MODIFICATION TO: MULTIRATE INFUSOR SV )WITH AND WITHOUT PCM), MULTIRATE INFUSOR LV (WITH AND WITHOUT PCM)

K051253 · Baxter Healthcare Corp · MEA · Jul 15, 2005 · General Hospital

Device Facts

Record IDK051253
Device NameMODIFICATION TO: MULTIRATE INFUSOR SV )WITH AND WITHOUT PCM), MULTIRATE INFUSOR LV (WITH AND WITHOUT PCM)
ApplicantBaxter Healthcare Corp
Product CodeMEA · General Hospital
Decision DateJul 15, 2005
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.5725
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the Multirate Infusor with Patient Control Module includes slow, continuous, subcutaneous or epidural administration of pain medications. It may also include the slow, continuous infusion of pain medications directly into an intraoperative site, subcutaneously for postoperative pain management or the continuous infusion of a local anesthetic near a nerve for regional anesthesia. The Patient Control Module allows for intermittent bolus doses of pain medication on patient demand.

Device Story

Single-use disposable elastomeric infusion pump; delivers medication at constant flow rate (1-12mL/hr) with patient-controlled bolus capability. Modification integrates Patient Control Module (PCM) into Multirate Infusor line using PVC 'Y' connectors and inline glass flow restrictors. Fluid path allows continuous flow via flow control module and bypass for bolus delivery. Used in clinical settings for pain management; operated by healthcare providers or patients (bolus). Device provides continuous infusion and on-demand bolus doses to manage pain; benefits include consistent medication delivery and patient-controlled analgesia.

Clinical Evidence

Bench testing only. No clinical data provided. Design verification tests were performed based on FMEA risk analysis to ensure modifications met acceptance criteria.

Technological Characteristics

Elastomeric infusion pump; materials include PVC and glass flow restrictors. Mechanical flow control via integrated modules. Single-use, disposable. No software or electronic components.

Indications for Use

Indicated for patients requiring slow, continuous subcutaneous, epidural, or intraoperative administration of pain medications, or regional anesthesia via local anesthetic near a nerve. Includes patient-controlled intermittent bolus dosing.

Regulatory Classification

Identification

An infusion pump is a device used in a health care facility to pump fluids into a patient in a controlled manner. The device may use a piston pump, a roller pump, or a peristaltic pump and may be powered electrically or mechanically. The device may also operate using a constant force to propel the fluid through a narrow tube which determines the flow rate. The device may include means to detect a fault condition, such as air in, or blockage of, the infusion line and to activate an alarm.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ KOS1253 510(k) Notification: Device Modification # JUL 1 5 2005 ### 510 (K) SUMMARY Baxter Healthcare Corporation Multirate Infusor with PCM | Preparer/Contact | Nanette Hedden<br>Project Manager<br>Global Regulatory Affairs<br>1620 Waukegan Rd.<br>McGaw Park, IL 60085<br>Telephone: (847) 473-6281<br>Fax: (847) 785-5116<br>Baxter Healthcare Corporation<br>One Baxter Parkway<br>Deerfield, Illinois 60015 | |----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Manufacturer | Manufacturing Location: 2025695<br>Baxter Healthcare Corporation<br>17511 Armstrong Avenue<br>Irvine, California 92614<br>Phone: (949) 474-6306<br>Fax: (949) 474-6375 | | Date summary was prepared: | May 6, 2005 | | Name(s) of the device: | Multirate Infusor (LV and SV) with<br>PCM | | Identification of predicate device(s): | Baxter Multirate Infusor (LV and SV)<br>Baxter PCM | #### Description of the device and modifications: Baxter's Multirate Infusor devices with attached PCM, are single-use disposable infusion pumps designed to deliver solution at a constant flow rate of 1-12mL/hr, depending upon the device configuration and settings, as well as allow for intermittent bolus doses of medication on patient demand. The modifications made to the Multirate Infusor (K011317) line to incorporate a PCM (K002739) include the following: - Inserting the same PVC "Y" connector from the single rate Infusors (K041738, . K884505) before the Multirate Flow Control Module so the fluid going to the PCM can bypass the Multirate Flow Control Module. (See Section 7.4.4 for drawings showing the Multirate Infusor with PCM.) {1}------------------------------------------------ ## 510(k) Notification: Device Modification - Providing an inline flow restrictor (the same glass flow restrictor already utilized in . this line) that provides a single flow rate for filling the PCM reservoir. - Another PVC "Y" connector is added at the output of the Multirate Flow Control ● Module and connects with the output of the Patient Control Module to allow for a single line to the patient which provides continuous flow as well as bolus doses when needed (the same as for the single rate Infusors with PCM attached). All of the changes above incorporate the same technology and materials that are already incorporated in the Baxter line of elastomeric infusion devices. Intended Use: The intended use of the Multirate Infusor with Patient Control Module includes slow, continuous, subcutaneous or epidural administration of pain medications. It may also include the slow, continuous infusion of pain medications directly into an intraoperative site, subcutaneously for postoperative pain management or the continuous infusion of a local anesthetic near a nerve for regional anesthesia. The Patient Control Module allows for intermittent bolus doses of pain medication on patient demand. #### Evaluation of Design Modifications As the basis for Baxter Healthcare's device evaluation studies and overall process for managing medical device risk, the company has performed a risk analysis using procedures based on ISO 14971 (2000) "Medical Devices - Application of Risk Management to Medical Devices". The risk analysis method used to assess the impact of the modification was Failure Modes and Effects Analysis (FMEA). Design verification tests based on the result of risk analysis and design input were performed to verify those modifications. All test results meet the acceptance criteria, and proved that those modifications to be appropriate. ### Conclusion The modified device, the Multirate Infusor with PCM, is substantially equivalent to the Multirate Infusor and PCM devices sold as stand alone devices. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health and Human Services (HHS). The logo features a stylized eagle with three stripes forming its body and wings. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES . USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 JUL 1 5 2005 Ms. Nanette Hedden Project Manager, Global Regulatory Affairs Baxter Healthcare Corporation 1620 Waukegan Road McGraw Park, Illinois 60085 Re: K051253 Trade/Device Name: Multirate Infusor (LV and SV) with Patient Control Module Regulation Number: 21 CFR 880.5725 Regulation Name: Infusion Pump Regulatory Class: II Product Code: MEB, MEA Dated: June 17, 2005 Received: June 20, 2005 Dear Ms. Hedden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register. {3}------------------------------------------------ Page 2 – Ms. Hedden Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR I-art 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Syrita Y. Michael Davis Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use . 510(k) Number (if known): Kos 1253 Device Name: Multirate Infusor LV and SV with Patient Control Module Indications For Use: maloutions i or 500. The intended use of the Multirate Infusor with Patient Control Module includes slow, The intendou doe of throus or epidural administration of pain medications. It may also include the slow, continuous infusion of pain medications directly into an intra-operative nitiations the olon, continuous and management or the continuous infusion of a local anesthetic near a nerve for regional anesthesia. The Patient Control Module allows for intermittent bolus doses of pain medication on patient demand. Prescription Use _____________________________________________________________________________________________________________________________________________________________ AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Rill C. Aym for me (Division Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices 510(k) Number: K051253
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...