← Product Code [OXO](/productcode/OXO) · K051754

# ORTHOSCAN, ORTHOSCAN HD (K051754)

_Orthoscan, Inc. · OXO · Aug 9, 2005 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051754

## Device Facts

- **Applicant:** Orthoscan, Inc.
- **Product Code:** [OXO](/productcode/OXO.md)
- **Decision Date:** Aug 9, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The OrthoScan mobile c-arm is designed to provide the physician with general fluoroscopic visualization of the patient including, but not limited to, surgical orthopedic procedures and critical and emergency care procedures. The intended users of this product are doctors, surgeons, radiologists and technologists who will use this product in a hospital or clinical environment. It is anticipated that the majority of users will be orthopedic doctors, podiatrists and pediatricians, among others. It is anticipated that this product will be used on a daily basis by such users. Users must be schooled in their medical fields and the use of x-ray equipment. They will be trained by OrthoScan specialists and/or qualified site personnel in the proper use of this product. The device labeling stipulates that only properly trained persons may operate this equipment.

## Device Story

OrthoScan/OrthoScan-HD is a mini C-arm mobile imaging system for real-time fluoroscopic visualization. System comprises a mobile workstation with an articulating arm for manual positioning of x-ray components. Input: x-ray radiation; Output: live fluoroscopic images displayed on a monitor. Operated by trained clinicians (surgeons, radiologists, technologists) in hospitals/clinics. System includes x-ray generator (up to 80 kVp), image intensifier, and digital processing circuitry. Features include automatic exposure rate control and interfaces for DICOM, thermal printers, and video recording. Provides clinicians with immediate visual feedback during orthopedic and emergency procedures to assist in surgical guidance and clinical decision-making. Benefits include portable, real-time imaging capability for point-of-care diagnostics and intraoperative support.

## Clinical Evidence

Bench testing only. No clinical data provided. Device performance is supported by compliance with recognized safety and performance standards for diagnostic x-ray imaging systems, including electrical safety, electromagnetic compatibility, and radiation protection standards.

## Technological Characteristics

Mobile C-arm system; manual positioning. Power: 90-260 VAC. X-ray generator: 40-80 kVp, 20-100 uA. Components: video monitor, isolation transformer, high voltage supply, x-ray tube, image intensifier. Connectivity: 100 Base T Ethernet, DICOM, NTSC video. Standards: 21 CFR 1020.30-32, NFPA 99, NFPA 70, UL 187, CSA-C22.2 No. 601.1-M90, IEC 601-1, IEC 601-1-2, IEC 601-1-3, IEC 601-2-7.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- GE/OEC Medical Systems series 6800 mini view digital mobile imaging device
- Hologic/FluoroScan Imaging System Premier Encore mobile imaging device

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

AUG 9 - 2005

#### thoscan .............................................................................................................................................................................. of Or � 651754

### 510 (k) Summary

This summary of 510 (k) safety and effectiveness information is being submitted in This Sunmal y of 510 (x) Sareery and SMDA 1990 and 21 CFR, Part 807.92.

| Date:<br>Name of Submitter: | June 14, 2005<br>OrthoScan, Inc.<br>Suite 900<br>8585 E. Hartford Dr.<br>Scottsdale, AZ 85255<br>480-503-8010                                                                                                                                |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Corresponding Official:     | Arlen Issette<br>Chief Executive Officer                                                                                                                                                                                                     |
| Device Proprietary Name:    | OrthoScan, OrthoScan-HD                                                                                                                                                                                                                      |
| Classification Name:        | System, X-ray, Fluoroscopic, Image-Intensified                                                                                                                                                                                               |
| Common/Usual Name:          | OrthoScan Imaging System                                                                                                                                                                                                                     |
| Substantial equivalence:    | The OrthoScan is substantially equivalent to the<br>following systems which are currently marketed:                                                                                                                                          |
|                             | 1. GE/OEC Medical Systems series 6800<br>mini view digital mobile imaging device.<br><br>2. Hologic/FluoroScan Imaging System<br>Premier Encore mobile imaging device                                                                        |
| Similarities/Differences:   | Refer to the attached Specification Comparison,<br>(Similarities/Differences). This chart compares the<br>functional characteristics of the OrthoScan with<br>these similar devices which are currently marketed.                            |
| Intended Use:               | The OrthoScan mobile c-arm is designed to provide<br>the physician with general fluoroscopic<br>visualization of the patient including, but not<br>limited to, surgical orthopedic procedures and<br>critical and emergency care procedures. |

{1}------------------------------------------------

## orthoScanHD

### The intended users of this product are doctors, User Characteristics: surgeons, radiologists and technologists who will use this product in a hospital or clinical environment. It is anticipated that the majority of users will be orthopedic doctors, podiatrists and pediatricians, among others. It is anticipated that this product will be used on a daily basis by such users. Users must be schooled in their medical fields and the use of x-ray equipment. They will be trained by OrthoScan specialists and/or qualified site personnel in the proper use of this product. The device labeling stipulates that only properly trained persons may operate this equipment. The OrthoScan is a mini C-arm mobile imaging General Description: system. The OrthoScan is used for processing and capturing live fluoroscopic images. Interfaces are provided for external peripheral devices such as controls, thermal printers, video cassette recorders and DICOM. The device is compatible with NTSC video outputs and 100 Base T Ethernet. A number of algorithms are used in the image processing functions performed by the system. A detailed description of these algorithms is provided in the image processor functions specification attached hereto as Attachment A. The descriptions and formal definitions of these types of algorithms are well known. References on the subject would include:

Jain. Anil K., Fundamentals if Digital Image Processing, Prentice Hall, Inc. 1989. Castleman, Kenneth R., Digital Image Processing, Prentice Hall, Inc., 1979.

{2}------------------------------------------------

# orthoscan H

The OrthoScan has the following physical characteristics:

- All components are contained in one mobile workstation .
- An articulating arm is attached to the workstation and extends out from the main . support to position the x-ray imaging components.
- All mechanical positioning of the workstation and articulating arm is manual . (non-motorized).
- Power ratings are between 90VAC and 260 VAC and 3A to 6A. The system is . powered by a detachable power cord.
- System power is isolated from input power by an isolation transformer. .
- Internal x-ray generator produces voltages up to 80 kVp. .
	- Fluoroscopic operation:
	- --- 40 to 80 kVp
	- --- 20 to 100 uA (.020 to .100 mA)
	- --- Automatic Exposure Rate Control
- Major components of the system include: ●
	- --- Video monitor
	- --- Input isolation transformer
	- --- Circuitry for digital processing and x-ray control
	- --- High voltage power supply
	- --- X-ray tube
	- --- Image Intensifier

### Standards:

In addition to complying with the Federal Performance Standard for Diagnostic X-ray Imaging Systems (21 CFR Part 1020.30-32), the OrthoScan mobile C-arm is designed in accordance with guidelines established in the following standards:

NFPA 99, Standard for Health Care facilities NFPA 70, National Electrical Code UL 187, Standard for X-ray Equipment CSA-C22.2 No. 601.1-M90, Medical Electrical Equipment IEC 601-1, Medical Electrical Equipment, General Requirements for Safety IEC 601-1-2, Medical Electrical Equipment, General Requirements for Safety, Electromagnetic Compatibility IEC 601-1-3, Medical Electrical Equipment, Radiation Protection in Diagnostic X-ray Equipment IEC 601-2-7, Medical Electrical Equipment, Safety of HV/X-av Generators 93/42/eec - Annex 1, Essential Requirements of the European Union Medical Device Directive

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for OrthoScanHD. The logo consists of a series of dots arranged in a grid pattern, followed by the word "OrthoScanHD" in a stylized font. The letters are bold and have a slightly futuristic look. There is a small black circle below the letters.

This concludes this 510 (k) Summary.

Attachment: OrthoScan Specification Comparison to Competitive Systems.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the department's name around the perimeter. Inside the circle is a stylized image of an eagle with three lines extending from its head, representing the department's mission to protect the health of all Americans.

#### Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

AUG 9 - 2005

Ms. Arlen Issette American C-Arm, Inc., a.k.a. OrthoScan, Inc. 8585 E, Hartford Drive, Suite 900 SCOTTSDALE AZ 85255

Re: K051754

Trade/Device Name: OrthoScan, OrthoScan HD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulation Number: 21 CFR 892.1720 Regulation Name: Mobil x-ray system Regulatory Class: II Product Code: JAA and IZL Dated: July 27, 2005 Received: July 27, 2005

Dear Ms. Issette:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling. and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{5}------------------------------------------------

This letter will allow you to begin marketing your device as described in your Section 510(k) rms letter watification. The FDA finding of substantial equivalence of your device to a legally premativel notified on " results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

### Indications for Use

510(k) Number (if known): 尾০51754 Not Available

Device Name:

OrthoScan, OrthoScan HD

Indications for Use:

The OrthoScan mobile c-arm is designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to, surgical orthopedic procedures and critical and emergency care procedures.

The intended users of this product are doctors surgeons, radiologists and technologists who will use this product in a hospital or clinical environment. It is anticipated that the majority of user will be orthopedic doctors, podiatrists and pediatricians, among others.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

David R. Legum

(Division Sign-Off) Division of Reproductive. Abdominal, and Padlological Devices 511)(k, Number

OrthoScan B-1

Page 1 of

---

**Source:** [https://fda.innolitics.com/device/K051754](https://fda.innolitics.com/device/K051754)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
