K051628 · Draeger Medical Systems, Inc. · MHX · Jul 20, 2005 · Cardiovascular
Device Facts
Record ID
K051628
Device Name
INFINITY MONITORS WITH SCIO MODIFICATIONS
Applicant
Draeger Medical Systems, Inc.
Product Code
MHX · Cardiovascular
Decision Date
Jul 20, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1025
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infinity Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the Infinity Network.
Device Story
Infinity Modular Monitors provide multi-parameter physiological monitoring in clinical settings. The system integrates a Scio gas module for continuous measurement of CO2, N2O, O2, and anesthetic agents (halothane, isoflurane, enflurane, sevoflurane, desflurane). The modification adds a sample gas return kit, allowing gas return to the breathing circuit, identical to the Vamos device. The system processes inputs from various sensors (ECG, pressure, temperature, gas sampling) to display real-time data and derived parameters. It generates visual and audible alarms when physiological values exceed preset limits. Healthcare professionals (physicians, nurses, technicians) operate the device to assess patient status and guide clinical interventions. Data can be recorded via the R50 Bedside recorder or the Infinity Network. The device is not MRI compatible.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Modular patient monitor with gas sampling module. Features sample gas return capability. Connectivity via Infinity Network and MIB protocol converter. Not MRI compatible.
Indications for Use
Indicated for adult, pediatric, and neonatal patients for multi-parameter monitoring including heart rate, respiration, invasive/non-invasive pressure, arrhythmia, temperature, cardiac output, SpO2, pulse rate, apnea, ST segment, tcpO2/tcpCO2, EEG, and FiO2. Gas module (Scio) measures CO2, N2O, O2, and anesthetic agents. Cardiac output, ST segment, and arrhythmia are for adult/pediatric only. tcpO2 is for neonates only (not under gas anesthesia). Used by healthcare professionals in clinical settings.
Regulatory Classification
Identification
The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.
{0}------------------------------------------------
Special 510(k) Infinity Monitors with Scio Modifications
JUL 2 0 2005
germedical
A Dräger and Siemens Company
K051628
510(k) SUMMARY as required per 807.92(c)
| Submitter's Name and Address: | Draeger Medical Systems, Inc.<br>16 Electronics Avenue<br>Danvers, MA 01923 |
|-------------------------------|-----------------------------------------------------------------------------------------------|
| Contact Person: | Penelope H. Greco<br>Regulatory Affairs Manager<br>Tel: (978) 907-7503<br>Fax: (978) 750-6879 |
| Date submission was prepared: | June 17, 2005 |
| Device Name: | |
| Common Name: | Monitor, Physiological, Patient<br>(with arrhythmia detection or alarms) |
| Classification Name: | MHX |
| Regulation Number: | 21 CFR 870.1025 |
| Class: | 2 |
Legally Marketed Device Identification: Infinity Modular Monitors
# Device Description:
The intent of this 510(k) is to submit a modification specific to Scio, a gas module used exclusively with Draeger Medical Systems' Infinity monitors. The modification is the addition of a sample gas return kit identifical to that of Draeger Medical's Vamos, 510(k) K040847. The modification allows the user the option of returning the sample gas from the Scio gas module to the breathing circuit.
#### Intended Use:
The Infinity Modular Monitors are intended for multi-parameter patient monitoring. The devices will produce visual and audible alarms if any of the physiological parameters monitored vary beyond preset limits and timed or alarm recordings will be produced. These devices will connect to an R50 Bedside recorder, either directly or via the Infinity Network.
### Predicate Devices:
Infinity Monitors with Scio: 510(k) K031340, K033957, K040188 Vamos: K040847
### Substantial Equivalence:
Scio with sample gas return was tested in accordance with internal design control procedures and was determined to be as safe and effective for its intended use as the predicate device.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page 1 of 1
# COMPANY CONFIDENTIAL
Draeger Medical Systems, Inc. 16 Electronics Avenue Danvers, MA 01923 Tel: (978) 907-7500 Fax: (978) 750-6879
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized eagle or bird-like figure.
JUL 2 0 2005
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Ms. Penelope H. Greco Regulatory Affairs Manager DRAEGER Medical Systems, Incorporated 16 Electronics Ave. Danvers, Massachusetts 01923
Re: K051628
Trade/Device Name: Infinity Monitors with SCIO Modifications Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia detector and alarm (including ST-segment measurement and alarm) Regulatory Class: II Product Code: MHX Dated: June 17, 2005 Received: June 20, 2005
Dear Ms. Greco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
# Page 2 - Ms. Greco
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Ficase oe advised that I Dris issuation sevice complies with other requirements of the Act that I Dri has made a astatutes and regulations administered by other Federal agencies. You must of ally I coleral statutes and registents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set CTN in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic form in the qualisy by ovisions (Sections 531-542 of the Act); 21 CFR 1000-1050. product radiation control provisions (Section ) vice as described in your Section 510(k) This letter will anow you to ough finding of substantial equivalence of your device to a legally prematics notineation. The a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acrioliary at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain Misoranding by reference to presenters willities under the Act from the Division of Small other general mionnational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Senette y. Michael Ond.
Chiu Lin, PH.D. Director Division of Anesthesiology, General Hospital Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
510(k) Number (if known):_
Device Name: Infinity Modular Monitors with Scio Modifications
Indications for Use:
The INFINITY Modular monitors are capable of monitoring:
- Heart rate . . . . . . . . . . . ●
- Respiration rate .
- Invasive pressure .
- Non-invasive pressure .
- Arrhythmia .
- Temperature .
- Cardiac output .
- Arterial oxygen saturation .
- Pulse rate .
- Apnea .
- . ST Segment Analysis
- 12-Lead ST Segment Analysis .
- tcp02/tcpC02 .
- EEG signals .
- FiO2 .
> Prescription Use_ (Per 21 CFR 801.109)
OR
Over-The-Counter Use_
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sukite Y. Michael Om.d
(Division Sign-Off) Division of Anesthesiology, General Ho Infection Control, Dental
510(k) Number: K051628
Page 1 of 2
{4}------------------------------------------------
### Indications for Use
The SCIO module samples breathing gases from adults and pediatrics. The gas modules continuously measure the content of CO2, N2O, O2 and one of the anesthetic agents, halothane, isoffurane, Enflurane, Sevoflurane and Desflurane in any mixture and communicates real time and derived gas information to the INFINITY monitors.
With etCO2 the monitors can measure end tidal carbon dioxide, inspired carbon dioxide, and respiration rate in either mainstream or side-stream measurement mode; and with etCO2+Respiratory Mechanics, spirometry and carbon dioxide can be monitored. The monitors can interface with specific third party devices via an MIB protocol converter.
The devices are intended to be used in the environment where patient care is provided by Healthcare Professionals, i.e. Physicians, Nurses, and Technicians, who will determine when use of the device is indicated, based upon their professional assessment of the patient's medical condition.
The devices are intended to be used on Adult, Pediatric and Neonatal populations, with the exception of the parameter Cardiac Output, ST Segment Analysis, and arrhythmia which are intended for use in the adult and pediatric populations only; and tcpO2 which is to be used in the neonatal population only when the patient is not under gas anesthesia.
#### MRI Compatibility Statement:
ﻌﺘﻤ ﺴ
The INFINITY Modular Monitors are not compatible for use in a MRI magnetic field.
Page 2 of 2 .
Auy Sulleron
(Division Sign-Off (Division Sign-Off) \
Division of Anesthestelogy General Hospital, Infection Control, Dental D
510(k) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.