AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
Device Story
Acetabular liner component for total hip arthroplasty; manufactured from enhanced UHMWPE material; designed to replace standard UHMWPE liners in existing AcuMatch A-Series hip systems. Device functions as bearing surface within acetabular shell; provides articulation for femoral head. Intended for use by orthopedic surgeons in clinical settings. Material enhancement intended to reduce wear while maintaining mechanical integrity. System includes 5 additional sizes to expand clinical utility. Device is terminally sterilized via gamma irradiation.
Clinical Evidence
No clinical data provided. Substantial equivalence is based on material property comparisons and bench testing to demonstrate that the enhanced polyethylene maintains suitable mechanical and physical properties compared to the predicate UHMWPE liners.
Indicated for skeletally mature patients undergoing primary total hip replacement for osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degeneration, or proximal femoral fractures. Also indicated for revision of failed reconstructions with sufficient bone stock and to restore mobility after fusion. Contraindicated in patients with active infection, insufficient bone stock, neuromuscular disorders preventing hip control, or patients with weight, age, or activity levels predisposing to early failure.
Regulatory Classification
Identification
A hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis is a device intended to be implanted to replace a hip joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across the joint. This generic type of device has a femoral component made of a cobalt-chromium-molybdenum (Co-Cr-Mo) alloy or a titanium-aluminum-vanadium (Ti-6Al-4V) alloy and an acetabular component composed of an ultra-high molecular weight polyethylene articulating bearing surface fixed in a metal shell made of Co-Cr-Mo or Ti-6Al-4V. The femoral stem and acetabular shell have a porous coating made of, in the case of Co-Cr-Mo substrates, beads of the same alloy, and in the case of Ti-6Al-4V substrates, fibers of commercially pure titanium or Ti-6Al-4V alloy. The porous coating has a volume porosity between 30 and 70 percent, an average pore size between 100 and 1,000 microns, interconnecting porosity, and a porous coating thickness between 500 and 1,500 microns. The generic type of device has a design to achieve biological fixation to bone without the use of bone cement.
AcuMatch® A-Series and MCS Acetabular Shells & Liners (K040613)
AcuMatch® A-Series 36mm Acetabular Liner, 12/14 Femoral Head and Press-Fit Femoral Stems (K041906)
Submission Summary (Full Text)
{1}------------------------------------------------
K051556 p1/3
SEP - 6 2005
# Summary of Safety and Effectiveness
# Exactech® AcuMatch® A-Series Enhanced Polyethylene Acetabular Liner
| Sponsor: | Exactech® Inc.<br>2320 N.W. 66th Court<br>Gainesville, Florida 32653 |
|----------------------|--------------------------------------------------------------------------------|
| Phone: | (352) - 377 - 1140 |
| Fax: | (352) - 378 - 2617 |
| FDA Establishment #: | 1038671 |
| Contact: | Diana Taylor<br>Regulatory Representative |
| Date: | August 26, 2005 |
| Trade Name: | Exactech® AcuMatch® A-Series Connexion GXL<br>Enhanced UHMWPE Acetabular Liner |
| Common Name: | Total Hip Prosthesis Acetabular Component |
| Device Description: | Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis | C.F.R. | 888.3358 |
|---------------------|----------------------------------------------------------------------------------------------|---------|----------|
| Product Code: | LPH | Section | |
| Device Description: | Hip joint metal/polymer semi-constrained cemented prosthesis | C.F.R. | 888.3350 |
| Product Code: | JDI | Section | |
| Device Description: | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis | C.F.R. | 888.3353 |
| Product Code: | MEH | Section | |
| Device Description: | Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis | C.F.R. | 888.3353 |
| Product Code: | LZO | Section | |
| Device Description: | Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis | C.F.R. | 888.3360 |
| Product Code: | LWJ | Section | |
Section 4
Page 1 of 3
{2}------------------------------------------------
### Summary of Safety and Effectiveness
Exactech® AcuMatch® A-Series Enhanced Polyethylene Acetabular Liner
### Intended Use / Indications:
AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
### Contraindications:
AcuMatch® A-Series Enhanced Polyethylene Acetabular liners are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
### Description:
This Special 510(k) premarket notification describes a material change from the predicate AcuMatch® A-Series UHMWPE polyethylene acetabular liners to acetabular liners manufactured from Enhanced Polyethylene. Additionally 5 new sizes are being introduced to the A-Series system.
### Packaging and Sterilization:
AcuMatch® A-Series Enhanced UHMWPE Acetabular Iiners are sealed in two Tyvek® pouches. The pouches are enclosed in a poly bag and placed into the outer carton. A tamper-resistant seals on the outer box contain the statement "sterile unless seal is broken". An outer product label is attached to the outer carton. The carton is then shrink-wrapped.
Products are terminally sterilized by a precision gamma irradiation dose of 25.2 - 30.8 kGy. The Sterility Assurance Level (SAL) is 10°.
> Section 4 Page 2 of 3
{3}------------------------------------------------
## Summary of Safety and Effectiveness
## Exactech® AcuMatch® A-Series Enhanced Polyethylene Acetabular Liner
### Legally Marketed Predicate Devices
| Model | Manufacturer | 510(k) # | Clearance<br>Date | Product<br>Code(s) |
|---------------------------------------------------------------------------------------------------|--------------|----------|-------------------|--------------------------------|
| AcuMatch® A-Series Porous<br>Coated Acetabular<br>Component | Exactech | K993082 | 11/19/1999 | LPH & JDI |
| AcuMatch® A-Series and<br>MCS Acetabular Shells &<br>Liners | Exactech | K040613 | 06/4/2004 | JDI, LPH &<br>MEH |
| AcuMatch® A-Series 36mm<br>Acetabular Liner, 12/14<br>Femoral Head and Press-Fit<br>Femoral Stems | Exactech | K041906 | 09/10/2004 | LPH, LZO,<br>MEH, LWJ<br>& JDI |
### Conclusion:
The proposed AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners are substantially equivalent to the predicate AcuMatch® A-Series Acetabular Liners of the same design, which have not had additional processing of the material. The proposed liners manufactured from enhanced material are expected to exhibit reduced wear, while maintaining suitable mechanical and physical properties.
> Section 4 Page 3 of 3
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a stylized human figure with outstretched arms.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
SEP - 6 2005
Ms. Diana Taylor Manager, Regulatory Affairs Exactech®, Inc. 2320 N.W. 66th Court Gainesville, Florida 32653
Re: K051556 Trade/Device Name: AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners Regulation Number: 21 CFR 888.3358, 21 CFR 888.3350, 21 CFR 888.3353, 21 CFR 888.3360 Regulation Name: Hip joint metal/polymer/metal semi-constrained porous-coated uncemented prosthesis, Hip joint metal/polymer semi-constrained cemented prosthesis, Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis, Hip joint femoral (hemi-hip) metallic cemented or uncemented prosthesis. Regulatory Class: II Product Code: LPH, JDI, LZO, MEH, LWJ Dated: August 09, 2005 Received: August 10, 2005
Dear Ms. Taylor:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave roved your corrent for the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for associated in the May 28, 1976, the enactment date of the Medical Device Amendments, or to econimer to may been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and Cosmetic Pror (110) that to device, subject to the general controls provisions of the Act. The I va may, dierely mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can may be subject to suen addrents regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2- Ms. Diana Taylor
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that I DA s issuated of a budenian and other requirements of the Act ulat IDA has made a decormination and Journer Federal agencies. You must of ally if ederal statutes and regulations and limited to: registration and listing (21 comply will an the 11ct 5 requirements, as a manufacturing practice requirements as set CTN Fart 607), adomig (21 OFR Part 820); and if applicable, the electronic forth in the quality systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) I his letter will anow you to begin halkening of substantial equivalence of your device to a legally premarket nothleadon: "The PDF misms sification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific acritorior your as not 100 100 - Also, please note the regulation entitled, Comaci the Office of Compilance and (21CFR Part 807.97). You may obtain Misoraliding by releveloc to premainter withilities under the Act from the Division of Small other general international and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
signature
Mark N. Melkerson Acting Director Division of General, Restorative, and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Exactech®, Inc.
### AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners
### Indications for Use
510(k) Number:
Kossrlo
Device Name: AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners
### Indications for Use:
AcuMatch® A-Series Enhanced Polyethylene Acetabular Liners are indicated for use in skeletally mature individuals undergoing primary surgery for total hip replacement due to osteoarthritis, rheumatoid arthritis, osteonecrosis, ankylosing spondylitis, congenital hip dysplasia, post-traumatic degenerative problems of the hip, and for treatment of proximal femoral fractures where prosthetic replacement is determined by the surgeon as the preferred treatment. Components of Exactech Hip Systems are also potentially indicated for revision of failed previous reconstructions where sufficient bone stock is present and to restore mobility resulting from previous fusion.
### Contraindications:
AcuMatch® A-Series Enhanced Polyethylene Acetabular liners are contraindicated in patients with active infection, patients without sufficient bone stock to allow appropriate insertion and fixation of the prosthesis, in neuromuscular disorders that do not allow control of the hip joint, and in patients whose weight, age, or activity level would cause the surgeon to expect early failure of the system.
| Prescription Use | X | or | Over the Counter Use | |
|------------------|---|----|----------------------|--|
|------------------|---|----|----------------------|--|
Please do not write below this line - use another page if needed.
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
| 510(k) Number | K051556 |
|---------------|-----------------------|
| | Section 3 Page 1 of 1 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.