PRECEDENCE SPECT/CT IMAGING SYSTEM

K051351 · Adac Laboratories · KPS · Jun 8, 2005 · Radiology

Device Facts

Record IDK051351
Device NamePRECEDENCE SPECT/CT IMAGING SYSTEM
ApplicantAdac Laboratories
Product CodeKPS · Radiology
Decision DateJun 8, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1200
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

Precedence is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. The x-ray computed tomography subsystems may consist of a whole body multi-slice CT or cardiovascular CT. The cardiovascular subsystem provides coronary imaging that is intended to produce cross-sectional images of the heart, cardiovascular, and peripheral vascular systems by computer reconstruction of x-ray transmission data taken at difference angles and planes. Precedence may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data for anatomical localization of the nuclear medicine data. Precedence may be used either as a separate single photon system, a separate CT system or as a combined CT and single photon system. The nuclear medicine and CT images may be transferred to other systems such as a radiation therapy planning system. The Precedence Imaging System should only be used by trained healthcare professionals.

Device Story

Hybrid SPECT/CT system; combines Skylight SPECT and Brilliance CT subsystems (6, 16, or 64 slice or CV configuration) in integrated housing. Inputs: radiopharmaceutical distribution (SPECT) and x-ray transmission data (CT). Operation: sequential acquisition of SPECT and CT; automatic coregistration and fusion of functional and anatomical data. Output: fused SPECT/CT images, non-attenuation/attenuation-corrected nuclear images, and CT cross-sectional images. Used in clinical settings by trained healthcare professionals. Dual consoles (acquisition room and shielded control room). Output used for anatomical localization of nuclear medicine data and comprehensive diagnosis. Benefits: faster diagnostic decision-making compared to separate systems; improved diagnostic accuracy via anatomical detail.

Clinical Evidence

Bench testing only. Performance verified through compliance with radiation safety standards (21 CFR 1020.30, 1020.33), electrical/mechanical safety standards (IEC 60601-1), and NEMA NU-1 imaging performance standards. No clinical data presented.

Technological Characteristics

Hybrid SPECT/CT system. Subsystems: Skylight SPECT and Brilliance CT (6, 16, 64-slice or CV configuration). Energy: Ionizing radiation (x-ray and gamma). Connectivity: Dual acquisition consoles (in-room and control room). Standards: 21 CFR 1020.30/1020.33, IEC 60601-1, NEMA NU-1. Software: Integrated acquisition and fusion processing.

Indications for Use

Indicated for imaging of the heart, cardiovascular, and peripheral vascular systems using the CV configuration CT subsystem; general nuclear medicine and CT imaging for anatomical localization. Contraindicated for non-cardiac and non-vascular structures (soft tissue and bone) when using the CV configuration CT subsystem.

Regulatory Classification

Identification

An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # Section B – Administrative Information r051351 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS General Information | A. | Submitted By: | ADAC Laboratories<br>540 Alder Dr.<br>Milpitas, CA 95035 | Contact: Coleen Coleman<br>Tel: (408) 468-3051<br>Fax: (408) 468-3050 | |----|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------| | B. | Device Trade Name: | Precedence SPECT/CT Imaging System | | | | Common Name: | Single Photon Emission Computed Tomography<br>Computed Tomography X-Ray | | | | Classification Name: | System, Emission Computed Tomography, (892.1200)<br>System, Computed Tomography X-Ray, (892.1750) | | | | Device Class: | 21CFR 892.1200, Class II<br>21 CFR 892.1750, Class II | | | | Product Code: | 90 KPS and 90 JAK | | | C. | Date prepared: | May 6, 2005 | | | D. | Predicate Device: | Precedence SPECT/CT (cleared as Griffin) (K041218)<br>Predicate Philips Plus CT Scanner (K033326)<br>Brilliance CT, Private Practice CV configuration<br>CT Scanner (K042293) | | #### E. Performance Standards - 21 CFR 1020.30 1020.33 as applicable for Ionizing Radiation Emitting . Products (Applicable Sections) - . NEMA NU-1 #### F. Intended Use: Precedence is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. The x-ray computed tomography subsystems may consist of a whole body multi-slice CT or cardiovascular CT. The cardiovascular subsystem provides coronary imaging that is intended to produce cross-sectional images of the heart, cardiovascular, and peripheral vascular systems by computer reconstruction of x-ray transmission data taken at difference angles and planes. Precedence may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT transmission data may be used to produce attenuation corrected nuclear medicine images. The nuclear medicine images and the CT images may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data for anatomical localization of the nuclear medicine data. Precedence may be used either as a separate single photon system, a separate CT system or as a combined CT and single photon system. The nuclear medicine and CT images may be transferred to other systems such as a radiation therapy planning system. The Precedence Imaging System should only be used by trained healthcare professionals. {1}------------------------------------------------ Contraindications for Use: The Precedence CV Configuration CT subsystem is not indicated for imaging structures that are defined as non-cardiac and non-vascular (including soft tissue and bone). - F. Device Description: The Precedence SPECT/CT Imaging System is a hybrid SPECT/CT system for performing CT studies, general nuclear medicine studies, or SPECT/CT sequentially (dual-modality studies) wherein the SPECT and CT studies may be automatically coregistered and displayed in fused form. Because the natures of the imaging modalities, they provide different information; the SPECT study yields functional information about metabolic processes and the CT study yields structural or anatomical information. As radionuclides become more tissue specific, diagnoses from nuclear images alone will be more difficult without the general anatomical detail less specific agents provide. Thus fused SPECT and CT images will provide the information required for accurate and comprehensive diagnoses. Precedence is constructed from two existing systems, the Skylight SPECT Imaging System and the Brilliance CT with subsystem options of 6, 16, or 64 slice or the CVCT configuration. The Precedence has two acquisition consoles. One console is placed in the acquisition room itself, consistent with the SPECT convention, and the other console is placed in the shielded scanner control room, as required for CT. The acquisition stations provide a single user interface for both SPECT and CT patient acquisition set-up. The SK Ylight and Brilliance system gantries remain intact as major subsystem components located within a common integrated housing. The combined Precedence SPECT-CT Imaging System is designed so that the system can operate in three modes: CT only, SPECT only, and combined SPECT/CT performed sequentially. No modifications have been made to either system, which would affect system performance. Precedence is intended for use primarily as an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. The clinical protocols and procedures are that are available on the modified Precedence Imaging System are the same as those on the Predicate Devices Precedence SPECT/CT (cleared as Griffin) (K041218), Predicate Philips Plus CT Scanner (K033326), and the Brilliance CT, Private Practice CV configuration CT Scanner (K042293). The intended use of the CV configuration CT Scanner is limited to the heart, cardiovascular, and peripheral vascular systems. Acquired SPECT and CT images on the Precedence may be registered and displayed in a fused format (overlaid in the same orientation) to provide combined single photon and anatomical data to provide anatomical localization of the nuclear medicine image. The Precedence CV Configuration CT subsystem is not indicated for imaging structures that are defined as non-cardiac and non-vascular (including soft tissue and bone). {2}------------------------------------------------ #### Comparison to Predicate Device: G. The modified Precedence SPECT/CT Imaging System and the Precedence predicate (K041218) are similar in that all the devices consist of a full functional SPECT and CT system. The patient may have a diagnostic SPECT and CT scan performed consecutively without having to move the patient. The Precedence provides a mean to reach the diagnostic decision faster than the conventional way of imaging patients with both SPECT and CT systems in separate locations. The combined Precedence SPECT-CT Imaging System is designed so that the system can operate in three modes: CT only, SPECT only, and combined SPECT/CT performed sequentially. The major difference is that the modified Precedence has added the CT subsystem options of the 64 slice and the CV configuration CT Scanner. - H. System Performance Test: - Radiation safety by compliance and certification to the performance standards for . ionizing radiation emitting product 21 CFR 1020.30 and 21 CFR 1020.3333. The radiation safety product report will be filed in accordance with 21 CFR 1002.10 with the Center for Device and Radiological Health. - Electrical and mechanical safety is assured as the system is designed to applicable . voluntary standards in the IEC 60601-1 series. The device performance was measured in accordance with the NEMA-NU-1 standard. - I. Conclusion: The Precedence Imaging System is substantially equivalent to the predicate devices, the Precedence SPECT/CT (cleared as Griffin) (K041218), Predicate Philips Plus CT Scanner (K033326), and the Brilliance CT, Private Practice CV configuration CT Scanner (K042293) based upon similar intended use, technological comparison, and system performance. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features the department's name encircling a stylized caduceus symbol. The caduceus is depicted as a staff with a serpent entwined around it, representing healing and medicine. The seal is in black and white. ### Public Health Service JUN 8 - 2005 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ADAC Laboratories, Inc. % Mr. Morten Simon Christensen Staff Engineer & FDA Office Coordinator Underwriters Laboratories, Inc. 1655 Scott Boulevard SANTA CLARA CA 95050-4169 Re: K051351 Trade/Device Name: Precedence SPECT/CT Imaging System Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: KPS and JAK Dated: May 19, 2005 Received: May 24, 2005 Dear Mr. Christensen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced we have re review your boother evere is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to the cholors) to regard ment date of the Medical Device Amendments, or to devices that have been May 20, 1970, and charment with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the do not require approval or or a provisions of the Act. The general controls provisions of the Act device, subjoct to the general registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket If your de roo be subject to such additional controls. Existing major regulations affecting your Approvation of the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling / (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {4}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) This letter will anow you to begin managements . premarket notification. The FDA finding of substantial equivalence of your device to a legally premarket notification: "The PDF maing of our device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please If you desire specific advice tor your de read on the following numbers, based on the regulation number at the top of this letter: | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Also, please note the regulation entired, formation on your responsibilities under the Act from the 807.97). You may only only concretional and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use 510(k) NUMBER (if known): KOS/351 DEVICE NAME: Precedence SPECT/CT Imaging System ## INDICATIONS FOR USE: Precedence is an imaging system combining the acquisition of single photon nuclear Precedence is an imaging system conforming consumed tomography system. The x-ray medicine images and images from an x-ray computer hads multi-slice CT or medicine images and mages from an x ray to f a whole body multi-slice CT or computed tomography subsystem may consisted a movides coronary imaging that is cardiovascular CI. The cardiovascular subsystem por cardiovascular, and peripheral intended to produce cross-sectional images of the hearsions asts taken at vascular systems by computer reconstruction of x-ray transmission data taken at vascular systems by computer reconstruction of its and annon-attenuation corrected and difference and planes. Frecedints into proceding harmaceuticals in the body as attenuation corrected images of the can concession as area to produce well as x-ray transmission images. The CT transmission data may be used to produce well as x-ray transmission mages. The CT manimablem maker medicine images and the CT attenuation corrected nuclear medicine images. The necicin' in the same attenuation corrected includine mings in a fused format (overlaid in the same images may be registered and displayed in a fused formation' data for natom images may be registered and displayed in rused formatoryical data for anatomical orientation) to provide combined single procedence may be used either as a separate localization of the nuclear medicine data. Precedence with and single photon localization of the nuclear medically data. I researce and single photos single photon system, a separate C. System or as a beatransfered to other systems such as system. The nuclear medicine and CT images may be transfered to other systems should system. The nuclear medicine and CT integos may or acked only be used by trained healthcare professionals. The Precedence CV Configuration CT subsystem is only for the imaging of the cardiovascular system. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Colin m. Pollard ivision Sign-Off
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%