The Medstrat echoeSYSTEM™ is intended to enable the communication, storage, viewing and manipulation of diagnostic medical images and data. The echoeSYSTEM™ can show images on workstations locally and/or across computer networks at widely distributed locations. The echoeSYSTEM™ software is intended for assisting healthcare professionals in preoperative planning and postoperative evaluation of orthopedic surgery. The device allows for overlaying of prosthesis templates on radiological images, and includes tools for performing measurements on the image and positioning the templates. The typical echoeSYSTEM™ users are trained professionals such as orthopedic surgeons, referring and collaborating physicians, and other authorized medical professionals. It is the user's responsibility to bear in mind the display monitor's quality, ambient light conditions and the tradeoff between image compression ratio and clinical effectiveness for specific image study indications.
Device Story
echoeSYSTEM™ is a client-server software system for viewing, storing, and manipulating diagnostic medical images. It ingests DICOM 3.0 data from imaging modalities. Operated by orthopedic surgeons and healthcare professionals on off-the-shelf PCs, it provides tools for image measurement and overlaying prosthesis templates onto radiological images to assist in surgical planning and evaluation. The system supports secure remote access via web interfaces and enables collaboration through saved states of annotations and measurements. By facilitating precise template positioning and preoperative assessment, the device aims to improve surgical planning accuracy and clinical workflow efficiency.
Clinical Evidence
No clinical data provided; bench testing only. Substantial equivalence is based on descriptive comparisons of technological features and intended use.
Technological Characteristics
Client-server software system; runs on off-the-shelf hardware; utilizes DICOM 3.0 protocol for image acquisition; supports JPEG image compression; TCP/IP communication; includes measurement and digital templating tools; web-based authentication and secure protocols.
Indications for Use
Indicated for healthcare professionals, including orthopedic surgeons and referring physicians, for preoperative planning and postoperative evaluation of orthopedic surgery using radiological images.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
5/20/2005
# B. ADMINISTRATIVE INFORMATION
JUL 1 4 2005
#### 510(k) Summary of safety and effectiveness 1.
#### Manufacturer:
| Address: | Medstrat, Inc.<br>1901 Butterfield Road<br>Suite 600<br>Downers Grove, IL 60515<br>Phone: 630-960-8700<br>Fax: 630-960-9787 |
|-----------------|-----------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Robert J. Bishop |
| Summary Date: | April 25, 2005 |
| Device Name: | |
| Trade name: | echoeSYSTEM™ |
| Common name: | Medstrat ECHOESTM PACS |
Classification name: Picture archiving and communication system
#### Predicate Device:
eFILM Workstation with Modules, K020995, eFilm Medical Inc.
Amicas Web/Intranet Image Server, K970064, Autocyt Group, Inc.
#### Device Description:
ECHOES™ is a client-server software system designed to allow viewing access to medical images on a personal computer by authorized medical professionals. This product is designed to function with off-the-shelf hardware and software products including standard communications products. It does not require specialized, or nonstandard, devices of any type.
Image acquisition is via the industry standard DICOM 3.0 protocol allowing the images to be produced from the digital data originated by the imaging modality.
The software will run on standard off-the-shelf hardware and system configurations,
#### Intended Use:
The Medstrat echoeSYSTEM™ is intended to enable the communication, storage, viewing and manipulation of diagnostic medical images and data. The echoeSYSTEM™ can show images on workstations locally and/or across computer networks at widely
#### Medstrat. Inc.
{1}------------------------------------------------
distributed locations. The echoeSYSTEM™ software is intended for assisting healthcare professionals in preoperative planning and postoperative evaluation of orthopedic surgery. The device allows for overlaying of prosthesis templates on radiological images, and includes tools for performing measurements on the image and positioning the templates. The typical echoeSYSTEM™ users are trained professionals such as orthopedic surgeons, referring and collaborating physicians, and other authorized medical professionals.
It is the user's responsibility to bear in mind the display monitor's quality, ambient light conditions and the tradeoff between image compression ratio and clinical effectiveness for specific image study indications.
| Feature | Efilm<br>Workstation<br>with Modules | Amicas<br>Web/Intranet<br>Server | EchoeSYSTEM |
|--------------------------------|--------------------------------------|----------------------------------|-------------------------|
| Indications for Use | Similar | Similar | Similar |
| Target Population | Health<br>Professionals | Health<br>Professionals | Health<br>Professionals |
| Uses Off-the-Shelf<br>Monitors | yes | yes | yes |
| Lossy Image Compression | yes | yes | yes |
| TCP/IP Communications | yes | yes | yes |
| Software Only | yes | yes | yes |
| Image Measurements | yes | yes | yes |
| DICOM image data | yes | yes | yes |
| Pixel-for-Pixel Zoom | yes | yes | yes |
| User Login/Authentication | NA | yes | yes |
| Digital templates | yes | no | yes |
| Editable state saves | no | no | yes |
#### Substantial equivalence:
ECHOES™ has Indications for Use similar to other medical image devices such as to eFILM Workstation with Modules [K020995], eFilm Medical Inc. 510(k) Holder and Amicas Web/Intranet Image Server [K970064], Autocyt Group, Inc. 510(k) Holder. ECHOESTM also shares with these devices a Target Population that is competent health professionals. Also equivalent to these devices, the ECHOES™ design will operate with off-the-shelf hardware and systems.
Like the Amicas Web/Intranet Image Server [K970064], Autocyt Group, Inc. 510(k) Holder, ECHOES™ employs Image Compression to remove redundant or unimportant information in the original image data. The compression methods are believed to conform to the voluntary recognized consensus JPEG standard as declared in Section A.8 herein.
Like the eFILM Workstation with Modules [K020995], eFilm Medical Inc. 510(k) Holder, ECHOESTM provides image viewing and manipulation in a diagnostic setting affording users access to various image processing and measurement tools to assist them in viewing images. Users can overlay templates on medical images to aid in operative and non-operative treatment planning.
Medstrat, Inc.
Page 10
Image /page/1/Picture/11 description: The image shows a black and white photograph of what appears to be a cluster of trees or bushes. The image is somewhat grainy, and the contrast is high, making it difficult to discern fine details. The composition is focused on the right side of the image, where the cluster is most dense, while the left side is more sparse and faded.
{2}------------------------------------------------
ECHOES™ provides secure Internet and extranet interfaces through authentication mechanisms, web accessible authentication servers and access for authorized users through secure protocols to web image servers. ECHOES™ provides functions for communicating, storing and editing state save data of annotations, measurements and image manipulation states intended to be used for supporting collaboration, clinical data collection and education among health professionals.
----------------------------------------------------------------------------------------------------------------------------
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JUL 1 4 2005
Dated: June 21, 2005 Received: June 29, 2005
4 2005
Medstrat, Inc. % Ms. Silvia Ankova Project Manager Underwriters Laboratories, Inc. 333 Pfingsten Rd. NORTHBROOK IL 60062
Re: K051745 Trade/Device Name: echoeSYSTEM™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ
Dear Ms. Ankova:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Image /page/3/Picture/12 description: The image shows a black and white logo. The logo features a stylized depiction of an eagle or bird of prey with three distinct lines forming its body and wings. The bird is positioned within a circular border, with text partially visible around the top and left edges of the circle. The text appears to be part of the organization's name or motto.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section 510(k) This leter will anow you'to organ matters of substantial equivalence of your device to a legally prematice notification: "The PDF macinsing for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific active for a one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, prease note the regulation on your responsibilities under the Act from the 807.77). I ou may oouall ourers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Hogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## 2. FDA Indication for Use Form
## Indications for Use
K651745 510(k) Number (if known): __
・.
. .
Device Name: echoeSYSTEM™
| Indications for Use: | The Medstrat echoeSYSTEM™ is intended to enable the communication, storage, viewing and manipulation of diagnostic medical images and data. The echoeSYSTEM™ can show images on workstations locally and/or across computer networks at widely distributed locations. The echoeSYSTEM™ software is intended for assisting healthcare professionals in preoperative planning and postoperative evaluation of orthopedic surgery. The device allows for overlaying of prosthesis templates on radiological images, and includes tools for performing measurements on the image and positioning the templates. The typical echoeSYSTEM™ users are trained professionals such as orthopedic surgeons, referring and collaborating physicians, and other authorized medical professionals. |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | It is the user's responsibility to bear in mind the display monitor's quality, ambient light conditions and the tradeoff between image compression ratio and clinical effectiveness for specific image study indications. |
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| | <div> <img alt="Signature" src="signature.png"/> </div> |
|---------------------------------------------------------------|---------------------------------------------------------|
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, and Radiological Devices | |
| 510(k) Number | K051745 |
Medstrat, Inc.
Page 12
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.