The NL3000 CereTom is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 25cm field of view, primarily head and neck.
Device Story
The NL3000 CereTom is a portable, 8-row, 32cm bore CT scanner. It features a translating gantry that moves over a stationary patient, allowing for imaging in clinical settings. The system includes a solid-state x-ray generator, 3264-element detector array, and a brushless DC servo drive for rotation and translation. It is powered by internal dry sealed batteries or line power. The device is operated via a workstation (Sony Vaio laptop) running third-party CT viewing software. The system provides high-resolution CT images, enabling clinicians to perform diagnostic imaging at the point of care. By allowing the gantry to move while the patient remains still, the device facilitates imaging for patients who may be difficult to transport to a traditional radiology suite, potentially improving clinical workflow and patient management.
Clinical Evidence
Bench testing only. The device was evaluated for compliance with Federal Diagnostic X-ray Performance Standards (21 CFR 1020.30 and 1020.33) and medical electrical equipment safety standards (UL60601-1, IEC 60601-1, EN 60601-1). No clinical data was provided.
Technological Characteristics
8-row, 32cm bore CT system. Components: solid-state x-ray generator, 3264-element solid-state detector array, power slip-ring, brushless DC servo drive. Power: dry sealed batteries/line power. Form factor: mobile translating gantry with 100mm caster wheels. Connectivity: communications link to workstation. Safety: emergency stop, x-ray indicators, interlocks, patient alignment laser. Standards: 21 CFR 1020.30/1020.33, UL60601-1, IEC 60601-1, EN 60601-1.
Indications for Use
Indicated for x-ray computed tomography imaging of anatomy within a 25cm field of view, primarily the head and neck.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Siemens SOMATOM Plus 4 with sliding Gantry Option (K991600)
Submission Summary (Full Text)
{1}------------------------------------------------
K051765
JUL 1 4 2005
NeuroLogica
# 510(k) SUMMARY For NeuroLogica Corporation NL3000 CereTom™ Computed Tomography System
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirements of 21 CFR Part 807.87(h)
| (1) | Submitter: | NeuroLogica Corporation |
|-----|------------|-------------------------|
| | | 14 Electronics Avenue |
| | | Danvers, Ma, 01923 |
Establishment Registration number: Not yet applied for
Contact person: Brahim Hadri Manager for Regulatory Affairs and Quality Systems Telephone 978-564-8520 FAX 978-564-8677 e-mail: bhadri@neuro-logica.com
## Date this summary was prepared: June 1, 2005
#### (2) Device Name:
Proprietary or Trade Name: CereTom
Device Model: NL3000
Classification Name: Computed Tomography X-ray System
Product code: 90JAK
Device classification: Class II
Regulation number: 21 CFR 892.1750
NeuroLogica Corporation NL3000 CereTom Computed Tomography System 510(k) Summary
{2}------------------------------------------------
### (3) Predicate device:
The legally marketed devices to which substantial equivalencies are being claimed are as follows:
- Philips Tomoscan M/EG CT. This predicate device was cleared under pre-. market Notification K964890.
- . GE LightSpeed Ultra CT. This predicate device was cleared under premarket Notification K000300
- Siemens SOMATOM Plus 4 with sliding Gantry Option. This predicate . device was cleared under pre-market Notification K991600
#### (4) Device Description:
The NL3000 CereTom is a high resolution, 8 row, 32 cm bore, 25cm field of view, Computed Tomography System. The lightweight translating gantry consists of a rotating disk with a solid state x-ray generator, 3264-element solid state detector array, collimator, control computer, communications link, power slip-ring, data acquisition system, reconstruction computer, touch screen LCD display, power system, brushless DC servo drive system (disk rotation), and stepper drive system (translation). The power system consists of dry sealed batteries which provide system power while unplugged from the charging outlet. When plugged into an outlet the system charges the batteries with a line power drain of less than 1300 watts. The work station consists of a Sony Vaio laptop computer. The workstation software consists primarily of a certified off-the-shelf CT viewing system from a third party vendor. In addition the system has the necessary safety features such as emergency stop switch, x-ray indicators, interlocks, patient alignment laser, and 110 percent x-ray timer. It requires very little x-ray power in comparison to large bore systems to obtain equivalent CT image quality. The gantry has retractable 100mm diameter rotating caster wheels so the system can be moved easily to different locations.
#### (5) Intended Use:
The NL3000 CereTom is intended to be used for x-ray computed tomography applications for anatomy that can be imaged in the 25cm field of view, primarily head and neck.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "NeuroLogica" in a bold, sans-serif font. The "O" in Logica is stylized with a circle around it. The image has some noise and artifacts, possibly from scanning or compression.
## Comparison of Technological Characteristics with the predicate device: (6)
NeuroLogica Corporation believes that the NL3000 Computed Tomography Noteriogrea Corporated use, is of comparable type in design, material, System, for tto meetidor and is substantially equivalent to the following cleared predicate devices: Philips Tomoscan M/EG (also named ANATOM 2009, ref: predicate dovices: 1 milling ref: K000300, and Siemens SOMATOM Plus4 with Translating Gantry option, ref: K991600
- Material: The CereTom uses similar material to the above listed o scanners such as solid state detectors, x-ray generator, slip ring, data acquisition ICs, rotational bearing, and motion control systems. It also uses batteries and wheels similar in material to the Tomoscan M/EG.
- Design: The CereTom is similar in general design principle to all of o the above listed CT systems. Specifically it is similar to the 1) Lightspeed in multi-slice (8 Row) image quality, 2) Tomoscan M/EG and Somatom Plus 4 as to the translating gantry (patient stays still while the gantry translates), and 3) Tomoscan M/EG as to batteries and wheels.
# 7) General Safety and Effectiveness Concerns:
All components of the NL3000 CereTom system subject to Federal Diagnostic An component Performance Standard and applicable regulations of 21 CFR Part 1020.30 and 1020.33 are certified to meet those requirements.
aller 1020.55 are certired to the Part 1002.10 with the Center for Device and Radiological Health (CDRH).
and Radiological Heath (ODTC1).
To minimize electrical, mechanical and radiation hazards, Neurologica adheres to recognized and established industry practices. The NL3000 CereTom system is designed to meet UL60601-1, IEC 60601-1 and EN 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety.
{4}------------------------------------------------
NeuroLogica
## 8) Conclusion
Based upon the above considerations, NeuroLogica Corporation believes that the NL3000 CereTom Computed Tomography System is of comparable type in design, material, functionality, technology and is, for its intended use, substantially equivalent to the following cleared predicate devices: Philips Tomoscan M/EG (also named ANATOM 2000, ref: K964890), GE LightSpeed Ultra, ref: K000300, and Siemens SOMATOM Plus4 with Translating Gantry option, ref: K991600. A product report, according to 21 CFR 1002.10, will be submitted to the FDA prior to first delivery of the NL3000 CereTom.
Use of the NL3000 CereTom does not result in any new potential safety risks. The equipment performs as well in its intended use as devices currently on the market.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NeuroLogica Corporation % Mr. Neil E. Devine, Jr. Responsible Third Party Intertek Testing Services NA, Inc. 70 Codman Hill Road BOXBOROUGH MA 01719
Re: K051765 Trade/Device Name: NL3000 CereTom Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: June 29, 2005 Received: June 30, 2005
Dear Mr. Devine:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
JUL 1 4 2005
{6}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 5 10(k) This letter will anow you to ocgin hancellig your antial equivalence of your device to a legally premarket notification. The PDA Intentig of Sacolantial of Prour device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire spectic advice for your device of our laboring regional (
contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation childer, frinscrange your responsibilities under the Act from the 807.97). You may obtain other general missistance at its toll-free number (800) Division of 201) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
NeuroLogica Corporation NeuroEogica Corporation
NL3000 CereTom Computed Tomography System 510(k) Submission
Section C
# Indications for Use
| 510(k) Number (if known): | <span style="text-decoration: line-through;">Not known</span> K051765 |
|---------------------------|-----------------------------------------------------------------------|
|---------------------------|-----------------------------------------------------------------------|
..............................................................................................................................................................................
NL3000 CereTom___ Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications for Use:
The NL3000 CereTom is intended to be used for x-ray computed tomography
. In the comment in the counted in the 25em field of view, primar The NL3000 Cerel om is meended to be used for x Tay other in the may of the may be and and neck.
X Prescription Use (Part 21 CFR 801 Subpart D
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE
) THE CONTINUE BELOW TENFEDED IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C Brogdon
(Division Sign-Off) Division of Reproductive, Ab and Radiological Devices 510(k) Number
Page *1* of *1*
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.