25 | Hemo-Dop |
|----------------------------------------------------------------|------------------------------------------------------|-----------------------------------------------------------------|
| Uses 8mHz CW Doppler<br>Sound to locate<br>Hemorrhoidal artery | Same | Uses 8 mHz Pulsed<br>Doppler to locate<br>Haemorrrhoidal artery |
| 8 mHz Doppler probe<br>is removable | 8 mHz Doppler probe<br>is built into the proctoscope | 8 mHz probe is built<br>in to the proctoscope |
| Proctoscope is disposable | Proctoscope is not disposable | Proctoscope is not disposable |
12
{2}------------------------------------------------
500H 510k Submission
500H Doppler Guided Proctoscope Multigon Industries, Inc.
| Proctoscope has fiber<br>Optic illumination | Has incandescent<br>bulb illumination<br>built in scope | Has bulb illumination<br>built in scope |
|---------------------------------------------|---------------------------------------------------------|-----------------------------------------|
| | | C |
Proctoscope has a port Same for hemorrhoidal artery ligation
Same
## 4. DEVICE DESCRIPTION
The model 500H Doppler guided proctoscope consists of two parts . The first part is a Doppler probe in a proctoscope that are used integrally with each other and are not intended to be used separately. The Doppler guided proctoscope has a removable fiber optic light pipe, and a removable 8 mHz CW Doppler probe. The second part is a combination 8 mHz continuous wave (CW) Doppler detector with loudspeaker and a light source.
The Doppler guided proctoscope is inserted in the rectum and rotated until one of the branches of the hemorrhoidal artery is located. The artery is located when the user hears the classical arterial blood flow sound. The Doppler guided proctoscope is rotated such that the arterial sound is maximized thus insuring that the hemorrhoidal artery is under the Doppler probe and in the Doppler guided proctoscope's rectangular slit. The physcician or surgeon can then ligate the artery . The Doppler guided proctoscope is then rotated until the next hemorrhoidal artery branch is detected. The process is repeated until all detected hemorrhoidal arteries are located and ligated. After ligation the Doppler guided proctoscope can be rotated 360 degrees to make sure that all of the hemorrhoidal arterial branches have been ligated.
#### 5. PERFORMANCE STANDARDS
No performance standards have been established for the 500H Doppler Guided Proctoscope under section 514 of the Federal Food and Drug Act. However the 500H Doppler Guided Proctoscope has been designed to meet the following standards: UL 2601-1 Safety Requirements for Medical Equipment AIUM/NEMA UD 2 Standard for Real Time Dsplay of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment AIUM/NEMA UD 3 Standard for Real Time Display of Thermal and Mechanical Output Indices on Diagnostic Ultrasound Equipment IEC 1157 Declaration of Acoustic Power IEC60601-1-2 IEC60601-2-37
{3}------------------------------------------------
## 6. INDICATIONS FOR USE
The 500H is a Doppler guided proctoscope which is used to detect blood vessels supplying hemorrhoids and for performing HAL (Hemorrhoid Arterial Ligation) for Class II and Class III hemorrhoids. It is to be used by physicians in hospitals, clinics, and physicians offices by prescriptions or doctors orders.
### 7. CONTRA-INDICATIONS
None known at this time.
## 8. COMPARISON TO PREDICATE DEVICES
The 500H Doppler guided proctoscope has the same device characteristics as the approved predicate devices listed in item 3 above with the commonality of 8 mHz Doppler , principles of operation, and a Doppler guided proctoscope.
### 9. TEST DATA
The Model 500H Doppler guided proctoscope has been subjected to extensive safety, performance testing, and validation before release. Final testing of the 500H included various performance tests designed to ensure that the device met all of its functional specifications. Safety tests have been performed to ensure the device complies with applicable industry and safety standards.
The Model 500H Doppler guided proctoscope device labeling includes instructions for safe and effective use, warnings, cautions and guidance for use. It has therefore shown to be safe and effective.
### 10. ACOUSTIC OUTPUT
This is a summary table of the acoustic output of the 500H. The 500H acoustic output is below preamendment levels for vascular applications and is a Track 1 device.
{4}------------------------------------------------
500H Doppler Guided Proctoscope Multigon Industries, Inc.
| | Acoustic Output 500H | |
|----------------|------------------------------------|-------------------------|
| System: | 500H | |
| Probes: | 8 mHz ( center frequency 7.98 mHz) | |
| | Three probes were tested. | |
| | MI | Ispta.3 (mW/cm squared) |
| Mean value | .0075 | 13.5 |
| Std. Dev. (Sx) | 0.0003 | 1.3 |
| Limit(X) | 0.089 | 19.1 |
### 11. SOFTWARE
The 500H has no software or firmware associated with it.
## 12. LITERATURE REVIEW
A review of the literature pertaining to the safety of the 500H Doppler guided proctoscope has been conducted and appropriate safeguards have been incorporated in the design of the 500H Doppler guided proctoscope.
# 13. CONCLUSIONS
The conclusion drawn from these tests is that the 500H Doppler guided proctoscope with an 8 mHz CW Doppler transducer is substantially equivalent in safety and efficacy to the predicate devices listed in item 3 above.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
AUG 2 9 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Public Health Service
Mr. William Stern Official Correspondent Multigon Industries, Inc. 1 Odell Plaza YONKERS NY 10701
Re: K052067
Trade Name: Model 500H Doppler Guided Proctoscope Regulation Number: 21 CFR 892.1540 Regulation Name: Nonfetal ultrasound monitor Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic ultrasonic transducer Regulatory Class: II Product Code: JAF and ITX Dated: July 28, 2005 Received: August 3, 2005
Dear Mr. Stern:
We have reviewed your Section 510(k) premarket notification of intent to market the device we have revewed your Section 510(K) prothantee is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosury to regarly manated to of the Medical Device Americal Device commerce provision in May 20, 1770, the encordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls Cosment Act (Act. The general controls provisions of the Act include requirements for annual provisions of the Act. "The general controls provider for the reactice, labeling, and prohibitions against misbranding and adulteration.
This determination of substantial equivalence applies to the following transducers intended for use I ins determination of Substantial equily and
#### Transducer Model Number
## 8 MHz Transducer Model 5008
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classince (see above) into existing major regulations affecting your device can be may be subject to such additional controllars, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that Flease be advised that 1 DAY's issualled of aevice complies with other requirements of the Act or any FDA hia statutes and regulations administered by other Federal agencies. You must comply with all
{6}------------------------------------------------
Page 2 -- Mr. Stern
the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); the Act STEQuirements, mortainly, but not miner mine requirements as set forth in the quality labeling (21 CFR Part 601), good manufacturing praviles ble, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This determination of substantial equivalence is granted on the condition that prior to shipping the I his device, you submit a postclearance special report. This report should contain complete information, including acoustic output measurements based on production line devices, requested in information, including acousine output measurenter 30, 1997 "Information for Manufacturers Seeking Appendix G, (Cherosed) of the Contr & Bound Systems and Transducers." If the special report is Marketing Clearance of Diagnostic Onluss (e.g., acoustic output greater than approved levels), then Incomplete of contains maccopable variable (org, assubts which as a result may be considered adulterated or misbranded.
The special report should reference the manufacturer's 510(k) number. It should be clearly and The special reports.com and should be submitted in duplicate to:
> Food and Drug Administration Center for Devices and Radiological Health Document Mail Center (HFZ-401) 9200 Corporate Boulevard Rockville, Maryland 20850
This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific administrat (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, general information on your Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html
If you have any questions regarding the content of this letter, please contact Mr. Rodrigo C. Perez at (301) 594-1212.
Sincerely vours,
Nancy C Brogdon
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure(s)
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): 1052067
Device Name: Guided Proctoscope Model 500H Dippler The Model 500H Doppler Guided Proctoscope Indications For Use: is used to detect blood vessels supplying hemorrhoids and for performing HAL ( Hemorrhoid Arterial Ligation) for Class II and Class III hemorrhoids. It is to be used by physicians in hospitals, clinics, and physician's offices by prescriptions or doctor's orders.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number K022067
Page 1 of
{8}------------------------------------------------
#### Diagnostic Ultrasound Indications for Use Form
# Fill out one form for each ultrasound system and each transducer.
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| | | Mode of Operation | | | | | | | | | |
|----------------------------------------------------------------------------------------------------|---|-------------------|---|-----|-----|------------------|----------------------|------------------------------|-----------------------|--------------------|--|
| Clinical Application | A | B | M | PWD | CWD | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(specify) | Other<br>(specify) | |
| Ophthalmic | | | | | | | | | | | |
| Fetal | | | | | | | | | | | |
| Abdominal | | | | | | | | | | | |
| Intraoperative (specify) | | | | | | | | | | | |
| Intraoperative Neurological | | | | | | | | | | | |
| Pediatric | | | | | | | | | | | |
| Small Organ (specify) | | | | | | | | | | | |
| Neonatal Cephalic | | | | | | | | | | | |
| Adult Cephalic | | | | | | | | | | | |
| Cardiac | | | | | | | | | | | |
| Transesophageal | | | | | | | | | | | |
| Transrectal | | | | | | | | | | | |
| Transvaginal | | | | | | | | | | | |
| Transurethral | | | | | | | | | | | |
| Intravascular | | | | | | | | | | | |
| Peripheral Vascular | | | | | N | | | | | | |
| Laparoscopic | | | | | | | | | | | |
| Musculo-skeletal<br>Conventional | | | | | | | | | | | |
| Musculo-skeletal Superficial | | | | | | | | | | | |
| Other (specify) | | | | | | | | | | | |
| N= new indication; P= previously cleared by FDA; E= added under Appendix E | | | | | | | | | | | |
| Additional Comments: 8 MHz Transducer Model 5008 for the<br>Model 1500H Doppler Guided Proctoscope | | | | | | | | | | | |
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Brogdon
(Division Sign-Off) Division of Reproductive, Abdominal, and Radiological Devices
Prescription Use (Per 21 CFR 801.109)
510(k) Number F-3
. . . .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.