FREE-STYLE FREEDOM BLOOD GLUCOSE MONITORING SYSTEM, MODEL 11001
K051839 · Abbott Diabetes Care, Inc. · NBW · Mar 9, 2006 · Clinical Chemistry
Device Facts
Record ID
K051839
Device Name
FREE-STYLE FREEDOM BLOOD GLUCOSE MONITORING SYSTEM, MODEL 11001
Applicant
Abbott Diabetes Care, Inc.
Product Code
NBW · Clinical Chemistry
Decision Date
Mar 9, 2006
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1345
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The FreeStyle Freedom™ Blood Glucose Monitoring System is specifically indicated for use on the finger, forearm, upper arm, thigh, calf and hand. The FreeStyle Freedom™ Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.
Device Story
FreeStyle Freedom BGMS measures glucose in whole blood samples using coulometric biosensor technology. Device utilizes test strips containing glucose dehydrogenase (GDH) enzyme to recognize glucose; electrochemical measurement occurs via biosensor. System provides quantitative glucose concentration readings. Used by healthcare professionals and patients at home for diabetes management. Features include 4 programmable alarms and LCD display. Output allows patients/providers to monitor diabetes control program effectiveness. Differences from predicate include modified LCD size, addition of alarm modes, factory-set measurement units, and reduced assay time (average 5 seconds).
Clinical Evidence
Performance evaluated via laboratory studies and clinical accuracy studies with lay users. Clinical data demonstrates that lay users obtain results correlating well with laboratory reference glucose values. No specific sensitivity/specificity metrics provided in summary.
Technological Characteristics
Coulometric electrochemical biosensor. PQQ-dependent glucose dehydrogenase enzyme. Operating range: 5-40°C, 5-90% humidity, 15-65% hematocrit. Concentration range: 20-500 mg/dL. Sample size: 0.3 uL. Connectivity: PC download via interface cable. Calibration: Lot-specific code entry. Software: Internal algorithm for charge-to-concentration conversion based on temperature and enzyme activity.
Indications for Use
Indicated for quantitative blood glucose measurement in people with diabetes mellitus (home use) and healthcare professionals. Sites: finger, forearm, upper arm, thigh, calf, hand. Contraindications: neonates, arterial blood, diagnosis or screening of diabetes.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k051839
B. Purpose for Submission:
Premarket Notification 510(k) of the intention to market the FreeStyle Freedom Blood Glucose Monitoring System
C. Measurand:
Glucose
D. Type of Test:
Quantitative - Coulometric electrochemical sensor, PQQ Glucose Dehydrogenase
E. Applicant:
Abbott Diabetes Care
F. Proprietary and Established Names:
FreeStyle Freedom Blood Glucose Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR §862.1345, Glucose test system
2. Classification:
Class II
3. Product code:
NBW, LFR
{1}
4. Panel:
Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use (below).
2. Indication(s) for use:
The FreeStyle Freedom™ Blood Glucose Monitoring System is specifically indicated for use on the finger, forearm, upper arm, thigh, calf and hand.
The FreeStyle Freedom™ Blood Glucose Monitoring System is intended for use in the quantitative measurement of glucose in whole blood. It is intended for use by healthcare professionals and people with diabetes mellitus at home as an aid in monitoring the effectiveness of a diabetes control program. It is not intended for the diagnosis of or screening for diabetes mellitus, and it is not intended for use on neonates or arterial blood.
3. Special conditions for use statement(s):
This product is intended for over-the-counter use.
Patients should not test on alternate sites (palm, back of the hand, forearm, upper arm, thigh, or calf) when they think their blood glucose is rapidly falling, such as within two hours of exercise or a rapid-acting insulin injection or insulin pump bolus. Testing with a fingertip sample may identify a hypoglycemic level (low blood sugar) sooner than a test with a forearm or palm sample.
Patients should not test on alternate sites when it has been less than two hours after a meal, a rapid-acting insulin injection or insulin pump bolus, physical exercise, or they think their glucose level is changing rapidly.
Patients should not test on alternate sites when they are concerned about the possibility of hypoglycemia.
4. Special instrument requirements:
The FreeStyle Freedom™ Blood Glucose Meter
2
{2}
3
I. Device Description:
The FreeStyle Freedom™ Blood Glucose Monitoring System is an electrochemical biosensor consisting of a glucose-oxidizing enzyme on a disposable test strip (the electrochemical sensor) and a hand-held current measuring device. Software internal to the hand-held device converts the measured current into glucose concentration using an algorithm that depends on the ambient temperature and the activity of the enzyme on the test strip. The user has the ability to validate the operation of the system by using glucose control solutions provided with the system.
The FreeStyle Freedom™ Blood Glucose Monitoring System consists of a hand-held blood glucose meter, test strips, and two levels of control materials. Each lot of test strips has a code number implying lot-specific calibration. The strip characteristics associated with each code number are pre-programmed and permanently stored in the hand-held meter.
The meter is turned on by strip insertion. The user must select a code number corresponding to their vial of strips via the meter's interface. After selecting the correct code, the user applies blood or a drop of control solution to the strip. The meter completes the glucose assay in 5 or more seconds. The meter's software converts the charge read off the test strip into a plasma glucose concentration and displays the value on the meter's LCD screen. The user has the ability to validate the operation of the system by using glucose control solutions provided with the system.
J. Substantial Equivalence Information:
1. Predicate device name(s):
FreeStyle 600 Blood Glucose Monitoring System
2. Predicate 510(k) number(s):
k050500
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Enzyme | Same | PQQ dependent glucose dehydrogenase |
| Detection Method | Same | Coulometric electrochemical sensor |
| Analyte | Same | D-glucose |
| Minimum Sample Size | Same | 0.3 uL |
| Test Strip | Same | Cleared in predicate submission |
| Humidity Range | Same | 5% - 90% |
{3}
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Hematocrit | Same | 15% - 65% |
| Operating Temperature Range | Same | 5 °C – 40 °C |
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Blood sample | Capillary whole blood, Fresh venous blood (within 30 minutes of draw) for professional use only | Venous, capillary, arterial, and neonatal whole blood |
| Concentration Range | 20-500 mg/dL | 20-600 mg/dL |
| Measurement time | 5 seconds average | 15 seconds |
# K. Standard/Guidance Document Referenced (if applicable):
CLSI EP05-A2: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline-Second Edition
CLSI EP06-A: Evaluation of the Linearity of Quantitative Analytical Methods; Approved Guideline
CLSI EP07-A2: Interference Testing in Clinical Chemistry; Approved Guideline- Second Edition
CLSI EP09-A2: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline - Second Edition
ISO 15197: In vitro diagnostic test systems — Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus
FDA Guidance "Guidance for Industry: In Vitro Diagnostic Glucose Test System"
Available at: http://www.fda.gov/cdrh/ode/glucose.pdf
FDA Guidance "General Principles of Software Validation; Final Guidance for Industry and FDA Staff"
Available at: http://www.fda.gov/cdrh/comp/guidance/938.pdf
FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices"
Available at: http://www.fda.gov/cdrh/ode/guidance/337.pdf
# L. Test Principle:
{4}
The FreeStyle Freedom™ meter utilizes coulometric biosensor technology to quantify glucose concentrations. Specifically, it correlates the total charge produced by the glucose oxidation reaction to a glucose concentration. Since complete oxidation of glucose in a sample can take in excess of 100 seconds, the total charge derived from the reaction is estimated from a partial reaction profile.
A mediator molecule shuttles electrons between a glucose-oxidizing enzyme and the working electrode on the strip. This rapid electron transport by the mediator maintains the enzyme in an oxidized state capable of reacting with glucose. Unlike the predicate, the re-oxidation of the mediator is accelerated and allows the glucose oxidation to proceed faster.
Software internal to the hand-held device uses the charge collected during the measurement time and the measured ambient temperature to project an estimated time course for the reaction. The software converts the estimate of the total amount of glucose to a concentration which is then displayed to the user.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
The company demonstrated the repeatability of their meter by performing within run precision studies on venous blood. One venous blood sample was adjusted to five different concentrations of glucose spanning the range claimed by the meter (20 mg/dL – 500 mg/dL). The company conducted the within-run precision study following guidelines in ISO 15197 and CLSI EP5-A2. The YSI was used as the reference standard for comparison. The company multiplied the YSI measurements by 1.12 to determine the plasma equivalent glucose values.
For the within-meter, within-vial precision studies, the company made 10 replicate measurements at each glucose concentration for each meter-vial pair. The company used vials from 3 different manufacturing lots in this study. A summary of the results is presented in the table below:
{5}
| Strip Lot# | YSI*1.12 | Mean | SD, within vial | CV, within vial |
| --- | --- | --- | --- | --- |
| 1 | 35.8 | 36.3 | 2.6 | 7.0 |
| | 90.9 | 84.4 | 3.0 | 3.6 |
| | 136.6 | 129.1 | 2.9 | 2.2 |
| | 206.9 | 186.1 | 5.8 | 3.1 |
| | 339.4 | 300.3 | 8.5 | 2.8 |
| 2 | 33.5 | 35.4 | 2.0 | 5.5 |
| | 93.7 | 89.8 | 2.4 | 2.6 |
| | 141.1 | 138.5 | 3.5 | 2.5 |
| | 201.9 | 192.4 | 5.8 | 3.0 |
| | 345.2 | 314.9 | 11.0 | 3.5 |
| 3 | 43.5 | 45.0 | 2.0 | 4.3 |
| | 95.1 | 92.9 | 3.0 | 3.2 |
| | 143.9 | 139.1 | 3.0 | 2.1 |
| | 206.1 | 192.7 | 5.2 | 2.7 |
| | 342.4 | 313.4 | 8.3 | 2.6 |
The average standard deviation observed in the within-meter, within-vial study was $2.5\mathrm{mg / dL}$ for glucose concentrations below $100\mathrm{mg / dL}$ with a corresponding CV of $4.4\%$ . For glucose concentrations above $100\mathrm{mg / dL}$ , the within-meter, within-vial average standard deviation was $6.0\mathrm{mg / dL}$ with a corresponding CV of $2.7\%$ .
The company demonstrated their within-lot precision of their device by measuring 5 different glucose concentrations 10 times across 16 meters. A summary of the results is presented in the table below:
| Strip Lot# | YSI*1.12 | Mean | SD, within lot | CV, within lot |
| --- | --- | --- | --- | --- |
| 1 | 35.8 | 36.3 | 3.1 | 8.7 |
| | 90.9 | 84.4 | 4.4 | 5.2 |
| | 136.6 | 129.1 | 5.4 | 4.2 |
| | 206.9 | 186.1 | 8.9 | 4.8 |
| | 339.4 | 300.3 | 14.3 | 4.8 |
| 2 | 33.5 | 35.4 | 2.6 | 7.4 |
| | 93.7 | 89.8 | 3.2 | 3.5 |
| | 141.1 | 138.5 | 5.3 | 3.8 |
| | 201.9 | 192.4 | 7.8 | 4.0 |
| | 345.2 | 314.9 | 15.2 | 4.8 |
| 3 | 43.5 | 45.0 | 2.8 | 6.3 |
| | 95.1 | 92.9 | 3.7 | 4.0 |
| | 143.9 | 139.1 | 4.1 | 2.9 |
| | 206.1 | 192.7 | 6.1 | 3.2 |
| | 342.4 | 313.4 | 10.2 | 3.3 |
{6}
The average standard deviation for measurements made within the same lot of strips at concentrations below 100 mg/dL was 3.3 mg/dL with a corresponding CV of 5.9%. For measurements above 100 mg/dL, the average standard deviation for measurements made within the same strip lot was 8.6 mg/dL with a corresponding CV of 4.0%.
The company assessed their day-to-day precision using measurements on 3 different concentrations of control solutions. The study involved 2 measurements per meter per day for 20 days using 3 lots of strips. A summary of these measurements:
| Control Solution | Average meter Reading (mg/dL) | Strip Lot 0526364 | Strip Lot 0526401 | Strip Lot 0526332 | Average Std. dev. | Average %CV |
| --- | --- | --- | --- | --- | --- | --- |
| Low | 55.1 | 1.6 | 1.7 | 2.1 | 1.8 | 3.3 |
| Normal | 104.1 | 3.1 | 3.2 | 3.2 | 3.2 | 3.0 |
| High | 328.6 | 10.8 | 11.6 | 12.8 | 11.7 | 3.6 |
b. Linearity/assay reportable range:
The company followed CLSI EP6-A in demonstrating the linear response of their device. They adjusted the unpooled blood of three different donors to 9 different YSI glucose concentrations. The average YSI glucose values obtained were as follows: 28, 78, 140, 192, 265, 320, 376, 438, 489 mg/dL. Each concentration of each blood sample was measured 3 times on 2 different lots of strips across 6 meters for a total of 972 measurements. Readings from the YSI were scaled by 1.12 for comparison. The proposed device demonstrated a linear relationship to the YSI with a slope of 0.8266, intercept of 9.6057 and a r-squared value of 0.9909.
In the clinical studies using fresh capillary blood, a linear regression of fingerstick measurements using the Freedom device vs. a YSI capillary fingerstick measurement yielded a line with a slope of 0.971, an intercept of 8.620 mg/dL, and an r-value of 0.981. The samples ranged from 50 mg/dL to 500 mg/dL by the YSI method.
In the clinical studies using fresh venous blood, a linear regression of 198 venous blood measurements using the Freedom device vs. a YSI venous measurement yielded a line with a slope of 0.922, an intercept of 19.063 and an r-value of 0.994. The samples ranged from 65 mg/dL to 490 mg/dL by the YSI method.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Data on expected values, traceability, and stability for the glucose control solutions used with this device was supplied in a previous submission, k050500.
{7}
# d. Detection limit:
See linearity study above. Another study was performed by repeated testing using out-of-range blood samples with 3 lots of strips. Pooled blood from 3 donors was adjusted to an out-of-range low glucose concentration of $13\mathrm{mg / dL}$ . The results are summarized below:
For the low glucose limit, the company tested 3 lots of strips:
| Average glucose (mg/dL) | Strip Lot | Total tests | "LO' results | Numeric results |
| --- | --- | --- | --- | --- |
| 13 | 1 | 36 | 34 | 2 |
| | 2 | 36 | 35 | 1 |
| | 3 | 36 | 30 | 6 |
| | Total | 108 | 99 | 9 |
| | | | 91.7% | 8.3% |
# e. Analytical specificity:
# Hematocrit Sensitivity
The company assessed the sensitivity of their device to variations in Hematocrit by testing on blood from three donors. They used 6 different meters and 3 lots of strips to measure 5 glucose concentrations spanning the claimed concentration range of the meter. At each glucose concentration, they adjusted the sample Hematocrit to one of 5 levels spanning the claimed Hematocrit range for the device. Each glucose and hematocrit combination was measured twice.
The table below summarizes the company's findings for the average bias over the 3 lots and 6 meters:
| Average Glucose (mg/dL) | Hematocrit | | | | |
| --- | --- | --- | --- | --- | --- |
| | 15% | 25% | 40% | 50% | 65% |
| 23.7 | -2.3 mg/dL | -2.8 mg/dL | -2.0 mg/dL | -0.8 mg/dL | 0.3 mg/dL |
| 132.9 | 13.6% | 5.8% | 0.7% | 1.0% | 2.3% |
| 249.6 | 16.6% | 11.8% | -0.1% | -1.3% | -1.6% |
| 362.4 | 15.3% | 12.1% | 1.3% | -4.6% | -4.7% |
| 470.7 | 2.8% | 3.0% | -0.7% | -7.2% | -9.2% |
# Temperature Sensitivity
The company assessed the sensitivity of their device to variations in temperature by
{8}
testing on blood from three donors. They used 6 different meters and 3 lots of strips to measure 5 glucose concentrations spanning the claimed concentration range of the meter. At each glucose concentration, they determined the observed glucose for three temperatures: 1) the claimed lower limit of the meter, $5^{\circ}\mathrm{C}$ , 2) the claimed upper limit of the meter, $40^{\circ}\mathrm{C}$ , and 3) $25^{\circ}\mathrm{C}$ .
The table below summarizes the company's findings for the average bias over the 3 lots and 6 meters:
| Average Glucose (mg/dL) | 5°C | 25°C | 40°C |
| --- | --- | --- | --- |
| 23.2 | -3.8 mg/dL | -3.3 mg/dL | -3.0 mg/dL |
| 137.0 | 5.1% | 0.7% | 6.2% |
| 259.6 | 4.2% | 2.7% | 9.3% |
| 367.3 | 3.2% | -0.7% | 7.8% |
| 481.7 | -1.1% | 0.2% | 4.5% |
# Impact of Humidity
The company assessed the sensitivity of their device to variations in humidity by testing on blood from three donors. They used 6 different meters and 3 lots of strips to measure 5 glucose concentrations spanning the claimed concentration range of the meter. At each glucose concentration, they determined the observed glucose at a relative humidity of $5\%$ , $50\%$ , and $90\%$ . Measurements were made in duplicate. All humidity studies were conducted at $25^{\circ}\mathrm{C}$ .
The table below summarizes the company's findings for the average bias over the 3 lots and 6 meters:
| [Glucose], mg/dL | Observed Bias relative to 25°C/50% Relative Humidity | | |
| --- | --- | --- | --- |
| | 5% | 50% | 90% |
| 23.2 | -2.3 mg/dL | -2.1 mg/dL | -1.8 mg/dL |
| 137.0 | 1.3% | 2.9% | 2.5% |
| 259.6 | 1.6% | 4.7% | 2.3% |
| 367.3 | 2.8% | 4.2% | 1.4% |
| 481.7 | 1.1% | 3.3% | -0.7% |
# Effect of Blood pH
The company assessed the sensitivity of their device to variations in blood pH by varying the carbon dioxide content of a sample of venous blood. They used 6 different meters and 3 lots of strips to measure 5 glucose concentrations spanning the claimed concentration range of the meter. At each glucose concentration, they determined the observed glucose at a pH of 7.15, 7.4, and 7.61. Measurements were
{9}
made in duplicate.
The table below summarizes the company's findings for the average bias over the 3 lots and 6 meters:
| [Glucose], mg/dL | Average | | |
| --- | --- | --- | --- |
| | pH = 7.4 | pH = 7.15 | pH = 7.61 |
| 26.8 | -1.8 mg/dL | 3.1 mg/dL | 0.5 mg/dL |
| 134.1 | -2.4% | -0.5% | -0.8% |
| 253.7 | 1.4% | 1.7% | 2.6% |
| 381.4 | 2.6% | 1.2% | 4.2% |
| 499.2 | -1.7% | -3.1% | 0.3% |
## Altitude Effect
The company assessed the variability in their device as a function of altitude by performing a linearity study at sea level and at a site at approximately 10,000 feet in elevation. The company used 6 different meters and 3 lots of strips to measure 9 glucose concentrations spanning the claimed concentration range of the meter. Measurements were made in duplicate.
The following graph illustrates the performance of the submitted device at sea level:

The following graph illustrates the performance of the submitted device at an altitude of 10,000 feet:
{10}

The following graph illustrates the average bias from sea level over the 3 lots and 6 meters:
| Glucose value mg/dL | Bias from Sea Level | | |
| --- | --- | --- | --- |
| | Lot 1 | Lot 2 | Lot 3 |
| 25.5 | -4.1 mg/dL | -4.6 mg/dL | -2.8 mg/dL |
| 78.5 | 6.7 mg/dL | 7.3 mg/dL | 6.7 mg/dL |
| 148.4 | 10.6 % | 10.7 % | 10.5 % |
| 211.6 | 10.1 % | 10.6 % | 12.8 % |
| 269.9 | 8.4 % | 9.9 % | 12.6 % |
| 334.6 | 4.3 % | 12.9 % | 8.8 % |
| 395.8 | 4.8 % | 12.2 % | 6.4 % |
| 455.6 | 2.3 % | 11.5 % | 1.8 % |
| 499.5 | 2.8 % | 6.6 % | 2.2 % |
# Impact of Chemical Interference
The company assessed the impact of endogenous and exogenous chemicals on the performance of their device. The company used 6 different meters and 3 lots of strips to measure 1 approximately normal glucose concentration from a venous blood sample. Chemicals tested for interference were added by spiking. Measurements were made in duplicate.
{11}
The following graph illustrates the performance of the submitted device at sea level:
| Interfering Agent | Concentration tested, mg/dL | Observed Difference in Reading (%) |
| --- | --- | --- |
| Acetaminophen | 20 | 1.1% |
| Ascorbic acid | 3 | 2.9% |
| Bilirubin | 20 | -0.7% |
| Cholesterol | 500 | 4.3% |
| Creatinine | 30 | 0.2% |
| Dopamine | 13 | 1.4% |
| Ephedrine | 10 | 0.1% |
| Ibuprofen | 40 | 3.5% |
| Lactic acid | 60 | 1.9% |
| L-dopa | 5 | 1.1% |
| m-dopa | 2.5 | 2.0% |
| Salicylic acid | 50 | -2.0% |
| Tetracycline | 4 | -1.1% |
| Tolazamide | 100 | 0.2% |
| Tolbutamide | 100 | 3.2% |
| Triglyceride | 3000 | -1.7% |
| Uric acid | 20 | 3.5% |
| Carbohydrates | | |
| Galactose | 100 | 94.9% |
| Lactose | 100 | 67.5% |
| Maltose | 100 | 61.7% |
| Xylose | 100 | 101.9% |
The company found that high levels of sugars other than glucose, which are often found in dialysis solutions, interfere with their device. The company added a warning to their user manual explicitly cautioning users about this problem.
## Double Dipping
In the concentration range of 90-100 mg/dL, measurements of a single application of blood had a median distance of 5.0 mg/dL from the scaled YSI measurement. In the concentration range of 90-100 mg/dL, measurements made using two sequential applications of blood (a double dip) had a median distance of 4.0 mg/dL from the scaled YSI measurement.
In the concentration range of 360-370 mg/dL, measurements of a single application of blood had a median distance of 40.7 mg/dL from the scaled YSI measurement. In the concentration range of 360-370 mg/dL, measurements made using two sequential applications of blood (a double dip) had a median distance of 36.7 mg/dL from the scaled YSI measurement.
f. Assay cut-off:
{12}
Not applicable for a device of this type.
# 2. Comparison studies:
# a. Method comparison with predicate device:
The company performed a clinical study to assess the ability of lay users to use the device to measure their glucose. A total of 186 subjects divided across 3 sites participated in the study. For each subject, the lay user performed a single finger stick measurement. For comparison, a trained operator also performed a single finger stick measurement. The trained operator also obtained additional fingerstick blood for measurement by a YSI 2300 STAT Plus Blood Glucose analyzer. Measurements on the YSI were performed twice. Six subjects were dropped from the subsequent data evaluation. Four were dropped due to missing YSI reference measurements. Two were excluded due to varying YSI reference measurements.
A linear least squares regression of the lay users' measurements versus the YSI 2300 STAT reference had a slope of 0.92, an intercept of $14.3\mathrm{mg / dL}$ , and an r-value of 0.99. A linear least squares regression of the trained operator measurements versus the YSI 2300 STAT reference had a slope of 0.95, an intercept of $12.9\mathrm{mg / dL}$ , and an r-value of 0.97. A linear least squares regression of the lay user measurements using the Freedom meter versus the trained operator measurements using the Freedom meter had a slope of 0.97, an intercept of $2.7\mathrm{mg / dL}$ , and a r-value of 0.98.
Number and % of results within YSI reference (all values were $\geq 75\mathrm{mg / dL}$ )
| User | Within ± 5% | Within ± 10% | Within ± 15% | Within ± 20% |
| --- | --- | --- | --- | --- |
| Lay user | 98/180 | 154/180 | 173/180 | 177/180 |
| | 54.4% | 85.6% | 96.1% | 98.3% |
| Trained operator | 93/180 | 148/180 | 163/180 | 172/180 |
| | 51.7% | 82.2% | 90.6% | 95.6% |
# b. Matrix comparison:
Not applicable for a device of this type.
# 3. Clinical studies:
# a. Clinical Sensitivity:
Not applicable for a device of this type.
# b. Clinical specificity:
Not applicable for a device of this type.
# c. Other clinical supportive data (when a. and b. are not applicable):
{13}
The company conducted a clinical study to demonstrate the accuracy of meter when used on the finger and forearm. 229 subjects in two sites participated in the study. A total of 7 patients were dropped from the study due to deviations in protocol. Measurements made by a medical technologist were compared to capillary and venous YSI plasma results. Two different lots of strips were used at one site. Three different lots of strips were used at the other. The level of testing varied at the two sites. After exclusions, a total of 955 fingerstick measurements and 190 measurements of the forearm were performed. In addition, a venous blood sample was obtained.
A linear regression of fingerstick measurements using the Freedom device vs. YSI capillary fingerstick measurement yielded a line with a slope of 0.971, an intercept of 8.620 mg/dL, and an r-value of 0.981. The samples ranged from 50 mg/dL to 500 mg/dL by YSI method.
A linear regression of forearm measurements using the Freedom device on subjects who had not eaten in more than 2 hours vs. YSI fingerstick measurement yielded a line with a slope of 0.98, an intercept of 1.893 mg/dL, and an r-value of 0.97.
A linear regression of forearm measurements using the Freedom including subjects who had recently eaten (72 of 190 forearm measurements) vs. YSI fingerstick measurement yielded a line with a slope of 0.931, an intercept of 8.634 mg/dL, and an r-value of 0.968.
A linear regression of 198 venous blood measurements using the Freedom device vs. YSI venous measurement yielded a line with a slope of 0.922, an intercept of 19.063 and an r-value of 0.994.
The company tested the performance of their device on the alternate sites claimed in their user manual. A YSI 2300 served as the reference method. After exclusions, 51 subjects participated in the trial. Subjects were tested by a technician first on a fingertip, then palm, forearm, upper arm, calf, thigh, and then an additional finger tip. For testing on the palm, participants were evenly divided for testing on the thenar region and hypothenar region.
The company used a Passing Bablok regression to calculate the slope and intercept of their alternate site data using scaled YSI measurements as a reference. The results of this analysis is as follows:
14
{14}
| Alternate Site | Slope | Intercept | R-value | N |
| --- | --- | --- | --- | --- |
| Upper Arm | 0.999 | 1.897 | 0.821 | 51 |
| Thenar | 1.056 | -2.208 | 0.891 | 24 |
| Forearm | 1.067 | -2.735 | 0.806 | 51 |
| Hypothenar | 1.069 | -2.427 | 0.946 | 27 |
| Thigh | 1.154 | -16.380 | 0.756 | 51 |
| Calf | 1.185 | -17.825 | 0.786 | 51 |
To evaluate the use of the back of the hand (base of the thumb and fore finger) as an alternate site, glucose values from 52 subjects were compared to scaled YSI measurements, ranging from $77\mathrm{mg / dL}$ to $246\mathrm{mg / dL}$ . 50/52 or $96\%$ of the values were within $20\%$ of the scaled YSI values.
# 4. Clinical cut-off:
Not applicable for a device of this type.
# 5. Expected values/Reference range:
The normal fasting glucose range for a non-diabetic adult is 70 to $110\mathrm{mg / dL}$ . (3.9 to 6.1 mmol/L) $^{1}$ . One to two hours after meals, normal glucose values should be less than 120 mg/dL $(6.7\mathrm{mmol / L})^2$ .
$^{1}$ Burtis CA Ashwood ER, eds: Tietz Textbook of Clinical Chemistry. $2^{\text{nd}}$
$^{2}$ Krall LP and Beaser RS: Joslin Diabetes Manual. Lea and Febiger. Philadelphia 1989. p. 138.
# N. Instrument Name:
FreeStyle Freedom Blood Glucose Monitoring System
# O. System Descriptions:
# 1. Modes of Operation:
Test strips can only be used once. Users must replace the strip before taking an additional reading. The meter is designed to exclude the feature that allows a user to change the unit of measure.
# 2. Software:
FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:
Yes ☐ X ☐ or No ☐
{15}
16
3. Specimen Identification:
There is no sample identification function with this device. Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
This device is intended for use with fresh capillary whole blood. Since the sample is applied directly to the test strip there are no special handling or storage issues.
5. Calibration:
Each bottle of test strips has a code number which is used to calibrate the meter. The user confirms that the code number on the test strip bottle matches the code number in the instrument. If the bottle and meter codes do not match, the user must change the meter's code setting by depressing the "c" and "m" buttons on the meter until the correct number is displayed. No further calibration is required of the user.
6. Quality Control:
The sponsor provides two levels of glucose control solutions with this device. To mark a test result as a control, the user depresses the "c" button on the meter for 2 seconds. Measurements marked as a control are excluded from supplemental data analysis supported by the meter (e.g., 14 day average glucose). An acceptable range of measurement for each control level is printed on the test strip vial label. If control results fall outside these ranges, the user is referred to a list of troubleshooting steps and the customer care line.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
The meter supports memory for storing 250 measurements and supplemental data analysis, such as the 14 day average glucose concentration mentioned above. The user can program the meter with up to 4 timed alarms. By inserting an interface cable into the meter, users can download their measurement history to a personal computer. Verification and validation information supplied by the company supports these performance claims.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.