← Product Code [MEB](/productcode/MEB) · K051474

# SYMBIOS DISPOSABLE INFUSION PUMP KIT (K051474)

_Symbios Medical Products, LLC · MEB · Jul 1, 2005 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051474

## Device Facts

- **Applicant:** Symbios Medical Products, LLC
- **Product Code:** [MEB](/productcode/MEB.md)
- **Decision Date:** Jul 1, 2005
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5725
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management.

## Device Story

Symbios Disposable Infusion Pump Kit is a convenience kit for temporary liquid infusion. Components include an elastomeric pump, catheter needle/introducer, infusion catheter, and bandage. Principle of operation: inflatable elastomeric reservoir mechanically pressurizes fluid, driving it through tubing and a small restrictor to achieve a pre-set infusion rate. Used in hospital or home settings for post-operative pain management; typically managed by clinicians or patients following physician instructions. Device is provided empty; drug selection is determined by the healthcare provider. Benefits include continuous delivery of topical anesthetics, facilitating outpatient recovery and reducing the need for frequent clinical visits for infusion management.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

System consists of an inflatable elastomeric reservoir, infusion catheter, catheter needle/introducer, and bandage. Operates via mechanical pressure from the elastomeric reservoir through a flow-restricting orifice. Non-electronic, passive infusion system. Provided non-sterile or sterile (implied by kit nature).

## Regulatory Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 051474

# Exhibit 6 Summary of Safety and Effectiveness

111 1 - 2005

Trade Name: Disposable Infusion Pump Kit

Common Name: Disposable Infusion Pump Kit

### Classification Name and Reference:

Pump

Pump. Infusion. Elastomeric, External (21 CFR 880.5725) Device Classification for the subject and/or predicate devices: Class II Device Panel Code: 80 Device Product Code: MEB

Catheter Needle/Introducer Catheter, Intravascular, short term (21 CFR 880.5200) Device Classification for the subject and/or predicate devices: Class II Device Panel Code: 80 Device Product Code: FOZ

Infusion Catheter

Catheter, Conduction, Anesthetic (21 CFR 868.5120) Device Classification for the subject and/or predicate devices: Class II Device Panel Code: 73 Device Product Code: BSO

## Device description:

Infusion of liguids in to a patient in the general hospital setting as well as at home is frequently required in medical treatment. For example, in orthopedics a disposable device is often indicated after outpatient arthroscopic surgery in order to infuse topical anesthetics for several days following the patient's return home. The infusion catheter is usually removed when the patient returns to the physician's office for a follow-up visit.

The Symbios Disposable Infusion Pump Kit is a convenience kit that includes the components necessary to provide temporary infusion of a liquid into a patient. The components of the system are an elastomeric pump, a catheter needle/introducer, an infusion catheter and a bandage. The pump is a disposable, self-contained, infusion system utilizing an inflatable elastomeric reservoir to mechanically pressurize a fluid and drive it through tubing to a small restrictor to provide infusion at a pre-set rate.

The device is provided empty and no specific drug references are made in the labeling. The device is not intended for delivery of blood, blood products, lipids or fat emulsions.

#### Intended use:

The Symbios Disposable Infusion Pump Kit is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a pre-set rate for post-operative pain management.

{1}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

JUL 1 - 2005

Mr. Jeffery J. Alholm CEO Symbios Medical Products, LLC 6131 West 80th Street Indianapolis, Indiana 46278

Re: K051474 Trade/Device Name: Disposable Infusion Pump Kit Regulation Number: 21 CFR 880.5725, 880.5200, 868.5120 Regulation Name: Infusion pump, Intravascular catheter, Anesthesia conduction catheter Regulatory Class: II Product Code: MEB, KGZ, BSO Dated: June 2, 2005 Received: June 7, 2005

Dear Mr. Alholm:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{2}------------------------------------------------

Page 2- Mr. Jeffery J. Alholm

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu Lin, Ph.D.

Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use

510(k) Number (if known): K051474

Device Name: Disposable Infusion Pump Kit

Indications For Use:

The Disposable Infusion Pump is a disposable, self-contained infusion system utilizing an inflatable elastomeric reservoir to mechanically provide percutaneous infusion of prescribed solutions at a proset rate for post-operative pain management.

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Auy Sulin

(Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

Page 1 of __

510(k) Number:

---

**Source:** [https://fda.innolitics.com/device/K051474](https://fda.innolitics.com/device/K051474)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
