SMISSON-CARTLEDGE THERMAL INFUSION SYSTEM, MODEL TIS-1200
K052055 · Smisson-Cartledge Biomedical, LLC · LGZ · Oct 26, 2006 · General Hospital
Device Facts
Record ID
K052055
Device Name
SMISSON-CARTLEDGE THERMAL INFUSION SYSTEM, MODEL TIS-1200
Applicant
Smisson-Cartledge Biomedical, LLC
Product Code
LGZ · General Hospital
Decision Date
Oct 26, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5725
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Full range from slow feed to rapid, high flow infusion of: crystalloid, colloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery warmed fluid to rewarm patients after surgery or for hypothermia warmed fluid for irrigation in urology procedures.
Device Story
Portable tabletop/pole-mount volumetric infusion pump; used in ambulances, helicopters, ERs, and ORs. Inputs: IV bags or cardiotomy equipment via sterile single-use disposable cassette kit. Operation: continuous infusion (10 mL/hr to 1200 mL/min) or discrete bolus delivery (up to 1 L). Features: footswitch for hands-free start/stop; AC power for fluid warming; lithium-ion battery for transport (heating disabled). Output: controlled delivery of warmed or ambient fluids to patient. Healthcare providers use device to manage fluid resuscitation and temperature regulation; benefits include rapid volume replacement and hypothermia prevention.
Clinical Evidence
Bench testing only. Performance testing verified mechanical characteristics and infusion capabilities against predicate devices. Biocompatibility testing conducted per ISO-10993.
Technological Characteristics
Volumetric infusion pump; tabletop/pole-mount form factor. Materials: biocompatible (ISO-10993 compliant). Energy: AC power (heating enabled) and lithium-ion battery (transport mode). Sterilization: Ethylene Oxide (EO) for disposable components. Connectivity: standalone.
Indications for Use
Indicated for patients suffering from blood loss due to trauma or surgery requiring volume replacement (crystalloid, colloid, or blood products), patients requiring rewarming after surgery or for hypothermia, and patients requiring warmed irrigation fluid for urology procedures.
Regulatory Classification
Identification
An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.
Smiths Level 1 H-1200 Fast Flow Fluid Warmer System (BK020043)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 16, 2021
Smisson-Cartledge Biomedical LLC Julie Stephens Correspondent 111 Laurel Ridge Drive Alpharetta, Georgia 30004
Re: K052055
Trade/Device Name: Smisson-Cartledge TIS-1200 Thermal Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion pump Regulatory Class: Class II Product Code: LGZ, FRN, FPA
Dear Julie Stephens:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated January 10, 2017. Specifically, FDA is updating this SE Letter because FDA has identified an additional product code to more clearly categorize your device technology.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact CAPT Alan Stevens, OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices, (301) 796-6294, Alan.Stevens@fda.hhs.gov.
Sincerely,
Alan M.
Stevens -S3
CAPT Alan Stevens Acting Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
U.S. Food & Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 www.fda.gov
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a bird in flight, composed of three curved lines.
JAN 1 0 2017
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Smisson-Cartledge Biomedical, LLC C/O Ms. Julie Stephens Consultant for Smisson-Cartledge Biomedical, LLC Regulatory Resources Group. Incorporated 111 Laurel Ridge Drive Alpharetta, Georgia 30004
Re: K052055
Trade/Device Name: Smisson-Cartledge TIS-1200 Thermal Infusion System Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: Class II Product Code: LGZ, FRN Dated: October 18, 2006 Received: October 19, 2006
Dear Ms. Stephens:
This letter corrects our substantially equivalent letter of October 26, 2006.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{2}------------------------------------------------
Page 2 - Ms. Julie Stephens
CFR Part 807); labeling (21 CFR Parts 801 and 809 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809] ), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Erin I. Keith -S
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known): K052055
Device Name: Smisson-Cartledge TIS-1200 Thermal Infusion System
Indications For Use:
Full range from slow feed to rapid, high flow infusion of:
- crystalloid, colloid, or blood product, including packed red blood cells, as volume . replacement for patients suffering from blood loss due to trauma or surgery
- warmed fluid to rewarm patients after surgery or for hypothermia .
warmed fluid for irrigation in urology procedures .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
{4}------------------------------------------------
## Exhibit J - 510(k) SUMMARY
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements under 21 CFR 807.92.
| Submitted By: | Smisson-Cartledge Biomedical LLC<br>502 Mulberry Street, Second Floor<br>Macon, GA 31201<br>Phone: (478) 744-9992 |
|---------------|-------------------------------------------------------------------------------------------------------------------|
|---------------|-------------------------------------------------------------------------------------------------------------------|
Contact Person: Julie Stephens, President/Consultant Regulatory Resources Group, Inc.
510(k) Number: K052055
Date Prepared: July 26, 2005
Device Name and Classification:
| Trade/Proprietary Name: | Smisson-Cartledge TIS-1200 Thermal Infusion System |
|-------------------------|----------------------------------------------------|
| Common Name: | Infusion Pump |
| Classification Name: | Infusion Pump |
| Product Code: | LGZ: FRN: FPA |
## Legally Marketed Predicate Device:
Belmont Fluid Management System (FMS2000) - 510(k) # K032778. K032674. K992672. K983975. K972284 Smiths Level 1® H-1200 Fast Flow Fluid Warmer System - 510(k) # BK020043
## Device Description:
The Smisson-Cartledge TIS-1200 Thermal Infusion System is a portable tabletop or polemount device intended for use in the ambulance, medical helicopter, hospital emergency room and operating room environments. (See Exhibits A and B for photographs and drawings) The system consists of an infusion device and a compatible single-use sterilized disposable set with supply lines capable of interfacing with intravenous (IV) bags or optional-use cardiotomy equipment. It also includes a footswitch to allow hands-free user-controlled stopping and starting of fluid delivery. The Smisson-Cartledge TIS-1200 Thermal Infusion System is a volumetric pump capable of continuous infusion (up to approximately 100 L at a rate of from 10 mL per hour to 1200 mL per minute) and discrete bolus delivery. When the system is set to Bolus mode, the user can select a predetermined delivery volume and a default or adjustable rate and deliver a fixed bolus of up to 1 L. When connected to alternating current (AC) power, the Smisson-Cartledge TIS-1200 Thermal Infusion System can deliver fluids at body temperature in certain modes. It can also be set to run on battery power with heating capabilities disabled to allow transport of the patient. A lithium-ion battery pack provides power backup. The unit (infusion pump) and Large Volume Reservoir Holder are provided non sterile. The Disposable Cassette Kit and Large Volume Reservoir are provided sterile, non-pyrogenic and are single-use only. They are sterilized by Ethylene Oxide (EO) sterilization method.
{5}------------------------------------------------
## Exhibit J - 510(k) SUMMARY
## Indications for Use:
Full range from slow feed to rapid, high flow infusion of crystalloid, or blood product, including packed red blood cells, as volume replacement for patients suffering from blood loss due to trauma or surgery. Full range from slow feed to rapid, high flow infusion of warmed fluid to rewarm patients after surgery or for hypothermia. Full range from slow feed to rapid, high flow infusion of warmed fluid for irrigation in urology procedures.
## Similarities and Differences to the Predicate Devices:
## Similarities
The Smisson-Cartledge TIS-1200 Thermal Infusion System has the same basic mechanical characteristics to infuse and warm liquid products into a patient as the predicate devices, and the indications for use are the Belmont predicate device.
## Differences
The Smisson-Cartledge TIS-1200 Thermal Infusion System utilizes different materials from the predicate devices; however, the materials are biocompatible.
## Summary of Testing:
The Smisson-Cartledge TIS-1200 Thermal Infusion System has the same indications for use, principles of operation, and mechanical characteristics as the predicate devices that were previously cleared for market under a 510(k). These conclusions were verified in performance / bench testing as summarized within the 510(k).
The Smisson-Cartledge TIS-1200 Thermal Infusion System device differs only in its materials used. All biocompatibility testing, as required, complies with ISO-10993 "Biological Evaluation of Medical Devices".
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.