← Product Code [NDQ](/productcode/NDQ) · K051373

# CRONUS GUIDEWIRE, MODEL 001-001470-1 (K051373)

_Stereotaxis, Inc. · NDQ · Dec 9, 2005 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K051373

## Device Facts

- **Applicant:** Stereotaxis, Inc.
- **Product Code:** [NDQ](/productcode/NDQ.md)
- **Decision Date:** Dec 9, 2005
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1290
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Telstar® Magnetic Navigation System [TMNS] is intended to navigate a magnetic-tipped device through tissue to designated target sites in the right and left heart, coronary, peripheral, and neurovasculature by orienting the device tip in a desired direction. The Stereotaxis Endovascular Guide Wire is intended to introduce and position over-the-wire catheters and other over-the-wire therapeutic devices within the neurovasculature, coronary and peripheral vasculature during PTCA or other intravascular interventional procedures.

## Device Story

The Stereotaxis Telstar® Magnetic Navigation System (TMNS) is an interventional workstation used in clinical settings to navigate magnetic-tipped guidewires through the vasculature. The system integrates a bi-planar fluoroscopy system with a computer-controlled magnetic field generator. The Cronus® Modified guidewire, featuring an embedded magnet at its tapered distal tip, is inserted into the patient; the TMNS applies external magnetic fields to orient the guidewire tip, allowing the physician to steer the device to target sites. This system facilitates the delivery of over-the-wire catheters and therapeutic devices. By enabling precise, remote-controlled navigation, the system aims to improve access to complex vascular anatomy. The physician monitors the procedure via fluoroscopic visualization and controls the magnetic field orientation to guide the wire, which transmits axial force from the proximal end to the tip.

## Clinical Evidence

Evidence includes bench testing, pre-clinical studies, and clinical in vivo testing. Results demonstrate that the TMNS and Cronus® Modified perform in an equivalent manner to the identified predicate devices. No specific statistical metrics (e.g., p-values, AUC) were provided in the summary.

## Technological Characteristics

The TMNS uses a computer-controlled magnetic field generator and bi-planar fluoroscopy. The Cronus® Modified is a 0.014" (0.36 mm) hydrophilic-coated guidewire with an embedded magnet and tapered distal tip. Lengths range from 180-300 cm. Sterilization is performed using 100% ethylene oxide. The system is networked for imaging and magnetic control.

## Regulatory Identification

A steerable catheter control system is a device that is connected to the proximal end of a steerable guide wire that controls the motion of the steerable catheter.

## Predicate Devices

- Stereotaxis Telstar® Magnetic Navigation System [TMNS] ([K013484](/device/K013484.md))
- BrainLAB VectorVision2 (VV2) ([K983831](/device/K983831.md))
- Stereotaxis Cronus® Predicate ([K042854](/device/K042854.md))
- Boston Scientific TRANSEND™ EX Platinum Steerable Guide Wire ([K971254](/device/K971254.md))

## Reference Devices

- Telstar® Digital Imaging System [TIS] ([K013484](/device/K013484.md), [K022565](/device/K022565.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the Department of Health and Human Services (HHS). The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Public Health Service

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

MAR - 8 2006

Ms. Kelly Rowland, MA Regulatory Affairs Specialist Ouality and Regulatory Affairs STEREOTAXIS, INC. 4041 Forest Park Avenue St. Louis, MO 63108

Re: K051373

Trade/Device Name: Cronus Guidewire, Model 001-001470-1 Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter guide wire Regulatory Class: II Product Code: NDQ Dated: November 29, 2005 Received: December 1, 2005

Dear Ms. Rowland:

This letter corrects our substantially equivalent letter of December 9, 2005

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices. good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA´s issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Rowland

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Barbara Muelund
to:

Mark N. Melkerson Acting Director Division of General. Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

051373

## Appendix 2: Indications for Use Statement

Statement

The indications for Use Statement:

510(k) Number: K

Device Name: Stereotaxis Guidewire and Stereotaxis Telstar® Magnetic Navigation System [MNS]

Stereotaxis Guidwire:

The Stereotaxis Cronus® Endovascular Guidewire is intended to introduce and position over-the-wire catheters and other over-the-wire therapeutic devices within the neurovasculature, coronary and peripheral vasculature during PTCA or other intravascular interventional procedures.

The Stereotaxis Cronus® Endovascular Guidewire is not intended for use without a Stereotaxis Magnetic Navigation System [MNS]. Rotational atherectomy devices and any ferromagnetic interventional devices are contraindicated for use with the Stercotaxis Endovascular Guidewire.

Stereotaxis Telstar" Magnetic Navigation System [MNS]: The TMNS is intended to navigate a magnet-tipped device through tissue to designated target sites in the right and left heart, coronary, peripheral, and neurovasculature by orienting the device tip in a desired direction.

Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR (Part 21 CFR 801 Subpari D)

Over-The-Counter Use (Part 21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Alba EDE MADRY
Conference
DRU, Office of Device Evaluation (ODE)
XM
(Division Sign-Off)

and Neurological

Page い!

1051373

The information hority WIS. Inc. in accordance with the provisions and 1 38 and 21 CTR 88149

{3}------------------------------------------------

TMNS & Cronus® Modified 510(k) Stereotaxis, Inc.

, ****

:

.

: : : ・

DEC 9 2005
K051373

May 20, 2005 Page 45

|                                  | Appendix 1: 510(k) Summary of Safety and Effectiveness                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Page |
|----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Statement                        | Information supporting claims of substantial equivalence, as defined under<br>the Federal Food, Drug and Cosmetic Act, respecting safety and effectiveness<br>is summarized below.<br><br>For the convenience of the Reviewer, this summary is formatted in<br>accordance with the Agency's final rule "...510(k) Summaries and 510(k)<br>Statements..." (21 CFR §807) and can be used to provide a substantial<br>equivalence summary to anyone requesting it from the Agency.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |      |
| <b>Device<br/>description</b>    | The Telstar® Magnetic Navigation System [TMNS] is an interventional<br>workstation for the intravascular navigation of a magnetic-tipped device<br>through tissue to designated target sites in the right and left heart, coronary,<br>peripheral, and neurovasculature. It combines a bi-planar fluoroscopy system<br>with a computer controlled magnetic field generator, to provide both<br>visualization and control of a magnetically actuated guidewire. The system<br>employs magnetic fields to orient the guidewire.<br><br>The Stereotaxis Cronus® Modified is a steerable guidewire that has a nominal<br>diameter of 0.014 in/0.36 mm and a nominal length of 210 cm or 300 cm.<br>The guidewire is designed only for use in conjunction with a Stereotaxis<br>Telstar® Magnetic Navigation System [TMNS]. The wire is configured with<br>a tapered distal tip and an embedded magnet, which is used for navigating the<br>wire through the vasculature. This device is sterilized with 100% ethylene<br>oxide. |      |
| <b>Intended use</b>              | The Telstar® Magnetic Navigation System [TMNS] is intended to navigate a<br>magnetic-tipped device through tissue to designated target sites in the right<br>and left heart, coronary, peripheral, and neurovasculature by orienting the<br>device tip in a desired direction.<br><br>The Stereotaxis Endovascular Guide Wire is intended to introduce and<br>position over-the-wire catheters and other over-the-wire therapeutic devices<br>within the neurovasculature, coronary and peripheral vasculature during<br>PTCA or other intravascular interventional procedures.                                                                                                                                                                                                                                                                                                                                                                                                                                             |      |
| Continued                        | Page 2 of 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |      |
| Technological<br>characteristics | The Telstar® Magnetic Navigation System [TMNS] employs application of<br>magnetic fields to orient the distal tip of a magnetically actuated guidewire.<br>The Telstar® Digital Imaging System [TIS], previously cleared under<br>K013484 and K022565, provides visualization through standard fluoroscopy.<br><br>The Cronus® Modified is a conventional 0.014" (0.36 mm) hydrophilic-<br>coated endovascular guidewire modified to accommodate magnetic actuation<br>and control. It is designed to navigate within the vasculature to deliver a<br>suitable catheter or interventional device to a desired site. The finished<br>lengths of the Cronus® Modified are between 180 and 300 cm. A taper runs<br>32 cm proximal to the distal tip. The pushable shaft is a continuous wire that<br>allows axial force, applied at the proximal end, to be transmitted to the tip of<br>the guidewire.                                                                                                                        |      |
| Performance<br>data              | Bench testing and pre-clinical and clinical <i>in vivo</i> testing demonstrate that the<br>Stereotaxis Telstar® Magnetic Navigation System [TMNS] and the Cronus®<br>Modified perform in an equivalent manner to the Stereotaxis Telstar®<br>Magnetic Navigation System [TMNS], and the Stereotaxis Cronus® Predicate<br>and the Boston Scientific TRANSENDTM EX Platinum Steerable Guide Wire<br>predicate devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |      |
| Conclusion                       | The Stereotaxis Telstar® Magnetic Navigation System [TMNS] is<br>substantially equivalent to the Stereotaxis Telstar® Magnetic Navigation<br>System [TMNS] (K013484) and the BrainLAB VectorVision2 (VV2)<br>(K983831) predicate devices.<br><br>The Stereotaxis Cronus® Modified is substantially equivalent to the<br>Stereotaxis Cronus® Predicate (K042854) and the Boston Scientific<br>TRANSENDTM EX Platinum Steerable Guide Wire (K971254) predicate<br>devices.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |      |
| Contact                          | Kelly Rowland<br>Regulatory Specialist                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |      |
| Date                             | May 20, 2005                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |      |

The information herein is considered CONFIDENTIAL to STEREOTAXIS, Inc. in accordance with the provisions and 13 considered CONFIDENTIAN (21 CFR §812.38, and 21 CFR §814.9.

{4}------------------------------------------------

: 24

## Appendix 1: 510(k) Summary of Safety and Effectiveness,

The information herein is considered CONFIDENTIAL to STEREOTAXIS, Inc. in accordance with the provisions and expectations of 21 CFR §20.61, 21 CFR §812.38, and 21 CFR §814.9.

---

**Source:** [https://fda.innolitics.com/device/K051373](https://fda.innolitics.com/device/K051373)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
