← Product Code [MQN](/productcode/MQN) · K051946

# MODUS MODULAR DISTRACTION OSTEOGENESIS SYSTEM (K051946)

_Medartis, Inc. · MQN · Aug 22, 2005 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051946

## Device Facts

- **Applicant:** Medartis, Inc.
- **Product Code:** [MQN](/productcode/MQN.md)
- **Decision Date:** Aug 22, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.4760
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Modus Modular Distraction Osteogenesis System is a subcutaneous distractor system intended for use in fractures, osteotomies and arthrodeses including: treatment of hemifacial microsomia, mandibular hypoplasia, microglossia, defects of the alveolar ridge, atrophic bone segments in the mandible and maxilla, mandibular symphysis widening, growth disturbances of the ascending corpus/ascending ramus, LeFort I and LeFort II/III advancements.

## Device Story

Subcutaneous distraction osteogenesis system; used for maxillofacial bone reconstruction, fractures, and osteotomies. System components: titanium plates, distraction cylinders (uni/bi-directional), and flexible extensions. Surgeon implants plates and cylinders to facilitate gradual bone distraction; allows vector adjustment via intra-operative bending. Used in clinical settings by surgeons. Provides mechanical force for bone lengthening and correction of skeletal deformities; benefits patients by addressing congenital or acquired maxillofacial growth disturbances and defects.

## Clinical Evidence

No clinical data.

## Technological Characteristics

Implant plates/ARS distractors: CP titanium Grade 4 (ASTM F 67). Distraction cylinders/extensions: implant steel (ASTM F 138 or F 139). System includes plates, distraction cylinders, and flexible extensions. Mechanical distraction principle. No software or electronic components.

## Regulatory Identification

A bone plate is a metal device intended to stabilize fractured bone structures in the oral cavity. The bone segments are attached to the plate with screws to prevent movement of the segments.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k): Summary

Modus® Modular Distraction Osteogenesis System

# 510(k): Summary

### ADMINISTRATIVE INFORMATION

| Manufacturer Name:         | Medartis, Inc.<br>127 W Street Road, Suite 203<br>Kennett Square, PA 19348<br>Telephone (610) 961-6101<br>FAX (610) 961-6108                             |
|----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Contact:          | Kate Gehret                                                                                                                                              |
| Representative/Consultant: | Floyd G. Larson<br>PaxMed International, LLC<br>11234 El Camino Real, Suite 200<br>San Diego, CA 92130<br>Telephone (858) 792-1235<br>FAX (858) 792-1236 |

#### Classification Name: External Mandibular Fixator and/or Distractor Modus Modular Distraction Osteogenesis System Trade/Proprietary Name: Mandibular Fixator and Distractor Common Name:

## ESTABLISHMENT REGISTRATION NUMBER

Medartis, Inc. has submitted an Establishment Registration to FDA. The Establishment Registration number has not yet been assigned. The owner/operator number for Medartis AG, the parent company of Medartis, Inc., is 9033581.

#### DEVICE CLASSIFICATION

FDA has classified bone fixation plates as Class II devices (21 CFR 872.4760). The product code for external mandibular fixator and/or distractor is MQN. This device classification is reviewed by the Dental Devices Branch.

#### INTENDED USE

The Modus Modular Distraction Osteogenesis System is a subcutaneous distractor system intended for use in fractures, osteotomies and arthrodeses including: treatment of hemifacial microsomia, mandibular hypoplasia, microglossia, defects of the alveolar

{1}------------------------------------------------

K 05 1946

#### 510(k): Summary

Modus® Modular Distraction Osteogenesis System

ridge, atrophic bone segments in the mandible and maxilla, mandibular symphysis widening, growth disturbances of the ascending corpus/ascending ramus, LeFort I and LeFort II/III advancements.

#### DEVICE DESCRIPTION

#### Modus Plates

The Modus plates are used with the Modus Distraction Cylinders for various applications of the maxillofacial bone structures. The Modus plates are offered in a variety of sizes and are designed to fit the anatomy of the mandible and/or maxilla and allow adjustment of the vector by intra-operative bending if indicated.

#### Modus Distraction Cylinders

The Modus Distraction Cylinders are used with the Modus plates and screws for gradual bone distraction osteogenesis and for corrective osteotomies of the mandible and maxilla. The Modus distraction cylinders are available in uni-directional or bi-directional with multiple distraction lengths.

#### Modus Flexible Extensions

Flexible extensions are to be used with distraction cylinders if needed. The extensions allow additional access for distraction of the surgical site.

#### Material Composition

The implant plates and ARS distractors are made of CP titanium Grade 4 and comply with ASTM F 67 standards for unalloyed titanium for surgical implant applications. The MDO distraction cylinders and flexible extensions are made of implant steel that complies with ASTM F 138 or ASTM F 139.

#### EQUIVALENCE TO MARKETED PRODUCT

For the purposes of FDA's regulation of medical devices, the Modus Distraction System is substantially equivalent in indications and design principles to the predicate devices, that have been determined by FDA to be substantially equivalent to pre-amendment devices.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three lines forming its body and wings. The eagle is enclosed within a circle that contains the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" in a circular arrangement. The text is written in all capital letters.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 2 2 2005

Medartis, Incorporated C/O Mr. Floyd G. Larson PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, California 92130

Re: K051946

Trade/Device Name: Modus Modular Distraction Osteogenesis System Regulation Number: 21 CFR 872.4760 Regulation Name: Bone Plate Regulatory Class: II Product Code: MQN Dated: July 15, 2005 Received: July 19, 2005

Dear Mr. Larson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice. labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Mr. Larson

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susan Runner

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): _ K 051946

Device Name: Modus Modular Distraction Osteogenesis System

Indications for Use:

The Modus® Modular Distraction Osteogenesis System is a subcutaneous distractor system intended for use in fractures, osteotomies and arthrodeses including: treatment of hemifacial microsomia, mandibular hypoplasia, microglossia, defects of the alveolar ridge, atrophic bone segments in the mandible and maxilla, mandibular symphysis widening, growth disturbances of the ascending corpus/ascending ramus, and LeFort I, and LeFort II/III advancements.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

SkenRunger

Division Sian-Off) Division of Anesthesiology, General Hospital. Infection Control, Dental Devic

510(k) Number. K051946

---

**Source:** [https://fda.innolitics.com/device/K051946](https://fda.innolitics.com/device/K051946)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
