MODIFICATION TO OLYMPUS INTEGRATED ENDOSURGERY SYSTEM ENDOALPHA (CONTROL UNIT FOR ENDOSURGERY UCES-2) SOFTWARE VERSION 3

K051613 · Olympus Medical Systems Corporation · ODA · Aug 15, 2005 · Gastroenterology, Urology

Device Facts

Record IDK051613
Device NameMODIFICATION TO OLYMPUS INTEGRATED ENDOSURGERY SYSTEM ENDOALPHA (CONTROL UNIT FOR ENDOSURGERY UCES-2) SOFTWARE VERSION 3
ApplicantOlympus Medical Systems Corporation
Product CodeODA · Gastroenterology, Urology
Decision DateAug 15, 2005
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment.

Device Story

Remote control unit for endosurgical ancillary equipment; enables central operation, display of active states, and memory of setup values. Inputs include touch-panel monitor, remote controller, and voice commands (via wired or wireless microphones). Transforms inputs into control commands for ancillary equipment (e.g., AV devices, OR beds, OR lights, DVD recorders). Used in OR environments by surgical staff. Does not contact patients. Benefits include streamlined workflow through centralized control and automation of initial settings.

Clinical Evidence

Bench testing only. Verification testing confirmed that nurse control panel, surgeon controller, and voice operation functions operate correctly according to specifications following software modifications.

Technological Characteristics

Remote control unit for ancillary medical equipment. Interfaces via touch-panel monitors (12-inch or 15/19-inch), remote controllers, and voice control (wired/wireless microphones). Connectivity includes integration with hospital AV control systems. Software-based control logic. No patient contact; no sterilization required.

Indications for Use

Indicated for use with Olympus endoscopes and ancillary equipment to provide central operation, display, automatic initial settings, and interlocking control of equipment during endoscopic procedures. Patient population is defined by the specific indications of the connected ancillary devices.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{1} DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Olympus Medical Systems Corp. % Ms. Laura Storms-Tyler Executive Director, RA & QA Olympus America, Inc. Two Corporate Center Drive Melville, NY 11747-3157 JUL 27 2015 Re: K051613 Trade/Device Name: Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: ODA, GCI Dated (Date on orig SE ltr): July 21, 2005 Received (Date on orig SE ltr): July 25, 2005 Dear Ms. Storms-Tyler, This letter corrects our substantially equivalent letter of August 15, 2005. We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be {2} Page 2 - found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the *Federal Register*. Please be advised that FDA’s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act’s requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH’s Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Benjamin R. Fisher -S Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3} # Indications for Use 510(k) Number (if known): 05/6/3 Device Name: Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 Indications For Use: The Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment. Prescription Use ☑ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) ![img-0.jpeg](img-0.jpeg) {4} AUG 15 2005 K051613 83/73 # 510(k) SUMMARY ## Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 May 31, 2005 ## 1 General Information - **Applicant** Olympus Medical Systems Corp. 2951 Ishikawa-cho, Hachioji-shi, Tokyo, Japan 192-8507 Establishment Registration No.: 8010047 - **Official Correspondent** Laura Storms-Tyler Executive Director, Regulatory Affairs & Quality Assurance Olympus America Inc. Two Corporate Center Drive, Melville, NY 11747-9058, USA Phone: 631-844-5688 FAX: 631-844-5554 Email: Laura.storms-tyler@olympus.com Establishment Registration No.: 2429304 - **Manufacturer** Olympus Medical Systems Corp. Hinode Plant 34-3 Hirai Hinode-machi, Nishitama-gun, Tokyo, Japan 190-0182 Establishment Registration No.: 3003637092 ## 2 Device Identification - **Device Name** Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 - **Common Name** Endosurgery System - **Regulation No:** 21 CFR 876.1500 - **Regulation Name:** Endoscope and accessories - **Regulatory Class:** II - **Product Code:** 78 KOG - **Prescription Status:** Prescription device - **Performance Standards:** None established under Section 514 of FDCA. 1/3 {5} K051613 pg 2 of 3 ## 3 Predicate Device Information - Device Name: Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 2 - 510(k) No: K041494 - Decision Date: 07/01/2004 ## 4 Device Description The Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) is the remote control unit of legally marketed ancillary equipment. The new clinical functions are not added to the ancillary equipment by the subject device. The subject device enables the remote control of the ancillary equipment, the display of their active states and the memory of the previous set-up values. The remote control is achieved by a touch-panel monitor, remote controller, and voice control. The voice control function enables the subject device to control the ancillary equipment by voice. The subject device does not come in contact with patients and is not subject to sterilization. The only modifications that were made are: - Add the operating function for the new ancillary equipment, Panasonic DVD recorder LQ-MD800. - Enable the voice control of the ancillary equipment with the wireless microphone unit ew152G2 or Infrared Wireless Microphone System, in addition to the conventional wired microphone. - Enable the control of the ancillary equipment with the ELO 15inch/19inch touch panel monitor 1526L/1926L, in addition to the conventional 12inch-touch panel monitor. - Enable control of Audio Visual (AV) devices, Operating Room (OR) beds, and OR lights via the hospital's integrated audio-visual control system. The subject device sends commands for controlling these devices to the integrated audio-visual control system by remote control or voice control. - Add a simplified main screen for the control panel monitor, in addition to the conventional GUI screen. ## 5 Intended Use The Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 has been designed to be used with an Olympus endoscope and ancillary equipment for central operation, central display, automatic initial setting, and interlocking operation of the ancillary equipment. The intended use of the EndoALPHA as stated above is to enable a central system to control various pieces of ancillary equipment. However, the approved indications for use for each separate ancillary device dictate the type of procedures that may be performed. This information is included in the instruction manual for each ancillary piece of equipment. This intended use is identical to the previously cleared one for the Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 2. 2/3 {6} K051613 Pg 3 of 3 # 6 Conclusion The risk analysis was carried out in accordance with established in-house acceptance criteria based on ISO 14971:2000. The design verifications that were performed as a result of this risk analysis assessment are described below. | Modification | Test performed | Acceptance criteria | | --- | --- | --- | | Operating function for additional ancillary equipment | 1. Nurse's control panel operation test: We confirmed whether the Nurse's control panel operation is performed correctly. 2. Surgeon's controller operation test: We confirmed whether the Surgeon's controller operation is performed correctly. 3. Voice operation test: We confirmed whether the voice operation is performed correctly. | 1. All the operations that are in specification operate. 2. All the operations that are in specification operate. 3. All the operations that are in specification operate. | The Olympus Integrated Endosurgery System EndoALPHA (Control Unit for Endosurgery UCES-2) Software Version 3 has the following similarities to the predicate device: - has the same intended use, - uses the same operating principle. In summary, the subject device described in this submission is, in our opinion, substantially equivalent to the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...