LNOP BLUE OXIMETRY SENSOR

K051439 · Masimo Corporation · DQA · Jul 15, 2005 · Cardiovascular

Device Facts

Record IDK051439
Device NameLNOP BLUE OXIMETRY SENSOR
ApplicantMasimo Corporation
Product CodeDQA · Cardiovascular
Decision DateJul 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2700
Device ClassClass 2
AttributesPediatric

Indications for Use

The LNOP Blue oximetry sensors are intended for the continuous nomitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for pediatric, infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments.

Device Story

Disposable pulse oximetry sensor; connects to Masimo SET compatible monitors. Uses non-invasive optical assessment (emitters/detectors) to measure tissue oxygenation. Designed specifically for pediatric, infant, and neonatal patients with congenital cyanotic cardiac lesions. Operates in hospitals, mobile, and home environments. Provides continuous SpO2 and pulse rate data to clinicians/caregivers. Output enables real-time assessment of arterial hemoglobin saturation, facilitating clinical decision-making regarding oxygenation status and patient stability.

Clinical Evidence

Clinical study on pediatric, infant, and neonatal patients with congenital cyanotic cardiac lesions. Compared sensor SpO2 values against arterial blood CO-Oximetry. Results: accuracy <4% SpO2 ARMs (60-80% SaO2 range) and <3% SpO2 ARMs (80-100% SaO2 range).

Technological Characteristics

Disposable optical sensor; non-invasive emitters/detectors; compatible with Masimo SET monitors. Materials identical to LNOP Inf-L sensors; biocompatible (non-toxic, non-irritating, non-sensitizing).

Indications for Use

Indicated for continuous monitoring of functional oxygen saturation (SpO2) and pulse rate in pediatric, infant, and neonatal patients with congenital cyanotic cardiac lesions in hospital, mobile, and home settings.

Regulatory Classification

Identification

An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) SUMMARY K051439 JUL 1 5 2005 e Image /page/0/Picture/3 description: The image shows the logo for Masimo, a medical technology company. Below the logo is the address "40 Parker Irvine, CA 92618". The image also includes the phone number 949-297-7000 and fax number 949-297-7001. | Submitted by: | Masimo Corporation<br>40 Parker<br>Irvine, CA 92618<br>949-297-7000<br>FAX 949-297-7001 | |---------------------------------------|-----------------------------------------------------------------------------------------| | Company Contact: | James J. Cronin, Vice President, Regulatory Affairs/Quality Assurance | | Date Summary Prepared: | May 23, 2005 | | Trade Name | LNOP Blue Oximetry Sensors | | Common Name | Oximeter Sensor | | Classification Name and Product Code: | Oximeter (74DQA) (870.2700) | | Substantially Equivalent Devices: | Masimo SET® Rad 57 Pulse CO-Oximeter<br>510(k) Number - K042536. | ### Device Description The LNOP Blue Oximetry Sensors are fully compatible disposable sensors for use with Masimo SET compatible pulse oximeter monitors. The LNOP Blue sensor is substantially equivalent to Masimo's LNOP Inf-L sensor except it is designed for use on patients with congenital cyanotic cardiac lesions. #### Intended Use The LNOP Blue oximetry sensors are intended for the continuous nomitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for pediatric, infant, and neonatal patients in hospitals, hospital-type facilities, mobile, and home environments. {1}------------------------------------------------ # 510(k) SUMMARY ### Technology Comparison The LNOP Blue oximetry sensors are equivalent in intended use, design, principles of operation, materials, and performance to predicate sensors and operate on identical principles of non-invasive optical assessment of tissue oxygenation using emitters and detectors. The LNOP Blue oximetry sensors are designed, configured, and manufactured for full compatibility with Masimo SET and Masimo SET compatible pulse oximetry sensors are constructed of similar materials and components as used in the predicate devices. ### Performance Testing #### Biocompatibility All the patient contacting materials used in the LNOP Blue sensors are the same materials that are used in Masimo's LNOP Inf-L sensors. Test results demonstrated that the materials were non-toxic, non-irritating, and non sensitizing. ### Clinical Testing Clinical studies were performed using the LNOP Blue Disposable oximetry sensors with Masimo SET Radical Pulse Oximeters on pediatric, infant, and neonatal patients with congenital cyanotic cardiac lesions and measuring the arterial hemoglobin saturation value with the instruments against the arterial hemoglobin oxygen determined from arterial blood samples with a CO-Oximeter. Clinical testing of the LNOP Disposable sensors resulted in an accuracy of less than 4% SpO2 ARMs in the range of 60%-80% SaO2 and less than 3% SpO2 ARMS in the range of 80%-100% for pediatrics, infants and neonates on patients with congenital cyanotic lesions. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH AND HUMAN SERVICES, USA" around the perimeter. Inside the circle is a stylized symbol that resembles three abstract human figures or a caduceus-like design, representing health and well-being. Public Health Service JUL 1 5 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Mr. James J. Cronin Vice President, Regulatory Affairs/Quality Assurance Masimo Corporation 40 Parker Irvine, California 92618 Re: K051439 Trade/Device Name: LNOP Blue Oximetry Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximetry Sensor Regulatory Class: II Product Code: DQA Dated: May 31, 2005 Received: June 02, 2005 Dear Mr. Cronin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. Cronin Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Sujite J. Michie, M.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indications for Use #### 510(k) Number (if known): LNOP Blue Device Name: Indications For Use: The LNOP Blue Sensors are indicated for the continuous nonitoring of fifictional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate (measured by an SpO2 sensor) for use with pediatric, infant, and neonatal patients with congenital cyanotic cardiac lesions in hospital-type facilities, mobile, and home environments. Prescription Use _ X (Per 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use _ (Per 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ain Sshom (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devic 510(k) Number:_
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...