K051865 · Terumo Medical Corp. · MEG · Sep 27, 2005 · General Hospital
Device Facts
Record ID
K051865
Device Name
TERUMO SURGUARD 2 SAFETY NEEDLE OR SIMILAR
Applicant
Terumo Medical Corp.
Product Code
MEG · General Hospital
Decision Date
Sep 27, 2005
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TERUMO® SurGuard2™ SAFETY NEEDLE device is intended for use in the aspiration and injection of fluids for medical purposes. The Terumo Safety Needle is compatible for use with standard luer slip and luer lock syringes. Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
Device Story
Hypodermic needle with hinged safety sheath attached to connector hub; optionally includes hypodermic syringe. Used for fluid aspiration/injection. Manual operation: user presses sheath against firm surface to lock over needle post-withdrawal. Prevents sharps injury. Used in clinical settings by healthcare professionals. Output is physical fluid delivery/aspiration; safety feature provides physical barrier. Benefits patient/clinician by reducing needlestick risk.
Clinical Evidence
Bench testing only. Performance testing included activation/deactivation force, puncture resistance, sheath/collar removal force, sheath radial force, protector fit, adhesive hold, and simulated use studies. Syringe testing included leakage, plunger gasket fit, nominal graduation capacity, deadspace, and conical fitting. Biocompatibility testing performed per ISO-10993.
Technological Characteristics
Hypodermic needle (18-30 gauge, 3/8"-1 1/2" length) with hinged safety sheath. Materials identical to predicates. Sterilization via electron beam (ANSI/AAMI/ISO 11137). Externally communicating, blood path indirect, limited duration (<24 hr). Manual operation; no energy source.
Indications for Use
Indicated for aspiration and injection of fluids for medical purposes using standard luer slip or luer lock syringes. Includes integrated manual safety shield to minimize accidental needlestick risk post-withdrawal.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
Predicate Devices
TERUMO® SURGUARD2™ SAFETY NEEDLE (18 to 25g) (K031453)
TERUMO® SURGUARD2™ SAFETY NEEDLE (26 to 30g) (K040531)
{0}------------------------------------------------
## SUMMARY OF SAFETY AND EFFECTIVENESS INFORMATION PERTAINING TO SUBSTANTIAL EQUIVALENCE
#### Device Name A.
### Proprietary Name
TERUMO® SurGuard2™ Safety Needle or similar proprietary name
### Classification Name
Hypodermic Single Lumen Needle (880.5570) with antistick
Product Code: 80 FMI / 80 MEG
Classification: Class II
### Common Name
Hypodermic needle with safety sheath or needle with needle protection device
#### B. Intended Use
The TERUMO® SurGuard2™ SAFETY NEEDLE device is intended for use in the aspiration and injection of fluids for medical purposes. The Terumo Safety Needle is compatible for use with standard luer slip and luer lock syringes.
Additionally, after withdrawal of the needle from the body, the attached needle safety shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
{1}------------------------------------------------
### C. Device Description
The SurGuard2™ Safety Needle consists of a hypodermic needle with a hinged safety sheath attached to the connector hub with or without an attached hypodermic syringe. The safety sheath contains a locking mechanism which is activated when the sheath is manually pressed over the needle immediately after use and just prior to disposal to minimize the possibility of sharps injury. The safety sheath is activated with a one-handed operation by pressing the sheath against a firm surface. The locking mechanism is located at a designated position within the body of the short or long sheath appropriate for the needle size it is to contain. The needle gauge sizes fall within the 18 to 30gauge range and the needle lengths are 3/8" to 1 ½". The hinge feature allows the user to set the sheath to the desired position for use. For user convenience, when the needle is in the "bevel up" position the sheath is located to the right. The SurGuard2™ Safety Needle will be individually packaged and sterilized by electron beam as a safety needle only or as a safety needle with attached Terumo syringe. The syringe sizes are 1. 3,5, and 10cc/ml.
The specification of the needle and safety feature portions are identical to the SurGuard2TM Safety Needle cleared under K031453 and K040531. manufactured by Terumo Medical Corporation, Elkton, MD, USA. The attached 3, 5, and 10cc/ml syringes are identical to those syringe cleared under K023271. manufactured by Terumo Philippines Corporation. Information provided in this 510k will show that the 1cc/ml syringe manufactured by Terumo Philippines Corporation is substantially equivalent to the 1cc/ml manufactured by Terumo Medical Corporation, USA, cleared under K 771205.
#### D. Substantial Equivalence
The SurGuard2TM Safety Needle manufactured by Terumo Philippines Corporation is substantially equivalent to:
- 1. K031453 TERUMO® SURGUARD2™ SAFETY NEEDLE (18 to 25g) with or without syringe manufactured by Terumo Medical Corporation, Elkton, Maryland.
- 2. K040531 TERUMO® SURGUARD2™ SAFETY NEEDLE (26 to 30g) with or without syringe manufactured by Terumo Medical Corporation. Elkton, Maryland.
- 3. K023271 TERUMO® SYRINGE manufactured by Terumo (Philippines) Corporation, Binan, Laguna, Philippines
- K771205 TERUMO® HYPODERMIC SYRINGE (1cc/ml) manufactured by ধ: Terumo Medical Corporation, Elkton, Maryland
{2}------------------------------------------------
All of these cleared devices serve as predicates for the device which is subject of this 510k.
#### E. Principle of Operation and Technology
The Terumo SurGuard2™ Safety Needle device with and without syringe manufactured by Terumo (Philippines) Corporation and Terumo Medical Corporation, USA (K040531 and K031453) and all referenced predicate devices are operated manually.
#### F. Materials
The type of materials used for the hypodermic needle, safety feature and syringe of the SurGuard2™ Safety Needle device are identical to the materials used for the cleared SurGuard2TM Safety Needle cleared under K031453 and K040531. Any differences in materials between the SurGuard27M Safety Needle manufactured by Terumo (Philippines) Corporation and the SurGuard2TM predicate manufactured by Terumo Medical Corporation, USA, raise no new issues of safety and effectiveness.
#### G. Specifications
| Product Descriptions |
|---------------------------------------|
| 18gauge x 1" (25mm) safety needle |
| 18gauge x 1 1/2" (38mm) safety needle |
| 19gauge x 1" (25mm) safety needle |
| 19gauge x 1 1/2" (38mm) safety needle |
| 20gauge x 1" (25mm) safety needle |
| 20gauge x 1 1/2" (38mm) safety needle |
| 21gauge x 1" (25mm) safety needle |
| 21gauge x 1 1/2" (38mm) safety needle |
| 22gauge x 1" (25mm) safety needle |
| 22gauge x 1 1/2" (38mm) safety needle |
| 23gauge x 1" (25mm) safety needle |
| 23gauge x 1 1/2" (38mm) safety needle |
| 25gauge x 5/8" (16mm) safety needle |
| 25gauge x 1" (25mm) safety needle |
| 25gauge x 1 1/2" (38mm) safety needle |
| 26gauge x 1/2" (13mm) safety needle |
| 27gauge x 1/2" (13mm) safety needle |
{3}------------------------------------------------
| 30 gauge x 1/2" (13mm) safety needle |
|------------------------------------------------------------|
| 1cc/mL syringe with 25gauge x 5/8" (16mm) safety needle |
| 1cc/mL syringe with 26gauge x 3/8" (9mm) safety needle |
| 1cc/mL syringe with 27gauge x 1/2" (13mm) safety needle |
| 3cc/mL syringe with 20gauge x 1" (25mm) safety needle |
| 3cc/mL syringe with 20gauge x 1 1/2" (38mm) safety needle |
| 3cc/mL syringe with 21gauge x 1" (25mm) safety needle |
| 3cc/mL syringe with 21 gauge x 1 1/2" (38mm) safety needle |
| 3cc/mL syringe with 22gauge x 1" (25mm) safety needle |
| 3cc/mL syringe with 22gauge x 1 1/2" (38mm) safety needle |
| 3cc/mL syringe with 23gauge x 1" (25mm) safety needle |
| 3cc/mL syringe with 25gauge x 5/8" (16mm) safety needle |
| 3cc/mL syringe with 25gauge x 1" (25mm) safety needle |
| 5cc/mL syringe with 20gauge x 1" (25mm) safety needle |
| 5cc/mL syringe with 20gauge x 1 1/2" (38mm) safety needle |
| 5cc/mL syringe with 21gauge x 1 1/2" (38mm) safety needle |
| 10cc/mL syringe with 20gauge x 1" (25mm) safety needle |
| 10cc/mL syringe with 20gauge x 1 1/2" (38mm) safety needle |
| 10cc/mL syringe with 21gauge x 1" (25mm) safety needle |
#### H. Performance
The following tests were performed on the SurGuard2™ Safety Needle:
- Activation Force .
- Deactivation Force ●
- Puncture Resistance .
- Sheath Removal Force .
- Collar Removal Force ◆
- . Sheath Radial Force
- Protector Fit ◆
- Adhesive Hold .
- Simulated Use Study .
The following tests were performed on the Terumo Philippines 1cc/ml syringe:
- Leakage (Aspiration and Injection) ●
- Plunger Gasket Fit .
- Nominal Graduation Capacity .
{4}------------------------------------------------
KOSIECT
5 of 6
- . Deadspace
- . Conical Fitting
None of the data raises any new issues of safety and effectiveness. Additionally, a risk analysis was conducted and there were no new issues of safety and effectiveness.
The performance of the SurGuard27M Safety Needle submitted in this 510k is substantially equivalent in intended use, design, technology/principles of operation, materials and performance to the legally marketed predicate devices.
#### l. Additional Safety Information
Manufacturing controls include visual, functional, and sterility tests.
The sterility of the device is assured using a sterilization method validated in accordance with ANSI/AAMI/ISO 11137-1994 Medical Devices -- Validation and Routine Control of Radiation Sterilization. The SurGuard2TM Safety Needle is sterilized to provide a Sterility Assurance Level (SAL) of 10th.
The Terumo SurGuard2TM Safety Needle is classified as Externally Communicating Device, Blood Path Indirect, Limited Duration of Contact (< 24 hr). The device's blood contacting materials were tested in accordance with the tests recommended in the FDA General Program Memorandum #G95-1 (5/1/95): Use of International Standard ISO-10993. "Biological Evaluation of Medical Devices Part-1: Evaluation and Testing". Results of the testing demonstrate that the blood contacting materials are biocompatible.
#### J. Conclusion
The SurGuard2™ Safety Needle manufactured by Terumo Philippines Corporation is substantially equivalent to:
- I . K031453 TERUMO® SURGUARD2™ SAFETY NEEDLE (18 to 25g) with or without syringe manufactured by Terumo Medical Corporation. Elkton, Maryland.
- K040531 TERUMO® SURGUARD2™ SAFETY NEEDLE (26 to 30g) with or 2. without syringe manufactured by Terumo Medical Corporation. Elkton, Maryland.
- K023271 TERUMO® SYRINGE manufactured by Terumo (Philippines) rri
- Corporation, Binan, Laguna, Philippines
- 4. K771205 TERUMO® HYPODERMIC SYRINGE (1cc/ml) manufactured by Terumo Medical Corporation, Elkton, Maryland
{5}------------------------------------------------
K051865
6 of 6
Differences between the devices do not raise any significant issues of safety or effectiveness.
Terumo's statement of substantial equivalence is done solely to comply with the requirements of the Federal Food, Drug and Cosmetic Act and is not intended whatsoever to be the basis for a patent infringement action.
| Date Prepared: | July 8, 2005 |
|----------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
| Prepared By: | Barbara Smith<br>Sr. Regulatory Affairs Specialist<br>Terumo Medical Corporation<br>Phone: 410-392-7241<br>Fax: 410-398-6079 |
| Prepared For: | Terumo (Philippines) Corporation<br>#124 East Main Avenue<br>Laguna Technopark<br>Binan, Laguna 4026<br>Philippines<br>Phone: 011-63-49-541-2111 |
{6}------------------------------------------------
Image /page/6/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol consists of a stylized caduceus, a staff with two snakes entwined around it, which is a traditional symbol of medicine and healthcare.
SEP 2 7 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Barbara Smith Senior Regulatory Affairs Specialist Terumo Medical Corporation Regulators Affairs Department 125 Blue Ball Road Elkton, Maryland 21921
Re: K051865
K021662
Trade/Device Name: Terumo Surguard 2 Safety Needle or Similar Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: MEG Dated: July 6, 2005 Received: July 11, 2005
Dear Ms. Smith:
We have reviewed your Section 510(k) premarket notification of intent to market the device We have reviewed your becalent 910(x) promised is substantially equivalent (for the referenced above and have determined and to legally marketed predicate devices marketed in
indications for use stated in the enclosure) to legally marketed projects indications for use stated in the enactment date of the Medical Device interstate collinered prof to triay 20, 1778, and existed in accordance with the provisions of Amendments, of to devices that have oben recident do not require approval of a premarket the Federal Food, Drug, and Cosmene Tree (110) therefore, subject to the general approval application (1117). 1 our . general controls provisions of the Act include confors provisions of the rec. "The genting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) of class III If your device is classifica (sec above) into controls. Existing major regulations affecting (PMA), it may be subject to such additional controllar controllations, Title 21, Parts 800 to 898. In the Foorel your device can be found in the Code of Pederal regional of the Federal Register.
{7}------------------------------------------------
Page 2 - Ms. Smith
Please be advised that FDA's issuance of a substantial equivalence determination does not Please oe advised that PDA s Issualles of a sacomments with other requirements mean that IDA has made a deceminations administered by other Federal agencies. of the Act of ally rederal statues and regarments, including, but not limited to: registration You must comply with an the Fict b requires (21 CFR Part 801); good manufacturing practice and listing (21 CFR Part 607), labeling (21 crss (QS) regulation (21 CFR Part 820); and if requirements as set forth in the quality of sections (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to ocean mailer.ing your substantial equivalence of your device to a premarket notification. The PDF Imaling or colusiin for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), il you desire specific advice for your de at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the may obtain other general micrimation of the mational and Consumer Assistance at its toll-free Division of Binas 2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Clus
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center of Devices and Radiological Health
Enclosure
{8}------------------------------------------------
K051865
1 of 1
# Indications For Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name:_______TERUMO® SurGuard2™ Safety Needle
Indications For Use:
The TERUMO® SurGuard 2™ SAFETY NEEDLE device is intended for use in the The TEROMO® Surouard 2 - ORRETTFREDDurposes. The Terumo Safety Needle is aspiration and injoction standard luer slip and luer lock syringes.
Additionally, after withdraw of the needle from the body, the attached needle safety Additionally, and withdraw of the noodle from the beay,
shield can be manually activated to cover the needle immediately after use to minimize risk of accidental needlestick.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR 807 Subpart C)
(Please do Not WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Anthen D. nuk
(Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital, Division of Anostrol, Dental Devices
510(k) Number: k951865
Page 1 of
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.