K051315 · Sectra Imtec AB · LLZ · Jun 15, 2005 · Radiology
Device Facts
Record ID
K051315
Device Name
MODIFICATION TO: SECTRA IDS5 WORKSTATION
Applicant
Sectra Imtec AB
Product Code
LLZ · Radiology
Decision Date
Jun 15, 2005
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
The Sectra IDS5 device is intended for the manipulation and displaying of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards. Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing. Lossy compressed mammographic images are not intended for diagnostic review. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images. Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
Device Story
Sectra IDS5 Workstation is a software-based PACS client running on Windows PCs; processes and displays digital medical images from various modalities. Operates in clinical environments (hospitals, remote clinics, home) by radiologists, clinicians, and technicians. Inputs include DICOM-compliant medical images; outputs are visualized images on high-resolution monitors. Features include mammography reading, 3D reconstruction, angiography, and prosthesis CAD. Healthcare providers use the workstation to review images and reports to support clinical decision-making. System requires server-side connectivity; supports telecommunications and teleconferencing. Patient benefit derived from efficient image access and diagnostic review by trained professionals.
Clinical Evidence
No clinical data provided. Bench testing only; device complies with ISO 9001:2000 and ACR/NEMA DICOM 3.0 standards.
Technological Characteristics
Software-based PACS workstation; runs on Windows 2000/XP PCs. Supports multi-monitor configurations (up to four). Connectivity via DICOM 3.0. No patient-contacting components. Software-based image processing and visualization.
Indications for Use
Indicated for trained professionals (physicians, radiologists, nurses, technicians) to manipulate and display medical images, including mammograms, from various modalities for diagnostic review and clinical assessment.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
K051315
JUN 1 5 2005
# 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92c)
Date Prepared:
May 12, 2005
# Submitter's Information:
Sectra Imtec AB Teknikringen 20 SE-583 30 Linköping Sweden Phone: +1 46 13 23 52 00 Fax: +1 46 13 21 21 85
#### Trade Name, Common Name, Classification:
| Trade Name: | Sectra IDS5 Workstation |
|----------------------|---------------------------------------------------|
| Software version: | 10.2 |
| Common Name: | Picture Archiving and Communications System |
| Classification Name: | Image Processing System (LLZ) (21 CFR § 892.2050) |
### Predicate Device:
| Applicant: | Sectra Imtec AB |
|----------------|-------------------------|
| 510(k) Number: | K050196 |
| Device: | Sectra IDS5 Workstation |
## Device Description:
The Sectra IDS5 Workstation is mainly a software product. It is used for visualization and processing of digital medical images. The IDS5 is used as a client together with a Sectra provided server (Class I Exempt). The system runs on PCs under the Windows operating systems. Most notably two or more monitors are used.
{1}------------------------------------------------
The Sectra IDS5 Workstation is in fact a family of devices, including several workstations or types of workstations, e.g. the following:
- Primary diagnostics workstation and the most powerful version of IDS5 -. IDS5/dx.net. It contains tools for assisting advanced users, e.g. the radiologists, in making a diagnosis.
- Dedicated workstation for mammography IDS5/mx.net. It has all functionality as an . IDS5/dx.net but with an additional mammography package. Reading of mammographic images shall only be conducted with IDS5/mx.net.
- Quality assurance workstation, mainly used by the technologists to prepare the images . for the reviewing radiologist.
- Clinicians workstation used by the clinicians within the hospital to view the medical . images and to read the report.
- . "Web" workstation that can be used by remote clinics to view images and reports.
- "At-home" workstation that can be used by users, e.g. radiologists, over a low . bandwidth connection.
### Indications for Use:
The Sectra IDSS device is intended for the manipulation and displaying of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc .; and teleconferencing.
Lossy compressed mammographic images are not intended for diagnostic review. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
# Technological Characteristics:
The IDS5 Workstation runs under the Window 2000 and Windows XP operating system for PCs (as a minimum and depending upon system configuration)'. The requirements on hardware are quite ordinary for a system used for displaying images. Most notably up to four monitors can be used.
# Performance Data:
The subject device is developed according to ISO 9001:2000 and complies with ACR/NEMA Digital Imaging Communications in Medicine version 3.0.
<sup>|</sup> Windows NT is supported for IDS5/web and IDS5/cl.net only.
{2}------------------------------------------------
#### Conclusion:
Similar to the predicate device, the IDS5 Workstation does not contact the patient, nor does it control any life sustaining devices. Images and information being reviewed, processed, relayed, and or transmitted are interpreted by a physician or trained medical personnel, providing ample opportunity for competent human intervention. The device and the predicate device share the same certification or conformance to performance standards and both function as Image Processing System (LLZ). Device failures, which might result in partial or failed transmissions, images, or data, may be recovered from storage or re-transmission after correcting the problem(s). Passwords are required for operation and to protect against unauthorized use.
Based on the information supplied in this Special 510(k), we conclude that the subject device is safe, effective, and substantially equivalent to the predicate device.
Pat And
Peter Andersson Regulatory Affairs Officer Sectra Imtec AB Linköping, Sweden
{3}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. In the center of the seal is a stylized image of three overlapping human figures, which are meant to represent the department's mission of protecting the health of all Americans and providing essential human services.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
# JUN 1 5 2005
Sectra Imtec AB % Mr. Carl Alletto Official Correspondent OTech, Inc. 1600 Manchester Way CORINTH TX 76210
Re: K051315
Trade/Device Name: Sectra IDS5 Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: May 17, 2005 Received: May 23, 2005
Dear Mr. Alletto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (0S) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
This letter will allow you to begin marketing your device as described in your Section \$10(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | - | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address
http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Kostsiz 510(k) Number:
Device Name: IDS5 Workstation by Sectra Imtec AB
Indications For Use:
The Sectra IDS5 device is intended for the manipulation and displaying of medical images, including mammograms. It can show images from different modalities and interfaces to various image storage and printing devices using DICOM or similar interface standards.
Device options make possible mammography reading, telecommunications; fast demonstration; prosthesis CAD; 3-D and angiography, etc.; and teleconferencing.
Lossy compressed mammographic images are not intended for diagnostic review. Mammographic images should only be viewed with a monitor approved by FDA for viewing mammographic images.
Typical users of this system are trained professionals, including but not limited to physicians, radiologists, nurses, medical technicians, and assistants.
# (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
V Prescription Use (Per 21 CFR 801.109)
OR
Over -The-Counter Use ________________________________________________________________________________________________________________________________________________________
Daniel h. Syzmm
Division Sign-C sion of Reproduct
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.