WIESLAB CAP PR-3 ANCA

K051458 · Eurodiagnostica · MOB · Nov 22, 2005 · Immunology

Device Facts

Record IDK051458
Device NameWIESLAB CAP PR-3 ANCA
ApplicantEurodiagnostica
Product CodeMOB · Immunology
Decision DateNov 22, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5660
Device ClassClass 2

Intended Use

The Wieslab™ Cap PR3-ANCA is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to proteinase 3 (PR3) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Wegener's granulomatosis. The assay is intended for use in patients with signs and symptoms consistent with WG. It is not intended for screening a healthy population. The analysis should be performed by trained laboratory professionals.

Device Story

Wieslab™ Cap PR3-ANCA is an ELISA test kit for detecting IgG antibodies to proteinase 3 (PR3) in human serum. Microtiter wells are coated with monoclonal antibody to PR3 and PR3 antigen. Patient serum is incubated in wells; specific antibodies bind to antigen. After washing, alkaline phosphatase-labeled anti-human IgG conjugate is added. Substrate solution is added; color intensity (optical density) correlates to bound antibody concentration. Performed by trained laboratory professionals in clinical settings. Results are calculated against a calibrator curve in arbitrary units. Output aids physicians in diagnosing Wegener's granulomatosis in symptomatic patients.

Clinical Evidence

Clinical performance evaluated on 363 characterized sera. Sensitivity for Wegener’s granulomatosis (WG) was 96.0% (95% CI: 83.3-99.5%). Specificity in healthy controls was 100% (95% CI: 97.0-100%), SLE 100% (95% CI: 92.0-100%), and RA 100% (95% CI: 80.5-100%). Method comparison with predicate (n=180) showed 98.3% overall agreement. Analytical precision (intra-assay CV 3-13%, inter-assay CV 4-14%) and linearity (10-320 U/mL) were established.

Technological Characteristics

ELISA-based immunological test system. Components: microtiter strips coated with monoclonal antibody to PR3 and PR3 antigen. Detection: alkaline phosphatase-labeled anti-human IgG conjugate and substrate solution. Measurement: optical density (OD) via spectrophotometry. Quantitative output in arbitrary units.

Indications for Use

Indicated for patients with signs and symptoms consistent with Wegener’s granulomatosis (WG) to aid in diagnosis. Not intended for screening healthy populations.

Regulatory Classification

Identification

A multiple autoantibodies immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the autoantibodies (antibodies produced against the body's own tissues) in serum and other body fluids. Measurement of multiple autoantibodies aids in the diagnosis of autoimmune disorders (disease produced when the body's own tissues are injured by autoantibodies).

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K05/458 ## Summary of Safety and Effectiveness Information Capture PR-3 ANCA Test Kit NOV 2 3 2005 - I. Euro-Diagnostica AB Per Albin Hanssons vag 41, Medeon SE-205 12 Malmo Sweden Contact person: Sten Gershagen Telephone: 46-40-321100 Date of preparation: May 4, 2005 II. Description of Device: The Capture-PR3-ANCA test kit is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to proteinase 3 (PR3) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Wegener's granulomatosis. The assay is intended for use in patients with signs and symptoms consistent with WG. It is not intended for screening a healthy population. The analysis should be performed by trained laboratory professionals. The wells of the microtiter strips are coated with monoclonal antibody to proteinase 3 and subsequently proteinase 3. During the first incubation, specific antibodies in diluted serum, will bind to the antigen coating. The wells are then washed to remove unbound antibodies and other components. A conjugate of alkaline phosphatase-labeled antibodies to human IgG binds to the antibodies in the wells in this second incubation. After a further washing step, detection of specific antibodies is obtained by incubation with substrate solution. The amount of bound antibodies correlates to the color intensity and is measured in terms of absorbance (optical density (OD)). The absorbance is then calculated against a calibrator curve and the results are given in arbitrary units. ## III. Predicate Device The Capture PR-3 ANCA test is substantially equivalent to the PR-3 ANCA ELISA Kit. Equivalence is demonstrated by the following comparative results: {1}------------------------------------------------ | Control and<br>Disease groups | Total<br>number | Negative<br>< 20 units | Equivocal<br>21-30 units | Positive<br>> 30 units | |-------------------------------|-----------------|------------------------|--------------------------|------------------------| | Blood donors: | 120 | 120 | 0 | 0 | | WG: | 50 | 2 | 0 | 48 | | MP: | 34 | 17 | 2 | 15 | | SLE: | 45 | 44 | 1 | 0 | | RA: | 17 | 16 | 1 | 0 | | GBM: | 29 | 26 | 0 | 3 | Table 1. Clinical sensitivity and specificity. A total of 295 frozen retrospective sera with clinical characterisation were assayed. The following table summarises the results WG = Wegener's granulomatosis, SLE = systemic lupus erythematosus MP = microscopic polyangiitis RA = rheumatoid arthritis GBM = glomerular basement membrane Clinical sensitivity (Equivocal samples excluded from calculations) 95% CI = 86.3-99.5 % WG = 48/50 = 96.0 % 95% CI = 29.1-65.3 % MP == 15/32 = 46.9 % ## Clinical specificity (Equivocal samples excluded from calculations) 95% CI = 72.7-97.8 % GBM = 26/29 = 89.7 % 95% CI = 92.0-100 % SLE = 44/44 = 100 % 95% CI = 79.4-100 % = 16/16 = 100 % RA Normal = 120/120 = 100 % 95% CI = 97.0-100 % The 95% confidence interval (CI) was calculated using the exact method. {2}------------------------------------------------ Table 2. Wielisa capture PR3-ANCA kit compared to IIF-ANCA. A total of 78 frozen retrospective sera from systemic vasculitis were assayed. The following table summarises the results. | | IIF-ANCA | Positive | Equivocal | Negative | Total | |----------|----------|----------|-----------|----------|-------| | c-ANCA | | 52 | 0 | 1 | 53 | | p-ANCA | | 1 | 1 | 13 | 15 | | Negative | | 10 | 0 | 0 | 10 | | Total | | 63 | 1 | 14 | 78 | Table 3. Relative sensitivity and specificity of the Wielisa capture PR3-ANCA kit compared to an alternative ELISA. A total of 180 frozen retrospective sera were assayed. The following table summarises the results. | | Positive | Equivocal | Negative | Total | |--------------------------|----------|-----------|----------|-------| | Positive | 40 | 0 | 0 | 40 | | Alternative<br>Equivocal | 2 | 0 | 0 | 2 | | ELISA<br>Negative | 2 | 2 | 134 | 138 | | Total | 44 | 2 | 134 | 180 | Sera falling in the equivocal range were excluded from the following calculations. | Relative sensitivity | = 40/40 = 100% | 95% CI = 91.2-100 % | |----------------------|--------------------|----------------------| | Relative specificity | = 134/136 = 98.5 % | 95% CI = 94.8-99.8 % | | Relative Accuracy | = 174/176 = 98.9 % | 95% CI = 96.0-99.9 % | The 95% confidence interval (CI) was calculated using the exact method. Table 4. Batch to batch variation was determined by testing three different samples in duplicate. Results were obtained for five different batches. | Sample | Mean value | SD | CV % | |--------|------------|----|------| | PK | 76 units | 10 | 13 | | K1 | 45 units | 3 | 7 | | K3 | 88 units | 10 | 11 | {3}------------------------------------------------ Table 5. Inter-assay precision was determined by testing four different samples in duplicate. Results were obtained for four different runs. | Sample | Mean value | SD | CV % | |--------|------------|----|------| | PK | 85 units | 5 | 6 | | K1 | 49 units | 3 | 6 | | K2 | 77 units | 5 | 7 | | K3 | 138 units | 9 | 7 | Table 6. Intra-assay precision was determined by testing two samples in 24 wells. | Sample | A-4-4-4-4<br>Mean value | CITY<br>1 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Acres of can and a compare<br>77 01<br>/0 | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |--------|-------------------------|-------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | (<br>1 30 | 1000 - 1000 | | | | 1 | D<br>1 | 4 A<br>ﺑ | ----------------------------------------------------------------------------------<br>Annual Partic Participant Property Controller Collection Comers of Children Comers of Children Comments of Children Comments of Children Comments of Children Comments of Chil | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Table 7. Linearity. The values were determined for serial two-fold dilutions of three positive sera. The I able 7 Elizarion the has 2 of dilution by standard linear regression. The data indicates that the assay has a linear relationship with serum dilution. | Serum | neat | 1:2 | 1:4 | 1:8 | 1:16 | 1:32 | r | |-------|------|-----|-----|-----|------|------|------| | 1 | 394 | 297 | 188 | 74 | 37 | 16 | 0.95 | | 2 | 334 | 269 | 106 | 48 | | | 0.94 | | 3 | 324 | 219 | 100 | 36 | 22 | 8 | 0.98 | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in all caps. The text is in a bold, serif font. The text is centered horizontally in the image. The background of the image is white. Public Health Service Eurodiagnostica · c/o Mr. William L. Boteler Jr. Immuno Probe, Inc. 1306 Bailes Lane, Suite F Frederick, MD 21701 Food and Drug Administration 2098 Gaither Road Rockville MD 20850 ## NOV 2 3 2005 Re: k051458 Trade/Device Name: Wieslab™ Cap PR-3 ANCA Regulation Number: 21 CFR 866.5660 Regulation Name: Multiple autoantibodies immunological test system Regulatory Class: Class II Product Code: MOB Dated: June 1, 2005 Received: June 2, 2005 Dear Mr. Boteler: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. Image /page/4/Picture/12 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the eagle, indicating the department's name and national affiliation. The seal is simple and monochromatic. {5}------------------------------------------------ Page 2 - If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0484. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html Sincerely yours, loburtz Beckerh Robert L. Becker, Jr., MD, PK Director Division of Immunology and Hematology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ Page 1 of 1 510(k) Number: K051458 Device Name: Wieslab™ Cap PR3-ANCA Indications For Use: The Wieslab™ Cap PR3-ANCA is an enzyme-linked immunosorbent assay (ELISA) for detection and semi-quantitation of IgG antibodies to proteinase 3 (PR3) in human sera. The assay is used to detect antibodies in a single serum specimen. The results of the assay are to be used as an aid to the diagnosis of Wegener's granulomatosis. The assay is intended for use in patients with signs and symptoms consistent with WG. It is not intended for screening a healthy population. The analysis should be performed by trained laboratory professionals. PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) スマーニュレア | ------------------------------------------------------------------------------------------------------------------------------------------------------------------ Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use V (Per 21 CFR 801.109) OR Over-The Counter Use (Optional Format 1-2-96) Maria Chan Division Stop-On of In Vitro Diagnostic 510(k) K051454
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...